← Product Code [FMF](/productcode/FMF) · K161568

# Bone Solutions Mixing and Delivery System (K161568)

_Bone Solutions, Inc. · FMF · Sep 16, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161568

## Device Facts

- **Applicant:** Bone Solutions, Inc.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Sep 16, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Bone Solutions Mixing and Delivery System is intended to be used for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site. The Bone Solutions Mixing and Delivery System is intended to be used in a controlled operating room environment with compatible devices by qualified medical personnel.

## Device Story

Bone Solutions Mixing and Delivery System is a disposable piston syringe system; includes syringe barrel with female luer, plunger, and funnel. Used in orthopedic surgery to mix and deliver bone graft materials (allograft, autograft, or synthetic) to surgical sites. Operated by qualified medical personnel in controlled operating room environments. Device is provided empty and sterile for single use. Facilitates manual mixing and delivery of graft material; provides surgeons with a controlled method for graft application. Benefits include standardized delivery mechanism for bone graft materials.

## Clinical Evidence

No clinical data. Substantial equivalence is based on design, material, and intended use comparison to the predicate device.

## Technological Characteristics

Disposable piston syringe system; 14 mL syringe barrel with female luer, plunger, and funnel. Materials are identical to predicate device. Sterile, single-use. No electronic components or software.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- ETEX Mixing and Delivery System ([K141245](/device/K141245.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, arranged in a cascading manner, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 16, 2016

Bone Solutions, Inc. Mr. Drew Diaz President - CEO 5712 Colleyville Blvd. Suite 210 Collevville. Texas 76034

Re: K161568

Trade/Device Name: Bone Solutions Mixing and Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: August 31, 2016 Received: September 2, 2016

Dear Mr. Diaz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth 

{1}------------------------------------------------

in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Christopher J. Ronk-S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K161568

Device Name

Bone Solutions Mixing and Delivery System

Indications for Use (Describe)

The Bone Solutions Mixing and Delivery System is intended to be used for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 5 510(k) Summary

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Bone Solutions, Inc. is hereby submitting this 510(k) summary.

# Submitter [510(k) owner]

Bone Solutions, Inc.

5712 Colleyville Blvd., Suite 210

Colleyville, TX 76034

## Company Contact

Drew Diaz President-CEO P 817.809.8850 diaz@bonesolutions.net

## Submitted Device Information

Trade Name: Bone Solutions Mixing and Delivery System Common Name: Piston Syringe Classification Name: Syringe, Piston, General Hospital

## Classification Information

Classification: Class II Classification Regulation: 21 CFR 880.5860 Classification Product Code: FMF

## Legally Marketed Predicate Devices

The Bone Solutions Mixing and Delivery System manufactured by Bone Solutions, Inc. (BSI) is substantially equivalent to the following device currently in commercial use:

| Device:        | ETEX Mixing and Delivery System                         |
|----------------|---------------------------------------------------------|
| Manufacturer:  | ETEX Corporation                                        |
| Address:       | 675 Massachusetts Ave., 12th Floor, Cambridge, MA 02139 |
| 510(k) number: | K141245                                                 |

#### Submitted Device Description

The Bone Solutions Mixing and Delivery System is commercially available disposable medical piston syringe (syringe barrel with female luer, plunger) to facilitate mixing and a funnel to facilitate filling of the syringe barrel. The system with a 14 mL syringe barrel and will be provided empty.

#### Intended Use

The Bone Solutions Mixing and Delivery System, like the predicate, is intended to provide the surgeons with a means to mix and deliver graft material to an orthopedic surgical site.

{4}------------------------------------------------

The Bone Solutions Mixing and Delivery System is intended to be used in a controlled operating room environment with compatible devices by qualified medical personnel. The device is sterile, single use, with a 3 year expected shelf life.

#### Indications for Use:

The Bone Solutions Mixing and Delivery System is intended to be used for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site.

#### Substantial Equivalence

The Bone Solutions Mixing and Delivery System is substantially equivalent to the predicate device, in which the basic features and intended uses are the same. The identical mixing syringes, made of identical materials, have been provided to both manufacturers by the same qualified supplier, as verified in the design history file. The lack of gradation markings on the BSI syringe does not change the performance or biocompatibility compared to the predicate Additions from the accessory tray are considered minor and do not raise questions concerning safety and effectiveness.

The Bone Solutions Mixing and Delivery System is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the ETEX Mixing and Delivery System, and raises no new issues of safety or effectiveness.

Drew Diaz President-CEO 5712 Colleyville Blvd., Suite 210 Colleyville, TX 76034 P 817.809.8850

Date: August 31, 2016

---

**Source:** [https://fda-staging.innolitics.com/device/K161568](https://fda-staging.innolitics.com/device/K161568)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com