The bubble humidifier is intended to add moisture to breathing gases for administration to patients >1 month in homecare, hospital, extended care and hospice. The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases.
Device Story
Passive, non-heated, non-powered bubble humidifier; used in-line between breathing gas source (oxygen concentrator, cylinder, wall outlet) and patient circuit (face mask or cannula). Device consists of humidifier bottle, lid, and PVC diffuser tube. Gas enters bottle, passes through water via diffuser for humidification, and exits to patient. Includes 6 PSI (410 mbar) audible pressure relief safety valve for downstream occlusion. Used in homecare, hospital, extended care, and hospice; operated by clinicians or patients. Provides humidified gas to improve patient comfort and airway hydration. Material change from DEHP-plasticized PVC to DOTP-plasticized PVC for diffuser components.
Clinical Evidence
No clinical data. Bench testing only. Testing included bond strength verification for the new material, biocompatibility (ISO 10993-5, 10993-10, 10993-18), and a useful life study verifying performance up to 120 cleaning cycles.
Technological Characteristics
Passive, non-heated, non-powered humidifier. Materials: Polyethylene bottle, polystyrene lid, PVC (DOTP) diffuser tubing. Dimensions: Bottle 5.12" length, 2.75" base diameter. Connectivity: In-line between gas source and patient circuit. Safety: 6 PSI (410 mbar) brass pop-off valve. Non-sterile, single-patient use.
Indications for Use
Indicated for patients >1 month requiring humidification of high flow supplemental breathing gases in homecare, hospital, extended care, and hospice settings.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 21, 2017
Salter Labs Mara Caler Regulatory Affairs 2365 Camino Vida Roble Carlsbad, California 92011
Re: K161719
Trade/Device Name: Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Class II Product Code: BTT Dated: September 14, 2017 Received: September 18, 2017
Dear Mara Caler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161719
Device Name
Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve
#### Indications for Use (Describe)
The bubble humidifier is intended to add moisture to breathing gases for administration to patients >1 month in homecare, hospital, extended care and hospice.
The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="padding-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of substantial equivalence information is provided:
### 510 (k) Summary
#### A. Submitter:
Salter Labs 2365 Camino Vida Robles Carlsbad, CA 92011 Telephone: 760-795-7100 760-683-6797 Fax:
Contact Person: Mara Caler Regulatory Affairs
Date Prepared: 14 Sept 2017
#### Device Names: B.
| Trade Name | Common Name | Classification Name | Product Code |
|---------------------------------------------------------------------------------------|-------------------------------|-------------------------------|--------------|
| Salter Labs Bubble<br>Humidifier (6-15 LPM) with<br>6 PSI (410mbars) safety<br>valve. | Respiratory gas<br>humidifier | Respiratory gas<br>humidifier | BTT |
| Regulation Number: | 868.5450 |
|-----------------------|----------------------|
| Classification: | II |
| Classification Panel: | Anaesthesiology (73) |
#### C. Predicate Devices:
This submission demonstrates substantial equivalence to the predicate device: K113542, Salter Labs Bubble Humidifier
#### D. Device Descriptions
The Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410 mbars) safety valve is an empty, disposable, non-sterile, not made with natural rubber latex, device intended to
{4}------------------------------------------------
humidify breathing gas prior to delivery to a patient. The Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve is provided with a 6 pounds per square inch (PSI) safety valve and can operate within flow rates of 6 to 15 liters per minute (LPM). The device is used with various breathing gas sources (i.e., oxygen concentrators, gas cylinders and wall outlets) and provides connection for delivery of humidified breathing gas via face masks or cannulas, and use of optional oxygen tubing and water traps (face masks, 21 CFR 868.5580: nasal cannulas, 21 CFR 868.5340; oxygen tubing, 21 CFR 868.5860 and water traps, 21 CFR 868.5995 are 510(k) exempt).
This device is a passive device and is not a cascade humidifier, is not heated and is not prefilled.
The device is made of a humidifier bottle which is used to hold water during use, a lid which seals the humidifier bottle, an audible pressure relief mechanism to notify the user of a downstream occlusion and a diffuser located at the end of a PVC diffuser tube inside the humidifier bottle. The diffusor is designed to uniformly disperse the gas throughout the water. Both the bottle and lid are constructed to be easy to grip and reduce the chance of cross threading. The bottle is permanently marked with "minimum/maximum" water levels. The lid is marked with minimum source pressure, flow ranges and pressure value of the safety valve.
#### E. Indications for Use:
The bubble humidifier is intended to add moisture to breathing gases for administration to patients >1 month in homecare, hospital, extended care and hospice.
The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases.
## E.1: Usage
The bubble humidifier is indicated for use with oxygen concentrators or gas sources in homecare, hospital, extended care facilities and hospice environments. The bubble humidifier is a non-sterile device indicated for single-patient usage. The device is indicated for patients who require humidification of high flow supplemental breathing gases.
## E.2: Comparison of intended use to the predicate device and separate it from the Indications for Use Statement.
The predicate device is indicated for any patient population requiring humidification of high flow supplemental breathing gases. There are no excluded patient populations. For clarity, a more detailed description of the patient population is defined in the indications for use statement. This device is intended for the same intended population, including patients >1 month in homecare, hospital, extended care and hospice environments.
{5}------------------------------------------------
#### F. Substantial Equivalence:
| | | Comparison of Technological Characteristics with the Predicate Device |
|--|--|-----------------------------------------------------------------------|
| Specification | Predicate: Salter Labs<br>Bubble Humidifier (High<br>Flow) | Test device: Salter Labs<br>Bubble Humidifier (6-15<br>LPM) with 6 PSI<br>(410mbars) safety valve | Changes |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Device Use & General Characteristics | | | |
| Indications for<br>use | The bubble humidifier is a<br>device that is intended to<br>add moisture to breathing<br>gases for administration to a<br>patient. The bubble<br>humidifier is a non-sterile<br>device indicated for single-<br>patient usage. The device is<br>indicated for patients who<br>require humidification of<br>high flow supplemental<br>breathing gases. | The bubble humidifier is<br>intended to add moisture to<br>breathing gases for<br>administration to<br>patients >1 month in<br>homecare, hospital, extended<br>care and hospice.<br>The bubble humidifier is a<br>non-sterile device indicated<br>for single-patient usage. The<br>device is indicated for<br>patients who require<br>humidification of high flow<br>supplemental breathing<br>gases. | The patient<br>population<br>and the use<br>environment<br>has been<br>defined |
| Target Patient<br>Population | The device is indicated for<br>patients who require<br>humidification of high flow<br>supplemental breathing<br>gases. | The device is indicated for<br>patients >1 month who<br>require humidification of<br>high flow supplemental<br>breathing gases. | The patient<br>population<br>has been<br>defined |
| Intended use | To humidify breathing gases | To humidify breathing gases | Unchanged |
| Environment of<br>use | For use in homecare,<br>hospital, acute care facilities,<br>extended care facilities and<br>hospice environments. | For use in homecare,<br>hospital, acute care facilities,<br>extended care facilities and<br>hospice environments. | Unchanged |
| Use | Single patient use | Single patient use | Unchanged |
| Supplied as | Non-sterile | Non-sterile | Unchanged |
| Principal of<br>Operation/<br>Mechanisms of<br>Action | The bubble humidifier is<br>indicated for use with<br>oxygen concentrators or gas<br>sources to humidify (add<br>water to) the oxygen | The bubble humidifier is<br>indicated for use with<br>oxygen concentrators or gas<br>sources to humidify (add<br>water to) the oxygen | Unchanged |
| Technology | Air is channelled through a | Air is channelled through a | Unchanged |
| Specification | Predicate: Salter Labs<br>Bubble Humidifier (High<br>Flow) | Test device: Salter Labs<br>Bubble Humidifier (6-15<br>LPM) with 6 PSI<br>(410mbars) safety valve | Changes |
| Device Use & General Characteristics | | | |
| | water-containing bottle. The<br>air then becomes humidified<br>before exiting to the patient | water-containing bottle. The<br>air then becomes humidified<br>before exiting to the patient | |
| Flow rate | 6-15 liters per minute | 6-15 liters per minute | Unchanged |
| Pop-off safety<br>valve | At 6 PSI | At 6 PSI | Unchanged |
| Power Source | None - this is a passive<br>device | None - this is a passive<br>device | Unchanged |
| Prescription Use /<br>Over-The-<br>Counter Use | Prescription Devices | Prescription Devices | Unchanged |
| Sterile / Non-<br>Sterile | Non-Sterile | Non-Sterile | Unchanged |
| Use | Single patient use | Single patient use | Unchanged |
| Required<br>preparations prior<br>to use | None | None | Unchanged |
| Manufacturing<br>Process of Device<br>from Material | Molded (bottle, lid)<br>Extruded (diffuser tube) | Molded (bottle, lid)<br>Extruded (diffuser tube) | Unchanged |
| Patient-Contact<br>Components | Device does not touch the<br>patient. | Device does not touch the<br>patient. | Unchanged |
| Labeling | Instructions for Use are<br>unchanged except for the<br>removal of the DEHP<br>symbol. | Instructions for Use are<br>unchanged except for the<br>removal of the DEHP<br>symbol. | SE |
| Environmental<br>Limitations | Storage temperature:<br>-40°C (-40°F) to +60°C<br>(140 °F) | Storage temperature:<br>-40°C (-40°F) to +60°C<br>(140 °F) | Unchanged |
| Materials | | | |
| K113542 | PVC (DEHP) | PVC (DOTP) | SE |
| Biocompatibility | ISO 10993 compliant | ISO 10993 compliant | Unchanged |
{6}------------------------------------------------
{7}------------------------------------------------
### Similarities and Differences:
#### Similarities:
The similarities between the Salter Labs Bubble Humidifier and the modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve are:
Design: The exact same design is used; there are no changes to the Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve except the materials used in the line/diffusor.
- Both are passive, non-heated, non-powered humidifier bottles. There is no external ● energy source.
- Both are in-line between the air source and patient circuit (nasal cannula) ●
- Both are intended for single-patient use. ●
- Both are supplied non-sterile
- Both use identical pop-off safety valve (6 PSI) made from identical material (brass) ●
- Both have identical dimensions. ●
| | Parameter (inch) | Comparison to predicate |
|-------------------|------------------|-------------------------|
| Bottle | | |
| • Length | 5.12 | No change |
| • Diameter (base) | 2.75 | No change |
| • Diameter (top) | 2.84 | No change |
| Lid | | |
| • Height | 2.26 | No change |
| • Diameter | 3.08 | No change |
| Tubing | | |
| • Length | 5.50 | No change |
| • Diameter | 0.29 | No change |
| Diffusor | | |
| • Length | 0.51 | No change |
| • Diameter (Base) | 0.40 | No change |
| • Diameter (Top) | 0.33 | No change |
- Both have identical performance characteristics ●
- o Flow rate: 6-15 LPM
- o 6 PSA (410 mbars) for pop-off safety valve
### Materials
- . Both use identical materials: polyethylene bottle with a polystyrene lid (except as identified in the "differences" table below).
{8}------------------------------------------------
## Energy source
- Both use are non-heated, non-powered in-line humidifier bottles. There is no . external energy source.
### Differences
The differences between the Salter Labs Bubble Humidifier and the modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve is a material modification from PVC (DEHP) to PVC (DOTP). The materials used in the remainder of the device are unchanged.
| Features | Predicate<br>K113542 | Modified<br>device | Performance<br>Testing |
|----------------------------------------------------------------------------------------------------------|----------------------|--------------------|-------------------------------------|
| Material Formulation<br>of the diffusor tubing<br>and diffusor (all other<br>materials are<br>unchanged. | PVC, DEHP | PVC, DOTP | Biocompatibility<br>and Performance |
{9}------------------------------------------------
### Standards
Standards used for this submission:
| Standard | Test | Results |
|--------------|--------------------------------------------------------------------------------------------------------|------------|
| ISO 10993-5 | Biological Evaluation of Medical Devices<br>Part 5: Tests for in-vitro cytotoxicity | Acceptable |
| ISO 10993-10 | Biological evaluation of medical devices -<br>Part 10: Tests for irritation and skin<br>sensitization. | Acceptable |
| ISO 10993-18 | Biological evaluation of medical devices -<br>Part 18: Chemical characterization of<br>materials | Acceptable |
#### Performance Data G.
The modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve was tested to verify that the bond strength of the new material used for the different tube and diffusor met the pre-defined performance specifications.
The test results demonstrate that the modified Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve is substantially equivalent to the predicate devices. No clinical testing was required for this submission. A cleaning study was performed which verified the cleaning outlined in the instructions for use. A useful life study was performed which verified performance up to 120 cleaning cycles.
### H. Conclusions
The Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) safety valve data and test results (bond strength testing) demonstrate that the devices are substantially equivalent to the predicate devices. Based on the performance testing performed on the modified device safety valve and bond strength, Salter Labs Bubble Humidifier (6-15 LPM) with 6 PSI (410mbars) we have demonstrated that the modified device continue to perform substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.