← Product Code [NHA](/productcode/NHA) · K161655

# On1 Concept (K161655)

_Nobel Biocare AB · NHA · Feb 28, 2017 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161655

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Feb 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The On1™ devices are intended for use in the field of dentistry. They are intended to be used in the upper or lower jaw for supporting tooth replacements to restore chewing function and esthetics. The On1™ esthetic abutments in combination with the On1™ Base on Nobel Biocare Conical Connection endosseous implants are indicated for single-unit cement retained restorations.

## Device Story

On1 Concept is a 2-piece dental implant abutment system for prosthetic rehabilitation. System includes On1 Base, healing cap, temporary abutment, and esthetic abutments (titanium alloy or zirconium oxide). On1 Base attaches to Nobel Biocare internal conical connection implants and is intended to remain in place after placement. Clinical screw threads accommodate healing caps and prosthetic screws. Used by dentists in clinical settings to support single-unit cement-retained restorations. Provides stable platform for tooth replacement, restoring chewing function and esthetics. No software or algorithmic components.

## Clinical Evidence

Bench testing only. No clinical data. Fatigue testing performed per ISO 14801. Biocompatibility, sterilization (Gamma radiation per ISO 11137; moist heat per ISO 17665-1), and shelf-life data leveraged from predicate devices.

## Technological Characteristics

2-piece dental abutment system. Materials: Titanium vanadium alloy (ASTM F1472, ASTM F136) and Y-TZP zirconium oxide (ISO 6872, ISO 13356). Compatible with Nobel Biocare internal conical connection implants (NP, RP, WP). Sterilization: Gamma radiation or pre-vacuum/gravity moist heat. No software or electronic components.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- NobelProcera Angulated Screw Channel Abutment Conical Connection ([K132746](/device/K132746.md))

## Reference Devices

- NobelActive Wide Platform (WP) ([K133731](/device/K133731.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three intertwined figures. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002

February 28, 2017

Nobel Biocare AB % Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887

Re: K161655

Trade/Device Name: On1 Concept Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 1, 2017 Received: February 2, 2017

Dear Charlemagne Chua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely. Mary S. Runner - 100 A For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DE PARTENT OF FEATH AND HUNN SERVICES<br>Food and Drug Administration                                                                                                                                                                                                                                                                                                                      | Form Approved: OMB No. 0910-0170<br>piration Date: January 31, 2017                                                      |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| ndications for Use                                                                                                                                                                                                                                                                                                                                                                         | es PRA Statement below.                                                                                                  |
| 510(k) Number (if known                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                          |
| 166665                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                          |
| Onl Concept<br>Device Name                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                          |
| Indications for Use (Describe                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                          |
| it it belies is a premiument pirestly compensed to an endossous in and it it                                                                                                                                                                                                                                                                                                               |                                                                                                                          |
| intended for use in prosthetic rehabilitation.                                                                                                                                                                                                                                                                                                                                             |                                                                                                                          |
|                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                          |
| of Use (Select one or both, as applicable,                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                          |
| A BEAR FERRATE A BO A BEACH A NO JUMITHO - CONTINO TON OG BASSE<br>D                                                                                                                                                                                                                                                                                                                       | I Over The -Counter Use (S) est The ------------------------------------------------------------------------------------ |
|                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                          |
| ອນາງອານຸຣິນ (HAQC) ປາໄຣອH ໄຂວ່ເງິດໄທ່ເປັນ ປີ ປະຕິທິນ ປີ ຄ.ສ. 2017 ອົງມາອາການປີ<br>FOR FDA USE ONLY                                                                                                                                                                                                                                                                                         |                                                                                                                          |
| . WOJER SSAREDA JIAMA FRATE AAR BHT OT MRO ACT3J9MCO AUOY GM32 TOM OD<br>The suites and a personis of the superious of the supplies noires and                                                                                                                                                                                                                                             |                                                                                                                          |
| of this information collection, including suggestions for reducing this purder, to:<br>ədi birinci ildə məqsədə vəq sinə qalınmış si noğramoni in notibellər sinfi növündə sinfi növündə əsirlər.<br>time<br>r wie on and in and in information ship purposes mortune or any the respective<br>a more bebeen also and nistriam bus redtep , secures etsb priteixe , suoilounteni weiver of |                                                                                                                          |
| Paperwork Reduction Act (PRA) Staff<br>Department of Health and Human Services<br>Office of Chief Information Officer<br>Pagstaff@fdd.hhs.goov<br>Food and Drug Administration                                                                                                                                                                                                             |                                                                                                                          |
| to noticello a st production is nost of the "instructs to the "losured to resource"<br>information unners it disks a currently vila OMB number.                                                                                                                                                                                                                                            |                                                                                                                          |
| FORM FDA 3888 (11744<br>Page 1 of 1                                                                                                                                                                                                                                                                                                                                                        | PSC Publishing Services (301) 4436740<br>ਜ ਤੋਂ                                                                           |

านออนดว นับ โร อยิน Jo Σε อย่าง

{3}------------------------------------------------

#### 510(k) Summary

#### I. SUBMITTER

Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden

Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887

Contact Person: Charlemagne Chua. Senior Requlatory Affairs Manager E-mail: Charlemagne.chua@nobelbiocare.com Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348

Date Prepared: February 27, 2017

II. DEVICE

Name of Device: On1 Concept Common or Usual Name: Endosseous Dental Implant Abutment Classification Name: Endosseous Dental Implant (21 CFR 872.3630) Requlatory Class: II Product Code: NHA

III.PREDICATE DEVICE

Primary Predicate NobelProcera Angulated Screw Channel Abutment Conical Connection (K132746)

Reference Predicate NobelActive Wide Platform (WP) (K133731)

IV. DEVICE DESCRIPTION

The On1 Concept is a 2-piece dental implant abutment system. It is intended to be used with the Nobel Biocare dental implants that have the internal conical connection. The On1 Concept features the On1 Base that once placed on the dental implant is not intended to be removed. The top portion of the On1 Clinical Screw incorporates threads to accommodate the On1 Healing Cap and On1

{4}------------------------------------------------

Prosthetic Screw allowing the On1 Clinical Screw to remain in position after placement and securing of the On1 Base.

The On1 Concept system includes a healing cap, temporary abutment and esthetic abutments which all fit onto the On1 Base. The esthetic abutments are available in both titanium alloy and zirconium oxide.

# V. INDICATIONS FOR USE

The On1™ device is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation.

#### VI. Comparison of Technological Characteristics

| Technological<br>characteristics |                                   | Subject Device                                                                                                                                                                                                                                                             | Primary Predicate                                                                                                                                                | Reference Predicate                                                            |
|----------------------------------|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
|                                  |                                   | On1 Concept                                                                                                                                                                                                                                                                | NobelProcera Angulated Screw<br>Channel Abutment Conical<br>(K132746)                                                                                            | NobelActive Wide<br>Platform (WP)<br>(K133731)                                 |
| Design Features                  | Compatible<br>Implant<br>Platform | Nobel Biocare Internal Conical<br>Connection<br>- Narrow Platform (NP)<br>- Regular Platform (RP)<br>- Wide Platform (WP)                                                                                                                                                  | Nobel Biocare Internal Conical<br>Connection<br>- Narrow Platform (NP)<br>- Regular Platform (RP)                                                                | Nobel Biocare Internal<br>Connection / Internal<br>Hex<br>- Wide Platform (WP) |
|                                  | Device<br>Material                | On1 Base, Temporary<br>Abutment, On1 Healing Cap,<br>Esthetic Abutment Titanium,<br>Clinical Screws and Prosthetic<br>Screws - Titanium vanadium<br>alloy (ASTM F1472, ASTM<br>F136)<br>On1 Esthetic Abutment<br>Zirconia – Y-TZP zirconium<br>oxide (ISO 6872, ISO 13356) | Abutment base and clinical<br>screw - Titanium vanadium<br>alloy (ASTM F1472, ASTM<br>F136)<br>Abutment post – Y-TZP<br>zirconium oxide (ISO 6872, ISO<br>13356) | Titanium vanadium<br>alloy (ASTM F1472,<br>ASTM F136)                          |
|                                  | Abutment<br>Design                | 2 piece (base placed either at<br>time of implant placement or<br>with final abutment)<br>Abutment shape fixed                                                                                                                                                             | 2 piece (base placed with final<br>abutment)<br>Abutment shape personalized                                                                                      | Patient Specific Design                                                        |
|                                  | Abutment<br>height                | Combined base and post<br>height<br>Temporary Abut – 8.3, 9.0 mm<br>Esthetic Abut Ti – 8.2, 9.0 mm<br>Esthetic Abut Zi - 8.2, 9.0 mm                                                                                                                                       | Not applicable - abutment<br>shape personalized                                                                                                                  | Not applicable -<br>abutment shape<br>personalized                             |
|                                  | Abutment<br>width                 | At base<br>4.8, 5.3, 6.5 mm                                                                                                                                                                                                                                                | Not applicable - abutment<br>shape personalized                                                                                                                  | Not applicable -<br>abutment shape<br>personalized                             |
|                                  | Abutment<br>Angulation            | No abutment angulation                                                                                                                                                                                                                                                     | Variable screw channel and<br>abutment angulation based on<br>personalized design.                                                                               | Not applicable -<br>abutment shape<br>personalized                             |

{5}------------------------------------------------

| Technological<br>characteristics | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                   | Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Reference Predicate                                                                                       |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Intended use                     | The On1™ devices are<br>intended for use in the field of<br>dentistry. They are intended to<br>be used in the upper or lower<br>jaw for supporting tooth<br>replacements to restore<br>chewing function and<br>esthetics.<br>The On1™ esthetic abutments<br>in combination with the On1™<br>Base on Nobel Biocare<br>Conical Connection<br>endosseous implants are<br>indicated for single-unit<br>cement retained restorations. | Nobel Biocare's NobelProcera<br>ASC Abutment Zirconia is a<br>customized endosseous dental<br>abutment. The abutment<br>attaches directly to the dental<br>implants and provides a<br>platform for restoration. The<br>NobelProcera ASC Abutment<br>Zirconia is designed and made<br>individually to fit the individual<br>requirements for each patient.<br>The NobelProcera ASC<br>Abutment Zirconia is made out<br>of Zirconia and is delivered with<br>a titanium adapter. | Not Applicable –<br>reference predicate to<br>support abutment for<br>use with Wide platform<br>implants. |
| Indication for<br>Use            | The On1™ device is a<br>premanufactured prosthetic<br>component directly connected<br>to an endosseous implant and<br>it is intended for use in<br>prosthetic rehabilitation.                                                                                                                                                                                                                                                    | The NobelProcera Angulated<br>Screw Channel Abutment<br>Conical Connection are<br>premanufactured prosthetic<br>components directly connected<br>to endosseous dental implants<br>and are intended for use as an<br>aid in prosthetic rehabilitation.                                                                                                                                                                                                                          | Not Applicable –<br>reference predicate to<br>support abutment for<br>use with Wide platform<br>implants. |

Analysis of Differences Between Subject Device and Predicate

The On1 Concept and predicate NobelProcera Angulated Screw Channel Conical Connection (K132746) are both two piece endosseous dental implant abutments intended to fit the Nobel Biocare internal conical connection dental implants.

The On1 Concept base is designed to be placed either when the dental implant is placed or with the final abutment. Once placed on the dental implant the On1 Base is not intended to be removed. The predicate NobelProcera Angulated Screw Channel Conical Connection (K132746) base is intended to adapt the abutment post to the internal conical connection implant and is placed along with the post when the definitive abutment is placed.

The On1 Concept components are all fixed design components. The abutments do not have any angulation built in. The predicate NobelProcera Angulated Screw Channel Conical Connection (K132746) abutments are individually shaped and can be angulated.

{6}------------------------------------------------

### Summarv:

The design differences between the subject and predicate devices were evaluated through comparative performance testing. The documentation submitted in the premarket notification demonstrates that the On1 Concept is substantially equivalent to the predicate devices.

# VII. PERFORMANCE DATA

Summary of Non-Clinical Testing:

Since the subject device does not represent a new worst case, data from the predicate device was leveraged in the following aspects of the 510(k).

- Sterile Device Information
	- o The sterilization method for the subject device provided sterile is the same as the predicate. The sterilization method is Gamma radiation and has been validated in accordance with ANSI/AAMI/ISO 11137. Therefore, no additional testing was required.
	- o The sterilization methods for the subject devices provided non-sterile and to be end user sterilized are pre-vacuum and gravity moist heat sterilization and have been validated in accordance with ANSI/AAMI/ISO 17665-1.
- Device Packaging
	- o The packaging for the subject device is the same as the predicate. This is a thermoform tray with peel top lid. Therefore, no additional testing was required.
- Shelf Life -
	- The packaging for the subject device is the same as the predicate and o is labeled with a 3-year expiration date. Real time aging was used to determine the expiration dating. Therefore, no additional testing was required.
- Biocompatibility -
	- The subject device is manufactured from the same material using the o same manufacturing method as the predicate, has the same intended use, and the same patient contact type and duration. Therefore, no additional testing was required.

The fatigue limit of the On1 Concept was determined using the methods described in ISO 14801. The results of the testing were used to address questions related to substantial equivalence based on difference in design between the subject and predicate devices.

# VIII. CONCLUSIONS

The On1 Concept was evaluated for substantial equivalence using standard and/or comparative testing. In cases where the On1 Concept could be shown to not represent a worst-case with respect to the predicates, data from these predicate

{7}------------------------------------------------

devices was leveraged to support the subject device. Based on technological characteristics and non-clinical test data included in this submission, the On1 Concept has been shown to be substantially equivalent to the NobelProcera Angulated Screw Channel Conical Connection Abutment (K132746).

---

**Source:** [https://fda-staging.innolitics.com/device/K161655](https://fda-staging.innolitics.com/device/K161655)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com