MEG-TORQ

K161500 · Micro-Nx Co., Ltd. · EKX · Feb 22, 2017 · Dental

Device Facts

Record IDK161500
Device NameMEG-TORQ
ApplicantMicro-Nx Co., Ltd.
Product CodeEKX · Dental
Decision DateFeb 22, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Indications for Use

This product is a cordless motor handpiece system intended for tightening and loosening an abutment screw to fix and remove the abutment on a dental implant in prosthodontic treatment.

Device Story

MEG-TORQ is a cordless, AC-powered motor handpiece system used by dental professionals for prosthodontic procedures. The device functions as a torque driver to tighten or loosen dental implant abutment screws. It features adjustable speed (15, 30, 45, 60 rpm) and torque settings (up to 35 Ncm), with forward and reverse rotation capabilities. The system includes the torque driver, a charge adapter, and a charge cradle. Operators control the device to ensure precise torque application, preventing damage to the abutment screw. By providing controlled mechanical force, the device facilitates secure implant restoration, benefiting patients through reliable and efficient abutment fixation.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including performance testing per ISO 14457:2012, sterilization validation per ANSI/AAMI ST79 and ISO 17665/11138 series, and electrical safety/EMC testing per IEC 60601-1 and related standards.

Technological Characteristics

Cordless motor handpiece; AC-powered; push-button chuck; 2.35mm screwdriver shank (ISO 1797-1 Type 1). Max torque 35 Ncm. Dimensions 30x28x200mm. Sterilization per ANSI/AAMI ST79, ISO 17665-1/2, ISO 11138-1/3. Electrical safety per IEC 60601-1, IEC 60601-1-6, EN 60601-1-2, EN 61000-3-2, EN 61000-3-3.

Indications for Use

Indicated for use in dental prosthodontic treatment to tighten or loosen abutment screws on dental implants. No specific age or gender restrictions.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of the human profile in triplicate. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 22, 2017 MICRO-NX Co., Ltd % Ms. Priscilla Chung Regulatory Affairs Consultant Lk Consulting Group USA, Inc. 800 Roosevelt Ste 417 Irvine. California 92606 Re: K161500 Trade/Device Name: MEG - TORQ Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EKX Dated: January 24, 2017 Received: January 25, 2017 Dear Ms. Chung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161500 Device Name MEG-TORQ Indications for Use (Describe) This product is a cordless motor handpiece system intended for tightening and loosening an abutment screw to fix and remove the abutment on a dental implant in prosthodontic treatment. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CER 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (K161500) Date: Feb 14, 2017 #### 1. Applicant / Submitter: MICRO-NX Co., Ltd. 37, Dalseong2chadong 3-ro, Guji-myeon, Dalseong-gun, Daegu 711-891 REPUBLIC OF KOREA Tel: +82-53-611-0071 Fax:+82-53-611-0074 #### 2. Submission Correspondent: Priscilla Chung LK Consulting Group USA, Inc. 800 Roosevelt Ste 417, Irvine CA 92620 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com #### 3. Device: | Proprietary Name: | MEG-TORQ | |------------------------------|-------------------------------------| | Common Name: | Dental Implant Torque Driver | | Classification Name: | Handpiece, Direct Drive, AC-powered | | Classification: | Class I, 21 CFR 872.4200 | | Classification Product Code: | EKX | #### 4. Predicate Device: CORDLESS PROSTHODONTIC SCREWDRIVER WITH TORQUE CALIBRATION SYSTEM, MODEL ISD900 by NAKANISHI Co., Ltd. (K110278) #### 5. Device Description: This device is a cordless torque driver used for connecting and disconnecting dental implant abutments used in dentistry. The torque driver function includes adjustment of spinning direction {4}------------------------------------------------ (forward and reverse), speed and torque. The maximum torque is 35 Ncm. The device consists of a torque driver, a charge adapter and a charge cradle. #### 6. Indications for Use: This product is a cordless motor handpiece system intended for tightening and loosening an abutment screw to fix and remove the abutment on a dental implant in prosthodontic treatment. #### 7. Performance Data(Non-Clinical): The following properties were tested based on the referenced standards. All the test results support substantial equivalence to the predicate devices. - Performance test in accordance with ISO 14457:2012 Dentistry Handpieces and motors - . Sterilization validation testing in accordance with ANSVAAMI ST79, ISO17665-1, ISO 17665-2, ISO 11138-1, and ISO11138-3 - · Electrical safety and EMC test in accordance with IEC 60601-1, IEC 60601-1-6, EN 606011-1-2, EN 61000-3-2, EN 61000-3-3 #### Performance Data(Clinical): 8. No clinical performance testing was provided in the submission. #### 9. Substantial Equivalence MEG-TORQ is substantially equivalent to ISD900 (K110278). The following comparison table is presented to demonstrate substantial equivalence. The intended use of the subject device is the same as the predicate device. The subject device is also similar to the predicate device in terms of operational modes, type of chuck, patient contacting part, and shank type which confirms to ISO1797-1. The differences between the subject device and the predicate devices are device dimensions, speed in rpm, and torque options. However, the non-clinical performance test results provided in this submission support that the subject device is substantially equivalent to the predicate device. Especially, the maximum speed and torque test was performed to verify that the use of the subject device with the highest rpm does not shear off the abutment screw. The test results showed that this difference does not raise a concern in this regard. | | Submission Device | Predicate Device | |---------------|-------------------|------------------------------------------------------------------------------------------| | 510(k) Number | K161500 | K110278 | | Device Name | Meg-Torque | CORDLESS PROSTHODONTIC<br>SCREWDRIVER WITH TORQUE<br>CALIBRATION SYSTEM, MODEL<br>ISD900 | {5}------------------------------------------------ | Common Name | Handpiece, Direct Drive, AC-<br>Powered | Handpiece, Direct Drive, AC-<br>Powered | |------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | MICRO-NX Co., Ltd | NAKANISHI Co., Ltd. | | Intended Use | This product is a cordless motor<br>handpiece system intended for<br>tightening and loosening an<br>abutment screw to fix and remove<br>the abutment on a dental implant in<br>prosthodontic treatment. | This product is a cordless motor<br>handpiece system intended for<br>tightening and loosening an abutment<br>screw to fix and remove the abutment<br>on a dental implant in prosthodontic<br>treatment. | | Operational modes | Speed control, Torque control,<br>Rotate both forward/reverse<br>operation, Calibration | Speed control, Torque control, Rotate<br>both forward/reverse operation,<br>Calibration | | Air/water ports | N/A | N/A | | Fiberoptics | N/A | N/A | | Dimensions | 30×28×200(mm)<br>(Width×Length×Height) | 27×29×207(mm)<br>(Width×Length×Height) | | Type of chuck | Push Button Chuck | Push Button Chuck | | Coupling dimensions | N/A | N/A | | Chemical<br>composition of the<br>waterlines | N/A | N/A | | The patient-<br>contacting portions<br>of the device | The rotary instrument. | The rotary instrument. | | Light intensity | N/A | N/A | | Bur extraction force | N/A | N/A | | Maximum air/water<br>pressure | N/A | N/A | | Speed in rpms | 15, 30, 45, 60 rpm | 15, 20, 25rpm | | Torque | 5,15,20,25,30,35Ncm | 10-40Ncm in 1 or 5 Ncm increments | | Conformance with<br>standards for shanks | Screwdriver shank ø2.35mm<br>ISO1797-1 Type1 | Screwdriver shank ø2.35mm<br>ISO1797-1 Type1 | | Coupling dimensions | N/A | N/A | | Hose connections | N/A | N/A | #### 10. Conclusion: Based on the performance testing results, comparison of technology, and the indications for use, contained in the submission, MICRO-NX Co., Ltd. concludes that the MEG-TORQ is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...