K161566 · Daehan Medical Systems Co., Ltd. · GXZ · Dec 21, 2016 · Neurology
Device Facts
Record ID
K161566
Device Name
DMS Disposable Subdermal Needle Electrodes
Applicant
Daehan Medical Systems Co., Ltd.
Product Code
GXZ · Neurology
Decision Date
Dec 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1350
Device Class
Class 2
Indications for Use
The DMS Disposable Subdermal Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyograph (EMG) and nerve potential signals, and are intended for stimulation/recording with stimulation/recording equipment for electromyograph (EMG) and nerve potential signals.
Device Story
DMS Disposable Subdermal Needle Electrodes are sterile, single-use stainless steel needles; used to record biopotential signals (EEG, EMG, nerve potentials) or provide stimulation during neurophysiological monitoring. Needles feature touch-proof connectors (DIN 42 802 1.5mm) for interface with external recording/monitoring equipment. Operated by clinicians in clinical settings; device acts as a passive interface between patient tissue and diagnostic equipment. Benefits include standardized, disposable access to subdermal sites for neurological assessment, reducing cross-contamination risk.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization), performance testing (stiffness, breakage, bond, electrical properties), and shelf-life testing per ASTM F1980-07.
Technological Characteristics
Needle material: stainless steel; Lead wire: PVC insulated tin-plated copper; Connectors: DIN 42 802 1.5mm touch-proof; Sterilization: Ethylene Oxide (EO) to SAL 10^-6; Dimensions: 0.41 mm diameter, 12-16 mm length; Lead lengths: 0.5-2.5 m.
Indications for Use
Indicated for use in recording biopotential signals (EEG, EMG, nerve potential) and stimulation/recording of EMG and nerve potential signals. Intended for subdermal, nerve, or muscle tissue application.
Regulatory Classification
Identification
A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.
Predicate Devices
Disposable Monopolar and Subdermal Needle Electrodes (K130136)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2016
Daehan Medical Systems Co., Ltd. Dong Cheul Kim R&D Manager 250 Okgucheondong-ro Siheung City, 15084 KR
Re: K161566
Trade/Device Name: DMS Disposable Subdermal Needle Electrodes Regulation Number: 21 CFR 882.1350 Regulation Name: Needle Electrode Regulatory Class: Class II Product Code: GXZ, Dated: November 15, 2016 Received: November 21, 2016
Dear Dong Cheul Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -5 Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161566
Device Name
DMS Disposable Subdermal Needle Electrodes
Indications for Use (Describe)
The DMS Disposable Subdermal Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyograph (EMG) and nerve potential signals, and are intended for stimulation/recording with stimulation/recording equipment for electromyograph (EMC) and nerve potential signals.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Daehan Medical Systems Co., Ltd. The logo consists of the letters "DMS" in a stylized font, with a swooping line underneath. The text "Daehan Medical Systems Co., Ltd." is written in a smaller font to the right of the letters. The logo is in blue color.
# 510(k) Summary
[As required by 21 CFR 807.92]
### 1. Date Prepared [21 CFR 807.92(a)(a)]
Nov. 15, 2016
### Submitter's Information [21 CFR 807.92(a)[1]] 2.
| Name of Sponsor: | Daehan Medical Systems Co., Ltd. |
|------------------|-----------------------------------------------------------------------------|
| Address: | 250 Okgucheondong-Ro, Siheung-City, Gyeonggido,<br>Republic of Korea, 15084 |
| Contact Name: | Dong Cheul Kim / R&D Manager |
- Contact Name: Dong Cheul Kim / R&D Manager
- -Telephone No.: +82-31-432-6275
- +82-31-432-6276 -Fax No.:
- -Email Address: dckim@dmsleadwire.com
- . Establishment Registration No.: 3002893037
- Name of Manufacturer: ● Same as Sponsor Address: Same as Sponsor -
### 3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)]
- Trade Name: DMS Disposable Subdermal Needle Electrodes
- Regulation Name: Needle Electrode
- Classification:
●
| Classification Panel | Neurology |
|---------------------------|-----------------|
| Classification Regulation | 21 CFR 882.1350 |
| Product Code | GXZ |
| Device Class | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Daehan Medical Systems Co., Ltd. The logo features the letters "DMS" in a stylized font, with a swooping line encircling the letters. The text "Daehan Medical Systems Co., Ltd." is written in a smaller, sans-serif font to the right of the letters.
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate device within this submission is shown as follow;
- 510(k) Number: K130136
- Applicant: Technomed Europe
- Regulation Name: Needle Electrode ●
- Device Name: Disposable Monopolar and Subdermal Needle Electrodes
There are no significant differences between the DMS Disposable Subdermal Needle Electrodes and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, and technical characteristics.
#### Description of the Device [21 CFR 807.92(a)(4)] 5.
The DMS Disposable Subdermal Needle Electrodes are supplied sterile and intended for single use. This subdermal needle electrodes can be used to record the neurological and neurophysiological evoked potentials.
The connector is provided with touch-proof and cannot be connected to an AC outlet. This connector is designed for connecting the recording or monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyograph (EMG) and nerve potential signals, and are intended for stimulation/recording with stimulation/recording equipment for electromyograph (EMG) and nerve potential signals.
The DMS Disposable Subdermal Needle Electrodes are supplied sterile.
### 6. Intended Use [21 CFR 807.92(a)(5)]
The DMS Disposable Subdermal Needle Electrodes are intended for use with recording, monitoring equipment for the recording of biopotential signals including electroencephalograph (EEG), electromyograph (EMG) and nerve potential signals, and are intended for stimulation/recording with stimulation/recording equipment for electromyograph (EMG) and nerve potential signals.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Daehan Medical Systems Co., Ltd. The logo features the letters "DMS" in a stylized font, with horizontal lines running through the letters. A curved line wraps around the letters. The text "Daehan Medical Systems Co., Ltd." is written in a smaller font to the right of the logo.
## 7. Technological Characteristics [21 CFR 807.92(a)(6)]
The DMS Disposable Subdermal Needle Electrode are based on a technical feature comparison, the subject device was found to be similar to predicate device with regard to design, function, and technical characteristics.
| | Proposed Device | Predicate Device |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K161566 | K130136 |
| Model | DMS Disposable Subdermal Needle<br>Electrode | Disposable Monopolar and<br>Subdermal Needle Electrodes |
| Manufacturer | Daehan Medical Systems Co., Ltd. | Technomed Europe |
| Device Class | Class II | Class II |
| Product code | GXZ | GXZ and IKT |
| Intended Use | The DMS Disposable Subdermal<br>Needle Electrodes are intended for<br>use with recording, monitoring<br>equipment for the recording of<br>biopotential signals including<br>electroencephalograph (EEG),<br>electromyograph (EMG) and nerve<br>potential signals, and are intended<br>for stimulation/recording with<br>stimulation/recording equipment<br>for electromyograph (EMG) and<br>nerve potential signals. | Needle Electrodes for Neurological<br>purposes are intended for use with<br>recording, monitoring equipment<br>for the recording of biopotential<br>signals including<br>electroencephalograph (EEG),<br>electromyograph (EMG) and nerve<br>potential signals, and are intended<br>for stimulation/recording with<br>stimulation/recording equipment<br>for electromyograph (EMG) and<br>nerve potential signals. |
| Anatomical sites | Subdermal, nerve or muscle tissue | Subdermal, nerve or muscle tissue |
| Needle Diameter | 0.41 mm | 0.30 to 0.60 mm |
| Needle Length | 12, 13, 14 and 16 mm | 7 to 20 mm |
| Lead Wire<br>Length | 0.5, 1.0, 1.5, 2.0 and 2.5 m | 1.0, 1.5 and 2.5 m |
| Needle material | Stainless steel | Stainless steel |
| Lead wire | PVC insulated tin plated with<br>copper | PVC insulated tin plated with<br>copper |
| Connectors | DIN 42 802 1.5mm and Touch<br>proof connector | DIN 42 802 1.5mm and Touch<br>proof connector |
| Sterilization<br>Method | EO ethylene oxide | EO ethylene oxide |
| Sterility<br>assurance level | 10-6 | 10-6 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Daehan Medical Systems Co., Ltd. The logo features the letters "DMS" in a stylized, blue font with a swooping line underneath. To the right of the logo, the company name "Daehan Medical Systems Co., Ltd." is written in a smaller, sans-serif font, also in blue. The overall design is clean and professional, suggesting a company in the medical technology sector.
### Non-Clinical Test Summary:
- 1) Biocompatibility
- ISO 10993-5: Biological evaluation of medical devices - Part 5: tests for in vitro cytotoxicity
- ISO 10993-10: Biological evaluation of medical devices - Part 10: tests for irritation and skin sensitization
- 2) Performance Testing
- Stiffness Test ●
- Breakage Test ●
- Bond Test
- Electrical Properties Test ●
- Shelf-life Testing 3)
- ASTM F1980-07:2011, Standard guide for accelerated aging of sterile barrier systems ● for medical devices
#### Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] 8.
When compared to the predicate device (K130136), the DMS Disposable Subdermal Needle Electrode in this submission presented the substantial equivalence in terms of:
- Intended use ●
- Device design ●
- Components and materials
- Technological characteristics ●
#### 9. Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Daehan Medical Systems Co., Ltd., concludes that the DMS Disposable Subdermal Needle Electrodes are substantially equivalent to predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.