K161708 · Shenzhen Pump Medical System Co., Ltd. · DXN · Nov 22, 2016 · Cardiovascular
Device Facts
Record ID
K161708
Device Name
Arm automatic blood pressure monitor
Applicant
Shenzhen Pump Medical System Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Nov 22, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Arm automatic blood pressure monitor, model: BF3202,BF3212, BF6002,BF6102, BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204.BF1214. BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; is intended for measuring systolic, diastolic blood pressure and pulse rate of an adult over the counter.
Device Story
Device is an automatic arm blood pressure monitor for home/over-the-counter use. It uses an air pump to inflate a cuff around the upper arm, compressing the artery. A pressure sensor collects cuff pressure, converting it to a digital signal for the CPU. Software calculates systolic/diastolic blood pressure and pulse rate using an oscillometric algorithm. Output is displayed on a TN-LCD screen. The device helps users monitor blood pressure at home, facilitating self-management of cardiovascular health. It operates on battery or AC adapter power.
Clinical Evidence
No clinical data. The manufacturer states that because the subject device is technically identical to the predicate regarding blood pressure measurement technology, clinical validation per ANSI/AAMI/ISO 81060-2:2009 is unnecessary.
Technological Characteristics
Oscillometric measurement principle. Power: 4 AA batteries or AC adapter (100-240V). Display: TN-LCD. Cuff size: 22cm-44cm. Software: Moderate level of concern. Testing performed: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-30.
Indications for Use
Indicated for measuring systolic, diastolic blood pressure and pulse rate of an adult individual using a non-invasive oscillometric technique with an inflatable cuff (22cm-44cm) wrapped around the single upper arm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Arm automatic blood pressure monitor, Model: BE6034, BE6134 (K151258)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol consisting of three stylized human profiles facing to the right, with a ribbon-like element flowing beneath them.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
Shenzhen Pump Medical System Co., Ltd Ms. Migo Yang Clinical Registration Specialist 2/f. M-7 Sinosteel Building. Maqueling Estate. Hi-tech Industrial Park Nanshan District, Shenzhen, 518057 CN
Re: K161708
> Trade/Device Name: Arm automatic blood pressure monitor, Model: BF3202, BF3212, BF6002, BF6102, BF1205, BF1215, BF1213, BF6051, BF6151, BF6043, BF6143, BF1204,BF1214, BF6023L, BF6123L, BF6048L, BF6148L, BF6053, BF6153, BF6055 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 23, 2016 Received: June 21, 2016
Dear Ms. Migo Yang,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
Page 2 - Ms. Migo Yang
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mada Yellen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161708
#### Device Name
Arm automatic blood pressure monitor, Model: BF3202, BF3212, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6148L, BF6053,BF6153, BF6055;
#### Indications for Use (Describe)
It is a device intended to measure the systolic, diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable sizes: 22cm~44cm (8.7in~17.3in)) is wrapped around the single upper arm.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: _K161708
#### Submitter's Identification: 1.
SHENZHEN PUMP MEDICAL SYSTEM CO., LTD. 2/F, M-7 Sinosteel Building, Maqueling Estate, Hi-Tech Industrial Park, Nanshan District, Shenzhen, PEOPLE'S REPUBLIC OF CHINA
Contact: Xie Qiongyu Tel:86-0755-26710795 Fax:86-0755-26498210 E-mail:xieqy@bpump.com.cn
## 2. Name of the Device:
Trade Name: Arm automatic blood pressure monitor Model: BF3202,BF3212, BF6002,BF6102, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153,BF6055; Regulation Description: Non-invasive blood pressure measurement system. Regulation Number: 21 CFR 870.1130 Regulation Class: II Product Code: DXN Review Panel: Cardiovascular
## 3. Information for the 510(k) Cleared Device(Predicate device):
Arm automatic blood pressure monitor, Model: BE6034, BE6134, K151258, SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.
## 4. Description of the device
Arm automatic blood pressure monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with
{4}------------------------------------------------
pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.
The arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102, BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204.BF1214. BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; has the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.
## 5. Intended Use:
Arm automatic blood pressure monitor, model: BF3202,BF3212, BF6002,BF6102, BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204.BF1214. BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; is intended for measuring systolic, diastolic blood pressure and pulse rate of an adult over the counter.
## 6. Indications for Use:
Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102, BF1205,BF1215, BF1213, BF6051,BF6151, BF6043,BF6143, BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; is intended to measure the systolic, diastolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff (available sizes: 22cm to 44cm (8.7in to 17.3in)) is wrapped around the single upper arm.
## 7. Comparison to the 510(k) Cleared Device (Predicate device):
The subject device, Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002.BF6102. BF1205.BF1215. BF1213. BF6051.BF6151. BF6043.BF6143. BF1204,BF1214, BF6023L,BF6123L, BF6048L,BF6148L, BF6053,BF6153, BF6055; and the predicate device: Arm automatic blood pressure monitor, Model: BE6034, BE6134 uses the same oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. Upper arm cuff is inflated automatically by air pump and the pressures are transferred via tubing to a sensor in the unit.
They differ by the appearance, the way of energy supply as well as PCB. The appearance of subject device is different from predicate device; The only way of energy supply of predicate device is battery while the way of energy supply of subject device can be both battery and adapter. The PCB of subject device has been updated. But those differences do not affect the accuracy or normal use of this device because they use the same fundamental scientific technology based on clinical declaration of identity and the declaration of clinical identity.
The subject device: Arm automatic blood pressure monitor, Model: BF3202,BF3212, BF6002,BF6102,BF1205,BF1215,BF1213,BF6051,BF6151,BF6043,BF6143, BF1204,BF1214,BF6023L,BF6123L,BF6048L,BF6148L,BF6053,BF6153,BF6055;
{5}------------------------------------------------
uses the same oscillometric method as the predicate device: Arm automatic blood pressure monitor, Model: BE6034, BE6134 to determinate the systolic and diastolic blood pressure and pulse rate. They have the same function and intended use. Based upon the aforementioned information, these devices are substantially equivalent. Please refer to table below for comparison.
{6}------------------------------------------------
| (Subject Devices:BE6034,BE6134,BF3202,BF3212,BF6002,BF6102,BF1213) | | | | | | | | | | |
|--------------------------------------------------------------------|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------------|--|
| | | Model | | | | | | Comparison | | |
| No | Items | BE6034, BE6134 | BF3202 | BF3212 | BF6002 | BF6102 | BF1213 | Result | Note | |
| 1 | Manufa<br>cturer | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM<br>CO.,LTD. | Identical | NA | |
| 2 | Intende<br>d Use | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic,<br>diastolic blood<br>pressure and<br>pulse rate of an<br>adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available<br>sizes: 22cm to<br>44cm (8.7in to<br>17.3in)) is<br>wrapped<br>around the<br>single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult<br>individual by<br>using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure and<br>pulse rate of an<br>adult individual by<br>using a non-<br>invasive<br>oscillometric<br>technique in which<br>an inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic,<br>diastolic blood<br>pressure and<br>pulse rate of an<br>adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to<br>17.3in)) is<br>wrapped around<br>the single upper<br>arm. | Identical | NA | |
| 3 | Indicati<br>ons for<br>Use | It is intended for<br>measuring<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult over the<br>counter. | It is intended<br>for measuring<br>systolic,<br>diastolic blood<br>pressure and<br>pulse rate of an<br>adult over the<br>counter. | It is intended for<br>measuring<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult over the<br>counter. | It is intended for<br>measuring<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult over the<br>counter. | It is intended for<br>measuring<br>systolic, diastolic<br>blood pressure and<br>pulse rate of an<br>adult over the<br>counter. | It is intended<br>for measuring<br>systolic,<br>diastolic blood<br>pressure and<br>pulse rate of an<br>adult over the<br>counter. | Identical | NA | |
| | | | | | Model | | | Comparison | | |
| No | Items | BE6034, BE6134 | BF3202 | BF3212 | BF6002 | BF6102 | BF1213 | Result | Note | |
| 4 | Patient<br>populati<br>on | Adult | Adult | Adult | Adult | Adult | Adult | Identical | NA | |
| 5 | Environ<br>ment of<br>use | Home | Home | Home | Home | Home | Home | Identical | NA | |
| 6 | Measuri<br>ng<br>Principl<br>e | Oscillometric<br>Method | Oscillometric<br>Method | Oscillometric<br>Method | Oscillometric<br>Method | Oscillometric<br>Method | Oscillometric<br>Method | Identical | NA | |
| 7 | Measur<br>ement<br>Range | Pressure:<br>0mmHg~280mm<br>Hg<br>(0kPa~37.3kPa)<br>Pulse: 40 bpm<br>~180 bpm | Pressure:<br>0mmHg~280mm<br>Hg<br>(0kPa~37.3kPa)<br>Pulse: 40 bpm<br>~180 bpm | Pressure:<br>0mmHg~280mm<br>Hg<br>(0kPa~37.3kPa)<br>Pulse: 40 bpm<br>~180 bpm | Pressure:<br>0mmHg~280mm<br>Hg<br>(0kPa~37.3kPa)<br>Pulse: 40 bpm<br>~180 bpm | Pressure:<br>0mmHg~280mmH<br>g (0kPa~37.3kPa)<br>Pulse: 40 bpm<br>~180 bpm | Pressure:<br>0mmHg~280m<br>mHg<br>(0kPa~37.3kPa)<br>Pulse: 40 bpm<br>~180 bpm | Identical | NA | |
| 8 | Accurac<br>y | Pressure:<br>±3mmHg(±0.4kPa<br>)<br>Pulse: ±5% | Pressure:<br>±3mmHg(±0.4<br>kPa)<br>Pulse: ±5% | Pressure:<br>±3mmHg(±0.4kPa<br>)<br>Pulse: ±5% | Pressure:<br>±3mmHg(±0.4kPa<br>)<br>Pulse: ±5% | Pressure:<br>±3mmHg(±0.4kPa<br>)<br>Pulse: ±5% | Pressure:<br>±3mmHg(±0.4k<br>Pa)<br>Pulse: ±5% | Identical | NA | |
| 9 | Display | TN-LCD Digital<br>Display | TN-LCD<br>Digital Display | TN-LCD Digital<br>Display | TN-LCD Digital<br>Display | TN-LCD Digital<br>Display | TN-LCD<br>Digital Display | Identical | NA | |
| 10 | Memor<br>y | 2 Memory sets, 60<br>readings each set. | 2 Memory sets,<br>60 readings<br>each set. | 2 Memory sets, 60<br>readings each set. | 2 Memory sets, 60<br>readings each set. | 2 Memory sets, 60<br>readings each set. | 2 Memory sets,<br>60 readings<br>each set. | Identical | NA | |
| 11 | Power<br>Source | 4 AA Alkaline<br>battery | 4 AA Alkaline<br>battery or<br>AC Adapter<br>(Input: AC<br>100-240 V,<br>50/60Hz,400m<br>A; Output :<br>D.C. 6.0V,<br>0.5A;) | 4 AA Alkaline<br>battery or<br>AC Adapter<br>(Input: AC 100-<br>240 V,<br>50/60Hz,400mA;<br>Output : D.C.<br>6.0V, 0.5A;) | 4 AA Alkaline<br>battery or<br>AC Adapter<br>(Input: AC 100-<br>240 V,<br>50/60Hz,400mA;<br>Output : D.C.<br>6.0V, 0.5A;) | 4 AA Alkaline<br>battery or<br>AC Adapter<br>(Input: AC 100-<br>240 V,<br>50/60Hz,400mA;<br>Output : D.C.<br>6.0V, 0.5A;) | 4 AA Alkaline<br>battery or<br>AC Adapter<br>(Input: AC 100-<br>240 V,<br>50/60Hz,400m<br>A; Output :<br>D.C. 6.0V,<br>0.5A;) | Different | NA | |
| | Items | Model | | | | | | | Comparison | |
| No | | BE6034, BE6134 | BF3202 | BF3212 | BF6002 | BF6102 | BF1213 | Result | Note | |
| 12 | Broadca<br>st<br>Functio<br>n | BE6034: YES;<br>BE6134: No | No | No | No | No | No | Different | NA | |
| 13 | Operati<br>ng<br>Environ<br>ment | Temperature:<br>+5℃~+40℃;<br>Humidity: 15%-<br>93%<br>Pressure: 70.0kPa<br>~106.0kPa<br>Altitude: ≤ 3 000<br>m | Temperature:<br>+5℃~+40℃<br>Humidity :<br>15%-93%<br>Pressure:<br>70.0kPa~<br>106.0kPa<br>Altitude: < 3<br>000 m | Temperature:<br>+5℃~+40℃;<br>Humidity: 15%-<br>93%<br>Pressure: 70.0kPa<br>~106.0kPa<br>Altitude: ≤ 3 000<br>m | Temperature:<br>+5℃~+40℃;<br>Humidity: 15%-<br>93%<br>Pressure: 70.0kPa<br>~106.0kPa<br>Altitude: ≤ 3 000<br>m | Temperature:<br>+5℃~+40℃;<br>Humidity: 15%-<br>93%<br>Pressure: 70.0kPa<br>~106.0kPa<br>Altitude: ≤ 3 000<br>m | Temperature:<br>+5℃~+40℃<br>Humidity:<br>15%-93%<br>Pressure:<br>70.0kPa~<br>106.0kPa<br>Altitude: ≤ 3<br>000 m | Identical | NA | |
| 14 | Storage<br>and<br>Transpo<br>rt<br>Environ<br>ment | Temperature: -<br>25℃~+70℃;<br>Humidity: 10%<br>~95%<br>Pressure:50.0kPa<br>~106.0kPa | Temperature:<br>-25°C ~ +70°C<br>Humidity :<br>10%~95%<br>Pressure:50.0k<br>Pa~106.0kPa | Temperature: -<br>25°C ~ +70°C ;<br>Humidity: 10%<br>~95%<br>Pressure:50.0kPa<br>~106.0kPa | Temperature: -<br>25°C ~ +70°C ;<br>Humidity: 10%<br>~95%<br>Pressure:50.0kPa<br>~106.0kPa | Temperature: -<br>25℃~+70℃;<br>Humidity: 10%~<br>95%<br>Pressure:50.0kPa<br>~106.0kPa | Temperature: -<br>25℃~+70℃;<br>Humidity:<br>10%~95%<br>Pressure:50.0kP<br>a~106.0kPa | Identical | NA | |
| 15 | Weight | 360g (Without<br>batteries) | 275g(without<br>batteries) | 275g(without<br>batteries) | 293g(without<br>batteries) | 293g(without<br>batteries) | 314g(without<br>batteries) | Different | NA | |
| 16 | Size | 140<br>mm×116mm×81<br>mm | 138mm×110m<br>m×68mm | 138mm×110mm×<br>68mm | 140mm×110mm×<br>70mm | 140mm×110mm×<br>70mm | 150mm×110m<br>m×77mm | Different | NA | |
#### Table 1. Comparison of Product Specification (Subject Deviceses:BF3202,BF3212,BF6002,BF6102,BF1213)
{7}------------------------------------------------
{8}------------------------------------------------
## Table 2. Comparison of Product Specification (Subject Deviceses:BF1205,BF1215,BF6051,BF6151,BF6055)
| No | Items | Model | | | | | | Comparison | |
|----|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------|
| | | BE6034, BE6134 | BF1205 | BF1215 | BF6051 | BF6151 | BF6055 | Result | Note |
| 1 | Manufacturer | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM | SHENZHEN<br>PUMP MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP MEDICAL<br>SYSTEM<br>CO.,LTD. | SHENZHEN<br>PUMP<br>MEDICAL<br>SYSTEM | Identical | NA |
| | | Model | | | | | | Comparison | |
| No | Items | BE6034, BE6134 | BF1205 | BF1215 | BF6051 | BF6151 | BF6055 | Result | Note |
| | | CO.,LTD. | CO.,LTD. | CO.,LTD. | | | CO.,LTD. | | |
| 2 | Intended Use | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic,<br>diastolic blood<br>pressure and<br>pulse rate of an<br>adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available<br>sizes: 22cm to<br>44cm (8.7in to<br>17.3in)) is<br>wrapped<br>around the<br>single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult<br>individual by<br>using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure<br>and pulse rate of<br>an adult individual<br>by using a non-<br>invasive<br>oscillometric<br>technique in<br>which an<br>inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8.7in to 17.3in))<br>is wrapped around<br>the single upper<br>arm. | It is a device<br>intended to<br>measure the<br>systolic, diastolic<br>blood pressure and<br>pulse rate of an<br>adult individual by<br>using a non-<br>invasive<br>oscillometric<br>technique in which<br>an inflatable cuff<br>(available sizes:<br>22cm to 44cm<br>(8…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.