K161744 · Symmetry Surgical, Inc. · HAE · Sep 7, 2016 · Neurology
Device Facts
Record ID
K161744
Device Name
Symmetry Sharp Kerrison Rongeur
Applicant
Symmetry Surgical, Inc.
Product Code
HAE · Neurology
Decision Date
Sep 7, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4840
Device Class
Class 2
Indications for Use
The Symmetry Sharp Kerrison® Rongeur is indicated for cutting bony tissue of the spine. The Symmetry Sharp Kerrison® Rongeur is a manually operated instrument that consists of a Handle and Tips (available in single use or reusable).
Device Story
Manual surgical instrument; used for cutting bony tissue in spine. Consists of handle and detachable tips (single-use or reusable). Surgeon operates device by squeezing handle; mechanical compression closes cup-and-ball assembly; sliding shaft and stationary footplate cut bone. Tips available in 1-4mm sizes; handles in 5-7 inch lengths. Reusable components require cleaning and steam sterilization; single-use tips provided sterile via gamma sterilization. Device provides surgeon with various cutting configurations for spinal bone characteristics. Benefits include precise bone removal during spinal procedures.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included cleaning validation (residual protein/hemoglobin/carbohydrate), steam sterilization validation (SAL 10^-6), gamma sterilization validation (SAL 10^-6), durability testing (100 cuts for tips, 50 cycles for handles), corrosion resistance, and shelf-life studies (2 years) per ASTM standards.
Technological Characteristics
Manual rongeur; 420 stainless steel construction; German Stainless Steel 1.4301/1.4310 (equivalent to 304/303/301 SS) for mating components. Mechanical actuation via cup-and-ball assembly. Dimensions: 1-4mm cup sizes, 5-7 inch handle lengths. Sterilization: Steam (reusable components), Gamma (single-use tips). No software or electronics.
Indications for Use
Indicated for cutting bony tissue of the spine in patients requiring spinal surgery.
Regulatory Classification
Identification
A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.
Predicate Devices
Symmetry Kerrison Disposable Tip Rongeur (K130541)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2016
Symmetry Surgical Inc. % Ms. Michele Lucey Regulatory Consultant Lakeshore Medical Device Consulting LLC 128 Blye Hill Landing Newbury. New Hampshire 03255
Re: K161744
Trade/Device Name: Symmetry Sharp Kerrison® Rongeur Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: August 6, 2016 Received: August 8, 2016
Dear Ms. Lucey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena -5ி/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161744
#### Device Name Symmetry Sharp Kerrison® Rongeur
Indications for Use (Describe)
The Symmetry Sharp Kerrison Rongeur is indicated for cutting bony tissue of the Symmetry Sharp Kerrison Rongeur is a manually operated instrument that consists of a Handle and Tips (available in single use or reusable).
Type of Use (Select one or both, as applicable)
| <div> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|--------------------------------------------------------------------------------|
| <div> <span></span> Over-The-Counter Use (21 CFR 801 Subpart D) </div> |
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K161744
## Symmetry Surgical Inc Special
### 510(k) Summary
### This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92(c)
### Submitter Information:
| Submitter's<br>Name: | Symmetry Surgical Inc. |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Address: | 3034 Owen Drive, Antioch TN 37013 |
| Telephone: | 1-800-251-3000 |
| Fax: | 1-615-964-5567 |
| Submitter<br>Regulatory<br>Contact: | Michele Lucey<br>Lakeshore Medical Device Consulting, LLC<br>128 Blye Hill Landing<br>Newbury, NH 03255<br>603-763-3455 |
| Date Prepared: | September 7, 2016 |
| Device Trade<br>Name: | Symmetry Sharp Kerrison® Rongeur |
| Common/Usual<br>Name: | Manual Rongeur |
| Regulation<br>Name: | Manual Rongeur |
| Regulation<br>Number: | 21 CFR 882.4840 |
{4}------------------------------------------------
| Class | 2 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | HAE |
| Predicate<br>Device | K130541 Kerrison Disposable Tip Rongeur |
| Intended Use | The Symmetry Sharp Kerrison® Rongeur is indicated for cutting<br>bony tissue of the spine. The Symmetry Sharp Kerrison®<br>Rongeur is a manually operated instrument that consists of a<br>Handle and Tips (available in single use or reusable). |
| Device<br>Description | The Symmetry Sharp Kerrison Rongeur is designed to have<br>detachable tips that are disposable. The detachable tips are<br>provided as either sterile single use or as nonsterile reusable tips.<br>The reusable tips must be cleaned and sterilized before use. The<br>sterile, disposable tips are available in 4 cup sizes: (1, 2, 3, and 4<br>mm); which are identifiable by etchings on each device. The<br>Reusable Tips for the Symmetry Sharp Kerrison® Rongeur Handle<br>are available in 3 cup sizes: (2, 3, and 4 mm); which are<br>identifiable by etchings on each device. The reusable tips have a<br>mating code etched on each pair to ensure that only matched pairs<br>are used. Each tip is designed with a shaft key that fits into its<br>mating slot on the detachable instrument handle. The tip has cups,<br>one within the stationary footplate and the other within the upper<br>sliding shaft, performing the cutting of the bony tissue in the spine<br>area. The various cup sizes within the working tips provide the<br>surgeon various cutting configurations needed for the different<br>characteristics of the bony tissue within the spine. The Symmetry<br>Sharp Kerrison® Rongeur Handle operates with a cup and ball<br>assembly for smooth motion of the footplate when the handle is<br>squeezed. The detachable, reusable instrument handles of this<br>device are provided non-sterile and must be cleaned and sterilized<br>prior to use. The instrument handles are available in three shaft<br>lengths: (5, 6, and 7 inches). |
{5}------------------------------------------------
| Feature/Specification | Subject Device | Predicate Device | Comparison |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Symmetry Sharp Kerrison®<br>Rongeur | Symmetry Kerrison Disposable<br>Tip Rongeurs | Of Subject Device to<br>Predicate Device |
| | | K130541 | |
| General Description | Manual Rongeur | Manual Rongeur | Same |
| Intended Use | The Symmetry Sharp<br>Kerrison® Rongeur is<br>indicated for cutting bony<br>tissue of the spine. The<br>Symmetry Sharp<br>Kerrison®<br>Rongeur is a manually<br>operated instrument that<br>consists of a Handle and<br>Tips (available in single<br>use or reusable). | The Rongeur (handle and<br>tip) is indicated for cutting<br>bony tissue of the spine.<br>The Symmetry Kerrison<br>Disposable Tip Rongeur is<br>a manually operated<br>instrument that consists of<br>handle and disposable<br>detachable tips. | Same, except for the<br>removal of "disposable" to<br>allow for a general tip<br>description |
| Device Size and Geometry | Tips are provided in 4 cup<br>sizes, 1,2, 3, and 4 mm.<br>Handle is provided in three<br>shaft lengths, 5, 6, and 7<br>inches | Tips are provided in 4 cup<br>sizes, 1,2, 3, and 4 mm.<br>Handle is provided in three<br>shaft lengths, 5, 6, and 7 inches | Same, except for the 1mm<br>tip will not be marketed as<br>a reusable device |
| Materials | Tips are 420 stainless steel<br>Handle is primarily 420<br>stainless steel with mating<br>component parts made from<br>Mating components are<br>manufactured with German<br>Stainless Steel 1.4301, 1.4310<br>(equivalent to 304 SS, 303 SS,<br>and 301 SS, respectively) | Tips are 420 stainless steel with<br>an ABS insert<br>Handle is primarily 420<br>stainless steel with mating<br>component parts made from<br>German Stainless Steel 1.4301,<br>1.4310 (equivalent to 304 SS,<br>303 SS, and 301 SS) | Same except for the<br>removal of the ABS insert<br>which had no mechanical<br>or performance function, no<br>longer present in the current<br>disposable tip<br>design |
| Mechanism of action | Cutting shaft and footplate | Cutting shaft and footplate | Same |
| Feature/Specification | Subject Device | Predicate Device | Comparison |
| | Symmetry Sharp Kerrison®<br>Rongeur | Symmetry Kerrison Disposable<br>Tip Rongeurs<br>K130541 | Of Subject Device to<br>Predicate Device |
| | closed on tissue through<br>mechanical handle<br>compression | closed on tissue through<br>mechanical handle compression | |
| Limits of use | Reusable handle, disposable<br>tips, and reusable tips | Reusable handle and disposable<br>tips | Similar, both devices have<br>reusable and disposable<br>components |
| How Supplied | Disposable tips are sterile<br>Reusable tips and reusable<br>handle are nonsterile and must<br>be cleaned and sterilized<br>before first use | Disposable tips are sterile<br>Reusable handle is nonsterile<br>and must be cleaned and<br>sterilized before first use | Same, only difference is<br>inclusion of the reusable<br>tips |
| SAL | 10-6 | 10-6 | Same |
| Sterilization Method | Gamma for the disposable tips<br>Steam for the reusable handle<br>and disposable tips | Gamma for the disposable tips<br>Steam for the reusable handle | Same except for adding<br>steam for reusable tips |
Technological Characteristics of the Device Compared to the Predicate Device:
{6}------------------------------------------------
Principles of Operation:
Each tip has cups, one within the stationary footplate and the other within the upper sliding shaft, performing the cutting of the bony tissue in the patient's spinal area when the instrument handles are compressed by the surgeon.
Design verification and validation testing including Performance sterilization for both the reusable and disposable components, Data cleaning for the reusable components, cutting performance, durability, shelf life (disposable tips), and corrosion resistance. The following table provides details relative to the device performance testing used to support substantial equivalence:
{7}------------------------------------------------
| Summary of Performance Testing Conducted to Establish Substantial Equivalence | | | | |
|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Performance Test | Verification and<br>Validation<br>Activities | Acceptance Criteria | Results | Device used for<br>Testing |
| Validation of the<br>recommended<br>cleaning process<br>for reusable<br>components | Cleaning<br>challenge for<br>reuse to confirm<br>the device can<br>be effectively<br>cleaned. | Cleaning validation<br>results must meet the<br>following criteria<br>acceptance criteria:<br>residual proteins<br>residual hemoglobin<br>and residual<br>carbohydrates | For substantial<br>equivalence the<br>device must be<br>able to be<br>adequately<br>cleaned for reuse<br>Pass | Symmetry Sharp<br>Kerrison® Rongeur<br>Handle (worst case) |
| Validation of<br>the<br>recommended<br>steam<br>sterilization<br>cycles for<br>reusable<br>components | Sterilization<br>validation | Must meet the<br>requirements as<br>defined in the<br>applicable ISO<br>Standard for the<br>overkill method to<br>achieve an SAL of 10-6 | For substantial<br>equivalence the<br>reusable<br>components<br>must be able to<br>be effectively<br>sterilized with<br>steam<br>sterilization<br>Pass | Symmetry Sharp<br>Kerrison® Rongeur<br>Handle (worst case) |
| Validation of<br>gamma<br>sterilization cycle<br>for single use<br>components | Sterilization<br>validation to<br>confirm 10-6<br>SAL for<br>gamma<br>sterilization | Must meet the<br>requirements as<br>defined in the<br>Applicable ISO<br>standard for VDMax to<br>achieve an SAL of<br>10-6 | For substantial<br>equivalence the<br>single use<br>component must<br>be able to be<br>sterilized using<br>gamma<br>sterilization<br>Pass | Symmetry Sharp<br>Kerrison®<br>Rongeur Tip |
| Confirmation of<br>multiple use for<br>the reusable<br>components | Verification<br>testing<br>confirming<br>performance for<br>multiple uses | The detachable tips<br>must maintain<br>acceptable cutting<br>performance for up to<br>100 cuts.<br>Reusable handle must<br>maintain performance<br>for 50 reuse cycles | For substantial<br>equivalence the<br>reusable must be<br>sufficiently<br>robust to support<br>reuse<br>Pass | Detachable Tips for<br>the Symmetry Sharp<br>Kerrison® Rongeur<br>Handle<br>And the Symmetry<br>Sharp Kerrison®<br>Rongeur Handle |
| Summary of Performance Testing Conducted to Establish Substantial Equivalence | | | | |
| Performance Test | Verification and<br>Validation<br>Activities | Acceptance Criteria | Results | Device used for<br>Testing |
| Resistance to<br>corrosion for the<br>reusable<br>components | Corrosion<br>testing | Must not corrode<br>when exposed to a<br>corrosion challenge | For Substantial<br>Equivalence<br>device must be<br>resistant to<br>corrosion<br>Pass | Symmetry Sharp<br>Kerrison® Rongeur<br>Handle (same<br>material as the<br>reusable tips) |
| Packaging<br>validation/shelf<br>life for single<br>use components | Shelf Life<br>studies per<br>ASTM<br>standards | Package and product<br>must demonstrate<br>package and product<br>stability over the<br>claimed shelf life<br>period | Package and<br>product<br>performance<br>evaluations<br>conducted on the<br>single use<br>component for<br>shelf life confirm<br>a two year shelf<br>life<br>Pass | Symmetry Sharp<br>Kerrison®<br>Rongeur Tip |
{8}------------------------------------------------
Based on the performance testing and comparison the subject device
is substantially equivalent to the predicate. Conclusion:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.