Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module
K161469 · Hospira, Inc. · FRN · Mar 28, 2017 · General Hospital
Device Facts
Record ID
K161469
Device Name
Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play Module
Applicant
Hospira, Inc.
Product Code
FRN · General Hospital
Decision Date
Mar 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Plum 360™ Infusion System with Hospira MedNet™ is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
Device Story
Volumetric infusion system for parenteral, enteral, epidural therapies, and blood product administration. Uses dedicated administration sets; features LCD display, alphanumeric keypads, audible/visual indicators, nurse call interface, and network connectivity. Operates via MedNet software for communication between pump, server, and facility systems. Clinicians program therapy via predefined screens; system manages infusion delivery. Updates include bolus functionality, modified dose ranges, and cybersecurity enhancements. Used in clinical settings to deliver precise fluid/medication volumes, improving therapy management and patient safety.
Clinical Evidence
Bench testing only. No clinical data required. Evidence includes system verification and validation, human factors evaluations for the bolus feature, and electrical/electromagnetic compatibility testing per IEC 60601 standards.
Technological Characteristics
Volumetric infusion pump. Connectivity: Networked via MedNet server. Standards: IEC 60601-1 (safety), IEC 60601-2-24 (infusion pump performance), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms). Software: CPU software v15.10, Communication Engine v3.03.
Indications for Use
Indicated for patients requiring parenteral, enteral, or epidural therapies, or administration of whole blood and blood products.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Plum 360™ Infusion System with MedNet / Smart Card Plug 'n' Play (K141789)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 28, 2017
Hospira, Inc Keith Dunn Associate Director, Global Regulatory Affairs, Devices 275 North Field Drive Lake Forest, Illinois 60045
Re: K161469
Trade/Device Name: Plum 360™ Infusion System with MedNet/ Smart Card Plug 'n' Play Module Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN. PHC. FPA Dated: February 22, 2017 Received: February 28, 2017
Dear Keith Dunn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature followed by the name "Tina Kiang". Below the name is the letter "s" preceded by a hyphen. The signature is a series of loops and curves.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161469
Device Name
Plum 360™ Infusion System with Hospira MedNet™/ Smart Card Plug 'n' Play Module
Indications for Use (Describe)
Plum 360™ Infusion System with Hospira MedNet™ is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5: 510(k) Summary
A summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
| Submitter Information | |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | ICU Medical |
| Address | D-393<br>600 North Field Drive<br>Lake Forest, IL. 60045 |
| Phone number | 224-212-4898 |
| Fax number | 224-212-5402 |
| Establishment<br>Registration<br>Number | 3005579246 |
| Name of contact<br>person | Yuliya Matlin |
| Date prepared | March 28, 2017 |
| Name of device | |
| Trade or<br>proprietary<br>name | Plum 360TM Infusion System with MedNet/ Smart Card Plug 'n' Play Module |
| Common or<br>usual name | Infusion Pump |
| Classification<br>name | Infusion Pump |
| Classification panel | 80 |
| Regulation | 21 CFR 880.5725 |
| Product Code(s) | FRN, PHC, FPA |
| Legally marketed<br>device(s) to which<br>equivalence is<br>claimed | Infusion System:<br>Plum 360TM Infusion System with MedNet / Smart Card Plug 'n' Play (K141789) |
| Device description | The Plum 360TM Infusion System withMedNet / Smart Card Plug 'n' Play Module remains a<br>volumetric Infusion System intended for use in parenteral, enteral and epidural therapies and the<br>administration of whole blood and blood products.<br>Dedicated administration sets are used for delivery of infusion therapy to the patient. The device<br>has a user interface consisting of LCD display, front panel alphanumeric keypads and keys,<br>audible and visual indicators, audible indicator volume control, nurse call interface, network<br>interface connectors, the drug library and software that provides pre-defined display screens for<br>therapy programming.<br>The continued use of the MedNet Software allows communication between the Infusion Pump,<br>MedNet server and the facility's communication systems. |
| Indication for Use | The Plum 360TM Infusion System with Hospira MedNet TM is indicated for use in parenteral,<br>enteral, and epidural therapies and the administration of whole blood and blood products. |
{4}------------------------------------------------
| Summary of<br>Indications for Use<br>Compared to<br>Predicate Device<br>Summary of the | | | With exception to the system name, the Indications for Use is identical to the predicate device.<br>Plum 360™ Infusion System with MedNet/ Smart Card Plug 'n' Play employs the same |
|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics<br>Compared to<br>Predicate Device | fundamental scientific technology, and principles of operation and intended use as the predicate<br>device. In addition, Plum 360™ Infusion System with MedNet/ Smart Card Plug 'n' Play<br>Module has the following technological characteristic enhancements compared to predicate<br>device with the following changes related to the bolus feature: | | |
| | | Feature | Subject Plum 360 Infusion System with<br>Bolus Feature Compared to Predicate Plum<br>360 Infusion System |
| | | CPU Software | Updated to 15.10<br>Software differences included introduction of<br>bolus functionality and associated user<br>interface modifications, addition of the dosing<br>units and modifications to programmable dose<br>ranges, infusion log updates and cybersecurity<br>enhancements |
| | | Communication Engine (CE) | Updated to 3.03<br>Software differences included Cybersecurity<br>enhancements and changes to support<br>additional bolus functionality |
| | | | |
{5}------------------------------------------------
| Performance Data -<br>Non Clinical Testing | The following standards and associated methods were applied to demonstrate the Plum 360™<br>Infusion System with MedNet / Smart Card Plug 'n' Play Module is substantially equivalent to |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | the current device after the historical and subject changes are applied<br>· IEC 60601-1Medical Electrical Equipment-Part 1: General Requirement for Basic<br>Safety and Essential Performance, 2005/(R)2012/C1:2009/(R)2012, Including |
| | Amendment 1:2012 |
| | · IEC 60601-2-24 Medical electrical equipment - Part 2-24: Particular requirements for<br>the basic safety and essential performance of infusion pumps and controllers, Edition<br>2.0 2012 |
| | · IEC 60601-1-2 Medical electrical equipment- part 1-2: General requirements for basic<br>safety and essential performance- Collateral standard: Electromagnetic compatibility -<br>Requirements and tests, Edition 3:2007 |
| | · IEC 60601-1-8 Collateral Standard: General Requirements, test and guidance for alarm<br>systems in medical electrical equipment and medical electrical systems, Edition 2.0<br>2006 |
| | A risk analysis was applied to assess the impact of modifications necessary to include a bolus<br>feature with the system. The analysis demonstrates the changes do not introduce new risks to<br>the patient. All risks which are identified have been mitigated as low as reasobaly possible. |
| | The device safety assurance case was updated to include the bolus feature that is being added to<br>the Plum 360™ Infusion System with MedNet / Smart Card Plug 'n' Play Module, as<br>recommended in the FDA guidance document, Infusion Pumps Total Product Life Cycle |
| | |
| | The assurance case defined the device system, including the indications for use, system<br>specifications, use environments, and user population. The supporting assurance arguments<br>covered the following attributes: |
| | · All potential risks have been mitigated and residual risk is acceptable. |
| | · Design verification and validation of device specifications are adequate. |
| | · Device reliability is adequate. |
| | The following evidence was included in the assurance case: |
| | · Performance Test Summary: System verification and validation activities for Plum 360<br>Infusion System confirmed that the system meets user needs and design inputs. All the<br>testing met the acceptance criteria. |
| | · Human factors evaluations have been conducted to validate the effectiveness of the bolus<br>feature. |
| | · Software verification and validation as recommended in the following FDA guidance<br>documents: |
| | Infusion Pumps Total Product Life Cycle<br>O |
| | Content of Premarket Submissions for Software Contained in Medical Devices<br>O<br>Applying Human Factors and Usability Engineering to Medical Devices<br>O |
| | |
| | Electrical and Electromagnetic Compatibility testing were conducted. Updates due to the bolus<br>feature complies with following standard: |
| | IEC 60601-2-24 Medical Electrical Equipment - Part-24:<br>Infusion Pump Basic<br>Safety and Essential<br>Edition 2.0 2012-10<br>Performance |
| | |
| Performance Data –<br>Clinical Testing | Clinical evaluation is not required for this submission to support substantial equivalence.<br>Human Factors studies have been conducted on subject devices, demonstrating passing results. |
{6}------------------------------------------------
## CONCLUSIONS DRAWN FROM NON-CLINICAL DATA
The conclusions drawn from the non-clinical tests demonstrate that the Plum 360™ Infusion System with MedNet / Smart Card Plug 'n' Play Module is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.