SoundStar eco 8F 3D Diagnostic eco Ultrasound Catheters, SoundStar eco 8F G 3D Diagnostic eco Ultrasound Catheters
K161700 · Sterilmed, Inc. · OWQ · Nov 21, 2016 · Cardiovascular
Device Facts
Record ID
K161700
Device Name
SoundStar eco 8F 3D Diagnostic eco Ultrasound Catheters, SoundStar eco 8F G 3D Diagnostic eco Ultrasound Catheters
Applicant
Sterilmed, Inc.
Product Code
OWQ · Cardiovascular
Decision Date
Nov 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Reprocessed 3D Diagnostic Ultrasound eco Catheters are indicated for intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible Carto® 3 EP Navigation Systems (version 2.3 and higher) the reprocessed 3D Diagnostic Ultrasound eco Catheters provide location information.
Device Story
Reprocessed 3D Diagnostic Ultrasound eco Catheters are intra-cardiac echo (ICE) catheters; feature embedded ultrasound transducer and magnetic location sensor in catheter tip. Input: acoustic signals for real-time ultrasound imaging; magnetic field data for 3D navigation. Operation: connects to compatible ultrasound systems and Carto® 3 EP Navigation System; provides real-time integration of ultrasound images with electromagnetic maps. Used in clinical settings by physicians; assists in visualizing cardiac anatomy/physiology and other devices during procedures. Benefits: enables precise navigation and visualization within heart/great vessels. Reprocessed by Sterilmed, Inc. up to 3 times; includes cleaning, sterilization, and functional testing (joint strength, deflection, sensor calibration). Output: visual images and location data displayed on navigation system console; informs clinical decision-making during electrophysiology procedures.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via cleaning/sterilization validation, biocompatibility (ISO 10993-1), packaging/shelf-life validation, and mechanical/functional testing (joint bond strength, torsional strength, flexation/deflection fatigue, tip buckling, sensor calibration, EEPROM reset, acoustic verification, and thrombogenicity). Electrical safety verified per IEC 60601-1 (defibrillation, leakage current) and IEC 61161 (acoustic power).
Technological Characteristics
8F, 90 cm diagnostic ultrasound catheter. Features integrated ultrasound transducer and magnetic location sensor. Materials identical to OEM. Connectivity: compatible with Carto® 3 EP Navigation System and specific Siemens/GE ultrasound systems. Sterilization: Ethylene Oxide (ISO 11135). Standards: ISO 10993 (biocompatibility), ASTM D4169/F88/F2096 (packaging), ASTM 1980-99 (shelf life), IEC 60601-1 (electrical safety), IEC 61161 (acoustic power).
Indications for Use
Indicated for intra-luminal visualization of cardiac and great vessel anatomy and physiology, and visualization of other devices in the heart. Compatible with Carto® 3 EP Navigation Systems (v2.3+) for location information.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
November 21, 2016
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Sterilmed. Inc. Neelu Gibson Senior Director, Regulatory Affairs 5010 Cheshire Parkway, Suite 2 Plymouth, Minnesota 55446
Re: K161700
Trade/Device Name: Reprocessed 3D Diagnostic Ultrasound eco Catheters (See attached list of models) Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OWO Dated: October 7, 2016 Received: October 11, 2016
Dear Neelu Gibson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mucha Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# K161700
## Models intended for reprocessing
| Original<br>Manufacturer | OEM<br>Model<br>Number | Reprocessed<br>Product Code | Description |
|---------------------------|------------------------|-----------------------------|--------------------------------------------------------------------------------------------------------------------|
| Biosense Webster,<br>Inc. | 10439011 | R10439011 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with Siemens<br>ultrasound systems. |
| Biosense Webster,<br>Inc. | 10439236 | R10439236 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with GE ultrasound<br>systems. |
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161700
Device Name
Reprocessed 3D Diagnostic Ultrasound eco Catheters
Indications for Use (Describe)
The Reprocessed 3D Diagnostic Ultrasound eco Catheters are indicated for intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. When used with compatible Carto® 3 EP Navigation Systems (version 2.3 and higher) the reprocessed 3D Diagnostic Ultrasound eco Catheters provide location information.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Sterilmed. The logo consists of a stylized, partially-formed circle on the left, with the word "STERILMED" in all caps to the right of the circle. The circle is silver on the top and green on the bottom. The word "STERILMED" is in a simple, sans-serif font.
| Submitter and Manufacturer: | Neelu Gibson<br>Sterilmed, Inc.<br>5010 Cheshire Parkway, Suite 2<br>Plymouth, MN 55446 |
|---------------------------------|----------------------------------------------------------------------------------------------------------|
| Manufacturing Facility Address: | 11400 73rd Avenue North<br>Maple Grove, MN 55369 |
| Primary Contact: | Neelu Gibson<br>Sterilmed, Inc.<br>Tel: 908-705-3160<br>Fax: 763-488-3350<br>Email: ngibson9@its.jnj.com |
| Date of Submission: | 21 May, 2016 |
| Trade Name: | Reprocessed 3D Diagnostic Ultrasound<br>eco Catheters |
| Regulation Name: | Diagnostic Intravascular Catheter |
| Device Classification: | Class II |
| Product Code: | OWQ |
# 510(k) Summary - K161700
| Predicate<br>Device: | Biosense Webster, Inc.: SoundStar eco 8F G Ultrasound Catheter, and SoundStar<br>eco 8F Ultrasound Catheter (510(k) K140318) |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Secondary /<br>Reference<br>Predicates | Sterilmed Inc: Reprocessed 3D Ultrasound Diagnostic Catheter (510(k) K110076)<br>Sterilmed Inc: Reprocessed Ivus Imaging Catheter (510(k) K043453) |
| Device<br>Description: | Sterilmed Reprocessed 3D Diagnostic Ultrasound eco Catheters are specially<br>designed catheters that provide two-dimensional imaging using an ultrasound<br>transducer and three-dimensional location sensor. |
| | They are IntraCardiac Echo (ICE) Catheters with acoustic array and magnetic<br>location sensor equivalent to the currently cleared SoundStar eco 8F G Ultrasound<br>Catheter, and SoundStar eco 8F Ultrasound Catheters 510(k) K140318. The<br>magnetic location sensor (providing location information to the CARTO® 3 EP<br>Navigation System) and an ultrasound transducer (acquiring real time ultrasound<br>images) are embedded in the catheter tip. |
| | The Reprocessed 3D Diagnostic Ultrasound Catheters, when connected to the<br>corresponding Ultrasound Systems, will provide real-time integration of the<br>ultrasound images with CARTO® 3 electromagnetic acquired maps. |
| | Note: Only the catheter is the subject of this submission. Accessories and/or any<br>other related equipment are not included in the scope of this submission. The table<br>below lists the specific model numbers that are the subject of this submission. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for STERILMED. The logo features a stylized, partially circular graphic on the left, with the word "STERILMED" in a sans-serif font to the right of the graphic. The graphic is two-toned, with the upper portion in silver and the lower portion in green.
| Original<br>Manufacturer | OEM<br>Model<br>Number | Reprocessed<br>Product<br>Code | Description |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Biosense Webster,<br>Inc. | 10439011 | R10439011 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with Siemens<br>ultrasound systems. |
| Biosense Webster,<br>Inc. | 10439236 | R10439236 | SoundStar® eco Diagnostic<br>Ultrasound Catheter, 8F, 90 cm<br>length. For use with GE ultrasound<br>systems. |
| Indications for<br>Use: | The Reprocessed 3D Diagnostic Ultrasound eco Catheters are indicated for intra-<br>cardiac and intra-luminal visualization of cardiac and great vessel anatomy and<br>physiology as well as visualization of other devices in the heart. When used with<br>compatible Carto® 3 EP Navigation Systems (version 2.3 and higher) the<br>reprocessed 3D Diagnostic Ultrasound eco Catheters provide location information. | | |
| Technological<br>Characteristics: | The Reprocessed 3D Diagnostic Ultrasound eco Catheters are identical to the<br>predicate devices in design, materials of construction, and intended use. There are<br>no changes to the clinical applications, patient population, performance<br>specifications, or method of operation. | | |
| Functional and<br>Safety Testing: | Representative samples of reprocessed devices were tested to demonstrate<br>appropriate functional characteristics.<br>Process validation testing was performed to validate the cleaning and sterilization<br>procedures as well as device packaging. In addition, the manufacturing process<br>includes visual and validated functional testing of all products produced.<br>Where appropriate prior validation testing of previously cleared Reprocessed 3D<br>Diagnostic Ultrasound eco Catheter K110076 and Reprocessed AcuNav Imaging<br>Catheter K043453 was utilized for the same reprocessing procedures, equipment,<br>sterile barrier and sterilization processing. The Reprocessed 3D Diagnostic<br>Ultrasound eco Catheters are reprocessed no more than three (3) times. Each device<br>is marked and tracked through each reprocessing cycle. After the device has<br>reached the maximum number of reprocessing cycles (3), the device is rejected<br>from further reprocessing. Reprocessing is performed only by the manufacturer<br>Sterilmed, Inc. | | |
| Summary of<br>Non-Clinical<br>Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization<br>validation (ISO 11135, USP <71>), biocompatibility testing (ISO 10993-1),<br>ethylene oxide residual testing (ISO 10993-7), packaging validation (ASTM<br>D4169, ASTM F88, ASTM F2096), and shelf life validation (ASTM 1980-99). In<br>addition, functional performance (including both the ultrasound and 3D location<br>sensor components) and mechanical reliability were validated using bench and<br>laboratory testing. | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for STERILMED. The logo features a stylized, circular graphic on the left, with the word "STERILMED" in a simple, sans-serif font to the right of the graphic. The graphic is two-toned, with the top portion in silver and the bottom portion in green.
| | Catheters perform as originally intended. |
|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | Sterilmed concludes that the Reprocessed 3D Diagnostic Ultrasound eco<br>Catheters are safe, effective, and substantially equivalent to the predicate<br>devices, Biosense Webster, Inc. SoundStar eco 8F and eco 8F G Ultrasound<br>Catheters (K140318), as described in this premarket notification submission.<br>The Reprocessed 3D Diagnostic Ultrasound eco Catheters are substantially<br>equivalent to the listed predicate devices with respect to their indications for<br>use (intended use) and technical characteristics. The information and data<br>provided in this 510(k) submission identify no new safety or effectiveness<br>issues. |
| Verification<br>Testing | Description | Pass/<br>Fail |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Joint Bond<br>Strength<br>(SM5653/5840) | Test measured tensile strength of device joints to<br>confirm reprocessing has no effect on device. | Pass |
| Torsional Strength<br>(SM5652/5842) | Test measured the torsional strength of the flexible tip<br>and shaft sections to confirm reprocessing has no effect<br>on device. | Pass |
| Flexation Fatigue<br>(SM5604/5827) | Test measured flexation through maximum range of<br>deflection 100 times to confirm that reprocessing has no<br>effect on device. | Pass |
| Deflection Fatigue<br>(SM5605/5823) | Testing measured deflection in a simulated use<br>model through full articulation multiple times to<br>confirm that reprocessing has no effect on device. | Pass |
| Tip buckling<br>(SM4091/5923) | Testing measured force required to cause tip<br>buckling, to ensure reprocessing has effect on<br>device. | Pass |
| CCS check<br>(SM5523/5822) | Calibration check testing confirmed that<br>reprocessing has no effect on device's ability to<br>maintain sensor location. | Pass |
| Auto ID<br>(SM5800/5830) | Testing confirmed that reprocessed devices continue<br>to be identified as Soundstar eco 8F catheters | Pass |
| EEPROM reset<br>verification<br>(SM5802/5841) | Testing confirmed that EEPROM can be effectively<br>reset and continue to function as intended on<br>reprocessed devices. | Pass |
| AMCS Acoustic<br>Verification<br>(SM5522/5828) | Testing confirmed that Acoustic Manual Calibration<br>System functions for reprocessed devices. | Pass |
| Thrombogenicity<br>testing<br>SM6020/6828 | Testing confirmed that reprocessed devices do not<br>cause any new risk of thrombus formation. | Pass |
# Summary of Performance Testing
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for STERILMED. The logo features a stylized, circular graphic on the left, with the word "STERILMED" in a combination of bold and light gray text to the right. The circular graphic appears to be a stylized recycling symbol, with the top half in silver and the bottom half in green.
#### Biocompatibility testing .
Biocompatibility data demonstrates that there is no risk presented by Reprocessed SoundStar eco 8F and eco 8F G Ultrasound Catheters.
- Electromagnetic compatibility (EMC) testing was performed by Sterilmed per IEC 60601-1:2005 AI:2012 3nd Edition Medical electrical equipment - Part 1 General requirements for basic safety and essential performance were the defibrillation, leakage current and functional testing.
### Defibrillation
Test report SM 5839 documents the defibrillation protection compliance to the IEC 60601-1:2005 Clause 8.5.5.1 for defibrillation protection testing. IEC 60601-1:2005 Clause 8.5.5.1 for defibrillation protection testing defines a peak value of the differential signal (Yl -Y2) to be less than 1 volt (passing the test). Voltage reading over 1 volt is considered a fail.
### Current Leakage and Functional Test
To mitigate the concerns of possible damage to the device due to reprocessing the devices are:
- 100% in-process inspected for electrical insulation via Hi-pot (high potential) . testing, in which each device is subjected to 1500VDC, and electrical leakage is monitored. Reference test report SM5831
- 100% in-process inspected for functionality. ●
### IEC 60601-1-2:2007
No physical changes or parts replacement is conducted by Sterilmed during the reprocessing. The devices are unchanged and are identical to the OEM devices as evidenced in functional tests SM5828 - AMCS Check - SoundStar eco 8F within Error! Reference source not found..
### IEC 61161:2013
Total emitted acoustic power of ultrasonic transducers was verified and documented in SM5826.
### Summary
The Reprocessed 8F Catheters and their predicate devices, the OEM 8F Catheters, have the same intended use and indications for use and fundamental scientific technology. Based on performance testing for the Reprocessed 8F Catheters against the OEM devices discussed above, Sterilmed concludes that the Reprocessed 8F catheters are safe, effective, and substantially equivalent to the predicate OEM 8F catheters.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.