K161670 · Bison Medical Co, Ltd. · GEX · Nov 2, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161670
Device Name
Lucid Q-PTP / HWA 55
Applicant
Bison Medical Co, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LUCID O-PTP Nd : YAG Laser System is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatologic and general surgical procedures for coagulation and hemostasis. 532nm Wavelength (nominal delivered energy of 585 nm with optional dye hand pieces): Tattoo removal light ink (red, tan, purple, orange, sky blue, green), Removal of Epidermal Pigmented Lesions, Minor Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Post Inflammatory Hyper-Pigmentation, Treatment of Becker's Nevi, Freckles and Nevi Spilus 1064nm Wavelength: Tattoo removal: dark ink ( black, blue, brown) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle
Device Story
Q-switched Nd:YAG laser system; produces 1064nm and 532nm wavelengths; optional dye handpieces convert 532nm to 585nm and 650nm. Device consists of laser tube, resonator, articulated arm, multi-spot handpiece, cooling system, and LCD control panel. Operated by physician in clinical setting; laser emission activated via foot switch. Physician controls energy pulse and spot size to optimize treatment. Laser energy delivered to soft tissue for surgical/dermatologic procedures; enables coagulation, hemostasis, and tissue ablation. Benefits include precise removal of pigmented lesions, tattoos, and skin resurfacing.
Clinical Evidence
No clinical or non-clinical performance test data provided. Substantial equivalence supported by adherence to safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
Technological Characteristics
Q-switched Nd:YAG laser; 1064nm and 532nm wavelengths; optional dye handpieces for 585nm/650nm. Features articulated arm delivery, Xe-gas lamp, and cooling system. Non-contact mode. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1 standards.
Indications for Use
Indicated for soft tissue incision, excision, ablation, vaporization, coagulation, and hemostasis in general dermatology and surgery. Specific applications include tattoo removal (various colors), removal of epidermal pigmented lesions, vascular lesions, telangiectasias, lentigines, cafe-au-lait, seborrheic keratoses, post-inflammatory hyper-pigmentation, Becker's nevi, freckles, nevi spilus, nevus of ota, hair removal/lightening, melasma, and skin resurfacing for acne scars and wrinkles.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Bison Medical Co., Ltd % Young Chi President Bio-med Usa Inc 27 New England Dr Ramsey, New Jersey 07446
Re: K161670
Trade/Device Name: Lucid Q-ptp / Hwa 55 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 27, 2016 Received: October 4, 2016
Dear Young Chi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K161670
Device Name LUCID Q-PTP/ HWA 55
Indications for Use (Describe)
The LUCID O-PTP Nd : YAG Laser System is indicated for the incision, excision, ablation, vaporization of soft tissues for general dermatologic and general surgical procedures for coagulation and hemostasis.
532nm Wavelength (nominal delivered energy of 585 nm with optional dye hand pieces): Tattoo removal light ink (red, tan, purple, orange, sky blue, green), Removal of Epidermal Pigmented Lesions, Minor Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses, Post Inflammatory Hyper-Pigmentation, Treatment of Becker's Nevi, Freckles and Nevi Spilus
1064nm Wavelength:
Tattoo removal: dark ink ( black, blue, brown) Removal of Nevus of Ota Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma, Skin resurfacing procedures for the treatment of acne scars, wrinkle
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (K) Summary
As required by CFR 807.92(c)
## 1. Manufacturer.
Prepared Sept 26, 2016
Reg Nr: 3011555967
BISON Medical Co., Ltd. Ace High-end tower 6, #1801/1802 234 Beotkkot-ro. Gasan dong GeumCheon -gu, Seoul 153-798, Rep of Korea t: 822865 7121, f: 82 2 865 7131
# 2. Submitter and Contact person
Bio-Med USA Inc. Young Chi, President. 27 New England Drive, Ramsey, NJ 07446. U.S.A. t: 1-973 278 5222 f: 1 201 934 6030 e mail: biomedusa(@msn.com
# 3. Name of Device
| Trade name | LUCID Q-PTP |
|----------------------|------------------------------------|
| Classification name | Powered, Laser surgical instrument |
| Common name | Nd:YAG Q-switched Surgical Laser |
| Regulation | 878.4810 Class II |
| Classification Panel | General and Plastic Surgery. |
| Product Code | GEX |
| Submission type | Traditional |
# 4. Legally marketed Predicate Device
| K113588 | Spectra | Nd:YAG | Lutronic Corp |
|---------|---------|--------|---------------|
|---------|---------|--------|---------------|
LUCID Q-PTP Q-Switched Nd:YAG laser system produce same two wave length (1064nm, 532nm), and same characteristics such as Design, Construction, Energy rate, Pulse Duration, applied optional Dye hand piece, Cooling system and intended use as already cleared predicate device K113588 by Lutronics.
# 5. Device Description
The LUCID O-PTP O-Switched Nd: Y AG laser system produces a two pulsed beam, 1064 nm Infrared and 532nm long pulse laser, and optional 2 dye Handpieces are available that convert the 532nm wave length to 585nm and 650nm, using different Handpiece able to control various treatment fluence.
this device is non-contacted mode and consists of main function,
laser tube ; placed in the mixed crystals of copper pipe to the heater and produces a laser beam,
Resonator : amplifies the beam, through the Xe-gas contained lamp : Xe-gas contains high pressure lamp to increase specific laser beam lamp
{4}------------------------------------------------
This converted light energy creates the ND:YAG crystal and exhaust from the crystal is amplified into a specific wave length. Laser energy produced is delivered to the Tissue by means of an articulated arm and a specially designed multi spot Hand Piece.
The Physician can optimize the effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam, and is able to activate laser emission using Foot Switch. Laser parameters and other system features are controlled from a display panel located on the front of the power supply unit.
## This system also consist of
Optic main Bench assembly, Articulated Arm Hand pieces, LCD control panel, Cooling system, Foot Pedal Switch
#### 6. Performance test
Clinical and Non-Clinical performance test data was not provided in this submission. But, manufactured in accordance with both mandatory and voluntary standard
IEC60601-1 part 1 : General requirement for basic safety and essential performance. IEC60601-1-2: 2007 E M C test IEC60601-2-22 Part 2, Particular requirements for safety of diagnostic and Therapeutic laser IEC60825-1 :2nd ED, Equipment classification and requirement.
Lucid Q-PTP, demonstrates no significant different compare to the predicate device
#### 7. Indication for use
The LUCID Q-PTP Nd : YAG Laser System is indicated for the incision, ablation, vaporization of soft tissues for general dermatology, dermatologic and general surgical procedures for coagulation and hemostasis.
#### 532mm Wavelength
(nominal delivered energy of 585 nm and 650 nm with optional dye handpieces) Tattoo removal light ink (red, tan, purple, orange, sky blue, green), Removal of Epidermal Pigmented Lesions, Minor Vascular Lesions, Talangiectasias Treatment of Lentigines, Cafe-Au-Lait, Sebrorrheic Keratoses,
Post Inflammatory Hyper-Pigmentation,
Treatment of Becker's Nevi, Freckles and Nevi Spilus
1064nm Wavelength:
Tattoo removal: dark ink (black, blue, green )
Removal of Nevus of Ota
Removal or lightening of unwanted hair with or without adjuvant preparation Treatment of Common Nevi, Melasma,
Skin resurfacing procedures for the treatment of acne scars, wrinkle
{5}------------------------------------------------
# 8. Biocompatibility.
This device are non-contacted mode. Hand piece tips is made by same material as predicate device.
### 9.Conclusion.
LUCID Q-PTP, Nd: YAG laser system, in this submission, is substantially equivalent to several already cleared predicate device in respect to the Intended use, Main function, Technology, Principal operation and performance. And every Safety test report show it as safe and effective as predicate device and it does not raise any additional issues for safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.