K161573 · Maine Molecular Quality Controls, Inc. · PMN · Sep 2, 2016 · Microbiology
Device Facts
Record ID
K161573
Device Name
FilmArray RP EZ Control Panel M265
Applicant
Maine Molecular Quality Controls, Inc.
Product Code
PMN · Microbiology
Decision Date
Sep 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3920
Device Class
Class 2
Indications for Use
FilmArray RP EZ Control Panel M265 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/ Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae on the FilmArray RP EZ assay performed on the FilmArray systems. FilmArray RP Positive Control is composed of synthetic RNA transcripts specifically designed for and intended to be used solely with the FilmArray RP EZ assay. This product is not intended to replace manufacturer controls provided with the device.
Device Story
FilmArray RP EZ Control Panel M265 provides external positive and negative quality control materials for the FilmArray RP EZ assay. Positive control consists of synthetic RNA transcripts representing all respiratory pathogens detected by the assay; negative control consists of non-specific RNA in buffer. Controls are processed by laboratory personnel identically to patient nasopharyngeal swab samples in Viral Transport Media. The device is used on the FilmArray 2.0 instrument to monitor qualitative amplification, detection, and identification steps. Results are reviewed by clinicians to verify assay performance and ensure validity of patient test results.
Clinical Evidence
No clinical data. Bench testing only. Reproducibility study (3 sites, 182 controls) showed 100% agreement for positive controls and 97.78% for negative controls. Internal precision and lot-to-lot studies confirmed consistent performance across operators, lots, and days, with stable Crossing Point (Cp) values.
Technological Characteristics
Ready-to-use liquid control panel. Positive control: synthetic RNA transcripts in buffers, stabilizers, and preservatives. Negative control: non-specific RNA in buffers, stabilizers, and preservatives. Designed for use with FilmArray 2.0 EZ Configuration system. Monitors extraction, reverse transcription, amplification, detection, and identification steps.
Indications for Use
Indicated for use as external positive and negative quality control material for in vitro nucleic acid testing of respiratory pathogens (Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A/B, Parainfluenza, RSV, B. pertussis, C. pneumoniae, M. pneumoniae) using the FilmArray RP EZ assay. Intended for clinical laboratory use.
Regulatory Classification
Identification
An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.
Special Controls
An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K161573
B. Purpose for Submission:
To obtain a substantial equivalence determination for the FilmArray RP EZ Control Panel M265.
C. Measurand:
Multi-analyte quality control materials
D. Type of Test:
FilmArray RP EZ Control Panel M265 is intended for *in vitro* diagnostic use as external assayed quality control materials to monitor the qualitative amplification, detection and identification steps of the laboratory nucleic acid test, FilmArray RP EZ assay on the FilmArray 2.0 instrument, which detects respiratory pathogens: Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/ Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, *Bordetella pertussis*, *Chlamydophila pneumoniae*, and *Mycoplasma pneumoniae*.
E. Applicant:
Maine Molecular Quality Controls, Inc. (MMQCI)
F. Proprietary and Established Names:
FilmArray RP EZ Control Panel M265
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3920, Assayed quality control material for clinical microbiology assays
2. Classification:
Class II (Special Controls)
{1}
3. Product code:
PMN
4. Panel:
83- Microbiology
H. Indication(s) for use:
1. Indications for use(s):
FilmArray RP EZ Control Panel M265 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/ Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumoniae on the FilmArray RP EZ assay performed on the FilmArray systems. FilmArray RP Positive Control is composed of synthetic RNA transcripts specifically designed for and intended to be used solely with the FilmArray RP EZ assay. This product is not intended to replace manufacturer controls provided with the device.
2. Special conditions for use statement(s):
For in vitro diagnostic use only
For prescription use only
3. Special instrument requirements:
The FilmArray RP EZ Control Panel M265 was evaluated on FilmArray 2.0 instrument.
I. Device Description:
FilmArray RP EZ Control Panel M265 is a quality control panel consisting of two controls: FilmArray RP EZ Positive Control and FilmArray RP EZ Negative Control. The Positive Control contains non-infectious surrogate control material; a solution of synthetic RNA in buffers, stabilizers and preservatives. The RNA carries segments of all respiratory pathogens detected by the FilmArray RP EZ assay (see Table below) on the FilmArray instrument 2.0 EZ Configuration system. The RNA in the Negative Control is non-specific RNA in buffers, stabilizers and preservatives. Each liquid control of FilmArray RP EZ Control Panel M265 is processed separately according to FilmArray RP EZ assay manufacturer's Instructions for Use for patient samples (nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infection and placed in Viral Transport Media (VTM)).
2
{2}
The FilmArray RP EZ Positive Control M265 is prepared nucleic acid concentrations of 3X-10X LoD for each of the organisms detected by the FilmArray RP EZ assay.
| Respiratory pathogens detected by FilmArray RP EZ assay | |
| --- | --- |
| Adenovirus | Influenza A H1-2009 |
| Coronavirus | Influenza B |
| Human Metapneumovirus | Parainfluenza Virus |
| Human Rhinovirus/ Enterovirus | Respiratory Syncytial Virus |
| Influenza A | Bordetella pertussis |
| Influenza A subtype H1 | Chlamydophila pneumoniae |
| Influenza A subtype H3 | Mycoplasma pneumoniae |
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Amplichek II, Bio-Rad Laboratories
2. Predicate 510(k) number(s):
DEN150058
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | FilmArray RP EZ Control Panel M265 (K161573) | Amplichek II (DEN150058) |
{3}
| Intended Use | FilmArray RP EZ Control Panel M265 is intended for use as external positive and negative, surrogate assayed quality control materials to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Adenovirus, Coronavirus, Human Metapneumovirus, Human Rhinovirus/Enterovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype H1-2009, Influenza B, Parainfluenza Virus, Respiratory Syncytial Virus, Bordetella pertussis, Chlamydophila pneumoniae, and Mycoplasma pneumonia on the FilmArray RP EZ assay performed on the FilmArray systems. The control panel also contains a negative control. This product is not intended to replace manufacturer controls provided with the device. | Amplichek II is intended for use as an external assayed quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Methicillin Resistant Staphylococcus aureus, Methicillin Sensitive Staphylococcus aureus, Clostridium difficile and Vancomycin-resistant Enterococci performed on Cepheid GeneXpert Systems. This product is not intended to replace manufacturer controls provided with the device. This product is only for use with assays and instruments listed in the Representative Results Chart in this labeling. |
| --- | --- | --- |
| Similarities | | |
| --- | --- | --- |
| Item | FilmArray RP EZ Control Panel M265 | Amplichek II (DEN150058) |
| Directions for Use | Process like patient sample Same | Same |
| Number of targets monitored in one assay | Multiplex | Same |
| Differences | | |
| --- | --- | --- |
| Item | FilmArray RP EZ Control Panel M265 (K161573) | Amplichek II (DEN150058) |
| Composition | Synthetic RNA | Intact microorganisms |
| Assay Steps Monitored | Same except does not monitor extraction | Extraction, reverse transcription, amplification, detection, identification |
# K. Standard/Guidance Document Referenced (if applicable):
None were referenced
{4}
L. Test Principle:
Not applicable
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Reproducibility:
A multi-site reproducibility study was performed with the FilmArray RP EZ Control Panel M265 on FilmArray instrument 2.0 EZ Configuration system at two CLIA-waived clinical sites and one simulated CLIA-waived test site (e.g., testing was performed at a laboratory with users that were screened to have training and educational backgrounds consistent with those in the CLIA-waived testing environment). Testing consisted of three positive and three negative controls run per day, and spanned a period of 10 days (total of 60 control runs per site). Multiple operators participated in testing at each of the three locations. Three lots each of external control material and three lots of RP EZ pouches were tested across all sites. A total of 182 external controls were tested (91 positive and 91 negative). Two tests were excluded from the final data analysis due to Invalid results caused by internal pouch control failures. The results are shown in the tables below:
| FilmArray RP EZ Control Panel Summary of External Reproducibility Test Results | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Category | SITE | | | | | | All Sites | All Sites |
| | Site #1 | | Site #2 | | Site #3 | | Overall Percent Agreement | 95% Confidence Interval |
| | #expected results/# tested1 | % Agreement with Expected Result1 | #expected results/# tested | % Agreement with Expected Result1 | #expected results/# tested | % Agreement with Expected Result1 | | |
| FilmArray RP EZ Positive Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 100% 90/90 | 95.9% to 100% |
| FilmArray RP EZ Negative Control | 30/30 | 100% | 30/30 | 100% | 28/303 | 93.3% | 97.78% 88/90 | 92.3% to 99.4% |
1 Expected result for the FilmArray RP EZ Positive Control is positive. Expected result for the FilmArray RP EZ Negative Control is negative.
2 The 2 Invalid samples were re-tested according to BioFire instructions and were not included in the Percent Correct analysis.
3 Two false positives were observed with the FilmArray RP EZ Negative Control at site #3. The detected organisms were 1) Coronavirus CoV-NL63, 2) Coronavirus CoV-229E. The FilmArray RP EZ assay only reports Coronavirus as Detected or Not Detected.
{5}
| Crossing Point (Cp) data for FilmArray RP EZ Control Panel External Reproducibility Study | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Respiratory Pathogen Analyte | Site | | Site 2 | | Site | | All Sites | |
| | Mean Cp | SD | Mean Cp | SD | Mean Cp | SD | Mean Cp | SD |
| Adeno | 22.6 | 1.2 | 22.0 | 1.3 | 22.2 | 1.2 | 22.3 | 1.2 |
| Adeno2 | 21.0 | 1.1 | 20.6 | 1.2 | 20.7 | 1.0 | 20.8 | 1.1 |
| CoV-229E¹ | 20.9 | 0.7 | 20.5 | 1.1 | 20.5 | 0.8 | 20.6 | 0.9 |
| CoV-HKU1¹ | 19.8 | 1.1 | 19.3 | 1.1 | 19.4 | 0.8 | 19.5 | 1.0 |
| CoV-NL63¹ | 19.6 | 1.5 | 18.8 | 1.0 | 18.9 | 0.7 | 19.1 | 1.1 |
| CoV-OC43¹ | 19.4 | 0.8 | 19.0 | 1.2 | 18.9 | 0.6 | 19.1 | 0.9 |
| hMPV | 21.8 | 1.3 | 21.4 | 1.4 | 21.2 | 1.0 | 21.5 | 1.2 |
| Entero1 | 20.4 | 1.0 | 20.0 | 1.1 | 20.0 | 0.8 | 20.1 | 1.0 |
| Entero2 | 20.4 | 1.2 | 20.0 | 1.1 | 20.0 | 0.8 | 20.1 | 1.0 |
| HRV1 | 17.5 | 0.7 | 17.2 | 1.1 | 16.9 | 0.7 | 17.2 | 0.8 |
| HRV2 | 18.0 | 0.7 | 17.7 | 1.0 | 17.4 | 0.7 | 17.7 | 0.8 |
| HRV3 | 22.7 | 0.9 | 22.6 | 1.3 | 21.8 | 0.7 | 22.4 | 1.0 |
| HRV4 | 19.5 | 0.9 | 19.2 | 1.0 | 18.9 | 0.6 | 19.2 | 0.8 |
| FluA-H1-2009 | 18.7 | 1.2 | 18.2 | 1.1 | 18.2 | 0.8 | 18.4 | 1.0 |
| FluA-H1-pan | 18.8 | 1.1 | 18.3 | 1.1 | 18.4 | 0.8 | 18.5 | 1.0 |
| FluA-H3 | 18.1 | 0.8 | 17.7 | 1.2 | 17.6 | 0.9 | 17.8 | 1.0 |
| FluA-pan1 | 19.7 | 0.8 | 19.2 | 1.1 | 19.1 | 0.7 | 19.3 | 0.9 |
| FluA-pan2 | 20.0 | 1.0 | 19.5 | 1.1 | 19.5 | 0.6 | 19.7 | 0.9 |
| FluB | 19.3 | 0.7 | 18.9 | 1.2 | 18.7 | 0.7 | 19.0 | 0.9 |
| PIV1 | 20.5 | 1.4 | 19.7 | 1.3 | 19.9 | 0.9 | 20.0 | 1.2 |
| PIV2 | 20.1 | 0.9 | 19.8 | 1.1 | 19.6 | 0.6 | 19.8 | 0.9 |
| PIV3 | 18.8 | 0.9 | 18.4 | 1.1 | 18.2 | 0.6 | 18.5 | 1.1 |
| PIV4 | 19.2 | 1.2 | 18.6 | 1.3 | 18.5 | 0.8 | 18.8 | 1.1 |
| RSV | 18.8 | 0.8 | 18.4 | 1.2 | 18.1 | 0.7 | 18.4 | 0.9 |
| Bper | 20.9 | 0.8 | 20.5 | 1.0 | 20.5 | 0.7 | 20.6 | 0.8 |
| Cpne | 19.0 | 0.8 | 18.5 | 1.2 | 18.4 | 0.7 | 18.6 | 0.9 |
| Mpne | 19.5 | 1.2 | 18.7 | 1.2 | 18.6 | 0.7 | 18.9 | 1.0 |
The results suggest that there are no significant differences between different users and different sites on different days. External reproducibility studies for the FilmArray RP EZ Control Panel M265 are acceptable.
## b. Precision:
An internal precision study for the FilmArray RP EZ Control Panel M265 was conducted over multiple days by testing three FilmArray RP EZ Control Panel lots with four FilmArray RP EZ assay lots performed by one operator. The results are shown in the table below:
{6}
7
| FilmArray RP EZ Control Panel Summary of Internal Precision Test Results | | | | |
| --- | --- | --- | --- | --- |
| Category | Operator #1 | | Combined | |
| | #expected results/#tested¹ | % Agreement with Expected Result¹ | Overall Percent Agreement | 95% Confidence Interval |
| FilmArray RP EZ Positive Control | 39/39 | 100% | 100% | 91% to 100% |
| FilmArray RP EZ Negative Control | 49/49 | 100% | 100% | 92.7% to 100% |
| Crossing Point (Cp) data for FilmArray RP EZ Control Panel External Precision Study | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Respiratory Pathogen Analyte | Control Lot#: M40DEC15A | | Control Lot#: A28DEC15A | | Control Lot#: A06JAN16A | | All Lots | |
| | Mean Cp | SD | Mean Cp | SD | Mean Cp | SD | Mean Cp | SD |
| Adeno | 22.1 | 1.1 | 21.9 | 1.4 | 21.6 | 0.6 | 21.4 | 0.5 |
| Adeno2 | 20.2 | 0.7 | 19.9 | 1.1 | 19.7 | 0.4 | 19.5 | 0.3 |
| CoV-229E¹ | 19.7 | 0.6 | 19.5 | 0.6 | 19.4 | 0.5 | 19.2 | 0.1 |
| CoV-HKU1¹ | 18.5 | 0.7 | 18.4 | 0.6 | 18.2 | 0.4 | 18.0 | 0.2 |
| CoV-NL63¹ | 18.1 | 0.4 | 18.0 | 0.5 | 17.9 | 0.4 | 17.8 | 0.3 |
| CoV-OC43¹ | 18.1 | 0.3 | 17.9 | 0.5 | 17.9 | 0.4 | 17.8 | 0.4 |
| hMPV | 21.4 | 0.8 | 21.0 | 1.0 | 21.1 | 0.4 | 21.1 | 0.4 |
| Entero1 | 19.1 | 0.7 | 19.1 | 0.8 | 18.9 | 0.4 | 19.1 | 0.3 |
| Entero2 | 19.0 | 0.6 | 19.0 | 0.8 | 18.9 | 0.4 | 19.1 | 0.3 |
| HRV1 | 16.2 | 0.1 | 16.0 | 0.4 | 16.0 | 0.4 | 16.0 | 0.3 |
| HRV2 | 17.1 | 0.2 | 16.8 | 0.4 | 16.9 | 0.4 | 16.8 | 0.3 |
| HRV3 | 24.9 | 1.6 | 23.6 | 1.9 | 23.8 | 1.7 | 22.9 | 0.5 |
| HRV4 | 19.9 | 0.5 | 19.6 | 0.7 | 19.9 | 0.4 | 20.0 | 0.4 |
| FluA-H1-2009 | 17.2 | 0.6 | 17.1 | 0.6 | 16.8 | 0.4 | 16.6 | 0.3 |
| FluA-H1-pan | 18.1 | 0.8 | 17.8 | 0.8 | 17.5 | 0.5 | 17.3 | 0.4 |
| FluA-H3 | 17.6 | 0.5 | 17.2 | 0.6 | 17.3 | 0.3 | 17.2 | 0.3 |
| FluA-pan1 | 18.5 | 0.4 | 18.2 | 0.4 | 18.4 | 0.4 | 18.5 | 0.4 |
| FluA-pan2 | 19.0 | 0.5 | 18.7 | 0.4 | 18.6 | 0.3 | 18.5 | 0.3 |
| FluB | 18.1 | 0.4 | 17.9 | 0.4 | 17.8 | 0.5 | 17.6 | 0.4 |
| PIV1 | 18.8 | 0.7 | 18.8 | 0.8 | 18.3 | 0.3 | 18.2 | 0.3 |
| PIV2 | 19.0 | 0.3 | 18.8 | 0.4 | 18.8 | 0.6 | 18.4 | 0.4 |
| PIV3 | 17.9 | 0.5 | 17.4 | 0.5 | 17.6 | 0.4 | 17.5 | 0.4 |
| PIV4 | 19.2 | 0.7 | 18.7 | 0.9 | 18.9 | 0.5 | 18.9 | 0.7 |
| RSV | 17.6 | 0.2 | 17.4 | 0.5 | 17.5 | 0.4 | 17.4 | 0.4 |
| Bper | 19.9 | 0.4 | 19.7 | 0.8 | 19.7 | 0.4 | 19.6 | 0.4 |
| Cpne | 17.6 | 0.2 | 17.4 | 0.3 | 17.6 | 0.4 | 17.5 | 0.4 |
| Mpne | 17.8 | 0.2 | 17.6 | 0.4 | 17.7 | 0.5 | 17.6 | 0.4 |
{7}
There appears to be no significant differences in mean Cp value when testing different control lots on different days. Precision studies are acceptable.
c. Within-run Testing:
Within-run reproducibility was demonstrated in a separate study conducted by one operator testing two lots of FilmArray RP EZ Control Panel M265 with one lot of FilmArray RP EZ pouches on the FilmArray 2.0, each within one day. The results are shown in the tables below:
8
{8}
| FilmArray RP EZ Control Panel Summary of Within-run Reproducibility Results | | | | |
| --- | --- | --- | --- | --- |
| Control | Control Lot# | No. of Tests | Pouch Lot | Correct Results |
| FilmArray RP EZ Positive Control | A06JAN16A | 6 | 239115 | 6/6 (100%) |
| FilmArray RP EZ Negative Control | M41DEC15A | 5 | 239115 | 5/5 (100%) |
| Crossing Point (Cp) data for FilmArray RP EZ Control Panel Within-run Reproducibility | | |
| --- | --- | --- |
| Respiratory Pathogen Analyte | FilmArray RP EZ Positive Control (Lot A06JAN16A) | |
| | Mean Cp | SD |
| Adeno | 21.4 | 0.5 |
| Adeno2 | 19.5 | 0.3 |
| CoV-229E1 | 19.2 | 0.1 |
| CoV-HKU11 | 18.0 | 0.2 |
| CoV-NL631 | 17.8 | 0.3 |
| CoV-OC431 | 17.8 | 0.4 |
| hMPV | 21.1 | 0.4 |
| Entero1 | 19.1 | 0.3 |
| Entero2 | 19.1 | 0.3 |
| HRV1 | 16.0 | 0.3 |
| HRV2 | 16.8 | 0.3 |
| HRV3 | 22.9 | 0.5 |
| HRV4 | 20.0 | 0.4 |
| FluA-H1-2009 | 16.6 | 0.3 |
| FluA-H1-pan | 17.3 | 0.4 |
| FluA-H3 | 17.2 | 0.3 |
| FluA-pan1 | 18.5 | 0.4 |
| FluA-pan2 | 18.5 | 0.3 |
| FluB | 17.6 | 0.4 |
| PIV1 | 18.2 | 0.3 |
| PIV2 | 18.4 | 0.4 |
| PIV3 | 17.5 | 0.4 |
| PIV4 | 18.9 | 0.7 |
| RSV | 17.4 | 0.4 |
| Bper | 19.6 | 0.4 |
| Cpne | 17.5 | 0.4 |
| Mpne | 17.6 | 0.4 |
{9}
Within-run reproducibility studies for the FilmArray RP EZ Control Panel M265 are acceptable.
# d. Lot-to-Lot Testing:
Lot-to-lot reproducibility was demonstrated by testing three lots of FilmArray RP EZ Positive Control using the same pouch lot. Results are shown in the following tables.
| FilmArray RP EZ Control Panel Summary of Lot-to-Lot Reproducibility Results | | | |
| --- | --- | --- | --- |
| Control Lot # | Number of Tests | Pouch Lot | Correct Results |
| M40DEC15A | 4 | 238915 | 4/4 (100%) |
| A28DEC15A | 3 | 238915 | 3/3 (100%) |
| A06JAN16A | 5 (1 Invalid) | 238915 | 4/4 (100%) |
| Crossing Point (Cp) data for FilmArray RP EZ Control Panel Lot-to-Lot Reproducibility | | | | | |
| --- | --- | --- | --- | --- | --- |
| Respiratory Pathogen Analyte | M40DEC15A | A28DEC15A | A06JAN16A | All Lots | All Lots |
| | Mean Cp | Mean Cp | Mean Cp | Mean Cp | SD |
| Adeno | 21.6 | 20.8 | 21.7 | 21.4 | 0.5 |
| Adeno2 | 19.8 | 18.8 | 19.7 | 19.5 | 0.6 |
| CoV-229E1 | 19.3 | 18.8 | 19.4 | 19.2 | 0.4 |
| CoV-HKU11 | 18.1 | 17.7 | 18.4 | 18.1 | 0.4 |
| CoV-NL631 | 18.1 | 17.7 | 18.1 | 18.0 | 0.4 |
| CoV-OC431 | 18.3 | 17.5 | 18.2 | 18.0 | 0.5 |
| hMPV | 21.1 | 20.1 | 20.8 | 20.7 | 0.6 |
| Entero1 | 18.4 | 18.0 | 18.5 | 18.3 | 0.4 |
| Entero2 | 18.4 | 18.0 | 18.5 | 18.3 | 0.4 |
| HRV1 | 16.2 | 15.8 | 16.2 | 16.1 | 0.4 |
| HRV2 | 17.1 | 16.6 | 17.2 | 17.0 | 0.4 |
| HRV3 | 25.6 | 24.2 | 25.3 | 25.1 | 1.9 |
| HRV4 | 19.6 | 18.9 | 19.6 | 19.4 | 0.5 |
| FluA-H1-2009 | 16.9 | 16.4 | 16.8 | 16.7 | 0.3 |
| FluA-H1-pan | 17.8 | 17.1 | 17.7 | 17.5 | 0.4 |
| FluA-H3 | 17.4 | 16.6 | 17.4 | 17.2 | 0.5 |
| FluA-pan1 | 18.5 | 17.8 | 18.4 | 18.3 | 0.4 |
| FluA-pan2 | 18.7 | 18.3 | 18.6 | 18.5 | 0.3 |
| FluB | 17.9 | 17.4 | 18.0 | 17.8 | 0.4 |
| PIV1 | 18.5 | 18.0 | 18.3 | 18.3 | 0.4 |
| PIV2 | 19.1 | 18.4 | 19.3 | 19.0 | 0.5 |
| PIV3 | 17.8 | 17.1 | 17.8 | 17.6 | 0.4 |
| PIV4 | 19.2 | 17.9 | 18.8 | 18.7 | 0.7 |
| RSV | 17.7 | 17.1 | 17.7 | 17.6 | 0.4 |
| Bper | 19.7 | 19.1 | 19.8 | 19.6 | 0.5 |
| Bper2 | 19.7 | 19.1 | 19.8 | 19.6 | 0.5 |
| Bper3 | 19.7 | 19.1 | 19.8 | 19.6 | 0.5 |
| Bper4 | 19.7 | 19.1 | 19.8 | 19.6 | 0.5 |
{10}
| Crossing Point (Cp) data for FilmArray RP EZ Control Panel Lot-to-Lot Reproducibility | | | | | |
| --- | --- | --- | --- | --- | --- |
| Respiratory Pathogen Analyte | M40DEC15A | A28DEC15A | A06JAN16A | All Lots | All Lots |
| | Mean Cp | Mean Cp | Mean Cp | Mean Cp | SD |
| Cpne | 17.6 | 17.1 | 17.8 | 17.5 | 0.4 |
| Mpne | 17.9 | 17.3 | 18.0 | 17.8 | 0.5 |
Lot-to-Lot reproducibility studies for the FilmArray RP EZ Control Panel M265 are acceptable.
e. Linearity/assay reportable range:
Not applicable
f. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Not applicable
Stability:
Open Vial Stability: Not applicable because FilmArray RP EZ Control Panel M265 is packaged for single use.
Closed Vial Real-time Stability: An accelerated stability study was performed to establish the shelf life stability claims for FilmArray RP EZ Control Panel M265. Based on this study, the FilmArray RP EZ Control Panel M265 is expected to be stable until the expiration date (12 months) when stored frozen $(-20^{\circ}\mathrm{C}$ or colder) and unopened. This product is for single use.
Real-Time Stability Program: Real-time stability studies are ongoing to support product claims and to monitor potential assay modifications for which the FilmArray RP EZ Control Panel M265 is indicated for use. Real-time stability study protocols and acceptance criteria were reviewed and found to be acceptable.
Shipping Stability: MMQCI ships the FilmArray RP EZ Control Panel on dry ice with overnight delivery, ensuring that the control material remains frozen upon receipt. The frozen control material is then to be stored at $-20^{\circ}\mathrm{C}$ , as indicated in the FilmArray RP EZ Control Panel Package insert. A shipping study was performed to confirm the shipping process and to investigate the outcome of a possible shipping delay and subsequent arrival with no dry ice. The study evaluated two (2) lots of the FilmArray RP EZ Positive Control placed in dry ice and was stored for two (2) days at ambient temperature then tested using the FilmArray RP EZ Assay. To simulate a shipping delay, two (2) lots of the FilmArray RP EZ Positive Control M266 were
{11}
stored for up to 6 days at ambient temperature and tested using the FilmArray RP EZ Assay.
The study demonstrated that the FilmArray RP EZ Positive Control M265 is stable for two (2) days on dry ice in MMQCI's standard shipping box. FilmArray RP EZ Positive Control is stable after five (5) days at ambient temperatures of approximately $19 - 21^{\circ}\mathrm{C}$ . The FilmArray RP EZ Positive Control should be stored frozen $(-20^{\circ}\mathrm{C}$ or colder) as indicated in the package insert.
# Expected Values:
FilmArray RP EZ Control Panel M265 is a qualitative control and the expected results are listed in the tables below.
FilmArray RP EZ Positive Control FilmArray Result Summary
| Multiple Organisms Detected. Influenza A – Multiple Subtypes Detected | |
| --- | --- |
| Results Summary | |
| Detected | Not Detected |
| Adenovirus | |
| Coronavirus | |
| Human Metapneumovirus Human Rhinovirus | |
| Enterovirus | |
| Influenza A H1-2009 | |
| Influenza A H3 | |
| Influenza B Parainfluenza Virus | |
| Respiratory Syncytial Virus | |
| Bordetella pertussis | |
| Chlamydophila pneumoniae | |
| Mycoplasma pneumoniae | |
{12}
13
FilmArray RP EZ Negative Control FilmArray Result Summary
| Negative | |
| --- | --- |
| Results Summary | |
| Detected | Not Detected |
| | Adenovirus
Coronavirus
Human Metapneumovirus
Human Rhinovirus
Enterovirus Influenza A
Influenza B
Parainfluenza Virus
Respiratory Syncytial Virus
Bordetella pertussis
Chlamydophila pneumoniae
Mycoplasma pneumoniae |
Matrix Effects:
A study was performed to evaluate the effect of the FilmArray RP EZ Control Panel M265 in the presence of viral transport media (VTM). Equal volumes of the same concentration of inactivated Influenza A H1N1-2009 were spiked into 300μL of VTM as well as 300μL of FilmArray RP EZ Negative Control, which contains the identical matrix found in the FilmArray RP EZ Positive Control. Both sample types of the H1N1-2009 were tested in triplicate by the FilmArray RP EZ assay.
Results demonstrated that samples prepared with FilmArray RP EZ Control Panel M265 matrix generate equivalent results to samples prepared with VTM. Study results are shown in the table below:
| Respiratory Pathogen Analyte | Spiked BEI H1N1-2009 Crossing Point (Cp) Data | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | VTM | | | Mean Cp | MMQCI Matrix | | | Mean Cp |
| FluA-H1-2009 | 16.9 | 17.2 | 17.2 | 17.1 | 17.4 | 16.3 | 16.5 | 16.7 |
| FluA-H1-pan | 18.1 | 18.5 | 18.2 | 18.3 | 18.7 | 17.5 | 17.4 | 17.8 |
| FluA-pan1 | 13.9 | 13.6 | 14.5 | 14.0 | 12.4 | 12.8 | 12.7 | 12.6 |
| FluA-pan2 | 19.1 | 19.3 | 19.6 | 19.4 | 18.9 | 18.2 | 18.6 | 18.5 |
{13}
g. Detection limit:
Not applicable
h. Analytical Reactivity (Inclusivity):
Not applicable
i. Cross Reactivity:
Not applicable
j. Interference:
Not applicable
k. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
3. Clinical Studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not Applicable.
14
{14}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809 and the specials controls for this device type.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.