K161551 · BTL Industries, Inc. · PBX · Sep 13, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161551
Device Name
TR-100
Applicant
BTL Industries, Inc.
Product Code
PBX · General, Plastic Surgery
Decision Date
Sep 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TR-100 device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The TR-100 massage device is intended to provide a temporary reduction in the appearance of cellulite.
Device Story
TR-100 is a radiofrequency (RF) device with an integrated massager for non-invasive thermal therapy. The system comprises a microprocessor-controlled unit, an RF generator, a massage generator, and a handpiece with an 8.4" color touch screen. The operator uses the touch screen and handpiece buttons to control therapy; the treatment tip illuminates to indicate energy flow. The device delivers monopolar RF energy to heat tissue to 40–45 °C, while the massage component provides manual circular or automatic vertical massage. Skin temperature is monitored via patient feedback and an integrated IR thermometer. The device is used in a clinical setting by a healthcare professional. By elevating tissue temperature, the device aims to provide temporary pain relief, reduce muscle spasms, improve local circulation, and temporarily reduce the appearance of cellulite.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing demonstrating the device's ability to reach and maintain effective treatment temperatures (40–45 °C) and conformance to electrical safety and biocompatibility standards (ISO 10993-1, 5, 10; IEC 60601-1, 1-2, 2-2).
Technological Characteristics
Materials: Aluminum, plastic, stainless steel. Energy: Monopolar RF (0.5 MHz ± 50 kHz), sinusoidal waveform. Output: Up to 300 W (95.5 W/cm²). Interface: 8.4" color touch screen. Connectivity: Standalone. Safety: Class II, BF electrical protection. Standards: ISO 14971, IEC 62304, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 10993-1, 5, 10.
Indications for Use
Indicated for patients requiring elevated tissue temperature for temporary relief of pain, muscle spasms, and increased local circulation, or for temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 13, 2016
BTL Industries, Inc. Mr. Jan Zarsky Director 47 Loring Drive Framingham, Massachusetts 01702
Re: K161551
Trade/Device Name: TR-100 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX Dated: August 31, 2016 Received: September 1, 2016
Dear Mr. Zarsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth
{1}------------------------------------------------
in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Christopher J. Ronk -S
Fox Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) n/a K161551
Device Name
TR-100
Indications for Use (Describe)
The TR-100 device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The TR-100 massage device is intended to provide a temporary reduction in the appearance of cellulite.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with three interconnected blue diamonds. The letters "B", "T", and "L" are placed inside each of the diamonds, respectively. The logo has a clean and modern design.
# Section 5 - 510(k) Summary
## General Information
| Sponsor: | BTL Industries, Inc.<br>47 Loring Drive<br>Framingham, MA 01702<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant: | BTL Industries, Inc.<br>47 Loring Drive<br>Framingham, MA 01702<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
| Contact Person: | Jan Zarsky |
|-----------------|--------------------------------------------------------------------|
| | BTL Industries, Inc. |
| | Executive VP |
| | <span style="text-decoration: underline;">zarsky@btlnet.com</span> |
| Summary Preparation | |
|---------------------|-------------|
| Date: | 09 May 2016 |
### Device Names
| Trade/Proprietary Name: | TR-100 |
|------------------------------|---------------------------------------------------------|
| Primary Classification Name: | Electrosurgical cutting and coagulation and accessories |
| Common Name: | Combined high frequency and pulsed massager |
| Classification Regulation: | 878.4400 |
| Product Code: | PBX |
## Legally Marketed Predicate Devices
The TR-100 is substantially equivalent to the current product that is already cleared for USA distribution under the following 510(k) Premarket Notification number:
- . Exilis XP (K143040).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with three interconnected blue diamonds. The letters "B", "T", and "L" are placed inside each of the diamonds, respectively. The logo has a clean and modern design.
## Product Description
The TR-100 is a state-of-the-art radiofrequency device with integrated massager, that enables the application of therapy by a non-invasive, high-frequency field.
The control unit of the system is fitted with a color touch screen, to facilitate use of the device. The on-screen information guides the operator through the entire therapy. For easier control, the handpiece is equipped with buttons, enabling operation of the device during therapy. The energy flow is indicated by the illuminated treatment tip.
The TR-100 consists of the following main components:
- . microprocessor-driven control unit
- high-frequency electromagnetic energy generator
- . massage generator
- . user interface with 8.4" color touch screen
- handpiece
#### Indications for Use
The TR-100 RF device is intended to provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain. muscle spasms, and increase in local circulation. The TR-100 massage device is intended to provide a temporary reduction in the appearance of cellulite.
### Non-clinical Testing
The TR-100 device has been thoroughly evaluated for electrical safety. The TR-100 has been found to conform to applicable medical device safety standards. The device complies with the following standards:
- ISO 14971 Medical devices – Application of risk management to medical devices
- IEC 62304 Medical Device Software - Software Life Cycle Processes
- IEC 60601-1 General requirements for safety
- IEC 60601-1-2 Electromagnetic compatibility-Requirements and Tests
- IEC 60601-2-2 Particular requirements for the basic safety and essential performance of high frequency surgery equipment and high frequency surgical accessories
- ISO 10993-1 Evaluation and testing within a risk management process
- ISO 10993-5 Biological Evaluation of Medical Devices-Tests for In Vitro toxicity
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters BTL in a stylized font. The letters are arranged within three interconnected diamond shapes. The logo is a solid blue color and is set against a white background.
- ISO 10993-10 Biological Evaluation of Medical Devices—Test for Irritation and Skin Sensitization
## Clinical testing
The substantial equivalence determination for the TR-100 is based on clinical performance testing. The aim of the performance test was to prove that the device reaches the effective treatment temperature and maintain it for required time.
### Comparison with the Predicate Device
| Device Name | TR-100 | Exilis XP |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | BTL Industries, Inc. | BTL Industries, Inc. |
| 510(k) Number | n/a | K143040 |
| Regulation | General & Plastic Surgery<br>21 CFR 878.4400<br>Electrosurgical cutting and coagulation<br>and accessories | General & Plastic Surgery<br>21 CFR 878.4400<br>Electrosurgical cutting and coagulation and<br>accessories |
| Product Code | PBX | PBX |
| Indications for<br>Use | The TR-100 device is intended to<br>provide heating for the purpose of<br>elevating tissue temperature for selected<br>medical conditions such as temporary<br>relief of pain, muscle spasms, and<br>increase in local circulation. The TR-100<br>massage device is intended to provide a<br>temporary reduction in the appearance<br>of cellulite. | The Exilis XP device is intended to provide<br>heating for the purpose of elevating tissue<br>temperature for selected medical conditions<br>such as temporary relief of pain, muscle<br>spasms, and increase in local circulation.<br>The Exilis XP massage device is intended to<br>provide a temporary reduction in the<br>appearance of cellulite. |
| Device<br>Technologies | Application of the heat to the tissue via<br>RF energy. Massaging of body parts<br>with massage tip. | Application of the heat to the tissue via RF<br>energy. Massaging of body parts with<br>massage attachment. |
| Electrical<br>Protection | Class II, BF | Class II, BF |
| Color Touch<br>Screen | 8.4" (215mm) / 640x480 pixel | 8.4" (215mm) / 640x480 pixel |
| | | |
| Device Name | TR-100 | Exilis XP |
| Manufacturer | BTL Industries, Inc. | BTL Industries, Inc. |
| 510(k) Number | n/a | K143040 |
| RF Tip Surface in<br>contact with skin | 3.14 cm² | 1.13 - 2.54 cm² |
| Maximum Output<br>Power<br>(Density) | Up to 300 W<br>(95.5 W/cm2) | Up to 120 W<br>(47.2 - 106.2 W/cm²) |
| Effective<br>Treatment<br>Temperature | 40 - 45 °C (104 - 113 °F) | 40 - 45 °C (104 - 113 °F) |
| Skin Temperature<br>Monitoring | Based on Patient's Feedback.<br>Integrated IR thermometer. | Based on Patient's Feedback.<br>Separate IR thermometer. |
| Heating Energy<br>Type | Radiofrequency | Radiofrequency |
| Modes of<br>Operation | Monopolar | Monopolar |
| Output Frequency | 0.5 MHz ± 50 kHz | 3.25 MHz ± 50 kHz |
| Waveform | Sinusoid | Sinusoid |
| Massage Tip<br>Material | Metal | Grey Plastic material |
| Massage<br>application | Manual Circular,<br>Automatic Vertical | Manual Circular |
| Material of the<br>Generator Case | Aluminium, Plastic, Stainless Steel | Aluminium, Plastic, Stainless Steel |
| Handpiece Holder<br>Availability | YES | YES |
| Patch Electrode | YES | YES |
| Energy Source | 100 - 240 VAC, max 7A, 50-60 Hz | 110 - 240 V, max 4A, 50-60 Hz |
| Device Name | TR-100 | Exilis XP |
| Manufacturer | BTL Industries, Inc. | BTL Industries, Inc. |
| 510(k) Number | n/a | K143040 |
| Material of the<br>Generator Case | Aluminum, Plastic, Stainless Steel | Aluminum, Plastic, Stainless Steel |
| RF Energy<br>Emission Indicator | YES | YES |
| Dimensions<br>(W x H x D) | 600 × 1000 × 600 mm<br>(24" × 39" × 24") | 406 × 270 × 302 mm<br>(15.98" × 10.63" × 11.87") |
| Weight | 50.7 lb (23 kg) | 16 lb (7.3 kg) |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a blue logo with the letters B, T, and L. The letters are inside of three diamond shapes that are connected to each other. The diamond shapes are also blue. The logo is simple and modern.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the letters B, T, and L. The letters are in blue and are inside of three diamond shapes. The diamond shapes are connected to each other. The logo is simple and modern.
### Substantial Equivalence
Based upon the intended use and technical information provided in this pre-market notification, the TR-100 device has been shown to be substantially equivalent to currently marketed predicate device.
## Conclusion
Based on the aforementioned information, the TR-100 is safe and effective and substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.