BD Nexiva and BD Nexiva Diffusics Closed IV Catheter Systems
K161777 · Becton Dickinson Infusion Therapy Systems, Inc. · FOZ · Aug 29, 2016 · General Hospital
Device Facts
Record ID
K161777
Device Name
BD Nexiva and BD Nexiva Diffusics Closed IV Catheter Systems
Applicant
Becton Dickinson Infusion Therapy Systems, Inc.
Product Code
FOZ · General Hospital
Decision Date
Aug 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed. BD Nexiva Diffusics closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when access ports not suitable for use with power injectors are removed.
Device Story
Over-the-needle peripheral intravascular catheter systems; include radiopaque Vialon catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, Luer adapter, and vent plug. Nexiva Diffusics variant includes side holes at catheter tip to optimize power injection. Features BD Instaflash technology for immediate blood visualization. Closed system design contains blood during insertion; septum wipes needle surface upon withdrawal to reduce exposure risk; passive needle tip protection reduces needlestick injury risk. Used in clinical settings by healthcare providers for vascular access. Output is physical access to patient's vascular system for fluid/medication delivery, blood sampling, or pressure monitoring. Benefits include reduced blood exposure, improved insertion visualization, and power injection capability.
Clinical Evidence
Bench testing only. Performance testing included air leakage, liquid leakage, fluid leakage at venous pressure, septum assembly failure/separation at max pressure, cannula/hub bond strength, needle hub/tube union strength, and needle removal force (system drag). All pre-determined acceptance criteria were met.
Technological Characteristics
Over-the-needle peripheral IV catheter. Materials: Vialon catheter, copolyester septum canister, polyisoprene septum. Features: integrated extension set, passive needle shielding, Instaflash flashback visualization. Diffusics model includes tip side holes for power injection. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for any patient population requiring short-term vascular access for blood sampling, blood pressure monitoring, or fluid administration. Suitable for power injection (up to 300 psi for Nexiva; up to 325 psi for Nexiva Diffusics) when appropriate ports are used.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth. The graphic is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2016
Becton Dickinson Infusion Therapy Systems Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street North West Buffalo, Minnesota 55313
Re: K161777
Trade/Device Name: BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV Catheter Systems Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II, Product Code: FOZ Dated: August 12, 2016 Received: August 16, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D Acting Division Director Science and Policy Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number ( <i>if known</i> ) | |
| Device Name | BD Nexiva Closed IV Catheter System |
| Indications for Use (Describe) | BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to<br>sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with<br>consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of<br>therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300<br>psi (2068 kPa) when access ports not suitable for use with power injectors are removed. |
| Type of Use (Select one or both, as applicable) | <div> <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) Page 1 of 1
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number (if known) | |
|--------------------------------|-----------------------------------------------|
| Device Name | BD Nexiva Diffusics Closed IV Catheter System |
| Indications for Use (Describe) | |
BD Nexiva Diffusics closed IV catheter systems are intended to be inserted into a patient's vascular system for short term
use to sample blood, monitor blood pressure, or administer fluids. These devices may be used for any patient population
with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration
of therapy. These devices are suitable for use with power injectors set to a maximum pressure of 325 psi (2240 kPa) when
access ports not suitable for use with power injectors are removed.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label><br><label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
40
{4}------------------------------------------------
bd.com
Image /page/4/Picture/2 description: The image shows the BD logo. On the left is an orange circle with a white star-like shape inside. To the right of the circle are the letters "BD" in blue. The letters are large and bold.
## 510(k) Summary (21 CFR §807.92)
## BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV Catheter Systems
| Submitter<br>Information | Submitter Name:<br>Submitter Address: | Becton Dickinson Infusion Therapy Systems Inc.<br>9450 South State Street<br>Sandy, UT 84070 | Device<br>Description<br>( <i>BD Nexiva</i> ) | BD Nexiva closed IV catheter systems are over-the-needle, intravascular<br>catheters. These devices have a radiopaque BD Vialon™ catheter, needle, needle<br>shield, septum, stabilization platform, integrated extension tubing, clamp, Luer<br>adapter, vent plug and, with dual port configurations, a pre-attached BD Q-Syte<br>Luer Access Split Septum. The needle and catheter are protected by a needle<br>cover. For dual port configurations, a BD Q-Syte device or an end cap with<br>protective cover is provided in the unit package. |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Kimberly Geisler<br>Staff Regulatory Affairs Specialist<br>kimberly.geisler@bd.com<br>(801) 565-2422 (phone) | | The closed system is designed to keep blood contained within the device<br>throughout the insertion process. The septum is designed to wipe visible blood<br>from the needle surface as the needle is withdrawn from the catheter, further<br>reducing the risk of blood exposure. The needle tip is passively protected when<br>the needle is removed, reducing the risk of accidental needlestick injury. |
| | Date of Preparation: | July 28, 2016 | | These devices have BD Instaflash™ Needle Technology, allowing for immediate<br>visualization of blood along the catheter. Continuous blood return is seen in the<br>extension tubing. The vent plug prevents blood leakage from the extension tubing<br>during insertion. Both the stabilization platform and Luer connector are color coded<br>to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue,<br>20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). |
| Subject Device<br>(BD Nexiva) | Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Closed IV Catheter System<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | Device<br>Description<br>( <i>BD Nexiva<br/>Diffusics</i> ) | BD Nexiva Diffusics closed IV catheter systems are over-the-needle, intravascular<br>catheters. These devices have a radiopaque BD Vialon™ catheter with side holes<br>located near the tip of the catheter which are designed to optimize power injection<br>procedures. These devices also have a needle, needle shield, septum,<br>stabilization platform, integrated extension tubing, clamp, Luer connector, and vent<br>plug. The Luer connector displays gauge-specific maximum flow rate and the<br>maximum power injector pressure limit setting. The needle and catheter are<br>protected by a needle cover. An end cap with protective cover is provided in the<br>unit package. |
| Predicate<br>Device<br>(BD Nexiva) | Trade Name:<br>510(k) Reference:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Closed IV Catheter System<br>K102520<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | | The closed system is designed to keep blood contained within the device<br>throughout the insertion process. The septum is designed to wipe visible blood<br>from the needle surface as the needle is withdrawn from the catheter, further<br>reducing the risk of blood exposure. The needle tip is passively protected when<br>the needle is removed, reducing the risk of accidental needlestick injury. |
| Subject Device<br>(BD Nexiva<br>Diffusics) | Trade Name:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Diffusics™ Closed IV Catheter System<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | | These devices have BD Instaflash™ Needle Technology, allowing for immediate<br>visualization of blood along the catheter. Continuous blood return is seen in the<br>extension tubing. The vent plug prevents blood leakage from the extension tubing<br>during insertion. The stabilization platform and Luer connector are color coded to<br>indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA<br>(1.1 mm)=Pink, 18GA (1.3 mm)=Green). |
| Predicate<br>Device<br>(BD Nexiva<br>Diffusics) | Trade Name:<br>510(k) Reference:<br>Common Name:<br>Regulation Number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | BD Nexiva™ Diffusics™ Closed IV Catheter System<br>K123734<br>Peripheral Intravascular or IV Catheter<br>21 CFR §880.5200<br>Intravascular Catheter<br>II<br>FOZ<br>General Hospital | | |
{5}------------------------------------------------
bd.com
{6}------------------------------------------------
bd.com
| Indications for<br>Use<br>(BD Nexiva) | Modifications were made to the Indications for Use for the purpose of aligning the<br>Indications for Use across the BD Nexiva family of products and for clarification.<br><br>Modifications include: 1) removal of the regulation reference as it is not relevant to<br>the healthcare practitioner; 2) removal of device features and performance<br>information which does not determine intended use; 3) removal of "less than 30<br>days" since this statutory regulation language for 21 CFR §880.5200 regarding the<br>device classification could be confusing to end users as the duration of use is<br>dependent on a number of factors, including patient status, duration of treatment,<br>and institutional protocol; 4) addition of "fluids being infused" and "duration of<br>therapy" as considerations for clinicians to evaluate prior to use of these catheters<br>for a particular patient; and 5) minor wording changes. |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
None of these changes are critical to the intended use and do not raise new questions of safety and effectiveness of the devices when used as indicated.
| SUBJECT | PREDICATE (K102520) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| BD Nexiva™ Closed IV Catheter System | BD Nexiva™ Closed IV Catheter System |
| Indications for Use | Indications for Use |
| BD Nexiva closed IV catheter systems are intended to be inserted into a patient's vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed. | As indicated in 21 CFR Part 880.5200, The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure. The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made. |
{7}------------------------------------------------
| bd.com | |
|--------|--|
| | |
| | Indications for<br>Use<br>(BD Nexiva<br>Diffusics) | Modifications were made to the Indications for Use for the purpose of aligning the<br>Indications for Use across the BD Nexiva family of products and for clarification.<br>Modifications include: 1) addition of "short term use" in alignment with the device<br>classification; 2) addition of the intended patient population, which is consistent<br>with the predicate device, which does not indicate a specific patient population<br>and is intended for general use; 3) removal of the flow rate information as it<br>pertains to device performance and does not determine intended use (Note: flow<br>rate information is provided in the Instructions for Use); and 4) minor wording<br>changes. None of these changes are critical to the intended use and do not raise<br>new questions of safety and effectiveness of the devices when used as indicated. | | |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--|
| | SUBJECT<br>BD Nexiva™ Diffusics™ Closed IV<br>Catheter System<br>Indications for Use | PREDICATE (K123734)<br>BD Nexiva™ Diffusics™ Closed IV<br>Catheter System<br>Indications for Use | | |
| | BD Nexiva Diffusics closed IV catheter<br>systems are intended to be inserted into a<br>patient's vascular system for short term use<br>to sample blood, monitor blood pressure, or<br>administer fluids. These devices may be<br>used for any patient population with<br>consideration given to adequacy of vascular<br>anatomy, procedure being performed, fluids<br>being infused, and duration of therapy.<br>These devices are suitable for use with<br>power injectors set to a maximum pressure<br>of 325 psi (2240 kPa) when access ports not<br>suitable for use with power injectors are<br>removed. | The BD Nexiva™ Diffusics™ intravascular<br>catheter is inserted into a patient's<br>vascular system to sample blood, monitor<br>blood pressure or administer fluids. The<br>BD Nexiva Diffusics catheters are suitable<br>for use with power injectors when a direct<br>connection is made. The maximum flow<br>rate and maximum power injector pressure<br>setting for each catheter size are listed in<br>the table below: | | |
| | | Max Flow Rate<br>(mL/sec) | Max Injector Setting<br>(psi) | |
| | 24 GA x 0.75 IN | 3.0 | 325 | |
| | 22 GA x 1.00 IN | 6.5 | 325 | |
| | 20 GA x 1.00 IN | 10.0 | 325 | |
| | 20 GA x 1.25 IN | 10.0 | 325 | |
| | 18 GA x 1.25 IN | 15.0 | 325 | |
| | | | | |
| | Technological<br>Characteristics | Technological characteristics of the subject and predicate devices are equivalent.<br>The subject BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV Catheter<br>Systems achieve their intended use based on the same technology and principles<br>of operation as the predicate BD Nexiva™ and BD Nexiva™ Diffusics™ Closed IV<br>Cather Systems, respectively. Design modifications were made to the subject<br>devices to reduce the force required to withdraw the needle from the catheter<br>assembly. Performance testing performed to support the design modifications<br>included:<br>• Air Leakage<br>• Liquid Leakage<br>• Fluid Leakage at Venous Pressure<br>• Septum Assembly Failure / Separation at Max Pressure<br>• Cannula / Hub Bond Strength<br>• Strength of Union between Needle Hub and Needle Tube<br>• Force to Remove Needle from Catheter System (Average System Drag)<br>• System Drag - Peak Offset Force<br>All of the pre-determined acceptance criteria were met. | | |
{8}------------------------------------------------
## bd.com
| Attribute | SUBJECT<br>BD Nexiva™ Closed IV Catheter System | PREDICATE (K102520)<br>BD Nexiva™ Closed IV Catheter System |
|--------------------------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Same as predicate | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter |
| Fundamental<br>Scientific<br>Technology | Same as predicate | Closed peripheral intravascular catheter<br>system designed with an integrated extension<br>set incorporating either a single port or Y<br>(dual)-port injection site. Incorporates BD<br>Instaflash™ technology to assist with<br>flashback visualization. |
| Components /<br>Materials | Septum Canister:<br>Copolyester + grey<br>colorant* | Septum Canister:<br>Copolyester + white<br>colorant |
| | All other subject component materials are the same as the predicate device | |
| Physical /<br>Mechanical<br>Specifications | Same as predicate | Catheter Diameters<br>Catheter Lengths<br>18G, 20G, 22G, 24G<br>0.56", 0.75", 1.00",<br>1.25", 1.75" |
*The subject BD Nexiva™ Closed IV Catheter System and subject BD Nexiva™ Diffusics™ Closed IV Catheter System septum canister materials are identical. As such, the Nexiva™ Closed IV Catheter System (K123734) serves as the reference device supporting this modification to the BD Nexiva™ Closed IV Catheter System non-patient contacting septum canister material.
| Attribute | SUBJECT<br>BD Nexiva™ Diffusics™ Closed IV<br>Catheter System | PREDICATE (K123734)<br>BD Nexiva™ Diffusics™ Closed IV Catheter<br>System |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Same as predicate | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter |
| Fundamental<br>Scientific<br>Technology | Same as predicate | Closed peripheral intravascular catheter<br>system designed with an integrated extension<br>set incorporating either a single port or Y<br>(dual)-port injection site. Incorporates BD<br>Instaflash™ technology to assist with<br>flashback visualization.<br>In addition, the catheter tip includes diffuser<br>holes to reduce the velocity of contrast media<br>exiting the catheter tip during CT scans. |
| Components /<br>Materials | Septum Canister: Copolyester + grey<br>colorant (modified<br>resin to colorant ratio<br>compared to predicate)<br><br>Septum: Polyisoprene (no<br>lubricant) | Septum Canister: Copolyester + grey<br>colorant<br><br>Septum: Polyisoprene + Silicone<br>Lubricant |
| All other subject component materials are the same as the predicate device | | |
| Physical /<br>Mechanical<br>Specifications | Same as predicate | Catheter Diameters Catheter Lengths<br>18G, 20G, 22G, 24G 0.75", 1.00", 1.25" |
{9}------------------------------------------------
bd.com
| Summary of<br>Performance<br>Tests | The following performance testing was performed to demonstrate substantial<br>equivalence of the subject BD NexivaTM and BD NexivaTM DiffusicsTM Closed IV<br>Catheter Systems to the predicate BD NexivaTM Closed IV Catheter System<br>(K102520) and BD NexivaTM DiffusicsTM Closed IV Catheter System (K123734),<br>respectively, and applicable standards: |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Air Leakage Liquid Leakage Fluid Leakage at Venous Pressure Septum Assembly Failure / Separation at Max Pressure Cannula / Hub Bond Strength Strength of Union between Needle Hub and Needle Tube Force to Remove Needle from Catheter System (Average System Drag) System Drag - Peak Offset Force All of the pre-determined acceptance criteria were met. |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and performance<br>testing, the subject BD NexivaTM Closed IV Catheter System and BD NexivaTM<br>DiffusicsTM Closed IV Catheter System are demonstrated to be substantially<br>equivalent to the predicate BD NexivaTM Closed IV Catheter System (K102520)<br>and BD NexivaTM DiffusicsTM Closed IV Catheter System (K123734), respectively. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.