← Product Code [FLL](/productcode/FLL) · K161728

# Microlife Digital Infrared Forehead Thermometer, Model  FR1MN1-1 (K161728)

_Microfile Intellectual Property GmbH · FLL · Dec 5, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161728

## Device Facts

- **Applicant:** Microfile Intellectual Property GmbH
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Dec 5, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

## Device Story

Device is an electronic infrared forehead thermometer for home use. It utilizes a thermopile sensor and lens to detect infrared radiation from the forehead. An application-specific integrated circuit (ASIC) processes the signal using a firmware-based algorithm to calculate body temperature. Output is displayed on an LCD screen with a multi-color backlight (green, yellow, red) to indicate temperature status. The device is operated by the user via two keys (Start, O/I). It provides an auto-display of the last reading upon startup. The device aids users in monitoring body temperature and identifying potential fever, facilitating clinical decision-making regarding health status.

## Clinical Evidence

Clinical study conducted per ASTM E1965-98 and IEC 80601-2-56. Data evaluated clinical bias, clinical uncertainty, and clinical repeatability to validate the accuracy of the updated software algorithm.

## Technological Characteristics

Infrared thermopile sensor (OTP-668D2); LCD with multi-color backlight; powered by 2x AAA 1.5V batteries. Standards: IEC 60601-1, IEC 60601-1-2, ASTM E 1965-98, ISO 14971, ISO 10993 series, ISO 80601-2-56, IEC 60601-1-11. Software algorithm class: firmware-based clinical offset table.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- Microlife Digital Infrared Forehead Thermometer, Model FR1DM1 ([K033820](/device/K033820.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 5, 2016

Microfile Intellectual Property GmbH c/o Ms. Susan Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021

Re: K161728

Trade/Device Name: Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: November 2, 2016 Received: November 3, 2016

Dear Ms. Susan Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tina
Kiang -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161728

### Device Name

Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1

Indications for Use (Describe)

The Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

The assigned 510(k) number is: K161728

- 1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland

Date Summary Prepared: November 2, 2016

Contact: Mr. Gerhard Frick Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard.frick@microlife.ch

#### 2. Name of the Device:

Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1

Regulation Number: 21 CFR Part 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL

- 3. Information for the 510(k) Cleared Device (Predicate Device):
Microlife Digital Infrared Forehead Thermometer, Model FR1DM1, K033820, Microlife Intellectual Property GmbH

- 4. Device Description:
The Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the forehead. Microlife FR1MN1-1 especially enables the user to take measurements and be alerted for possible elevated temperature readings accordingly.

{4}------------------------------------------------

The Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1, consists of mainly seven parts:

- a) Thermopile Sensor
- b) Application-Specific Intergrated Circuit
- c) Erasable Programmable Read-Only Memory Integrated Circuit
- d) Lens
- e) LCD and Backlight
- f) 2 Keys (Start key, O/I key)
- g) 2 batteries AAA (LR03) 1.5V

The new Model FR1MN1-1 has the same intended use and temperature measurement fundamental algorithm as the 510(k) cleared devices.

Auto-Display memory The last reading is automatically displayed for 2 seconds when the unit is switched ON.

#### న. Indications for Use:

The Microlife Digital Infrared Forehead Thermometer, Model FRIMN1-1 is intended for the intermittent measurement and monitoring of human body temperature. The device is indicated for use by people of all ages in the home.

| 6.                                                    | Comparison to the 510(k) Cleared Devices (Predicate Devices):                                                                              |                                                                                                                                                                    |                              |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Item                                                  | Predicate device: Microlife<br>Digital Infared Forehead<br>Thermometer FR1DM1,<br>Microlife Intellectual Property<br>GmbH, K033820         | Modified device:<br>Microlife Digital Infrared<br>Forehead Thermometer FR1MN1-<br>1, Microlife Intellectual Property<br>GmbH                                       | Assessment of<br>Differences |
| Thermometer type                                      | Infrared thermometer                                                                                                                       | Infrared thermometer                                                                                                                                               | same                         |
|                                                       | Contact                                                                                                                                    | Contact                                                                                                                                                            | same                         |
| Intended use                                          | Intended for the intermittent<br>measurement and monitoring of<br>human body temperature in the<br>home, for use on people of all<br>ages. | Intended for the intermittent<br>measurement and monitoring of<br>human body temperature. The<br>device is indicated for use by<br>people of all ages in the home. | same                         |
| Device technology                                     | Infrared                                                                                                                                   | Infrared                                                                                                                                                           | same                         |
| Item                                                  | Predicate device: Microlife<br>Digital Infared Forehead<br>Thermometer FR1DM1,<br>Microlife Intellectual Property<br>GmbH, K033820         | Modified device:<br>Microlife Digital Infrared<br>Forehead Thermometer FR1MN1-<br>1, Microlife Intellectual Property<br>GmbH                                       | Assessment of<br>Differences |
| Measurement target                                    | Human                                                                                                                                      | Human                                                                                                                                                              | same                         |
| Measuring location                                    | Forehead                                                                                                                                   | Forehead                                                                                                                                                           | same                         |
| Appearance<br>(ID design)                             | Image: Forehead Thermometer                                                                                                                |                                                                                                                                                                    | different shape              |
| Physical dimension                                    | (141.0±0.5)mm (L)<br>*(26.0±0.5)mm (W)<br>*(20.0±0.5)mm (D)                                                                                | (157.0+0.5)mm (L)<br>*(53.0±0.5)mm (W)<br>*(36.0+0.5)mm (D)                                                                                                        | different                    |
| Power supply                                          | 1 battery CR2032 3.0V                                                                                                                      | 2 batteries AAA (LR03) 1.5V                                                                                                                                        | different                    |
| Measuring range<br>(Body Mode)                        | 34°C ~42.2°C<br>(93.2~108.0°F)                                                                                                             | 34°C ~42.2°C<br>(93.2~108.0°F)                                                                                                                                     | same                         |
| Measuring range<br>(Non-clinical mode<br>/laboratory) | 32~212°F (0~100 °C)                                                                                                                        | 80.6~109.4°F (27~43 °C)                                                                                                                                            | different                    |
| Accuracy<br>(Clinical Mode)                           | ±0.3 °C, 34.0 ~ 42.2 °C<br>(±0.5 °F, 93.2 ~ 108.0 °F)                                                                                      | 0.4 °F, 96.8 ~ 102.2 °F                                                                                                                                            | different                    |
| Display resolution                                    | 0.1°C or 0.1 °F                                                                                                                            | 0.1°C or 0.1°F                                                                                                                                                     | same                         |
| Operating<br>temperature<br>(Body Mode)               | 16.0°C-40°C (60.8°F~104.0°F)                                                                                                               | 16.0°C-40°C (60.8°F~104.0°F)                                                                                                                                       | same                         |
| Storage conditions                                    | -25°C~55°C (-13°F~131°F)                                                                                                                   | -25°C~55°C (-13°F~131°F)                                                                                                                                           | same                         |
| Display type                                          | LCD                                                                                                                                        | LCD                                                                                                                                                                | same                         |
| Memory                                                | 12 set, last measurement                                                                                                                   | last measurement                                                                                                                                                   | different                    |
| Item                                                  | Predicate device: Microlife<br>Digital Infared Forehead<br>Thermometer FR1DM1,<br>Microlife Intellectual Property<br>GmbH, K033820         | Modified device:<br>Microlife Digital Infrared<br>Forehead Thermometer FR1MN1 -<br>1, Microlife Intellectual Property<br>GmbH                                      | Assessment of<br>Differences |
| Power auto off                                        | Approx. 1 minute after the last<br>measurement has been taken                                                                              | Approx. 20 seconds, after the last<br>measurement has been taken                                                                                                   | different                    |
| Low/dead                                              | Low battery at 2.7V                                                                                                                        | Low battery at 2.6V                                                                                                                                                | different                    |
| battery indications                                   | Dead battery at 2.55V                                                                                                                      | Dead battery at 2.5V                                                                                                                                               | different                    |
| Backlight                                             | Single backlight                                                                                                                           | Green, yellow and red backlight<br>according to the measured<br>temperature                                                                                        | different                    |
| Reference site                                        | No                                                                                                                                         | No                                                                                                                                                                 | same                         |
| Beeper indication                                     | Yes                                                                                                                                        | Yes                                                                                                                                                                | same                         |
| Sensor type                                           | Same                                                                                                                                       | Same                                                                                                                                                               | same                         |
| Lens                                                  | Yes                                                                                                                                        | Yes                                                                                                                                                                | same                         |
| Labeling                                              | Microlife                                                                                                                                  | Microlife                                                                                                                                                          | same                         |
| Error                                                 | Display "Err" when the system<br>has a malfunction.                                                                                        | Display "Err" when the system has<br>a malfunction.                                                                                                                | same                         |
| IC                                                    | Same                                                                                                                                       | Same                                                                                                                                                               | same                         |
| Sensor                                                | OTP-668D2                                                                                                                                  | OTP-668D2                                                                                                                                                          | same                         |
| Software algorithm                                    | FW table "xyz" rev.1                                                                                                                       | FW table "xyz" rev. 1.1                                                                                                                                            | different                    |

{5}------------------------------------------------

{6}------------------------------------------------

The subject FR1MN1-1 device uses the same infrared technology, measures and monitor body temperature at the forehead site as the predicate Microlife FR1DM1.

The major differences are as below:

- Power supply, memory sets, low/dead battery indications, backlight, etc. These are mainly a) user- friendly features.
- b) Software Algorithm

The modified device uses slightly adjusted clinical offset tables for additional improved accuracy. Our algorithm was adjusted to improve accuracy. A clinical study was performed to validate this.

{7}------------------------------------------------

Based upon the aforementioned information, the two devices are substantially equivalent.

- 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating performance of the Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 in the intended environment of use is supported by testing that was conducted in accordance with Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers

The following testing was conducted to demonstrate performance as well as substantial equivalence to the predicate devices:

The following National and International Standards were utilized for testing the subject device:

| Item | Test name                                                                               | Pass/Fail Criteria                                 | Results                               |
|------|-----------------------------------------------------------------------------------------|----------------------------------------------------|---------------------------------------|
| 1.   | IEC 60601-1:2005+A1:2012                                                                | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |
| 2    | AAMI/IEC 60601-1:2005/(R)<br>2012 And<br>A1:2012,C1:2009/(R)2012 And<br>A2:2010/(R)2012 | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |
| 3    | IEC 60601-1-2:2007                                                                      | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |
| 4    | ASTM E 1965-98 (2009)                                                                   | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |
| 5    | ISO 14971:2007                                                                          | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |
| 6    | AAMI/ANSI/ISO 10993-1:2010                                                              | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |
| 7    | AAMI/ANSI/ISO 10993-5:2009                                                              | Pass/Fail Criteria is Outlined in this<br>Standard | Passed All<br>Testing<br>Requirements |

{8}------------------------------------------------

| 8  | AAMI/ANSI/ISO 10993-10:2010   | Pass/Fail Criteria is Outlined in this Standard | Passed All Testing Requirements |
|----|-------------------------------|-------------------------------------------------|---------------------------------|
| 9  | AAMI/ANSI/ISO 10993-12:2012   | Pass/Fail Criteria is Outlined in this Standard | Passed All Testing Requirements |
| 10 | AAMI/ANSI/ISO 80601-2-56:2009 | Pass/Fail Criteria is Outlined in this Standard | Passed All Testing Requirements |
| 11 | IEC 60601-1-11:2010           | Pass/Fail Criteria is Outlined in this Standard | Passed All Testing Requirements |

It was our conclusion that Microlife Digital Infrared Forehead Thermometer, Model FR1MN1-1 tested meets all relevant requirements of the aforementioned tests.

#### 8. Discussion of Clinical Tests Performed:

Controlled human clinical studies were conducted in accordance with ASTM E1965-98, IEC 80601-2-56 Test Report using the Microlife Digital Infrared Forehead Thermometer Model FR1MN1-1. Clinical data was presented evaluating clinical bias, clinical uncertainty and clinical repeatability per clinical validation for Microlife FR1MN1-1.

#### 9. Software Information:

Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

#### 10. Conclusions:

Conclusions drawn from the non-clinical and clinical tests demonstrate that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K161728](https://fda-staging.innolitics.com/device/K161728)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).