K161634 · Nobel Biocare AB · LLZ · Nov 10, 2016 · Radiology
Device Facts
Record ID
K161634
Device Name
IPS CaseDesigner
Applicant
Nobel Biocare AB
Product Code
LLZ · Radiology
Decision Date
Nov 10, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of cranio-maxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids.
Device Story
IPS CaseDesigner is a 3D image-based software tool for virtual planning of orthognathic surgery. Input consists of patient CBCT/CT data (DICOM). Surgeons use the software to plan osteotomies (Le Fort I, Sagittal Split, Ramus, Chin) by virtually cutting and repositioning bone fragments in 3D space. The software provides 2D/3D visualization, measurement tools, and soft tissue simulation to predict post-operative appearance. Output includes surgical splint order files sent to a manufacturing facility for production of patient-specific surgical aids. Used by oral and maxillofacial surgeons in clinical settings to facilitate surgical planning and improve patient outcomes through precise, individualized surgical guidance.
Clinical Evidence
No clinical data. Substantial equivalence supported by software verification and validation testing conducted in accordance with FDA guidance for software contained in medical devices.
Technological Characteristics
Software-based surgical planning tool; runs on Windows 64-bit and Mac platforms. Inputs: DICOM CT/CBCT data, generic 3D models (STL). Functions: 2D/3D visualization, distance/angle measurement, bone fragment manipulation (translation/rotation), soft tissue simulation, and surgical splint order file generation. Standalone software application.
Indications for Use
Indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner (e.g., CT) to support diagnostic and treatment planning for cranio-maxillofacial procedures, including orthognathic surgery.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 10, 2016
Nobel Biocare Ab % Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887
Re: K161634
Trade/Device Name: IPS CaseDesigner Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: October 19, 2016 Received: October 20, 2016
Dear Charlemagne Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Robert Oals
Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
| Indications for Use |
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below. |
| 510(k) Number (if known) | K161634 |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | IPS CaseDesigner |
| Indications for Use (Describe) | IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of cranio-maxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids. |
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (1/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
{3}------------------------------------------------
# 1. 510(k) Summary
### I. Submitter
Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x7830 Fax: (714) 998-9348
Submitted for: Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Date Prepared : October 19, 2016
# II. Device
Name of Device: IPS CaseDesigner Common or Usual Name: Picture Archiving and Communications System Classification Name: System, Image Processing, Radiological (21 CFR 892.2050) Regulatory Class: II Product Code: LLZ
# III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary Predicate
- Maxilim, Medicim NV, K052424 ●
Additional Predicates
- Surgicase Orthognathic Software Wizard, Materialise N.V., K111641 ●
{4}------------------------------------------------
# IV. Device Description
The IPS CaseDesigner software is a tool for the virtual planning of orthognathic surgery. The user is a (oral and maxillofacial) surgeon who needs to plan an orthognathic surgery, and will use the software to plan the osteotomy (bone cutting). This allows the surgeon to re-position the digitally cut pieces in the desired alignment and allows the visualization of how the patient aspect will look like, through soft tissue simulation. The software supports various types of osteotomy procedures (Le Fort I, Sagittal Split, Ramus and Chin).The selection of the appropriate osteotomy procedure for each patient case is at the discretion of the clinician.
The IPS CaseDesigner is a 3D image-based environment that allows the clinician to visualize a patient's anatomy. The creation of the 3D augmented patient model is derived from CBCT (Cone-Beam CT) data (DICOM). This allows the clinician virtually plan and simulate a craniomaxillofacial surgery. The IPS CaseDesigner software has the capability to save a surgical splint file that the clinician can send to the manufacturing facility for surgical splint creation based on the CBCT Data. The software does not have any functionality for designing.
# V. Indications for Use
IPS CaseDesigner is indicated for use as a software and image segmentation system for the transfer of imaging information from a scanner such as a CT scanner. It is also indicated to support the diagnostic and treatment planning process of cranio-maxillofacial procedures. IPS CaseDesigner facilitates the service offering of individualized surgical aids.
#### Discussion
The Indications for Use statement between the subject and predicate devices are equivalent; minor differences in wording do not alter the intended therapeutic use of the subject device nor do they affect the safety and effectiveness of the subject device relative to the predicate. Both the subject and predicate devices are software devices intended for the transfer of medical images and to support diagnostics and treatment planning of cranio-maxillofacial procedures.
#### VI. Comparison of Technological Characteristics
The subject and predicate devices are software based surgical planning tools which allow for transfer of medical images, and creation of 3D models.
A comparison of the subject and predicate devices is provided in the table below.
{5}------------------------------------------------
| | SUBJECT DEVICE | PRIMARY PREDICATE | ADDITIONAL<br>PREDICATE |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | IPS CaseDesigner | Maxilim<br>(K052424) | SurgiCase, SurgiCase CMF,<br>ProPlan CMF<br>(K111641) |
| Indications<br>for Use<br>Statement | IPS CaseDesigner is<br>indicated for use as a<br>software and image<br>segmentation system for the<br>transfer of imaging<br>information from a scanner<br>such as a CT scanner. It is<br>also indicated to support the<br>diagnostic and treatment<br>planning process of cranio-<br>maxillofacial procedures.<br>IPS CaseDesigner facilitates<br>the service offering of<br>individualized surgical aids. | Maxilim is indicated for use<br>as a software interface and<br>image segmentation system<br>for the transfer of imaging<br>information from a medical<br>scanner such as a CT<br>scanner. It is also indicated<br>for use as a planning and<br>simulation software for<br>surgical treatment,<br>specifically maxillofacial<br>procedures. | SurgiCase contains software<br>module for pre-operative<br>simulation of orthognathic<br>surgical treatment options,<br>based on imaging<br>information from a medical<br>scanner such as a CT scanner<br>or a Magnetic Resonance<br>Imaging (MRI) scanner. |
| Intended use | IPS CaseDesigner is a<br>software for the transfer and<br>visualization of imaging<br>information from equipment<br>such as a CT scanner to<br>support the diagnostic and<br>treatment planning process<br>in cranio-maxillofacial<br>regions. IPS CaseDesigner<br>facilitates the service<br>offering of individualised<br>surgical aids. | Maxilim is indicated for use<br>as a software interface and<br>image segmentation system<br>for the transfer of imaging<br>information from a medical<br>scanner such as a CT<br>scanner. It is also indicated<br>for use as a planning and<br>simulation software for<br>surgical treatment,<br>specifically maxillofacial<br>procedures. | Information not available |
| Clinical Use | Cranio-maxillofacial and<br>orthognathic treatment | Maxillofacial treatment | Cranio-maxillofacial and<br>Orthognathic treatment |
| Image Import | DICOM data format from<br>CT/ CBCT scanner<br>3D models in generic open<br>file format (STL) | DICOM data format from<br>CT/ CBCT scanner | CT/CBCT or MRI scan data |
| Software<br>Functions | 2D and 3D Visualization<br>Measuring tool,<br>Distance and angle | 2D and 3D Visualization<br>Measuring tool,<br>Distance and angle | 2D and 3D Visualization<br>Measuring tool |
| | SUBJECT DEVICE | PRIMARY PREDICATE | ADDITIONAL<br>PREDICATE |
| | IPS CaseDesigner | Maxilim<br>(K052424) | SurgiCase, SurgiCase CMF,<br>ProPlan CMF<br>(K111641) |
| | Bone fragments can be<br>moved in the 3D space<br>(translations and rotations) in<br>order to plan the ideal post-<br>operative position of each<br>fragment. | Bone fragments can be<br>moved in the 3D space<br>(translations and rotations) in<br>order to plan the desired<br>post-operative position of<br>each fragment. | Information not available |
| | 3D Surgical Model creation | 3D Surgical Model creation | 3D objects from medical<br>image data |
| | Soft tissue simulation | Soft tissue simulation and<br>3D photo mapping | Soft tissue simulation and<br>3D photo mapping |
| | Based on the created<br>orthognathic plan the user<br>can export surgical splint<br>order files. These files can<br>be used to calculate and<br>produce orthognathic splints<br>in the production backend | No | Patient-specific surgical<br>splints can be generated to<br>transfer the planned dental<br>occlusion to surgery. |
| Computer<br>System<br>Requirements | Windows 64bit,<br>Mac | Windows 32 or 64 bit. | Windows 64 bit. |
{6}------------------------------------------------
#### Performance Data VII.
The following performance data were provided or relied upon in support of the substantial equivalence determination.
# Software Validation
Software verification and validation testing was conducted on the subject device and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
{7}------------------------------------------------
# VIII. Conclusion
Although there are minor differences in the indications for use statements and technological features of the subject and predicate devices, the testing supports that these differences do not raise any new questions of safety and effectiveness. Therefore, it is concluded that the subject device is substantially equivalent to the cited predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.