# Arctic Sun Temperature Management System (K161602)

_Medivance, Inc. · DWJ · Sep 2, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161602

## Device Facts

- **Applicant:** Medivance, Inc.
- **Product Code:** [DWJ](/productcode/DWJ.md)
- **Decision Date:** Sep 2, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.5900
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Arctic Sun® Temperature Management System is a thermal regulating system, indicated for monitoring and controlling patient temperature in adult and pediatric patients of all ages.

## Device Story

Non-invasive thermal regulating system; monitors/controls patient temperature (32°C to 38.5°C). System comprises Arctic Sun 5000 Control Module and disposable ArcticGel Pads. Inputs: patient core temperature via external medical probe (naso-pharyngeal, bladder, rectal, or esophageal). Operation: control module recirculates temperature-controlled water to hydrogel-adhesive pads applied to patient skin. Modes: automatic (algorithm adjusts water temperature based on probe input) or manual (clinician-adjusted). Used in clinical settings; operated by healthcare professionals. Output: temperature regulation; EMR connectivity for data output. Benefits: precise patient temperature management.

## Clinical Evidence

Bench testing only. Hardware design verification and software validation were performed to verify new system and software requirements. Device retested for compliance with electrical safety and electromagnetic compatibility standards.

## Technological Characteristics

Thermal regulating system; recirculating water-based heat exchange. Components: Arctic Sun 5000 Control Module, disposable ArcticGel Pads with biocompatible hydrogel adhesive. Connectivity: EMR data output support. Software: algorithm-based temperature control. Sterilization: pads are non-sterile.

## Regulatory Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

## Predicate Devices

- Arctic Sun® Temperature Management System ([K142702](/device/K142702.md))

## Reference Devices

- ArcticGel Pads ([K002577](/device/K002577.md))
- ArcticGel Pads ([K142702](/device/K142702.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 02, 2016

Medivance, Inc. Stacci Cronk Regulatory Affairs Manager 321 South Taylor Ave, Suite 200 Louisville, Colorado 80027

Re: K161602

Trade/Device Name: Arctic Sun Temperature Management System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: August 2, 2016 Received: August 3, 2016

Dear Stacci Cronk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Hillebrand

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known) K161602

Device Name Arctic Sun® Temperature Management System

Indications for Use (Describe)

The Arctic Sun® Temperature Management System is a thermal regulating system, indicated for monitoring and controlling patient temperature in adult and pediatric patients of all ages.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1 FDA

PSC Publishing Servi

{3}------------------------------------------------

#### Page 23

K101092

# 510(k) SUMMARY

# Arctic Sun® Temperature Management System

| 510(k) Owner:                                                | Medivance, Inc.<br>A wholly owned subsidiary of C. R. BARD, Inc.<br>321 South Taylor Avenue, Suite 200<br>Louisville, CO 80027 USA |  |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person:                                              | Stacci Cronk, RAC<br>Regulatory Affairs Manager<br>Phone: 303-327-5151<br>E-mail: stacci.cronk@crbard.com                          |  |
| Date Prepared:                                               | June 2, 2016                                                                                                                       |  |
| Trade Name:                                                  | Arctic Sun® Temperature Management System (Model<br>5000 Control Module)                                                           |  |
| Common/Usual Name:                                           | Patient temperature management system                                                                                              |  |
| Classification Name:                                         | Thermal regulating system                                                                                                          |  |
| Regulation:                                                  | 21 C.F.R. §870.5900                                                                                                                |  |
| Classification:                                              | II                                                                                                                                 |  |
| Product Code:                                                | DWJ                                                                                                                                |  |
| Predicate Device                                             |                                                                                                                                    |  |
| Arctic Sun® Temperature Management System<br>Medivance, Inc. | K142702                                                                                                                            |  |

### Indications for Use

The Arctic Sun Temperature Management System is a thermal regulating system, indicated for monitoring and controlling patient temperature in adult and pediatric patients of all ages.

### Device Description

The Arctic Sun Temperature Management System is a non-invasive, thermal regulating system that monitors and controls patient temperature within a range of 32°C to 38.5°C (89.6°F to 101.3°F). The Arctic Sun Temperature Management System consists of the Arctic Sun 5000 Control Module and disposable non-sterile ArcticGel Pads. The control module recirculates temperature-controlled water to the ArcticGel Pads. A commercially-available medical temperature probe, such as naso-pharyngeal, bladder, rectal, or esophageal,

{4}------------------------------------------------

connected to the control module senses the patient's core temperature. A control algorithm automatically adjusts the water temperature (automatic mode) or the clinician can adjust the water temperature (manual mode) to obtain the desired patient temperature. The ArcticGel Pads come in various sizes to cover a broad range of patients and fit both males and females. Each pad has an inlet and an outlet connection that attaches to a fluid delivery line that is connected to the Arctic Sun 5000 Control Module. Up to six pads can be connected at one time. The pads adhere to the patient by the use of a biocompatible hydrogel adhesive.

Please note that this Special 510(k) Notice is specific to the Arctic Sun 5000 Control Module. The ArcticGel Pads are previously 510(k) cleared under K002577 and K142702 and are not the subject of the Special 510(k) Notice.

# Purpose of Special 510(k) Notice

The purpose of this 510(k) is to modify the cleared Arctic Sun Temperature Management System by incorporating minor component, software and associated labeling changes.

# Summary of Technical Characteristics Compared to the Predicate Device

The modifications to the Arctic Sun Temperature Management System do not change the intended use, indications for use, or fundamental scientific technology of the device. The Arctic Sun Temperature Management System (Model 5000) includes the following modifications:

- New alert and alarm (software change)
- Option to add two additional customizable protocols (software change) ●
- . Data output capability for Electronic Medical Records (EMR) connectivity support (software change)
- System default changes
- User preference changes to the user interface (software change)
- New/modified components ●
- Labeling (new caution, new IFU for temperature simulator key, and various changes ● for clarifications and/or safer use of the device)

### Performance Testing

The software for the Arctic Sun Temperature Management System was developed and is maintained in accordance with the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Device" (05/11/05) and "General Principles of Software Validation; Final Guidance for Industry and FDA Staff" (1/11/02). Hardware design verification and software validation of the Arctic Sun Temperature Management System was performed to verify all new system and software requirements, the result of which confirmed the function of the device modifications.

In addition to functional testing, the Arctic Sun Temperature Management System was retested, when applicable to confirm ongoing compliance with Electrical Safety and Electromagnetic Compatibility Standards.

{5}------------------------------------------------

### Substantial Equivalence

The Arctic Sun Temperature Management System has the same Indications for Use, Intended Use and fundamental scientific technology as the predicate Arctic Sun Temperature Management System. The Arctic Sun Temperature Management System was shown to be substantially equivalent in indications for use, design, technological characteristics, materials, and system features and functions to the predicate device.

### Conclusion

Based on the testing and comparison to the predicate device, the Arctic Sun Temperature Management System performs as intended, raises no new or different safety or effectiveness issues and is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K161602](https://fda-staging.innolitics.com/device/K161602)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).