Stryker's Precision Thin Reciprocating Blade, 0.010 inch (REF 5100-437-010) is a sterile, single use cutting accessory intended to cut bone in the posterior canal wall during an Otology procedure.
Device Story
Sterile, single-use reciprocating blade accessory for surgical drills; used in Otology procedures to cut bone in posterior canal wall. Operated by surgeon via electric handpiece, console, and footswitch/handswitch; system converts electrical signals to mechanical reciprocating motion. Blade thickness of 0.010 inches allows for narrow kerf width during bone removal. Used in clinical settings; provides precise bone cutting to facilitate surgical access or reconstruction. Benefits include improved surgical precision in confined anatomical spaces.
Clinical Evidence
No clinical studies performed. Evidence consists of bench testing (kerf width, life/simulated use, temperature) and design validation. Simulated use testing conducted with 3 ENT surgeons on fresh cadaveric human temporal bone and 8 ENT surgeons on PHACON temporal bone models. Biocompatibility testing per ISO 10993-1 confirmed device is non-sensitizing, non-irritating, and non-toxic.
Technological Characteristics
Stainless steel blade and arbor; vacuum brazed. Dimensions: 0.010 in. thickness, 0.124 in. shank. Powered by electric surgical drill/handpiece. Single-use; sterile (gamma irradiated, SAL 10^-6). Packaging: foil pouch with polybag.
Indications for Use
Indicated for use in Otology procedures to cut bone in the posterior canal wall. Intended for general patient population.
Regulatory Classification
Identification
An ear, nose, and throat bur is a device consisting of an interchangeable drill bit that is intended for use in an ear, nose, and throat electric or pneumatic surgical drill (§ 874.4250) for incising or removing bone in the ear, nose, or throat area. The bur consists of a carbide cutting tip on a metal shank or a coating of diamond on a metal shank. The device is used in mastoid surgery, frontal sinus surgery, and surgery of the facial nerves.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular text element and a symbolic graphic. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The graphic element features a stylized depiction of what appears to be an abstract human form or a symbol representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 17, 2017
Stryker Instruments Dr. Joanna McCarthy Senior Regulatory Affairs Specialist Stryker Corporation 4100 E. Milham Ave. Kalamazoo, MI 49009
Re: K161514
Trade/Device Name: Precision Thin Reciprocating Blade, 0.010in Regulation Number: 21 CFR 874.4140 Regulation Name: Ear, Nose, and Throat Bur Regulatory Class: Class I Product Code: EQJ Dated: December 13, 2016 Received: December 15, 2016
Dear Dr. McCarthy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
| Form Approved: OMB No. | 0910-0120 |
|--------------------------|------------------|
| Expiration Date: | January 31, 2017 |
| See PRA Statement below. | |
| 10(k) Number (if known) | 161514 |
|-------------------------|--------|
|-------------------------|--------|
ייעוץ (אוסורפ
-
Indications for Use (Describe)
นาย 1 ค.ศ. 1984 - 1992 - 1992 - 1992 - 1992 - 1992 - 1982 - 1982 - 1992 - 1992 - 1992 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1997 - 1
| Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| | <span> <input checked="true" type="checkbox"/> </span> |
| | <span> <input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
PSC Publishing Services (301) 443-6740 EF
r vəzi vəzi qot sudovi ilk və vəzi və qalınmışdır.
0
င - မ ခရမျ
นที่010.0 เอชธไช สูติเมื่อวันที่ 1 ที่ปี ที่ต่างๆ สอบอรมันที่มีความ สิ่ง 2012 โรคติดที่มีความ
{3}------------------------------------------------
#### 510(k) Number: K161514
#### Section 5.1 Submitter
510(k) Owner: Stryker Instruments,
4100 E. Milham Avenue,
Kalamazoo,
MI 49001, USA
Phone: 269-323-7700
Fax: 269-389-5412
Contact Person: Joanna McCarthy PhD, Senior Regulatory Affairs Specialist
Date Prepared: 27th May 2016
#### Section 5.2 Device Information
Trade Name: Precision Thin Reciprocating Blade, 0.010in. Device Common Name: ENT Bur Classification Name: Bur, Ear, Nose and Throat
| FDA Product Code | Device | Regulation Number | Class |
|------------------|----------------------------------|-------------------|-------|
| EQJ | <i>Bur, Ear, Nose and Throat</i> | 21 CFR 874.4140 | I |
Table 5-1. Classification Data for Precision Thin Reciprocating Blade, 0.010in.
{4}------------------------------------------------
## Section 5.3 Comparison against Regulation
Stryker currently markets 31 similar devices within the micro reciprocating saw blade family which are independently listed under the FDA product code EQJ. In this premarket notification all 31 devices covered under the EQJ requlation are compared to the subject device Precision Thin Reciprocating Blade, 0.010in based on their technological characteristics and general ENT indications. It is only the addition of the specific indication for the subject device Precision Thin Reciprocating Blade, 0.010in. that is being compared to the following regulation;
Regulation Description: Ear, Nose and Throat Bur Regulation Number: 21 CFR 874.4140
#### Section 5.4 Reference Device
| Device Name: | Micro Reciprocating Saw Blade (REF MR-0580) |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type: | Reference Device |
| 510(K) Number: | K131053- “ANSPACH XMAX, EMAX2 and EMAX2 Plus<br>System with Otologic Attachment System” cleared under<br>FDA Product Codes ERL and EQJ. |
| Description: | The primary comparator device (ENT Regulation 21CFR<br>874.4140) has successfully addressed decision points 1 to 4<br>in the 510(k) Decision Making Flowchart as per FDA)<br>Guidance for Industry and FDA Staff, The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket<br>Notifications [510(k)], dated July 28, 2014.<br><br>However, the technological characteristics (design and<br>material) of the Precision Thin Reciprocating Blade, 0.010in<br>is compared to Anspach Micro Reciprocating Saw Blade,<br>which is cleared through the 510(k) - K131053 |
The Anspach Effort, Inc .:
Table 5-2. Classification Data for Reference Device.
#### Section 5.5 Device Description
Stryker's Precision Thin Reciprocating Blade, 0.010in is a sterile, single use cutting accessory which is operated by a surgical drill (handpiece). This handpiece moves the Precision Thin Reciprocating Blade, 0.010in. in a reciprocating manner so that the blade removes bone material on the back stroke of its cut. The Precision Thin Reciprocating Blade, 0.010in. has a thickness of 0.010 inches and therefore can be used in surgical procedures where a narrow kerf width is required when cutting bone.
{5}------------------------------------------------
## Section 5.6 Indications for Use
Stryker's Precision Thin Reciprocating Blade, 0.010 inch (REF 5100-437-010) is a sterile, single use cutting accessory intended to cut bone in the posterior canal wall during an Otology procedure.
{6}------------------------------------------------
## Section 5.7 Comparison of Technological Characteristics against Regulation 21 CFR 874.4140 (ENT Burs) and reference device.
| Feature | Comparator ENT Bur (21 CFR 874.4140) | Subject Device: Precision Thin Reciprocating Blade,<br>0.010in. | Justification |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Product Class | Class I | Class I | Identical |
| Regulation | 21 CFR 874.4140 - Bur, ear, nose and throat | 21 CFR 874.4140 - Bur, ear, nose and throat | Identical |
| FDA Product Code | EQJ - Bur Ear, Nose and Throat | EQJ - Bur Ear, Nose and Throat | Identical |
| Intended Function | Intended to cut bone. | Intended to cut bone. | Identical |
| Patient Population | General | General | Identical |
| Indications For Use | An ear, nose, and throat bur is a device consisting<br>of an interchangeable drill bit that is intended for<br>use in an ear, nose, and throat electric or<br>pneumatic surgical drill (874.4250) for incising or<br>removing bone in the ear, nose, or throat area.<br>The bur consists of a carbide cutting tip on a<br>metal shank or a coating of diamond on a metal<br>shank. The device is used in mastoid surgery,<br>frontal sinus surgery, and surgery of the facial<br>nerves. | Stryker's Precision Thin Reciprocating Blade, 0.010 inch<br>(REF 5100-437-010) is a sterile, single use cutting accessory<br>intended to cut bone in the posterior canal wall during an<br>Otology procedure. | Similar |
| Contraindications | None known | None known | Identical |
{7}------------------------------------------------
| Feature | Comparator ENT Bur (21 CFR 874.4140) | Subject Device: Precision Thin Reciprocating<br>Blade, 0.010in. | Justification |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Conditions for Use | Single Use | Single Use | Identical |
| Principle of<br>operation<br>/Mechanism of<br>Action | Legally marketed micro reciprocating blades listed<br>under FDA product code EQJ (ENT Bur) are used in<br>conjunction with an electric reciprocating handpiece<br>(CORETM Reciprocating Saws), a CORETM Console<br>and a footswitch or handswitch. When the system is<br>assembled, the surgeon controls the footswitch; this<br>modifies the electrical signal to the motor, controlling<br>the speed of the cutting accessory. | Stryker's Precision Thin Reciprocating Blade, 0.010in.<br>is used in conjunction with an electric motor (CORETM<br>Reciprocating Saws), a CORETM Console and a<br>footswitch. When the system is assembled, the<br>surgeon controls the footswitch or handswitch; this<br>modifies the electrical signal to the motor, controlling<br>the speed of the cutting accessory. | Identical |
| Motor power supply | Powered by Electrical energy. | Electrical energy. | Identical |
| For use with a surgical<br>drill (handpiece) per 21<br>CFR 874.4140 | Legally marketed micro reciprocating blades - listed<br>under FDA product code EQJ (ENT Bur) are used with<br>the CORETM Reciprocating Saws. | Used with the CORETM Reciprocating Saws as<br>indicated on the product label. | Identical |
| Source of Activation | Footswitch and Handswitch. | Footswitch and Handswitch. | Identical |
| Feature | Subject Device: Precision Thin<br>Reciprocating Blade, 0.010in. | Device:<br>Anspach<br>Reference<br>Micro Reciprocating Saw Blade<br>(MR-0580) | Justification |
| Cutting Accessory<br>Locking Mechanism | Plug<br>Blade<br>locking<br>Tip<br>and<br>mechanism. | Chuck is tightened around shank. | Similar |
| Shank of the Cutting<br>Accessory | 0.125 - 0.124 inches | 0.124 inches | Identical |
| Cutting Accessory<br>Geometry | Cut Edge: 0.766 in. (19.5mm)<br>Blade Height: 0.137 in.<br>Cut Thickness: 0.010 in. | Cut Edge: 21.35 mm (0.84 in)<br>Cut Thickness: 0.25 mm (0.010 in) | Similar |
| Cutting Accessory<br>Length of exposure | 1.42 in. | 1.37 in. | Similar |
| Patient Contacting<br>Material | Blade Arbor - Stainless Steel | Blade Arbor - Stainless Steel. | Similar |
| | Blade - Stainless Steel<br>Vacuum Brazing Paste - NICROBRAZ<br>51 S ALLOY. | Blade - Stainless Steel.<br>Vacuum Brazing Paste - Unknown. | |
| Sterilization | Supplied sterile, gamma irradiated. | Supplied sterile, gamma irradiated. | Identical |
| Sterility Assurance<br>Level (SAL) | SAL of 10-6 | Minimum SAL of 10-6 | Identical |
| Shelf-Life | 5 Years | Unknown | Similar |
| Packaging<br>Configuration | Individually packaged in foil pouch with<br>Polybag. | Tyvek / Film pouch with the internal<br>packaging being a PETG clamshell<br>blister. | Similar |
{8}------------------------------------------------
Table 5-3: Comparative Device Justification Table.
{9}------------------------------------------------
## Section 5.8 Performance Data
The following verification tests were performed which demonstrates that the device meets the performance requirements under its indications for use conditions.
- Kerf Width Testing ●
- Life, Simulated Use Testing ●
- Temperature Testing ●
Simulated use of a canal wall reconstruction was completed with 3 ENT surgeons for the specific indication in fresh cadaveric Human temporal bone. To supplement this cadaveric study design validation was also completed for the specific indication using PHACON's Temporal Bone Model with 8 ENT surgeons. This model was selected as a suitable cut media for design validation following a technical evaluation along with expert medical opinion. Collectively, the results of these performance tests demonstrate the functionality, integrity, safety and effectiveness of Stryker's Precision Thin Reciprocating Blade, 0.010in. for its intended use and support a determination of substantial equivalence to the regulation 21 CFR 874.4140 (ENT bur).
# 5.8.1 Biocompatibility Testing
The subject device Precision Thin Reciprocating Blade, 0.010in. is classified as external communicating device: tissue/bone/dentin with limited patient contact (< 24 hours).
The Biocompatibility evaluation was conducted in accordance with;
- AAMI/ANSI/ISO ISO 10993-1:2009/(R)2013, Biological evaluation of medical . devices - Part 1: Evaluation and testing within a risk management process, and Guidance for Industry and FDA Staff.
- Draft Guidance for Industry and FDA Staff, Use of International Standard ISO-● 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," (Date April 23, 2013).
- FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-● 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," (Dated May 1, 1995).
Results of testing validate that the subject device is non-sensitizing, non-irritating and non-toxic (cytotoxic and systemic).
#### 5.8.2 Clinical Studies
No clinical studies were performed to support substantial equivalence.
{10}------------------------------------------------
## Section 5.9 Conclusions
The subject device Precision Thin Reciprocating Blade, 0.010in. is substantially equivalent in intended use, operating principle, fundamental technology, overall design and materials to ENT Burs as described in 21 CFR 874.4140.
Risk analysis for the subject device due to the addition of the specific indication reveals no new issues regarding safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.