K161503 · Yerim Engineering Co., Ltd. · GEX · Aug 26, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161503
Device Name
ORION
Applicant
Yerim Engineering Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
532nm The ORION is indicated for coagulation and hemostasis of vascular and cutaneous lesions in dermatology including, but not limited to, the following general categories: vascular lesions[angiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions (nevi, lentigines, chloasma, cafk-au-lait, tattoos (red and green ink); verrucae; skin tags; keratoses; plaques; cutaneouslesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size). 755 nm The ORION is indicated for stable long-term, or permanent hair reduction which is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I- VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles. 1064 nm The ORION is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus, spider veins and poikiloderma of civate; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as,but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques. The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent hair reduction which is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).
Device Story
Orion is a medical laser system for dermatology; utilizes 532nm, 755nm, and 1064nm wavelengths. System comprises main body, handpiece, and protective goggles. Physician-operated via LCD touch screen control panel; microprocessor manages laser emission. Device targets melanin in hair follicles or vascular/pigmented lesions to achieve coagulation, hemostasis, or hair reduction. Cold air cooling used for skin protection. Output is laser energy delivered to target tissue; clinical decision-making based on visual assessment of lesions or hair growth. Benefits include lesion size reduction, color lightening, and stable long-term hair reduction.
Long Pulse Nd:YAG, Alexandrite, and KTP laser system. Wavelengths: 532nm, 755nm, 1064nm. Pulse duration: 0.1-300ms. Spot size: 2-20mm. Cooling: Cold air. Interface: Microprocessor-controlled LCD touch screen. Power: AC220-240V/16A.
Indications for Use
Indicated for patients requiring coagulation/hemostasis of vascular/cutaneous lesions, treatment of benign pigmented lesions, wrinkles, pseudofolliculitis barbae, and permanent hair reduction. Applicable to all skin types (Fitzpatrick I-VI) including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Family of Altus Medical Modified Cool Glide Aesthetic Lasers (K014040)
{0}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with outstretched arms, resembling an eagle or bird in flight. The figure is composed of three overlapping profiles, creating a sense of depth and movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26, 2016
Yerim Engineering Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 800 Roosevelt Ste 417 Irvine, California 92620
Re: K161503
Trade/Device Name: Orion Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 23, 2016 Received: June 1, 2016
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known) K161503
Device Name
ORION
# Indications for Use (Describe)
### 532nm
The ORION is indicated for coagulation and hemostasis of vascular and cutaneous lesions in dermatology including, but not limited to, the following general categories: vascular lesions[angiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions (nevi, lentigines, chloasma, cafk-au-lait, tattoos (red and green ink); verrucae; skin tags; keratoses; plaques; cutaneouslesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
# 755 nm
The ORION is indicated for stable long-term, or permanent hair reduction which is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I- VI) including tanned skin. It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.
# 1064 nm
The ORION is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus, spider veins and poikiloderma of civate; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as,but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques.
The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles. Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent hair reduction which is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (K161503)
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: ____Aug 19, 2016____
#### 1. Applicant / Submitter:
YERIM ENGINEERING Co., Ltd. Ga-307, Techno-zone ,117, Hwanggeum-ro, Yangchon-eup, Gimpo-si, Gyeonggi-do 415-843 Republic of Korea Tel : +82-31-991-9841 Fax : +82-31-991-9845
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 800 Roosevelt Ste 417, Irvine CA 92620 Phone: 714-202-5789, Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### 3. Device:
| Proprietary Name: | ORION |
|------------------------------|--------------------------------------|
| Common Name: | Medical Laser System |
| Classification Name: | Instrument, Surgical, Powered, Laser |
| Classification: | Class II, 21 CFR 878.4810 |
| Classification Product Code: | GEX |
#### 4. Predicate Device:
■ Primary Predicate Device: Family of Altus Medical Modified Cool Glide Aesthetic Lasers by Altus Medical, Inc. (K014040) ■ Reference Predicate Device:
Cynosure Apogee Elite Laser by CYNOSURE, INC. (K034030)
CLARITY LPC Laser System by Lutronic Corporation (K130199)
{5}------------------------------------------------
#### 5. Device Description:
The subject device is composed of the main body and a handpiece which is an irradiation device and as an accessory part, protective goggles for protection of the worker.
It is controlled by a micro processor interfaced to a LCD touch screen control panel. The computer controls start and stop of the treatment. When the key switch of the system is turned clockwise, the main power will be inputted, which will be conveyed to the hand piece through the control board.
Meanwhile, the control board connected to the touch screen is connected to the lamp of the handpiece and controls the same, and controls the whole system through the data connected to the touch screen control panel. When the switch of the hand piece is pressed, the lamp will laser.
#### 6. Indications for Use:
## 532nm
For coagulation and hemostasis of vascular and cutaneous lesions in dermatology including, but not limited to, the following general categories: vascular lesions[angiomas, hemangiomas (port wine), telangiectasia (facial or extremities telangiectasias venous anomalies, leg veins)); benign pigmented lesions (nevi, lentigines, chloasma,cafk-au-lait, tattoos (red and green ink); verrucae; skin tags; keratoses; plaques; cutaneouslesion treatment (hemostasis, color lightening, flattening, reduction of lesion size).
## 755 nm
The ORION is indicated for stable long-term, or permanent hair reduction which is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I- VI) including tanned skin.
It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.
## 1064 nm
The ORION is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as,but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (sigmificant reduction in the intensity of black and/or blue/black tattoos) and plaques.
The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent hair reduction which is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).
{6}------------------------------------------------
#### Performance Data(Non-Clinical): 7.
The following properties were tested based on the referenced standards. All the test results support substantial equivalence to the predicate devices.
- . IEC 60601-1 Medical electrical equipment - part 1: general requirements for basic safety and essential performance
- . IEC 60601-1-6 Medical electrical equipment - part 1-6: general requirements for basic safety and essential performance - collateral standard: usability.
- IEC 62366-1 Medical devices part 1: application of usability engineering to medical devices. .
- IEC 60601-2-22 Medical electrical equipment part 2-22: particular requirements for basic . safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment. (Radiology)
- . IEC 60601-1-2 Medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests. (General II (ES/EMC))
- . IEC 60825-1 Edition 2.0 2007-03, safety of laser products - part 1: equipment classification, and requirements [including: technical corrigendum 1 (2008), interpretation sheet 1 (2007), interpretation sheet 2 (2007)]. (Radiology)
#### 8. Substantial Equivalence
The ORION is substantially equivalent to the Family of Altus Medical Modified Cool Glide Aesthetic Lasers by Altus Medical, Inc. (K014040), Cynosure Apogee Elite Laser by CYNOSURE, INC. (K034030) and CLARITY LPC Laser System by Lutronic Corporation (K130199).
The following comparison table is presented to demonstrate substantial equivalence.
| | | Subject Device | Primary Predicate<br>Device | Reference<br>Predicate Device | Reference<br>Predicate Device | |
|-----------------------------|---------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | | | | |
| Device name | | ORION | Family of Altus<br>Medical Modified<br>CoolGlide<br>Aesthetic<br>Lasers (Primary<br>Predicate) | Cynosure Apogee<br>Elite Laser<br>(Reference<br>Predicate) | CLARITY LPC<br>Laser System<br>(Reference<br>Predicate) | |
| Manufacturer | | YERIM<br>ENGINEERING<br>Co., Ltd. | Altus Medical, Inc. | CYNOSURE, INC. | Lutronic<br>Corporation | |
| 510(k) number | | K161503 | K014040 | K034030 | K130199 | |
| Technology | | Long Pulse<br>Nd:YAG | Long Pulse<br>Nd:YAG | Long Pulse<br>Nd:YAG | Long Pulse<br>Nd:YAG | |
| Power (CW Laser) | | AC220 - 240<br>V/16A | 110 - 120<br>VAC/20A<br>220 - 240<br>VAC/20A | 200/220 VAC/30A | 4.0kVA(AC220~2<br>30V) | |
| Energy | 1064nm | 100J | 100J | 100J | 100J | |
| (pulse and | 755nm | 50J | - | 50J | 55J | |
| super pulse) | 532nm | 10J | 10J | n/a | n/a | |
| If pulsed, how is this done | | Long pulse | Long pulse | Long pulse | Long pulse | |
| Frequency of<br>pulse | 1064 nm | 1Hz | 2Hz | 5Hz | Up to 10Hz | |
| | 755nm | 3Hz | - | 3Hz | Up to 10Hz | |
| | 532nm | 2Hz | 2Hz | | n/a | |
| Pulse train<br>duration | 1064 nm | 0.1~300ms | 0.3 - 300 ms | Adjustable<br>0.4- 300 msec | 0.35 ms - 300 ms | |
| | 755nm | 0.1~300ms | | Adjustable<br>0.5 – 300 msec | 0.35 ms - 300 ms | |
| | 532nm | 0.1~300ms | 0.3 - 300 ms | | | |
| Spot size at target | | 2~20mm | 10mm | 3~24mm | 2, 3, 5, 8, 10, 12,<br>15, 18, 20 mm | |
| Wavelength | | 1064 nm,<br>755nm,<br>532nm | 1064 nm,<br>532nm | 1064 nm,<br>755nm | 1064 nm,<br>755nm | |
| Aiming beam | | 650nm | 650 nm | 650nm | 650nm | |
| Energy source | | Nd:YAG,<br>Alexandrite, KTP | Nd-YAG, KTP | Nd:YAG,<br>Alexandrite | Nd:YAG,<br>Alexandrite | |
| Cooling method | | Cold air | Sapphire cooling | Cold air | Cold air | |
| Intended use | 532nm | The ORION is<br>indicated for<br>coagulation and<br>hemostasis of<br>vascular and<br>cutaneous lesions<br>in<br>dermatologyincludi<br>ng, but not limited<br>to, the following<br>general categories:<br>vascular<br>lesions[angiomas,<br>hemangiomas (port<br>wine),<br>telangiectasia<br>(facial or<br>extremities<br>telangiectasias,ven<br>ous anomalies, leg<br>veins)]; benign<br>pigmented lesions<br>(nevi, lentigines,<br>chloasma,cafk-au-<br>lait, tattoos (red<br>and green ink);<br>verrucae; skin tags;<br>keratoses; plaques;<br>cutaneouslesion<br>treatment | For coagulation<br>and hemostasis of<br>vascular and<br>cutaneous lesions<br>in<br>dermatologyincludi<br>ng, but not limited<br>to, the following<br>general categories:<br>vascular<br>lesions[angiomas,<br>hemangiomas (port<br>wine),<br>telangiectasia<br>(facial or<br>extremities<br>telangiectasias,ven<br>ous anomalies, leg<br>veins)]; benign<br>pigmented lesions<br>(nevi, lentigines,<br>chloasma,cafk-au-<br>lait, tattoos (red<br>and green ink);<br>verrucae; skin tags;<br>keratoses; plaques;<br>cutaneouslesion<br>treatment<br>(hemostasis, color<br>lightening, | | | |
| | | lightening,<br>blanching,<br>flattening,<br>reduction of lesion<br>size). | flattening,<br>reduction of lesion<br>size). | | | The CLARITY<br>LPC Laser System<br>is indicated for<br>temporary hair<br>reduction. Stable<br>long-term or<br>permanent<br>reduction through<br>selective targeting<br>of melanin in hair<br>follicles.<br>Permanent hair<br>reduction is<br>defined as the long-<br>term, stable<br>reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9,<br>and 12 months<br>after the<br>completion of a<br>treatment regime.<br>On all skin types<br>(Fitzpatrick 1-VI)<br>including tanned<br>skin. Treatment of<br>benign pigmented<br>lesions. Treatment<br>of wrinkles. The<br>photocoagulation<br>of dermatological<br>vascular lesions<br>(such as port-wine<br>stains,<br>hemangiomas,<br>telangiectasias). |
| | 755 nm | The ORION is<br>indicated for stable<br>long-term, or<br>permanent hair<br>reduction.<br>Permanent hair<br>reduction is<br>defined as long-<br>term stable<br>reduction in the<br>number of hair<br>regrowth after a<br>treatment regime. It<br>is used for all skin<br>types (Fitzpatrick<br>I- VI) including<br>tanned skin.<br>It is also indicated<br>for the treatment of<br>vascular lesions,<br>benign pigmented<br>lesions, and<br>wrinkles. | - | The Cynosure<br>Apogee Elite<br>Dermatological<br>Laser is indicated<br>for stable long-<br>term, or permanent<br>hair reduction.<br>Permanent hair<br>reduction is<br>defined as long-<br>term stable<br>reduction in the<br>number of hair<br>regrowth after a<br>treatment regime. It<br>is used for all skin<br>types (Fitzpatrick<br>I- VI) including<br>tanned skin.<br>It is also indicated<br>for the treatment of<br>vascular lesions,<br>benign pigmented<br>lesions, and<br>wrinkles. | | |
| | 1064 nm | The ORION is<br>intended for the<br>coagulation and<br>hemostasis of<br>benign vascular<br>lesions such as, but<br>not limited to, port<br>wine stains,<br>hemangiomas,<br>warts,<br>telangiectasia, | Dermatology:<br>The Altus Medical<br>Aesthetic<br>CoolGlide laser<br>systems are<br>intended for<br>thecoagulation and<br>hemostasis of<br>benign vascular<br>lesions such as, but<br>not limited toport<br>wine stains, | The Cynosure<br>Apogee Elite<br>Dermatological<br>laser is intended for<br>the coagulation and<br>hemostasis of<br>benign vascular<br>lesions such as, but<br>not limited to, port<br>wine stains,<br>hemangiomas,<br>warts, | | The CLARITY<br>LPC Laser System<br>is indicated for the<br>removal of<br>unwanted hair, for<br>stable long term or<br>permanent hair<br>reduction and for<br>treatment of PFB.<br>Permanent hair<br>reduction is<br>defined as the long- |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.