# CAVUX™ Cervical Cage (K161642)

_Providence Medical Technology, Inc. · ODP · Oct 7, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161642

## Device Facts

- **Applicant:** Providence Medical Technology, Inc.
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Oct 7, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

CAVUX™ Cervical Cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-C7) with accompanying radicular symptoms at one disc level. DDD is defined as discogenc pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six weeks of non-operative treatment with the device. Devices are intended to be used with autogenous bone graft and supplemental fixation, such as an anterior plating system.

## Device Story

CAVUX™ Cervical Cage is an intervertebral body fusion device; implanted into cervical spine (C3-C7) to treat degenerative disc disease. Device is hollow, titanium alloy (6AI4V-ELI) cage; filled with autogenous bone graft; features fenestrations (windows) for bone visualization and growth. Implanted via delivery instrument with physical stop to prevent over-insertion. Used by surgeons in clinical settings; requires supplemental fixation (e.g., anterior plating system). Provides structural support to interbody space; promotes fusion; benefits patients by stabilizing spine and reducing discogenic pain.

## Clinical Evidence

Bench testing only. Performed static/dynamic axial compression, static/dynamic torsion, subsidence, and expulsion testing per ASTM F2077-14, ASTM F2267-04, and draft ASTM F-04.25.02.02. Pyrogenicity testing conducted per FDA guidance. Results met all acceptance criteria and exceeded predicate device performance.

## Technological Characteristics

Material: Titanium-6AL-4V ELI alloy (ASTM F136). Design: Hollow cage with fenestrations; superior/inferior surfaces feature teeth, grit blasting, and acid etching. Delivery: Includes delivery instrument with physical stop. Sterilization: Sterile, single-use. Connectivity: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- PMT Cervical Cage ([K122801](/device/K122801.md))

## Reference Devices

- Titan Spine, LLC: Endoskeleton® TC ([K100889](/device/K100889.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 7, 2016

Providence Medical Technology, Inc. Mr. Edward Liou Chief Operating Officer 1331 North California Boulevard, Suite 320 Walnut Creek, California 94596

Re: K161642

Trade/Device Name: CAVUX™ Cervical Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: September 7, 2016 Received: September 9, 2016

Dear Mr. Liou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161642

Device Name CAVUX™ Cervical Cage

#### Indications for Use (Describe)

CAVUX™ Cervical Cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-C7) with accompanying radicular symptoms at one disc level. DDD is defined as discogenc pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six weeks of non-operative treatment with the device. Devices are intended to be used with autogenous bone graft and supplemental fixation, such as an anterior plating system.

| Type of Use (Select one or both, as applicable)                                                                |                                                                                                               |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span>                     ☑ Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

## Providence Medical Technology, Inc.'s CAVUX™ Cervical Cages

- Date Prepared: September 7, 2016
- Company: Providence Medical Technology, Inc. 1331 N. California Blvd., Suite 320 Walnut Creek, CA 94596
- Contact Person: Edward Liou ed@providencemt.com Phone: 415.923.9376 Facsimile: 415.923.9377
- CAVUX™ Cervical Cage Trade Name:
- Common Name: Cervical Cage

Classification Name: Intervertebral Fusion Device With Bone Graft, Cervical

Regulation Number: 21 CFR 888.3080

- ODP Product Code: Class: Class II
Predicate Devices:

Primary Predicate: PMT Cervical Cage (K122801, Cleared 5/24/13) (This device has not been subject to recall.)

Reference Device: Titan Spine, LLC: Endoskeleton® TC (K100889, Cleared 7/29/10)

#### Device Description:

The CAVUX™ Cervical Cage is a cervical intervertebral body fusion device. The system is comprised of a variety of implant sizes to accommodate various patient anatomies and pathology. All implantable components are manufactured from medical grade titanium alloy (6AI4V —ELI Ti). The center of the implant is hollow and is to be filled with autogenous bone material. The design incorporates "windows" through the implant to permit visualization of the graft material and, over time, formation of new bone.

{4}------------------------------------------------

#### Indications for Use:

CAVUX™ Cervical Cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C3-C7) with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received at least six weeks of non-operative treatment prior to treatment with the device. Devices are intended to be used with autogenous bone graft and supplemental fixation, such as an anterior plating system.

#### Technological Characteristics:

The CAVUX™ Cervical Cage consists of the following:

- CAVUX™ Cervical Cage Implant .
- . Delivery Instrument packaged with the Implant

The CAVUX™ Cervical Cage implants are manufactured from Titanium-6AL-4V ELI alloy, which conforms to ASTM F136 and are available in a variety of sizes and lordotic angles to accommodate patient anatomy. Superior and inferior surfaces of the implant feature teeth that provide bony contact with the endplates while a box shape in the center of the implant with fenestrations (windows) is intended to house autogenous bone. The superior and inferior surfaces of the implant are grit blasted and acid etched to improve fixation to adjacent bone.

The CAVUX™ Cervical Cage is held within the delivery instrument which facilitates insertion of the implant into the interbody space. The delivery instrument features a physical stop to prevent over-insertion.

Devices are supplied sterile and single use only.

#### Performance Testing

The following testing was performed using the subject device and results were compared to those of the predicate device:

#### Static and Dynamic Axial Compression Testing

- ASTM F2077-14 Standard Test methods for Intervertebral Body Fusion Devices .
#### Static and Dynamic Torsion Testing

- ASTM F2077-14 Standard Test methods for Intervertebral Body Fusion Devices .
#### Subsidence Testing

- . ASTM F2267-04 Standard Test Method for Measuring Load Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression

{5}------------------------------------------------

#### Expulsion Testing

- ASTM Draft Standard F-04.25.02.02 Static Push-out Test Method for Intervertebral . Body Fusion Devices
### Pyrogenicity Testing

- . Pyrogenicity Testing was conducted and in compliance with FDA Guidance Documents listed below:
	- "Submission and Review of Sterility Information in Premarket Notification o (510(k)) Submissions for Devices Labeled as Sterile: Guidance for Industry and Food and Drug Administration Staff"
	- "Pyrogen and Endotoxins Testing: Questions and Answers" O

In all instances, the CAVUX™ Cervical Cage functioned as intended and met all pre-determined acceptance criteria. Bench testing results of the CAVUX™ Cervical Cage exceeded those of the predicate PMT Cervical Cage. Therefore, the performance characteristics of the CAVUX ™ Cervical Cage are substantially equivalent to those of the PMT Cervical Cage.

#### Conclusions

The CAVUX™ Cervical Cage has the same indication for use, technological characteristics, and principles of operation as the predicate device. The differences between the subject and predicate devices are that the subject device includes a surface treatment and is available in different sizes than the predicate device. These minor differences between the CAVUX™ Cervical Cage and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the CAVUX™ Cervical Cage is as safe and effective as the PMT Cervical Cage. Thus, the CAVUX™ Cervical Cage is substantially equivalent to the PMT Cervical Cage.

---

**Source:** [https://fda-staging.innolitics.com/device/K161642](https://fda-staging.innolitics.com/device/K161642)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).