K161733 · Sorin Group Italia S.R.L. · DTN · Mar 8, 2017 · Cardiovascular
Device Facts
Record ID
K161733
Device Name
INSPIRE SVR 1200, INSPIRE 6F C, INSPIRE 8F C
Applicant
Sorin Group Italia S.R.L.
Product Code
DTN · Cardiovascular
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INSPIRE SVR 1200 with Ph.I.S.I.O. coating is intended for use in adult surgical procedures requiring cardiopulmonary bypass. It collects and handles venous blood from cardiotomy reservoir. The INSPIRE SVR 1200 is intended to be used for 6 hours or less.
Device Story
The INSPIRE SVR 1200 is a flexible PVC soft-shell venous reservoir bag used during cardiopulmonary bypass (CPB) procedures. It functions as a collection vessel for venous blood and suction blood from a cardiotomy reservoir. The device features a phosphorylcholine-based (Ph.I.S.I.O.) coating on blood-contacting surfaces to improve hemocompatibility. It connects to INSPIRE 6F C and 8F C oxygenator systems via a molded fitting. Operated by perfusionists in an OR setting, the device manages blood flow up to 8 L/min with a maximum volume of 1325 mL. By providing a stable reservoir for blood during extracorporeal circulation, it supports the maintenance of patient hemodynamics during surgery. The device is single-use and intended for short-term (≤6 hours) clinical application.
Clinical Evidence
Bench testing only. No clinical data provided. In vitro testing conducted per ISO 10993-1 and FDA guidance for extracorporeal blood circuit defoamers. Testing included biocompatibility (cytotoxicity, sensitization, toxicity, pyrogenicity, hemocompatibility, genotoxicity), blood trauma, coating leaching, surface integrity, filtration efficiency, air challenge, flow rate capacity, pressure drop, structural integrity, burst testing, and sterility. All tests met acceptance criteria.
Technological Characteristics
Soft-shell venous reservoir made of flexible PVC. Features Ph.I.S.I.O. (phosphorylcholine-based) surface coating. Includes 1/2" venous inlet and 3/8" blood outlet. Max flow rate 8 L/min; max volume 1325 mL. Includes mounting backplate. Sterilized via ethylene oxide. Single-use, non-pyrogenic fluid path.
Indications for Use
Indicated for adult and small adult patients undergoing surgical procedures requiring cardiopulmonary bypass, specifically for the collection and handling of venous and cardiotomy suction blood for durations of 6 hours or less.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
BMR 1900 Ph.I.S.I.O.Venous Reservoir Bag (K112771)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, arranged in a stacked, overlapping manner. The graphic is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 08, 2017
Sorin Group Italia S.r.l % Scott Light Regulatory Affairs Manager Sorin Group USA, Inc. 14401 W 65th Way Arvada, Colorado 80004
Re: K161733
Trade/Device Name: INSPIRE SVR 1200 Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: DTN Dated: February 3, 2017 Received: February 6, 2017
Dear Mr. Light:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
M.A. Hillemann
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K161733
Device Name
INSPIRE SVR 1200 (Softshell Venous Reservoir)
Tyne of I Ise (Select one or hoth as annlicable)
Indications for Use (Describe)
INSPIRE SVR 1200 with Ph.I.S.I.O. coating is intended for use in adult surgical procedures requiring cardiopulmonary bypass. It collects and handles venous blood from cardiotomy reservoir. The INSPIRE SVR 1200 is intended to be used for 6 hours or less.
| | <span> <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--|--------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="text-decoration: overline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
| SUBMITTER: | Sorin Group Italia<br>86, Via Statale 12 Nord<br>41037 Mirandola (MO) Italy |
|-----------------------|-------------------------------------------------------------------------------|
| CONTACT PERSON: | Luigi Vecchi<br>Phone: 39 0535 29811<br>Fax: 39 0535 25229 |
| DATE PREPARED: | March 8, 2017 |
| TRADE NAMES: | INSPIRE SVR 1200 soft shell venous reservoir. |
| COMMON NAMES: | Soft shell venous reservoir. |
| CLASSIFICATION NAMES: | Cardiopulmonary Bypass Blood Reservoir<br>(21 CFR 870.4400, Product Code DTN) |
| PREDICATE DEVICE: | BMR 1900 Ph.I.S.I.O.Venous Reservoir Bag (K112771) |
| REFERENCE DEVICE: | CBMVR800 Venous Reservoir Bag (K920774) |
## DEVICE DESCRIPTION:
The Sorin INSPIRE SVR 1200 (hereinafter referred to as INSPIRE SVR 1200) is a flexible PVC soft shell venous reservoir bag. This reservoir may be attached to the INSPIRE 6F C and INSPIRE 8F C oxygenator systems using a molded fitting connected to the gas exchange module. The blood contacting surfaces of this device are coated with a phosphorylcholine-based (Ph.I.S.I.O.) coating that provides a uniform surface. The device is provided with a 1/2" venous inlet connector and a 3/8" blood outlet connector. The venous inlet has an integral cardiotomy reservoir connector to allow receipt of blood from both venous cannulation and from a cardiotomy reservoir.
The INSPIRE SVR 1200 can be operated at flow rates up to 8 liters per minute (I/min) and has a maximum reservoir volume of 1325 mL.
#### INDICATION FOR USE:
INSPIRE SVR 1200 with Ph.I.S.I.O. coating is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects and handles venous blood and suction blood from cardiotomy reservoir. The INSPIRE SVR 1200 is intended to be used for 6 hours or less.
{4}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS:
The modified device has the same fundamental technological characteristics, principles of operation and control mechanisms as the predicate devices.
The INSPIRE SVR 1200 soft shell venous reservoir is similar to the predicate devices as it consists of a soft PVC material and is used to collect blood during cardiopulmonary bypass procedures lasting up to 6 hours. The intended use and principles of operation are consistent with the predicate devices. It differs from the predicate devices in volume and dimensions, length of purge lines, and it includes a mounting backplate.
The device is ethylene oxide sterilized, has a non-pyrogenic fluid path and is for single use only.
### NON CLINICAL TEST RESULTS:
Applicable tests were conducted in accordance with the requirements of ISO 10993-1. The current draft Guidance "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and its guidelines was also considered limited to external communicating device, circulating blood, limited contact duration.
In vitro testing was conducted on the INSPIRE SVR 1200, the subject of the present 510(k).
The tests performed on INSPIRE SVR 1200 demonstrate it is substantially equivalent to the predicate devices.
Testing was conducted in accordance with the relevant requirements of "Guidance for Extracorporeal Blood Circuit Defoamer 510(k) Submissions; Final Guidance for Industry and FDA, November 29, 2000".
The following list provides the performance tests conducted to demonstrate compliance of INSPIRE SVR 1200 to the product's performance specifications and predicate devices substantial equivalency.
The INSPIRE SVR 1200 successfully met all acceptance criteria for each of the following tests:
- Biocompatibility (cytotoxicity, sensitization, acute systemic toxicity, . pyrogenicity, hemocompatibility, and genotoxicity)
- Blood trauma
- Leaching of coating
- Surface modification integrity
- Surface modification coverage ●
- Filtration efficiency
- Air challenge performance ●
{5}------------------------------------------------
- Flow rate capacity .
- 2h continuous use at 0.5 LPM ●
- Maximum operating volume
- . Minimum operative volume at different flow rates
- Pressure drop
- . Structural integrity
- Blood pathway integrity ●
- Burst testing
- Sterility ●
- Package integrity .
## CONCLUSIONS:
The results of in vitro studies demonstrated that the INSPIRE SVR 1200 performs in a manner substantially equivalent to the BMR 1900 Ph.I.S.I.O. and CBMVR800 predicate devices with respect to the relevant functional parameters.
Test results of this study demonstrate the predicate devices substantial equivalence with respect to device function.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.