The QuantumCam is a gamma camera system designed to acquire data for whole body static, gated or dynamic and multi-slice images. The system is intended for use as diagnostic imaging device. When used with appropriate radio pharmaceuticals, images are produced representing the internal distribution of radioactivity in head or body. The system allows you to acquire data for high resolution three dimensional, static, gated or dynamic images of biochemical and metabolic processes using Tc-99m, TI-201, I-123, I-131, In-111, Ga-67, Co-57.
Device Story
QuantumCam is a dual-detector gamma camera system for nuclear medicine imaging; utilizes scintillation crystals and photomultiplier tubes based on Anger camera principles. System includes mechanical gantry for detector positioning, patient table, and PC-based acquisition station. Inputs are gamma rays emitted from patient-administered radiopharmaceuticals; system transforms these into functional images of internal organ/structure radioactivity distribution. Operated by medical personnel in clinical settings; images transferred to external workstation for interpretation. Supports diagnostic decision-making by visualizing biochemical/metabolic processes. Expanded energy range (40-400 keV) allows use of high-energy isotopes like I-131 via High Energy General Purpose (HEGP) collimator.
Clinical Evidence
Bench testing only. Performance verified via NEMA testing on the QuantumCam system using the HEGP collimator. Metrics reported: sensitivity (183 cpm/uCi), spatial resolution at 10cm (< 16.5 mm), and septal penetration (39.8%).
Technological Characteristics
Dual-detector Anger-type scintillation camera; NaI crystals; photomultiplier tubes; HEGP collimator (HEX shape, 3.0mm hole, 43mm length). Gantry/table constructed of steel and anodized aluminum. Energy range: 40-400 keV. Power: 200-240VAC. Connectivity: Networked via PC-based acquisition station. Software: Windows 7 64-bit, Microsoft SQL database.
Indications for Use
Indicated for patients requiring whole body static, gated, dynamic, or multi-slice nuclear medicine imaging to visualize biochemical and metabolic processes using radiopharmaceuticals (Tc-99m, TI-201, I-123, I-131, In-111, Ga-67, Co-57).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 20, 2016
DDD-Diagnostic A/S % Mr. Niels Sørensen RA & Product Engineer Dr. Neergaards Vej 5E Horsholm, 2970 DENMARK
Re: K161674
Trade/Device Name: QuantumCam 9SYS2070-B02 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: August 9, 2016 Received: June 16, 2016
Dear Mr. Sørensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
K161674
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
## 510(k) Number (if known)
Device Name QuantumCam, 9SYS2070-B02
## Indications for Use (Describe)
The QuantumCam is a gamma camera system designed to acquire data for whole body static, gated or dynamic and multislice images. The system is intended for use as diagnostic imaging device. When used with appropriate radio pharmaceuticals, images are produced representing the internal distribution of radioactivity in head or body. The system allows you to acquire data for high resolution three dimensional, static, gated or dynamic images of biochemical and metabolic processes using Tc-99m, TI-201, I-123, I-131, In-111, Ga-67, Co-57.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DDD Diagnostic. The logo consists of three capital "D"s, with the first two "D"s in black and the third "D" in purple. Below the "D"s is the word "Diagnostic" in a smaller font and in red.
510(k) Submission QuantumCam
1PMN3099-A01
### 510(k) Summary or 510(k) Statement 5.
#### 510(k) Summary 5.1
Ref. to 21 CFR 807.92
| 1 | Submitted by: | DDD-Diagnostic A/S<br>Dr. Neergaards Vej 5E<br>2970 Horsholm, Denmark<br>Tel: +45 45 768888<br>Fax: + 45 45 164659 |
|---|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact person: | Niels Sørensen<br>Tel: + 45 45 768888<br>Fax: + 45 45 164659<br>E-mail: niels.sorensen@ddd-diagnostic.dk |
| | Preparation date: | 02 May 2016 |
| 2 | Device Trade Name: | QuantumCam (Commercial name) 9SYS2070-B02<br>(BodyMD is the development name) |
| | Common Name: | Gamma Camera System |
| | Classification name: | Emission computed tomography system |
| 3 | Predicate Device: | QuantumCam 9SYS2070-A01,<br>DDD-Diagnostic A/S |
| | 510(k) Number: | K140206 |
| 4 | Device description: | The QuantumCam is a general purpose dual detector gamma camera<br>system comprised of a mechanical gantry allowing the Detectors to<br>be positioned and moved in close proximity to the patient for<br>scanning. The Gantry provides for positioning of the detectors in<br>location(s) suitable for patient to be brought in position for<br>tomographic, whole body scanning as well as planar scanning in<br>sitting or standing position |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DDD Diagnostic. The logo consists of three large letters, "DDD", with the first two letters in black and the third letter in purple. Below the letters, the word "Diagnostic" is written in purple.
1PMN3099-A01
---
| Functional description: | The Detectors are standard Nal (Sodium Iodide) / Photomultiplier based scintillation detectors designed following the Anger Camera principles first described by Hal Oscar Anger. This detector technology has been used effectively in Nuclear Medicine for decades. The Detectors are equipped with standard Collimators following the principles known to the industry and this submission is specific about the HEGP collimator. The image data from the detectors are collected by hardware in the detector and the data acquisition is controlled by software running on a standard personal computer with a suitable operating system. The data may subsequently be transferred to a Nuclear Medicine Workstation for processing and interpretation. This Workstation is not part of the QuantumCam system. |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5 Intended use: | The intended use of the QuantumCam gamma camera system is to perform general nuclear medicine imaging procedures. This is intended to be accomplished by imaging the distribution of a radiopharmaceutical within the human body. Gamma rays emitted from the radiopharmaceutical are detected by the gamma camera system and formed into images characterizing and showing the state of organs or structures in the form of functional images. When QuantumCam is connected to a nuclear medicine workstation, these images can be used in concert with other clinical data to assist in making clinical diagnoses by authorized medical personnel.<br>Indications for use. The QuantumCam is a gamma camera system designed to acquire data for whole body static, gated or dynamic and multi-slice images. The system is intended for use as diagnostic imaging device. When used with appropriate radio pharmaceuticals, images are produced representing the internal distribution of radioactivity in head or body. The system allows you to acquire data for high resolution three dimensional, static, gated or dynamic images of biochemical and metabolic processes using Tc-99m, TI-201, I-123, I-131, In-111, Ga-67, Co-57. |
| 6 a Summary of technological characteristics: | The submitted device has the same technological and functional characteristics as the predicate device. However the energy range is |
expanded for the submitted device:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DDD Diagnostic. The logo consists of three capital "D" letters, with the first two "D" letters in black and the third "D" letter in purple. Below the letters, the word "Diagnostic" is written in purple.
1PMN3099-A01
| | Submitted device:<br>QuantumCam, 9SYS2070-B02 | Predicate device:<br>QuantumCam, 9SYS2070-A01 |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design: | The QuantumCam is an open<br>gantry design supported on an<br>open base frame. The complete<br>system is comprised of: a gantry<br>base, detector tower, detector<br>gear box and two imaging<br>detectors, collimator cart, PC,<br>and electronics including a main<br>PSU. | The QuantumCam is an open<br>gantry design supported on an<br>open base frame. The complete<br>system is comprised of: a gantry<br>base, detector tower, detector<br>gear box and two imaging<br>detectors, collimator cart, PC, and<br>electronics including a main PSU. |
| Material: | The QuantumCam employs<br>detector of scintillation crystals<br>in a traditional Anger design. The<br>gantry and patient table employs<br>steel and anodized aluminum,<br>steel plate and aluminum covers. | The QuantumCam employs<br>detector of scintillation crystals in<br>a traditional Anger design. The<br>gantry and patient table employs<br>steel and anodized aluminum,<br>steel plate and aluminum covers. |
| Energy source: | Mains supply.<br>200VAC - 240VAC | Mains supply.<br>200VAC - 240VAC |
| Detector: | The QuantumCam has two<br>separate detectors. Each<br>detector casting is supported by<br>means of a U-shaped detector<br>arm in a balanced design<br>enabling manual tilt of the<br>detector. The complete detector<br>is comprised of: a detector<br>casting, a Nal detector crystal,<br>PM tubes, collimators and<br>electronics for data processing. | The QuantumCam has two<br>separate detectors. Each detector<br>casting is supported by means of a<br>U-shaped detector arm in a<br>balanced design enabling manual<br>tilt of the detector. The complete<br>detector is comprised of: a<br>detector casting, a Nal detector<br>crystal, PM tubes, collimators and<br>electronics for data processing. |
| Energy range: | 40 - 400 keV | 40 - 300 keV |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DDD Diagnostic. The logo consists of three large letters "D", with the first two "D"s in black and the third "D" in purple. Below the letters, the word "Diagnostic" is written in purple.
1PMN3099-A01
| Software: | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The QuantumCam acquisition station is based on a PC architecture with Windows 7 64bit operating system and Microsoft SQL for database handling. A dedicated S/W package for Graphic User Interface (GUI) - S/W package for acquisition setup and gantry/detector motion control. The package also includes means for data transmission via an external network. | The QuantumCam acquisition station is based on a PC architecture with Windows 7 64bit operating system and Microsoft SQL for database handling. A dedicated S/W package for Graphic User Interface (GUI) - S/W package for acquisition setup and gantry/detector motion control. The package also includes means for data transmission via an external network. |
6 b Description of how the non clinical test results have been collected:
Non clinical data has been obtained according to DDD quality system procedures. The HEGP Collimator is a late addition to the range of collimators for QuantumCam and verification is done according to standards as listed below, conducted and reported by qualified experts:
| Specifications: | Collimator hole:<br>Shape = HEX<br>Size = 3.0 mm<br>Septa = 1,75 mm<br>Length = 43 mm | Measured and compared to Unicorn drawings. |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensitivity:<br>183 cpm/uCi | | NEMA test on a QuantumCam system. |
| Spatial Resolution @<br>10cm, HEGP collimator:<br>< 16.5 mm | | NEMA test on a QuantumCam system. |
| Septal Penetration:<br>39,8 % | | NEMA test on a QuantumCam system. The<br>official specification is based on septa thickness<br>and is caluculated based on the absorption<br>characteristics of lead (4,8%). |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DDD Diagnostic. The logo consists of three capital "D" letters. The first two "D" letters are black, and the third "D" letter is purple. Below the three "D" letters, the word "Diagnostic" is written in purple.
510(k) Submission
QuantumCam
| Detector shielding: | The QuantumCam detector was mounted with a |
|---------------------|------------------------------------------------|
| Li = 6,4% | High Energy General Purpose collimator (HEGP). |
| LFi = 1,4% | A I-131 source (364 keV) in source holder. |
| LSi = 20% | NEMA test on a QuantumCam system. |
Rationale for substantial Energy Range is expanded for the submitted device to be able to use equivalence: high energy isotopes such as I-131. For this purpose a HEGP collimator is added to the range of existing collimators for QuantumCam. They all have similar characteristics and the choice of which collimator to use is the users, based on guidelines from professional societies. Adding the HEGP collimator still fulfill the intended use of the QuantumCam system.
The two devices are therefore considered substantial equivalent.
Conclusion: The submitted device has been evaluated for effectiveness, electrical and mechanical safety, and has been found in compliance with applicable medical device standards, as referred to in Appendix 2. Based on this DDD considers the submitted device, QuantumCam (expanded energy range) 9SYS2070-B02 to be substantially equivalent in terms of safety and effectiveness to the predicate device, QuantumCam 9SYS2070-A01.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.