K161629 · Sm Eng Co., Ltd. · GAM · Feb 10, 2017 · General, Plastic Surgery
Device Facts
Record ID
K161629
Device Name
CRYLREX
Applicant
Sm Eng Co., Ltd.
Product Code
GAM · General, Plastic Surgery
Decision Date
Feb 10, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CRYLREX® is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Device Story
CRYLREX® is a sterile, braided, coated, synthetic absorbable surgical suture composed of a copolymer of 90% glycolide and 10% L-lactide. It is coated with poly(glycolide-co-lactide) (30/70) and calcium stearate. The suture is available in dyed (D&C Violet No. 2) or undyed (natural) forms, with or without stainless steel (SUS 304) needles. It is intended for use by clinicians for soft tissue approximation and ligation. The device maintains approximately 75% tensile strength after 2 weeks and 50% after 3 weeks, with complete absorption occurring within 56 to 70 days. It is a single-use device.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included USP 35/36 standards for suture diameter, needle attachment, tensile strength, sterility, and pyrogenicity. Biocompatibility testing performed per ISO 10993 (genotoxicity, cytotoxicity, irritation, sensitization, systemic toxicity, and implantation effects). Real-time and accelerated aging studies supported shelf life.
Technological Characteristics
Multifilament braided suture; 90% glycolide/10% L-lactide copolymer; coated with poly(glycolide-co-lactide) (30/70) and calcium stearate. Stainless steel (SUS 304) needles. Sterilized via ethylene oxide (ISO 11135-1). Complies with USP 35 <861>, <871>, <881> and ASTM F88-09, F1929-98.
Indications for Use
Indicated for general soft tissue approximation and/or ligation, including ophthalmic procedures. Contraindicated for cardiovascular or neural tissue.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 10, 2017
SM ENG CO., Ltd % Mr. Sanglok Lee Wise Company, Inc. #306, 3 Daerim-ro 27ga-gil Yeongdeungpo-gu, 07413 KR
Re: K161629
Trade/Device Name: CRYLREX® Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: December 28, 2016 Received: January 5, 2017
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### K161629
Device Name CRYLREX®
Indications for Use (Describe)
CRYLREX® is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
## The assigned 510(k) Number: K161629
#### 01. Date of Submission: 2016.12.28
#### 02. Applicant
Company name: SM ENG CO., LTD Address: 46, Nakdong-daero 1302beon-gil, Sasang-gu, Busan, Korea TEL: +82 51 3058016 FAX: +82 513058021 Email: leesg@sm-eng.net
#### 03. Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. #306, 3 Daerim-ro 27ga-gil,Yeongdeungpo-gu, Seoul, Korea TEL: +82 7088123619 / +82 28313615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
## 04. Proposed Device Identification
Proprietary Name: CRYLREX® Common Name: Synthetic Absorbable Polyglactin Suture With or Without Needle Device Class: Class II Regulation Number:21 C.F.R. 878.4493 Product Code: GAM Device Classification Name: Suture, Absorbable, Synthetic, Polyglycolic Acid
#### 05. Indication for use
CRYLREX® is indicated for use in general soft tissue approximationand/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
#### 06. Predicate devices
Predicate device 01 510(k) Number:K130735 DeviceName: WEGO-PGLA Absorbable Surgical Suture Classification Name: Suture, Absorbable, Synthetic, Polyglycolic Acid Manufacturer: FOOSIN MEDICAL SUPPLIES INC.LTD
#### 07. Device Description
CRYLREX® is a sterilized medical device composed with multifilament absorbable, braided, coated suture, 90% glycolide and 10% L-lactide, with needle, stainless steel SUS 304.
CRYLREX® Suture is a synthetic absorbable, braided, coated suture composed of a copolymer made from 90% glycolide and 10% L-lactide and is available both dyed and
{5}------------------------------------------------
undyed. CRYLREX® Suture is coated with poly(glycolideco-lactide)(30/70) and calcium stearate. The suture is available undyed (natural) or dyed (D&C Violet No.2). Approximately 75% of tensile strength remain after 2 weeks, 50% of tensile strength remain after 3 weeks. Complete absorption in tissues takes around 56 to 70 days.
CRYLREX® Sutures are U.S.P. except for diameters in the following sizes:
| U.S.P.<br>SUTURE SIZE<br>DESIGNATION | MAXIMUM<br>OVERSIZE<br>(mm) |
|--------------------------------------|-----------------------------|
| 8-0 | 0.007 |
| 7-0 | 0.010 |
| 6-0 | 0.021 |
| 5-0 | 0.036 |
| 4-0 | 0.031 |
| 3-0 | 0.036 |
| 2-0 | 0.021 |
| 0 | 0.056 |
| 1 | 0.046 |
| 2 | 0.031 |
MAXIMUM SUTURE OVERSIZE IN DIAMETER (mm) FROM U.S.P.
## 08. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposeddevice complies with the following standards:
USP 35 <861> SUTURES - DIAMETER
USP 35 <871> SUTURES - NEEDLE ATTACHMENT
USP 35 <881> TENSILE STRENGTH
USP Absorbable Surgical Suture
Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA
ASTM F88-09. Standard Test Method for Seal Strength of Flexible Barrier Materials;
ASTM F1929-98(2004) Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
USP, <71> STERILITY TEST
USP 36, <151>; Pyrogen Test
ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide- Part 1: Requirements fordevelopment, validation and routing control of a sterilization process for medical devices;
USP <85> Bacterial Endotoxin Limit;
ISO 10993, Biological Evaluation of Medical Devices:
Including:
Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (ISO 10993-1:2009)
Biological evaluation of medical devices - Part 3:Tests for genotoxicity, carcinogenicity and reproductivetoxicity (ISO 10993-3:2014)
Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood (ISO 10993-4:2002, including Amd 1:2006)
{6}------------------------------------------------
Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (ISO 10993-5:2009)
Biological evaluation of medical devices - Part 6: Tests for local effects after implantation (ISO 10993-6:2007)
Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (ISO 10993-7:2008)
Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (ISO 10993-10:2010)
Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (ISO 10993-11:2006)
Real-time and accelerated aging stability testing was performed to support shelf life of CRYLREX®. Additionally, the residual strength and absorption rate studies were performed and the sutures were evaluated in accordance with the requirements outlined in FDA's Class II Special Controls Guidance Document: Surgical Sutures.
#### 09. Substantially Equivalent Conclusion
| Property | Proposed device<br>CRYLREX® | Predicate device 01<br>WEGO-PGLA Absorbable<br>Surgical Suture |
|----------------|-----------------------------|----------------------------------------------------------------|
| 510(k) Number | K161629 | K130735 |
| Product Code | GAM | Same |
| Regulation No, | 2 1 CFR 878.4493 | Same |
| Class | II | Same |
| Sterile | Yes | Same |
| Single Use | Yes | Same |
| Configuration | PGLA Suture and Needle | Same |
{7}------------------------------------------------
| Suture | | | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Raw Suture Material | Neosorb | Same | |
| Manufacturer of Raw Suture<br>Material | Samyang Biopharmaceuticals<br>Corporation | Same | |
| Final Material Composition | 90% glycolide and 10% L-<br>lactide (PGLA) | Same | |
| Coating material | Poly(glycolide-co-lactide)<br>(30/70)+ Calcium Stearate | Same | |
| Color (Colorant) | Undyed (natural) or Dyed (D&C<br>Violet No.2) | Same | |
| Absorbable/Nonabsorbable | Absorbable | Same | |
| Braided/Monofilament | Braided | Same | |
| Suture Size | The proposed device is<br>available in 8-0, 7-0, 6-0,5-0, 4-<br>0,3-0,2-0, 0, 1 and 2, which are<br>the sizes identified in the<br>currently recognized United<br>States Pharmacopoeia except<br>for diameter | The proposed device is<br>available in 6-0,5-0, 4-0,3-0,2-0,<br>0 and 1, which are the sizes<br>identified in the currently<br>recognized United States<br>Pharmacopoeia except for<br>diameter | |
| Length of Suture | 20cm, 30cm, 45cm, 60cm,<br>70cm, 75cm, 90cm, 100cm,<br>125cm, 140cm, 150cm, 250cm, | 30cm, 45cm, 60cm, 75cm,<br>90cm, 100cm, 120cm,<br>150cm,180cm, 200cm, 250cm,<br>280cm, 300cm, 320cm,<br>360cmand 390cm | |
| Diameter of Suture | Oversize | Oversize | |
| Tensile strength | The tensile strengths of<br>proposed device comply with<br>thetensile requirement listed in<br>USP 35 <881> TensileStrength | The tensile strengths of<br>proposed device comply with<br>thetensile requirement listed in<br>USP 35 <881> TensileStrength | |
| Needle Attachment | The bond between suture and<br>needle ofthe applicantdevice<br>meet the requirements defined<br>in USP 35 <871>. | The bond between suture and<br>needle ofthe applicantdevice<br>meet the requirements defined<br>in USP 35 <871>. | |
| Needle | | | |
| Material | Stainless Steel | Stainless Steel | |
| Needle type | Taper point, Reverse<br>Cutting, Conventional cutting,<br>Taper cutting, Spatula, Blunt<br>point | Taper, Spatula, Cutting, Blunt | |
{8}------------------------------------------------
The proposed device, CRYLREX® Suture, is determined to be SubstantiallyEquivalent (SE) to the predicate device, SYNTHETIC ABSORBABLE SUTURE (K130735) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.