K161636 · Lipogems International Spa · MUU · Nov 4, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161636
Device Name
Lipogems System
Applicant
Lipogems International Spa
Product Code
MUU · General, Plastic Surgery
Decision Date
Nov 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Lipogems System is a sterile medical device intended for the closed-loop processing of lipoaspirate tissue in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation.
Device Story
Lipogems System is a sterile, single-use, closed-loop device for processing lipoaspirate tissue. Input: lipoaspirate tissue harvested via legally marketed lipoplasty system. Processing: tissue passes through an ABS unit containing five stainless steel spheres and two stainless steel sieve filters to concentrate and wash the adipose tissue. Output: processed autologous adipose tissue ready for transfer. Used in various surgical specialties (orthopedic, neurosurgery, plastic, etc.) by physicians. The device facilitates minimal manipulation of fat for autologous transfer. Benefits include providing a standardized, closed-loop method for preparing adipose tissue for clinical use without additional manipulation.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical design, materials, and functional characteristics to the predicate device (K142682).
Technological Characteristics
ABS processing unit; 5 stainless steel spheres; 2 stainless steel sieve filters; Luer-lock connections; self-occluding valves. Available in 60cc and 240cc capacities. Sterilized via ethylene oxide gas (SAL 10^-6). Single-use, closed-loop system.
Indications for Use
Indicated for patients undergoing surgical procedures requiring the transfer of autologous adipose tissue, including orthopedic, arthroscopic, neurosurgical, gastrointestinal, urological, general, gynecological, thoracic, laparoscopic, and plastic/reconstructive surgeries. Contraindicated for use with non-legally marketed accessories or if additional manipulation of harvested fat is required.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection. The profiles are depicted in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 4, 2016
Lipogems International S.p.A % Scott Bruder, M.D., Ph.D. Principal, Bruder Consulting International, LLC 268 Glen Place Franklin Lakes, New Jersey 07417
Re: K161636
Trade/Device Name: Lipogems System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: September 27, 2016 Received: September 30, 2016
Dear Dr. Bruder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161636
Device Name Lipogems System
#### Indications for Use (Describe)
The Lipogems System is a sterile medical device intended for the closed-loop processing of lipoaspirate tissue in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Device Name: | The Lipogems System |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submission: | K161636 |
| Date of Submission: | June 11, 2016 |
| 510(k) Owner & Manufacturer: | Lipogems International S.p.A.<br>Viale Bianca Maria 24<br>20129 Milano (MI) Italy<br>Phone: +39 02 3707 2408/9<br>Fax: +39 02 3707 2410 |
| 510(k) Submitter and Contact: | Scott Bruder, MD, PhD<br>Principal<br>Bruder Consulting International, LLC<br>268 Glen Place<br>Franklin Lakes, NJ 07417<br>Phone: 201.874.9701<br>Email: scottbruder@me.com |
| FDA Product Code: | MUU |
| FDA Requlation Number: | 21 CFR 878.5040 |
| FDA Classification Name: | System, Suction, Lipoplasty |
| Classification Panel: | General and Plastic Surgery |
| Common Name: | Lipoplasty System |
| FDA Classification: | Class II |
| Predicate Device:<br>Reference Device: | K142682 The Lipogems System, Lipogems International, S.p.A.<br>K150156 Ranfac FATS Procedure Pack, Ranfac Corporation |
#### Indications for Use / Intended Use:
The Lipogems System is a sterile medical device intended for the closed-loop processing of lipoaspirate tissue in medical procedures involving the harvesting, concentrating and transferring of autologous adipose tissue harvested with a legally marketed lipoplasty system. The device is intended for use in the following surgical specialties when the transfer of harvested adipose tissue is desired: orthopedic surgery, arthroscopic surgery, neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and plastic and reconstructive surgery when aesthetic body contouring is desired. Only legally marketed accessory items, such as syringes, should be used with the system. If harvested fat is to be transferred, the harvested fat is only to be used without any additional manipulation.
#### Device Description:
The subject device is identical in every way to the predicate device previously cleared, and described in K142682, also submitted by Lipogems International, S.p.A.
{4}------------------------------------------------
The Lipogems System is a pre-assembled device consisting of:
- An ABS processing unit (Lipogems Processing Unit) containing 5 stainless steel spheres and ● 2 stainless steel sieve filters;
- An input washing line connected to the 'blue' end-cap the processing unit;
- An access port for the loading of material to be processed with a Luer-lock connection and self-● occluding valve (blue end-cap);
- . A drain line connected to the 'gray' end-cap of the processing unit;
- An access port for the discharge of processed material with a Luer-lock connection and self-● occluding valve (gray end-cap);
- . A bag for collecting waste material.
The Lipogems System Processing Unit is supplied sterile, for single use only, and manufactured in two size variants with the same functional characteristics:
- LGD 240: Lipogems Processing Unit with 240 cc capacity and standard exit sieve; .
- LGD 60: Lipogems Processing Unit with 60 cc capacity and standard exit sieve; ●
| Item | Reference device<br>(Ranfac) | Predicate device<br>(Lipogems) | Subject device<br>(Lipogems) | Similarity |
|------------------------------|--------------------------------------------------------------------|------------------------------------|------------------------------------|-----------------------------------------------------------------|
| Device name | Ranfac Fat Aspiration<br>Transfer Syringe (FATS)<br>Procedure Pack | The Lipogems System | The Lipogems System | Identical<br>to<br>predicate |
| Device<br>Manufacturer | Ranfac Corporation | Lipogems International<br>SpA | Lipogems International<br>SpA | Identical<br>to<br>predicate |
| 510(k) Reference | K150156 | K142682 | K161636 | N/A |
| FDA Product<br>Code | MUU | MUU | MUU | Identical |
| FDA Classification<br>Name | System, Suction,<br>Lipoplasty | System, Suction,<br>Lipoplasty | System, Suction,<br>Lipoplasty | Identical |
| FDA Regulation<br>Number | 878.5040 | 878.5040 | 878.5040 | Identical |
| System concept | Closed Loop | Closed loop | Closed loop | Identical |
| Device Size<br>Variations | None | Two: LGD 60, LGD 240 | Two: LGD 60, LGD 240 | Identical<br>to<br>predicate |
| Fill Volumes | 5ml to 25ml of fat | Up to 30 or 120 ml of fat | Up to 30 or 120 ml of fat | Identical<br>to<br>predicate;<br>Different<br>from<br>reference |
| Construction | Preassembled | Preassembled | Preassembled | Identical |
| Sterility | Supplied sterile for single<br>use | Supplied sterile for single<br>use | Supplied sterile for single<br>use | Identical |
| Sterilization<br>method | Ethylene oxide Gas | Ethylene oxide Gas | Ethylene oxide Gas | Identical |
| Sterility Assurance<br>Level | SAL = 10-6 | SAL = 10-6 | SAL = 10-6 | Identical<br>to<br>predicate |
#### Predicate and Reference device comparison table
{5}------------------------------------------------
| Predicate and Reference device comparison table | | | | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Item | Reference device<br>(Ranfac) | Predicate device<br>(Lipogems) | Subject device<br>(Lipogems) | Similarity |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Identical |
| Indications for Use<br>/ Intended Use | The Ranfac FATS<br>Procedure Pack is used in<br>medical procedures<br>involving the harvesting<br>and transferring of<br>autologous adipose tissue.<br>The FATS Procedure Pack<br>is for concentrating<br>adipose tissue harvested<br>with a legally marketed<br>lipoplasty system. The<br>device is intended for use<br>in the following surgical<br>specialties when the<br>concentration of harvested<br>adipose is desired:<br>Neurosurgery,<br>gastrointestinal surgery,<br>urological surgery, plastic<br>& reconstructive surgery,<br>general surgery, orthopedic<br>surgery, gynecological<br>surgery, thoracic surgery,<br>laparoscopic surgery,<br>arthroscopic surgery. | The Lipogems System is a<br>sterile medical device<br>intended for the closed-<br>loop processing of<br>lipoaspirate tissue for the<br>purpose of transferring<br>autologous adipose tissue<br>for aesthetic body<br>contouring (lipofilling) in<br>applications including<br>plastic and reconstructive<br>surgery, neurosurgery,<br>gastrointestinal and<br>affiliated organ surgery,<br>urological surgery, general<br>surgery, orthopedic<br>surgery, gynecological<br>surgery, thoracic surgery,<br>and laparoscopic surgery.<br>Only legally marketed<br>accessory items, such as<br>syringes, should be used<br>with the system. If<br>harvested fat is to be<br>reimplanted, the harvested<br>fat is only to be used<br>without any additional<br>manipulation. | The Lipogems System is a<br>sterile medical device<br>intended for the closed-<br>loop processing of<br>lipoaspirate tissue in<br>medical procedures<br>involving the harvesting,<br>concentrating and<br>transferring of autologous<br>adipose tissue harvested<br>with a legally marketed<br>lipoplasty system. The<br>device is intended for use<br>in the following surgical<br>specialties when the<br>transfer of harvested<br>adipose tissue is desired:<br>orthopedic surgery,<br>arthroscopic surgery,<br>neurosurgery,<br>gastrointestinal and<br>affiliated organ surgery,<br>urological surgery, general<br>surgery, gynecological<br>surgery, thoracic surgery,<br>laparoscopic surgery, and<br>plastic and reconstructive<br>surgery when aesthetic<br>body contouring is desired.<br>Only legally marketed<br>accessory items, such as<br>syringes, should be used<br>with the system. If<br>harvested fat is to be<br>transferred, the harvested<br>fat is only to be used<br>without any additional<br>manipulation. | Similar |
#### Substantial Equivalence Discussion:
The Subject and Predicate device are identical in every material and functional way, as there has been no change to the device materials, design, manufacturing, sterilization, packaging or any other feature since its prior clearance under K142682.
The subject device of this submission is indicated for concentration and transfer of autologous adipose tissue. The minimally manipulated nature of such tissue in the subject device, and the reference device, is reiterated in the IFUs and form the basis of substantial equivalence.
#### Conclusion:
Based on the information contained within this submission, Lipogems International S.p.A. believes that the Lipogems System is substantially equivalent to the identified predicate devices already cleared, and that the revised language of the IFU and other labelling in this submission is appropriate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.