K161599 · Genadyne Biotechnologies, Inc. · OMP · Apr 6, 2017 · General, Plastic Surgery
Device Facts
Record ID
K161599
Device Name
UNO Negative Pressure Wound Therapy System
Applicant
Genadyne Biotechnologies, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Apr 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Genadyne UNO is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of low to moderate exudates and infectious material. Appropriate wound types include: - Chronic - Acute - Traumatic - Subacute and dehisced wounds - Partial-thickness burns - Ulcers (such as diabetic or pressure) - Flaps and grafts Genadyne UNO is a single patient use device.
Device Story
Portable, battery-powered suction pump; applies negative pressure to wound sites to remove low-to-moderate exudates and infectious material; promotes wound healing. Single-patient use; operated by clinician or patient. System includes pump, disposable 70ml canisters with hydrophobic shut-off filters, and sterile dressings. Healthcare providers monitor wound progress; device output (negative pressure) facilitates exudate management. Benefits include accelerated healing and infection control in chronic, acute, and traumatic wounds.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on dressing kits (silicone foam, PU film strips). Pump performance verified via in-house protocols and compliance with IEC 60601-1 and IEC 60601-1-2 standards.
Technological Characteristics
Powered suction pump; 3V DC power via Alkaline-Manganese Dioxide AA batteries. Max vacuum 125 mmHg. Includes 70ml disposable canisters with hydrophobic shut-off filters. Sterile dressings (silicone foam, PU film). Dimensions 3" x 4.4" x 2.4"; weight 400g. Compliance: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing of chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic/pressure), flaps, and grafts. Contraindicated for necrotic tissue with eschar, untreated osteomyelitis, malignancy, exposed arteries/veins/organs, non-enteric/unexplored fistulas, anastomotic sites, emergency airway aspiration, pleural/mediastinal/chest tube drainage, and surgical suction.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
Pico Single Use Negative Pressure Wound Therapy System (K151436)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2017
Genadyne Biotechnologies, Inc. Chien Ming Goh (Andrew) Vice President, Regulatory Affairs 16 Midland Ave Hicksville, New York 11801
Re: K161599
Trade/Device Name: Uno Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: January 18, 2017 Received: February 28, 2017
Dear Mr. Goh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K161599
Device Name
Genadyne UNO Negative Pressure Wound Therapy
Indications for Use (Describe)
Genadyne UNO is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of low to moderate exudates and infectious material.
Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcers (such as diabetic or pressure)
- Flaps and grafts
Genadyne UNO is a single patient use device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510k Summary
Negative Pressure Wound Therapy
Genadyne Biotechnologies, Inc. 16 Midland Ave, Hicksville, NY 11801
E-mail: Andrew@genadyne.com (t) 516.217.0100 (f) 516.977.8974
Contact Person: Mr. Chien-Ming GOH (Andrew)
Date Prepared: April 4, 2017
Name of Device
Genadyne UNO Negative Pressure Wound Therapy System
## Common or Usual Name
Powered Suction Pump
#### Classification Name
OMP, Negative Pressure Wound Therapy Powered Suction Pump
21 C.F.R. § 878.4780
#### Predicate Device
Pico Single Use Negative Pressure Wound Therapy System, K151436
#### Device Description
The UNO Wound Vacuum System is portable, battery powered wound suction pump with the intention to apply negative pressure to the wound.
## Intended Use / Indications for Use
Genadyne UNO is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of low to moderate exudates and infectious material.
Appropriate wound types include:
- Chronic
- Acute
{4}------------------------------------------------
- Traumatic
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcers (such as diabetic or pressure)
- Flaps and grafts
It is a single patient use device.
# Technological Characteristics
# Table of Comparison to Predicate Devices:
| Comparative Information | | |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Device | New Device |
| Company | Smith and Nephew | Genadyne Biotechnologies |
| Device Name | Pico Single Use Negative Pressure<br>Wound Therapy System | UNO Single Patient Use Negative<br>Pressure Wound Therapy System |
| 510 (K) Number | K151436 | |
| Technical Data | | |
| Max Vacuum | 100 mmHg | 125 mmHg |
| Battery Type | Lithium AA (L91) | Alkaline-Manganese Dioxide AA<br>(QU1500) |
| Power (Battery) | 3V DC | 3V DC |
| Dimensions / Weight | 3.5" x 3.5" x 1.0" / <120g | 3" x 4.4" x 2.4" / 400g |
| Accessories | | |
| Canisters | N/A | Two 70 ml disposable canister with a<br>build-in hydrophobic shut off filter for<br>overflow protection |
| Reusable | No | No |
| Sterile | Pump, dressing, and secondary<br>fixation strips are sterile | Dressings provided are sterile |
| Accessories | | |
| | | 70ml Canisters |
| Dressings | 10cm x 20cm<br>10cm x 30cm<br>15cm x 15cm<br>15cm x 20cm | 10cm x 20cm<br>10cm x 30cm<br>10cm x 40cm<br>15cm x 15cm<br>15cm x 20cm<br>15cm x 30cm |
| | | 20cm x 20cm |
| | | 20cm x 25cm |
| | | 25cm x 25cm |
| | Fixation Strips | Fixation Strips |
| | Carrying Case | Carrying Case |
| | | |
| Indications for Use | | |
| | PICO is indicated for patients who<br>would benefit from a suction device<br>(negative pressure wound therapy)<br>as it may promote wound healing<br>via removal of low to moderate<br>levels of exudate and infectious<br>materials. Appropriate wound types<br>include:<br>- Chronic<br>- Acute<br>- Traumatic<br>- Subacute and dehisced wounds<br>- Partial-thickness burns<br>- Ulcers (such as diabetic or<br>pressure)<br>- Flaps and grafts<br>- Closed surgical incisions<br>PICO Single Use Negative<br>Pressure Wound Therapy System<br>is suitable for use in both hospital<br>and homecare setting | UNO is indicated for use in patients<br>who would benefit from negative<br>pressure wound therapy particularly as<br>the device may promote wound healing<br>by the removal of low to moderate<br>exudates and infectious material.<br>Appropriate wound types include:<br>- Chronic<br>- Acute<br>- Traumatic<br>- Subacute and dehisced wounds<br>- Partial-thickness burns<br>- Ulcers (such as diabetic or pressure)<br>- Flaps and grafts<br>Genadyne UNO is a single patient<br>use device. |
| | | |
| Contraindications | | |
| - | The Pico is contraindicated in the<br>presence of : | The Genadyne UNO is<br>contraindicated in the presence of: |
| - | Necrotic tissue with Eschar present | Necrotic tissue with Eschar present |
| - | Previously confirmed and untreated<br>osteomyelitis. | Untreated osteomyelitis |
| - | malignancy in the wound bed or<br>margins of the wound (except in<br>palliative care to enhance quality of<br>life) | Malignancy (with exception to<br>enhance quality of life) |
| - | Exposed arteries, veins, or organs | Exposed arteries, veins, or organs |
| - | Non-enteric and unexplored fistulas | Non-enteric and unexplored fistulas |
| - | Anastomotic sites | Anastomotic sites |
| - | Emergency airway aspiration | Emergency airway aspiration |
| - | Pleural, mediastinal or chest tube<br>drainage | Pleural, mediastinal or chest tube<br>drainage |
| - | Surgical suction | Surgical suction |
| | | |
| Compliance | | |
| | IEC 60601-1 | IEC 60601-1 |
| | IEC 60601-1-2 | JEC 60601-1-2 |
| UL 60601-1 | | |
| Can/CSA C22.2 | | |
| Storage / Transport | 5°C to +25°C (41°F to 77°F) | -18°C to +43°C (0°F to 110°F) |
| | Relative Humidity 10% to 75 % | Relative Humidity 10% to 95 % |
| | 700 - 1060 mbar Atmospheric pressure | 700 – 1060 mbar Atmospheric pressure |
| Operation | 5°C to 35°C (41°F to 95°F) | 18°C to 34°C (65°F to 94°F) |
| | Relative Humidity 10% to 95 % | Relative Humidity 10% to 95 % |
| | 700 - 1060 mbar Atmospheric pressure | 700 - 1060 mbar Atmospheric pressure |
{5}------------------------------------------------
{6}------------------------------------------------
- 15. Discussion of nonclinical and clinical testing
It appears that the non-clinical results are all showing that it is safe for patient use. The dressing kit, which contains a silicone foam dressing, 4 PU film strips, met the acceptance criteria for biocompatibility testing, and the pump has met the acceptance criteria for all applicable IEC testing. Bench testing was done with in house protocol to establish ensure the performance and outcome of is expected when using together with the device as a system.
The Genadyne UNO has similar design characteristics and provides similar functions to the PICO system. The intended use, indications and instructions for use for the subject and predicate devices are similar. The dressing materials and features are all similar as well. The Genadyne UNO does not raise any new issues of safety and effectiveness. Therefore, it is in our conclusion that the Genadyne UNO is substantially equivalent to the predicate device.
- 16. Conclusion & Determination of Substantial Equivalence
Based on the information presented above, it is concluded that the UNO is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.