Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with Chemotherapy Dr)
K161473 · Medline Industries, Inc. · LZA · Oct 19, 2016 · General Hospital
Device Facts
Record ID
K161473
Device Name
Medline Powder-Free Nitrile Exam Glove Extended Cuff (Tested for Use with Chemotherapy Dr)
Applicant
Medline Industries, Inc.
Product Code
LZA · General Hospital
Decision Date
Oct 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; extended cuff design. Used by healthcare professionals in clinical settings to provide a barrier against contamination between patient and examiner. Tested for resistance to chemotherapy drug permeation per ASTM D6978-05. Provides protection during handling of specific chemotherapy agents; breakthrough times vary by agent (e.g., Carmustine 40.4 min, Thiotepa 60.6 min). Single-use only; available in multiple sizes (S, M, L, XL).
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility evaluated per ISO 10993-10. Physical properties tested per ASTM D6319-10. Chemotherapy permeation tested per ASTM D6978-05.
Technological Characteristics
Material: Nitrile. Form factor: Extended cuff, blue color, non-sterile, single-use. Dimensions and physical properties (tensile strength, elongation) comply with ASTM D6319-10. Biocompatibility per ISO 10993-10. Permeation resistance tested per ASTM D6978-05.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner. Also indicated for use during handling of chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
EMG Blue Nitrile Examination Gloves Powder Free with Tested for Use with Chemotherapy Drugs (K141623)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 19, 2016
Medline, Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist One Medline Place Mundelein, Illinois 60060
Re: K161473
Trade/Device Name: Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: September 14, 2016 Received: September 16, 2016
Dear Jennifer Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161473
Device Name
Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs)
Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
The following chemicals have been tested with these gloves.
Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes
Carmustine 3.3mg/ml - 40.4 min. Cisplatin 1.0mg/ml - No breakthrough up to 240 min. Cyclophosphamide 20.0mg.ml - No breakthrough up to 240 min. Dacarbazine 2.0mg/ml - No breakthrough up to 240 min. Doxorubicin Hydrochloride 2/0mg.ml - No breakthrough up to 240 min. Etoposide 20.0mg/ml - No breakthrough up to 240 min. Fluoruouracil 50mg/ml - No breakthrough up to 240 min. Methotrexate 25mg - No breakthrough up to 240 min. Mitoxantrone 2.0mg/ml - No breakthrough up to 240 min. Paclitaxel 5.0mg/ml - No breakthrough up to 240 min. Thiotepa 10mg/ml - 60.6 min.
Please note that the following drugs have extremely low permeation times: Carmustine (3.3 mg/ml) has a minimum breakthrough time of 40.4 minutes; Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 60.6 minutes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the Medline logo, set against a blue background. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a stylized star or cross symbol. The symbol is white, contrasting with the blue background, and appears to be composed of four elongated triangles converging at a central point. A small gray rectangle is visible in the upper right corner of the image.
Aedline Industries, Inc. One Medline Place Mundelein, IL 60060
# K161473 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
#### Submitter / 510(k) Sponsor
Medline Industries, Inc. 1 Medline Place Mundelein, IL 60060
Registration Number: 1417592
#### Contact Person
Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: JAMason@medline.com
Summary Preparation Date September 14, 2016
Type of 510(k) Submission Traditional
#### Device Name / Classification
Name of Device: Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Proprietary Name: Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) Common Name: Patient Examination Glove Classification Name: Patient Examination Glove Product Code: LZA, LZC Classification Panel: General Hospital Regulatory Class: I Regulation #: 21 CFR 880.6250
#### Predicate Device
EMG Blue Nitrile Examination Gloves Powder Free with Tested for Use with Chemotherapy Drugs K141623
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is in white, bold, sans-serif font, and to the right of the word is a white starburst symbol. There is a gray bar on the right side of the image.
ledline Industries, Inc. One Medline Place Mundelein, IL 60060
#### Device Description
The Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) are single use only, disposable gloves intended for medical purposes to be worn on the hands or fingers of examiners. The gloves are powder-free and are made of nitrile with a blue colorant. The gloves are offered non-sterile and are available in small, medium, large and extra-large sizes.
Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. The subject device was tested according to ASTM D6978, Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Minimum breakthrough times were determined for a wide range of chemotherapy agents.
#### Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical Glove to Permeation by Chemotherapy Drugs.
| Chemotherapy Drugs | Breakthrough Times |
|------------------------------------|--------------------------------|
| Carmustine 3.3mg/ml | 40.4 min. |
| Cisplatin 1.0mg/ml | No breakthrough up to 240 min. |
| Cyclophosphamide 20.0mg/ml | No breakthrough up to 240 min. |
| Dacarbazine 2.0mg/ml | No breakthrough up to 240 min. |
| Doxorubicin Hydrochloride 2.0mg/ml | No breakthrough up to 240 min. |
| Etoposide 20.0mg/ml | No breakthrough up to 240 min. |
| Fluorouracil 50mg/ml | No breakthrough up to 240 min. |
| Methotrexate 25mg/ml | No breakthrough up to 240 min. |
| Mitoxantrone, 2.0mg/ml | No breakthrough up to 240 min. |
| Paclitaxel 5.0mg/ml | No breakthrough up to 240 min. |
| Thiotepa 10mg/ml | 60.6 min. |
The following chemicals have been tested with these gloves
Please note that the following drugs have extremely low permeation times: Carmustine (3.3 mg/ml) has a minimum breakthrough time of 40.4 minutes; Thiotepa (10.0 mg/ml) has a minimum breakthrough time of 60.6 minutes.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a sans-serif font, with a stylized white starburst symbol to the right of the text. The background is a solid dark blue color. The logo is simple and modern, and the colors are clean and professional.
Medline Industries, Inc. One Medline Place Mundelein, IL 60060
# Summary of Technological Characteristics
# TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
| Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Medline Powder-Free Blue<br>Nitrile Exam Glove Extended<br>Cuff (Tested for Use with<br>Chemotherapy Drugs) | EMG Blue Nitrile<br>Examination Gloves Powder<br>Free with tested for use with<br>chemotherapy drugs | Different |
| 510(k) Reference | K161473 | K141623 | N/A |
| Product Owner | Medline Industries, Inc. | Eco Medi Glove SDN BHD | Different |
| Product Code | LZA, LZC | LZA, LZC | Same |
| Intended Use | A powder-free patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or<br>finger to prevent<br>contamination between patient<br>and examiner and for use with<br>chemotherapy drugs. In<br>addition these gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 Standard Practice<br>for Assessment of Medical<br>Glove to Permeation by<br>Chemotherapy Drugs. | A powder-free patient<br>examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand or<br>finger to prevent<br>contamination between patient<br>and examiner and for use with<br>chemotherapy drugs. In<br>additional these gloves were<br>tested for use with<br>chemotherapy drugs in<br>accordance with ASTM<br>D6978-05 standards Practice<br>for assessment of Medical<br>Glove to Permeation by<br>Chemotherapy drugs. | Same |
| Regulation Number | 21 CFR 880.6250 | 21 CFR 880.6250 | Same |
| Materials | Nitrile | Nitrile | Same |
| Color | Blue | Blue | Same |
| Design Configurations | Small<br>Medium<br>Large<br>Extra-Large | Small<br>Medium<br>Large<br>Extra-Large | Same |
| Dimensions - Length | Complies with ASTM D6319-<br>10<br>270mm min. | Complies with ASTM D6319-<br>10<br>270mm min. | Same |
| Dimensions - Width | Complies with ASTM D6319-<br>10<br>Small - 80 ±10mm min.<br>Medium - 95 ±10mm min. | Complies with ASTM D6319-<br>10<br>Small - 85 ±5mm min.<br>Medium - 95 ±5mm min. | Same |
| | | | |
| | Large - 110±10mm min. | Large - 105±5mm min. | |
| | Extra-Large 120±10mm min. | Extra-Large 115±5mm min. | |
| Dimensions - Thickness | Complies with ASTM D6319-<br>10 | Complies with ASTM D6319-<br>10 | Same |
| | Palm - 0.10mm min. | Palm - 0.10mm min. | |
| | Finger - 0.10mm min. | Finger - 0.10mm min. | |
| Physical Properties -<br>Tensile | Complies with ASTM D6319-<br>10 | Complies with ASTM D6319-<br>10 | Same |
| | Before Aging - 14 MPa, min. | Before Aging - 14 MPa, min. | |
| | After Aging - 14 MPa, min. | After Aging - 14 MPa, min. | |
| Physical Properties -<br>Elongation | Complies with ASTM D6319-<br>10 | Complies with ASTM D6319-<br>10 | Same |
| | Before Aging – 500% min. | Before Aging – 500% | |
| | After Aging - 400% min. | After Aging - 400% | |
| Freedom from Holes | AQL 1.5 Inspection level G-2<br>Meets ASTM D6319-10 | AQL 2.5 Inspection level G-1<br>Meets ASTM D6319-00a | Similar |
| Residual Powder | Complies with ASTM D6319-<br>10 < 2.0 mg | Complies with ASTM D6319-<br>00a <2.0 mg | Same |
| Biocompatibility | ISO 10993-10:2010/(R) 2014<br>Under the conditions of the<br>test, not an irritant | Under the conditions of the<br>test, not an irritant | Same |
| | ISO 10993-10:2010/(R) 2014<br>Under the conditions of the<br>test, not a skin sensitizer | Under the conditions of the<br>test, not a skin sensitizer | |
| Prescription vs. OTC | OTC | OTC | Same |
| Sterile vs. Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Single Use vs. Reusable | Single use | Single use | Same |
| Tested for Use with<br>Chemotherapy Drugs | Yes ASTM 6978-05 | Yes ASTM D6978-05 | Same |
| Chemotherapy Drugs<br>Tested | Carmustine 3.3mg/ml<br>Cisplatin 1.0mg/ml | Carmustine 3.3mg/ml | Similar |
| | Cyclophosphamide 20.0mg/ml<br>Dacarbazine 2.0mg/ml | Cyclophosphamide 20.0mg/ml | |
| | Doxorubicin Hydrochloride<br>2mg/ml | Doxorubicin Hydrochloride<br>2mg/ml | |
| | Etoposide 20.0mg.ml | Etoposide 20.0mg.ml | |
| | Fluorouracil 50 mg/ml<br>Methotrexate 25mg/ml<br>Mitoxantrone 2.0mg/ml | Fluorouracil 50 mg/ml<br>Methotrexate 25mg/ml | |
| | Paclitaxel 5.0mg/ml<br>Thiotepa 10mg/ml | Paclitaxel 5.0mg/ml<br>Thiotepa 10mg/ml | |
| | | Cytarabine 100mg.ml | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the Medline logo against a blue background. The logo consists of the word "MEDLINE" in a sans-serif font, positioned to the left of a stylized starburst symbol. The starburst is white and appears to be composed of four elongated triangles converging at a central point. The overall design is clean and corporate, suggesting a professional and established brand identity.
Medline Industries, Inc. One Medline Place Mundelein, IL 60060
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo of Medline, a healthcare company. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a stylized star-like symbol. The star is white and has four points, with the bottom point extending further down than the others. The background is a solid dark blue color, providing a strong contrast to the white text and symbol.
Medline Industries, Inc. One Medline Place Mundelein, IL 60060
#### Summary of Non-Clinical Testing
The biocompatibility evaluation for the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) was conducted in accordance with ANSVAAMI/ISO 10993-1:2009 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. The Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) is classified as a surface contacting device with a limited contact duration of less than 24 hours.
The following tests were performed to evaluate the biocompatibility of the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs):
- ISO 10993-10: Primary Skin Irritation Test ●
- ISO 10993-10: Closed Patch Sensitization Test
#### Performance Testing (Bench)
Physical performance qualities were evaluated per ASTM D6319, Standard Specification for Nitrile Examination Gloves for Medical Application.
Permeation testing was conducted to support the addition of the labeling claim: Tested for use with chemotherapy drugs. The gloves were tested according to ASTM D6978, Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. Minimum breakthrough times were determined for a wide range of chemotherapy drugs.
## Summary of Clinical Testing
Not applicable.
#### Conclusion
Based on the non-clinical performance testing Medline Industries, Inc. concludes that the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) are as safe, as effective, and perform as well as the predicate device, EMG Blue Nitrile Medical Examination Gloves Powder Free with Tested for Use with Chemotherapy Drugs Labeling Claim (K141623). Therefore, the Medline Powder-Free Blue Nitrile Exam Glove, Extended Cuff (Tested for Use with Chemotherapy Drugs) are substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.