K161600 · Olympus Winter & Ibe GmbH · JOS · Sep 6, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161600
Device Name
Resection Electrodes
Applicant
Olympus Winter & Ibe GmbH
Product Code
JOS · General, Plastic Surgery
Decision Date
Sep 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications. The general indications include transcervical resection, vaporization, ablation, cutting and coagulation of tissue in the uterus in saline irrigation fluid. Specific indications: - transcervical diagnosis and treatment (resection, vaporization, ablation, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis) - lysis of intrauterine septa - endometrial ablation
Device Story
Resection Electrodes are single-use electrosurgical accessories for resectoscope systems; used in gynecological procedures. Device consists of active tip (loops, rollers, needles, or buttons), PTFE color code, insulator, guiding tube, telescope clip, and shaft. Operated by physicians in clinical settings; powered by external electrosurgical generator. Device delivers electrical current to tissue in saline irrigation fluid to perform resection, vaporization, biopsy, cutting, and coagulation. Output is physical tissue modification; enables minimally invasive treatment of intrauterine pathologies. Benefits include precise tissue removal and reduced surgical trauma.
Clinical Evidence
No clinical data. Substantial equivalence is based on identical design, manufacturing, and technological characteristics to the predicate device (K152092), relying on previously reviewed performance and validation data.
Technological Characteristics
Single-use electrosurgical electrodes; active tips include loops, rollers, needles, buttons. Materials include PTFE insulation. Operates via bipolar electrosurgery in saline irrigation fluid. Compatible with resectoscope systems. Sterilization per ISO 11135; packaging per ISO 11607-1.
Indications for Use
Indicated for gynecological endoscopic diagnosis and treatment in patients requiring transcervical resection, vaporization, ablation, cutting, or coagulation of intrauterine tissue, including myomas, polyps, synechias, endometrium, and lysis of intrauterine septa.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 6, 2016
Olympus Winter & Ibe GmbH % Dolan Mills Sr. Specialist, Regulatory Affairs Gyrus ACMI, Inc. 136 Turnpike Road Southborough, MA 01772
Re: K161600
> Trade/Device Name: Resection Electrodes (model numbers WA47505S, WA47506S. WA47507S, WA47540S, WA47551S, WA47555S, WA47560S, WA47566S) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: JOS, HIH Dated: June 8, 2016 Received: June 9, 2016
Dear Dolan Mills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K161600
#### Device Name
Resection Electrodes (model numbers WA47505S, WA47507S, WA47507S, WA47551S, WA47551S, WA47550S, WA47566S
Indications for Use (Describe)
Electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications.
The general indications include transcervical resection, vaporization, cutting and coagulation of tissue in the uterus in saline irrigation fluid.
Specific indications:
- transcervical diagnosis and treatment (resection, vaporization, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis)
- lysis of intrauterine septa
- endometrial ablation .
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness June 8, 2016
## 1. General Information
| Manufacturer: | Olympus Winter & Ibe GmbH<br>Kuehnstr. 61<br>22045 Hamburg<br>Germany<br>Establishment Registration No.: 9610773 |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Dolan Mills<br>Sr. Specialist, Regulatory Affairs<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 901.373.0236<br>Fax: 901.373.0220<br>Email: dolan.mills@olympus-osta.com<br>Establishment Registration No.: 3003790304 |
## 2. Device Identification
| Common Name: | Electrode, Electrosurgical / Hysteroscope (and accessories) |
|-------------------------|-----------------------------------------------------------------------------------------------|
| Regulation Number: | 878.4400 / 884.1690 |
| Classification | Electrosurgical cutting and coagulation device and accessories / Hysteroscope and accessories |
| Device Class: | II |
| Product Code: | JOS / HIH |
| Review Panel: | General & Plastic Surgery / Obstetrics/Gynecology |
| Proprietary/Trade Name: | Resection Electrodes |
| Model numbers: | WA47505S, WA47506S, WA47507S, WA47540S, WA47551S, WA47555S, WA47560S, WA47566S |
# 3. Predicate Devices
The subject devices have been previously cleared for urological indications under K152092 (product codes FAS / FJL).
{4}------------------------------------------------
The predicate devices have not been subject to a design-related recall.
No reference devices were used in this submission.
## 4. Product Description
The Olympus Resection Electrodes that are subject to this submission are for application in saline. Depending on the characteristics of electrical current, which is provided by the electrosurgical generator, electrosurgery can be used for coagulation, vaporization and cutting.
The subject HF-Resection Electrodes consist of an active tip, PTFE color code identification, an insulator between the electrode and electrode tube, a guiding tube, telescope clip and arm (shaft).
The design and dimensions of the electrodes vary to accommodate various procedural conditions. The active tips of the various electrodes may consist of loops, rollers, needles or buttons.
All subject Resection Electrodes are single-use electrodes and are delivered sterile and are used in combination with a reusable HF cable (the HF cable is not subject to this submission).
## 5. Indications for Use
Electrodes are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications.
The general indications include transcervical resection, vaporization, ablation, cutting and coagulation of tissue in the uterus in saline irrigation fluid.
Specific indications:
- transcervical diagnosis and treatment (resection, vaporization, ablation, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis)
- lysis of intrauterine septa -
- endometrial ablation
Although the subject and predicate devices have different indications for use statements, the intended use is the same. The subject device is indicated for gynecologic applications while the predicate device is indicated for urological indications. However, when the decision-making criteria specified in the FDA
{5}------------------------------------------------
guidance document, "General/Specific Intended Use," (issued November 4, 1998) are applied, the subject device indications are determined to fall within the scope of the intended use of the predicate device. The specific gynecology indications of the subject device do not represent a different intended use since there is an extensive knowledge base regarding the use of bipolar electrosurgery for gynecological applications, including the disease states listed in the specific gynecologic indications. Therefore, the intended use comparison supports substantial equivalence.
## 6. Comparison of Technological characteristics
The subject and predicate devices are identical in design and are based on the same technological principle with the same elements:
- Identical in design
- I Used in combination with a resectoscope system (identical to predicate)
- I Like the predicate electrodes, the subject device resection electrode series features loops, needles, rollers, and buttons as active tip shapes
- I Utilizing resection in saline as mode of ablation (identical to predicate)
## 7. Performance Data
The subject device is identical in design and manufacturing to the predicate device (K152092). Therefore, the current submission relies on performance testing previously reviewed and deemed acceptable in K152092.
## 8. Sterilization and Shelf Life
Sterilization is performed according to ISO 11135 and packaging conforms with ISO 11607-1. Since the subject device is identical in design and manufacturing to the predicate device, the current submission relies on sterilization and packaging validation testing previously reviewed and deemed acceptable in K152092.
## 9. Conclusion
The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.