← Product Code [JJX](/productcode/JJX) · K161526

# LIAISON CMV IgM Serum Control Set (K161526)

_DiaSorin, Inc. · JJX · Jun 30, 2016 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161526

## Device Facts

- **Applicant:** DiaSorin, Inc.
- **Product Code:** [JJX](/productcode/JJX.md)
- **Decision Date:** Jun 30, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The LIAISON® CMV IgM Serum Control Set (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® CMV IgM assay on the LIAISON® Analyzer family.

## Device Story

LIAISON® CMV IgM Serum Control Set consists of negative and positive control samples; used to monitor performance of LIAISON® CMV IgM chemiluminescent immunoassay on LIAISON® Analyzer family. Modification from previous predicate (K040290) involves transition from 5% serum/buffer matrix to human serum matrix and extension of open-use stability claim from 4 to 16 weeks. Controls are ready-to-use, stored at 2-8°C. Used by laboratory personnel to verify assay performance and detect reagent failure.

## Clinical Evidence

Bench testing only. Studies included commutability between samples and controls, precision equivalence, control value assignment, control range definition, and real-time stability testing (12-month shelf-life, 16-week open-use stability).

## Technological Characteristics

Liquid, ready-to-use human serum/defibrinated plasma matrix. Contains 0.1% ProClin® 300 and 0.09% sodium azide as preservatives. Designed for use with LIAISON® Analyzer family. Storage at 2-8°C. No software or electronic components.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- LIAISON® Control CMV IgM ([K040290](/device/K040290.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

SPECIAL 510(k): Device Modification OIR Decision Summary

To: THE FILE

RE: DOCUMENT NUMBER K161526

This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class I device requiring 510(k). The following items are present and acceptable:

1. The name and 510(k) number of the SUBMITTER'S previously cleared device.

Trade Name: LIAISON® Control CMV IgM 510(k) number: K040290

2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use and package labeling.

3. A description of the device MODIFICATIONS, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

This change was for: Modifications to the LIAISON® CMV IgM Control Set consist of a change from a buffer to a human serum matrix and extension of the open use stability claim.

4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics is shown in the table below.

|  Summary of Similarities and Differences: LIAISON® CMV IgM Control Set  |   |   |
| --- | --- | --- |
|  Characteristic | Predicate Device DiaSorin LIAISON® Control CMV IgM (K040290) | Modified Device DiaSorin LIAISON® CMV IgM Serum Control Set (K161526)  |
|  Intended Use | The LIAISON® CMV IgM controls (negative and positive) are used for monitoring substantial reagent failure of the LIAISON® CMV IgM chemiluminescent immunoassay (CLIA). The LIAISON® CMV IgM quality control material contains only a 5% serum matrix and may not adequately control the DiaSorin LIAISON® CMV IgM assay for serum specimens. | The LIAISON® CMV IgM Serum Control Set (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® CMV IgM assay on the LIAISON® Analyzer family.  |
|  Negative Control | 5% Human Serum/plasma not reactive for CMV IgM antibodies, diluted in PBS buffer, BSA, with ProClin® 300 as a preservative. | Human Serum/plasma non-reactive for CMV IgM antibodies, 0.1% ProClin® 300 and 0.09% sodium azide.  |
|  Positive Control | 5% Human Serum/plasma reactive for CMV IgM antibodies, diluted in PBS buffer, BSA, with ProClin® 300 as a preservative and an inert yellow dye. | Human Serum/plasma reactive for CMV IgM antibodies, 0.1% ProClin® 300 and 0.09% sodium azide.  |

{1}

|  Summary of Similarities and Differences: LIAISON® CMV IgM Control Set  |   |   |
| --- | --- | --- |
|  Characteristic | Predicate Device DiaSorin LIAISON® Control CMV IgM (K040290) | Modified Device DiaSorin LIAISON® CMV IgM Serum Control Set (K161526)  |
|  Reagent Configuration | 2 vials each level (negative and positive)
0.7 mL/vial, ready to use. | Same  |
|  Storage | Store at 2-8°C | Same  |
|  Open Use Stability | Once opened controls are stable for four (4) weeks when properly stored at 2-8°C between uses. | Once opened controls are stable for sixteen (16) weeks when properly stored at 2-8°C between uses.  |

5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modifications and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.

---

**Source:** [https://fda-staging.innolitics.com/device/K161526](https://fda-staging.innolitics.com/device/K161526)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).