K161726 · ETHICON, Inc. · GAM · Jul 12, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161726
Device Name
E-Pack Procedure Kit
Applicant
ETHICON, Inc.
Product Code
GAM · General, Plastic Surgery
Decision Date
Jul 12, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4493
Device Class
Class 2
Indications for Use
The modified E-PACK™ procedure kits are provided with the same package inserts with the same indication statements for each component of the kit, identical to the labeling provided with the individually marketed device.
Device Story
E-PACK™ Procedure Kit is a convenience kit containing two or more separate, legally marketed Ethicon finished surgical devices. Components are organized in a plastic sleeve, placed in a rigid PETG thermoform blister tray with a Tyvek lid, and housed in a carton dispenser box. Used in clinical settings (e.g., OR) by healthcare professionals. Kit classification is determined by the highest-class component (Class II). The device provides clinical utility by grouping necessary surgical tools for specific procedures, streamlining preparation for the clinician. Benefits include improved procedural efficiency and convenience. Sterilized via ethylene oxide.
Clinical Evidence
Bench testing only. Evidence includes design verification activities: stability testing, biocompatibility (ISO 10993-7), labeling verification, packaging/transit testing (ISO 11607-1), and sterilization validation (Ethylene Oxide). No clinical data was required or presented.
Technological Characteristics
Convenience kit containing multiple finished surgical devices. Packaging materials: rigid PETG thermoform blister, Tyvek lid, plastic organizer sleeve. Sterilization: Ethylene Oxide (validated for two cycles). Standards: ISO 10993-7 (biocompatibility/residuals), ISO 11607-1 (packaging), ISO 14971 (risk management).
Indications for Use
Indicated for the same patient populations and clinical conditions as the individual components contained within the kit, as specified in the labeling of each constituent device.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 12, 2016
Ethicon, Inc. Ms. Enza Deserio Regulatory Affairs Specialist Route 22 West, P.O. Box 151 Somerville, New Jersey 08876
Re: K161726 Trade/Device Name: E-PACKTM Procedure Kit Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM, GAQ, FZP, GAN, GAW, FTL, LDF Dated: June 20, 2016 Received: June 22, 2016
Dear Ms. Deserio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161726
Device Name E-PACK™ Procedure Kit
Indications for Use (Describe)
The modified E-PACK™ procedure kits are provided with the same indication statements for each component of the kit, identical to the labeling provided with the individually marketed device.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
k) Pre-Market Notification thicon, Inc.
Image /page/3/Picture/1 description: The image shows the logo for Ethicon, Inc., a Johnson & Johnson company. The word "ETHICON" is written in large, red, sans-serif font, with ",INC." in a smaller font to the right. Below the company name is the text "a Johnson & Johnson company" in a cursive font. The logo is simple and clean, with a focus on the company name.
# 510(k) Summary
| Submitter: | Ethicon, Inc. a Johnson & Johnson company<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151 |
|------------|---------------------------------------------------------------------------------------------------------|
|------------|---------------------------------------------------------------------------------------------------------|
- Contact Person: Enza Deserio Regulatory Affairs Specialist Ph: (908) 218-2662 Fax: (908) 218-2595 Email: Edeserio@its.jnj.com
Date Prepared: June 20, 2016
- Device Trade Name: E-PACK™ Procedure Kit
II
- Device Common Name: Convenience Kit
Class:
- Classification Name: Absorbable poly (glycolide/l-lactide) surgical suture (21 CFR 878.4493) Suture, Nonabsorbable, Steel, Monofilament And Multifilament, Sterile (21 CFR 878.4495) Clip, Implantable (21 CFR 878.4300) Suture, Absorbable, Synthetic (21 CFR 878.4830) Suture, Nonabsorbable, Synthetic, Polypropylene (21 CFR 878.5010) Mesh, Surgical, Polymeric (21 CFR 878.3300) Electrode, Pacemaker, Temporary (21 CFR 870.3680)
- GAM, GAQ, FZP, GAN, GAW, FTL, LDF Product Code:
{4}------------------------------------------------
# Predicate Device Information:
| Device | Company | Product<br>Code | 510(k)<br>Number |
|---------|---------------|-----------------|------------------|
| E-PACK™ | Ethicon, Inc. | GAM | K970317 |
# Device Description:
Ethicon, Inc. E-PACK™s are considered convenience kits because two or more separate types of Ethicon finished devices are packaged together for the convenience of the user. Ethicon, Inc. certifies that all components within the E-PACK™ procedural kit are legally marketed devices manufactured by or for Ethicon, Inc. The classification of the kit is based on the highest classification of the devices that are provided in the kit. Ethicon, Inc. E-PACK™ Procedure Kits highest device classification is Class II.
The modified package will consist of individual finished devices, with or without their original primary package, that are placed into a plastic sleeve organizer that is subsequently placed into a blister tray with a Tyvek lid.
# Indications For Use:
The modified E-PACK™ procedure kits are provided with the same package inserts with the same indication statements for each component of the kit, identical to the labeling provided with the individually marketed device.
# Packaging:
The modified E-PACK™ packaging will consist of a rigid PETG Thermoform Blister with a Tyvek lid. The modified packaging will include a Plastic Organizer sleeve where the individual devices are placed. The Thermoform blister tray/Tyvek lid is placed in carton dispenser box. The Thermoform blister tray will come in three different sizes to accommodate the different sizes of E-PACK™ s (small, medium, and large). New products cleared through the 510(k) process will be available subsequently for inclusion in the E-PACK™. The 510(k) submissions for these products will state that they will be included in the appropriate E-PACK™.
# Summary of Technological Characteristics and Performance Testing:
Ethicon, Inc. followed its internal procedures for Design Control and performed the following design control activities for E-PACK™ Procedure Kits:
- i. Stability- Confirmed the stability of the E-PACK™ Procedure Kits and the individual products within the E-PACKTM Procedure Kits.
- ii. Biocompatibility Confirmed that E-PACKTM Procedure Kits meet the ISO 10993-7, Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals, requirements.
{5}------------------------------------------------
- iii. Labeling Confirmed the E-PACK™ Procedure Kits meet labeling requirements for convenience kits.
- iv. Packaging/Transit Testing Confirmed E-PACK™ Procedure Kit weight does not exceed the transit test worst case and the E-PACK™ components fit within the limits of the current packaging system. The proposed package configuration passed all transit testing requirements found in ISO 11607-1. Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems and packaging systems.
- v. Sterilization- Confirmed that E-PACKTM Procedure Kit products can be sterilized with Ethylene Oxide and can be sterilized two times.
The safety and effectiveness of the modified E-PACK™ procedure kits and the substantial equivalence to the predicate device have been demonstrated via data collected during design verification. The results of these tests provide reasonable assurance that the modified device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing. The modified package configuration used for the proposed device meets the requirements ISO 14971:2007 Medical devices --Application of risk management to medical devices and ISO 11607-1: 2006 Packaging for terminally sterilized medical devices -- Part 1: Requirements for materials, sterile barrier systems and packaging systems.
# Summarv of Substantial Equivalence:
The proposed device, Ethicon, Inc. E-PACK™, is substantially equivalent to the predicate, Ethicon, Inc. E-PACK™ in that they share:
- the same fundamental scientific technology,
- . the same intended use,
- . the same design.
- the same materials.
- . similar packaging materials,
- the same labeling components.
- . the same sterilization process (Ethylene Oxide),
- . the same sterility assurance level (SAL) is 10-6 and
- the same shelf life.
In summary, the proposed device, Ethicon. Inc. E-PACK™, is substantially equivalent to the predicate Ethicon, Inc. E-PACK™, as further demonstrated in Table 2, Section 12: Comparison of Proposed and Predicate Devices.
# Conclusion:
Based on the intended use, technological characteristics, safety and performance testing, the modified package configuration of the proposed E-PACK™ has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate E-PACK™ (K970317).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.