DynaMatrix is intended for use in guided tissue regeneration and bone regeneration procedures. It may be used for bone regeneration and healing of periodontal defects, for gingival augmentation, to maintain or enhance alveolar ridges, or to contain or prevent migration of graft material. The device is provided sterile and is intended for one-time use.
Device Story
DynaMatrix/DynaMatrix Plus consists of layered sheets of bioabsorbable extracellular collagen membrane matrix derived from porcine Small Intestinal Submucosa (SIS). Device acts as passive, resorbable scaffold; guides epithelial tissue ingrowth and angiogenesis; maintains space around bony defects; prevents epithelial down-growth; allows new bone formation. Used in oral surgery settings by clinicians. Supplied sterile in Tyvek pouches; trimmed/shaped by clinician to fit defect. Implanted once; remains in body until resorbed. Benefits include guided tissue/bone regeneration and defect healing.
Clinical Evidence
No clinical trials performed. Evidence based on risk analysis, verification activities, and a swine pilot study demonstrating successful incorporation and lack of foreign body response for devices of similar thickness. Biocompatibility qualification performed.
Technological Characteristics
Material: Porcine small intestinal submucosa (collagen types I and III). Form: Lyophilized, layered sheets (100-800 µm nominal thickness). Sterilization: Ethylene oxide. Resorption: 16 weeks (DynaMatrix) to 26 weeks (DynaMatrix Plus). Implantable, bioabsorbable, single-use.
Indications for Use
Indicated for human patients requiring guided tissue regeneration or bone regeneration procedures, including healing of periodontal defects, gingival augmentation, alveolar ridge enhancement, or containment of graft material in oral surgery settings.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2017
Cook Biotech Incorporated Perry Guinn VP Regulatory Affairs And Quality Assurance 1425 Innovation Place West Lafayette, Indiana 47906
Re: K161762
Trade/Device Name: Dynamatrix/dynamatrix Plus Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: June 24, 2016 Received: June 27, 2016
Dear Perry Guinn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161762
Device Name DynaMatrix/DynaMatrix Plus
Indications for Use (Describe)
DynaMatrix is intended for use in guided tissue regeneration and bone regeneration procedures. It may be used for bone regeneration and healing of periodontal defects, for gingival augmentation, to enhance alveolar ridges, or to contain or prevent migration of graft material. The device is supplied sterile and intended for one-time use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
Submitted by: Perry Guinn, Vice President, Quality Assurance and Regulatory Affairs Cook Biotech Incorporated 1425 Innovation Place West Lafayette, IN 47906 (765) 497-3355 24 June, 2016
### Name of Device:
| Trade/Proprietary Names: | DynaMatrix™, DynaMatrix Plus™ |
|-------------------------------|-----------------------------------|
| Common/Usual Names: | Barrier, animal source, intraoral |
| Proposed Classification Name: | Bone Grafting Material |
| Product Code: | NPL |
| Device Class: | 21 CFR §872.3930, Class II |
Performance Standards: No performance standards that have been established under Section 514 of the Food, Drug and Cosmetic Act apply to this device.
#### Predicate Device:
The predicate device is DynaMatrix (K082058), cleared October 2, 2008.
### Reference Device:
The reference device is Surgisis Periodontal Membrane (K010952), cleared June 10, 2002.
### Intended Use:
DynaMatrix is intended for use in guided tissue regeneration and bone regeneration procedures. It may be used for bone regeneration and healing of periodontal defects, for gingival augmentation, to maintain or enhance alveolar ridges, or to contain or prevent migration of graft material.
The device is provided sterile and is intended for one-time use.
This intended use is identical to that of the predicate device cleared under K082058.
### Device Description:
DynaMatrix consists of layered sheets of bioabsorbable extracellular collagen membrane matrix derived from porcine Small Intestinal Submucosa (SIS). These sheets are lyophilized, packaged in a Tyvek pouch, and sterilized using ethylene oxide to an SAL of 10°. The device is available in sheets ranging from 2 cm to 12 cm2, and can be trimmed or shaped to the appropriate size to
{4}------------------------------------------------
fit the defect to be treated. When hydrated, the membrane maintains suture tear resistance. It is considered to be a permanent contact implanted device, and is fully biocompatible.
DynaMatrix is available by prescription only and is intended for use in an oral surgery setting for guided tissue regeneration and bone regeneration procedures. It may be used for bone regeneration and healing of periodontal defects, for gingival augmentation, to maintain or enhance alveolar ridges, or to contain or prevent migration of graft material. Dynamatrix achieves its intended use by providing a passive, resorbable scaffold to guide epithelial tissue ingrowth and angiogenesis of the epithelial layer in order to maintain space around the bony defect, prevent epithelial down-growth, and allow for new bone formation.
## Comparison to Predicate Device:
This Special 510(k) describes a dimensional change to supply additional DynaMatrix graft sizes (within the cleared size range of the reference device). The indication for use remains identical to that of the predicate device.
The subject device is identical to the predicate with regard to intended use, indications for use, target population, anatomical site, use setting, performance, materials, compatibility with the environment and other devices, chemical safety, design, standards met, biocompatibility, and sterility. The subject differs from the predicate device solely in regard to its dimension as detailed in the Substantial Equivalence Table (Table 5-1). The introduction of new device sizes does not affect the intended use or the substantial equivalence of the product.
# Summary of Non-Clinical Tests:
No additional testing was required.
### Substantial Equivalence:
Table 5-1 below provides a comparison of the subject device, its predicate and the reference device.
| Device | DynaMatrix/ DynaMatrix<br>Plus (Subject Device) | DynaMatrix<br>(Predicate Device) | SURGISIS Periodontal<br>Membrane<br>(Reference Device) |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Cook Biotech Inc. | Cook Biotech Inc. | Cook Biotech Inc. |
| 510(k) number | unassigned | K082058 | K010952 |
| Product Code | NPL | NPL | NPL |
| Material<br>composition | Porcine small intestinal<br>submucosa; primarily<br>collagen types I and III | Porcine small intestinal<br>submucosa; primarily<br>collagen types I and III | Porcine small intestinal<br>submucosa; primarily<br>collagen types I and III |
| Supplied sterile? | Yes | Yes | Yes |
| Sterilization method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Intended for single<br>use? | Yes | Yes | Yes |
| Device | DynaMatrix/ DynaMatrix<br>Plus (Subject Device) | DynaMatrix<br>(Predicate Device) | SURGISIS Periodontal<br>Membrane<br>(Reference Device) |
| Intended for<br>prescription use<br>only? | Yes | Yes | Yes |
| Target population | Human, oral, periodontal | Human, oral, periodontal | Human, oral, periodontal |
| Location of intended<br>application | Setting appropriate for oral<br>surgery | Setting appropriate for<br>oral surgery | Setting appropriate for<br>oral surgery |
| Frequency and<br>duration of use | Device is intended to be<br>implanted once and remain<br>in the body until resorbed<br>(implant with contact >30<br>days) | Device is intended to be<br>implanted once and<br>remain in the body until<br>resorbed (implant with<br>contact >30 days) | Device is intended to be<br>implanted once and<br>remain in the body until<br>resorbed (implant with<br>contact >30 days) |
| Materials required<br>for use | No additional materials are<br>required | No additional materials<br>are required | No additional materials<br>are required |
| Packaging<br>configuration | Tyvek® Pouch | Tyvek® Pouch | Tyvek® Pouch |
| Shelf life | 18 months | 18 months | 18 months |
| Intended Use | DynaMatrix™ is intended<br>for use in guided tissue<br>regeneration and bone<br>regeneration procedures. It<br>may be used for bone<br>regeneration and healing of<br>periodontal defects, for<br>gingival augmentation, to<br>maintain or enhance<br>alveolar ridges, or to contain<br>or prevent migration of graft<br>material. | DynaMatrix™ is intended<br>for use in guided tissue<br>regeneration and bone<br>regeneration procedures.<br>It may be used for bone<br>regeneration and healing<br>of periodontal defects, for<br>gingival augmentation, to<br>maintain or enhance<br>alveolar ridges, or to<br>contain or prevent<br>migration of graft<br>material. | SURGISIS® Periodontal<br>Membrane is a<br>bioabsorbable,<br>implantable material<br>intended to aid in the<br>treatment of periodontal<br>defects. |
| Sizes | 2 cm2 to 12 cm2 | 1.5 cm x 2.0 cm (3 cm2)<br>2.0 cm x 3.0 cm (6 cm2)<br>3.0 cm x 4.0 cm (12 cm2)<br>(3 cm2 to 12 cm2) | 0.5 cm2 to 50 cm2 |
| Thickness (µm) | 100-800 (nominal) | 100-300 (nominal) | 100-800 (nominal) |
| Resorption profile<br>(Fully resorbed<br>within) | DynaMatrix: 16 weeks<br>DynaMatrix Plus: 26 weeks | 16 weeks | 16 weeks |
Table 5-1. Substantial Equivalence Information
{5}------------------------------------------------
### Discussion of Similarities and Differences:
The subject device (DynaMatrix Plus) has been compared to the predicate device (DynaMatrix, K082058) on the bases of fundamental scientific technology and intended use. The intended use and material composition of both subject and predicate devices are identical. The sole difference is the addition of new sizes to the product line. Any potential new risks associated with the additional device sizes have been identified by appropriate risk analysis methods. These potential new risks have been addressed with verification activities in a manner
{6}------------------------------------------------
satisfactory to pre-determined acceptance criteria to ensure that no change to device substantial equivalence has occurred. Specifically, biocompatibility qualification was performed to ensure that no new risks of biocompatibility were present. Validation activities were comprised of a swine pilot study that showed successful incorporation and no foreign body response for devices matching the nominal thickness of the subject device, and FDA clearance of another SIS device with the same nominal thickness (K133306).
### Conclusion:
Successful risk analysis and completion of verification activities provide evidence to support the conclusion that the size modification does not introduce new risks and that the subject device performs comparably and is substantially equivalent to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.