K161468 · Transformair, LLC · FRA · Feb 8, 2017 · General Hospital
Device Facts
Record ID
K161468
Device Name
Transformair Indoor Air Purifier
Applicant
Transformair, LLC
Product Code
FRA · General Hospital
Decision Date
Feb 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6500
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation. Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions: [Table of log reduction data]. The Transformair Indoor Air Purifier is for Over-The-Counter use.
Device Story
In-duct air purification device; installed in building HVAC systems. Uses photo-electrocatalytic (PEC) technology to destroy airborne microorganisms. Input: ambient air containing bacteria, viruses, and mold spores. Process: air passes through a non-woven polyester pre-filter, then a catalytic filter coated with titanium dioxide (TiO2) and metal wire mesh; 6 UV-A lamps (320-400 nm) illuminate the catalytic surface, triggering a photochemical reaction that generates hydroxyl free radicals. Output: purified air with reduced microbial load. Operated by facility maintenance/HVAC personnel; intended for medical facility environments. Benefits patients/staff by reducing airborne pathogen concentrations.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including GLP-compliant lethality kinetics testing against Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger, and Bacillus globigii across temperature ranges (45°F-110°F), and UL electrical/fire safety testing.
Technological Characteristics
In-duct medical UV air purifier. Materials: non-woven polyester pre-filter, TiO2-coated polyester catalytic filter, metal wire mesh, metal housing. Energy: 120/220V AC, 6 UV-A lamps (320-400 nm). Dimensions: 28.5" x 21" x 12.25". Standards: ANSI/UL 1995-2011, CAN-CSA C22.2 No. 236-11, UL 2043.
Indications for Use
Indicated for destruction of airborne bacteria, viruses, and mold spores in medical facilities. Intended for use in HVAC ducts to reduce microbiological contaminants.
Regulatory Classification
Identification
A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
Predicate Devices
ABRACAIR Air Cleaners QTZ300-60 and TI 100-30P (K052732)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, forming a symbolic representation of the department's mission related to health and human well-being.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2019
Transformair, LLC % Dave Yungvirt CEO Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041
Re: K161468
Trade/Device Name: Transformair Indoor Air Purifier Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: January 31, 2017 Received: February 1, 2017
Dear Dave Yungvirt:
This letter corrects our substantially equivalent letter of February 8, 2017
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161468
Device Name Transformair Indoor Air Purifier
#### Indications for Use (Describe)
The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions:
| Organism | Name | Average Maximum log reduction /exposure time (hours) | | |
|---------------------|-------------------|------------------------------------------------------|-------------------|-------------------|
| | | Test temperature | | |
| | | 45°F | 72°F | 110°F |
| Bacteria | Staphylococcus | 4.88 / 24 hours | 4.02 / 0.33 hours | 4.20 / 0.33 hours |
| Bacteria | Escherichia coli | 4.31 / 24 hours | 4.79 / 24 hours | 4.40 / 0.33 hours |
| Virus | MS2 bacteriophage | 4.13 / 24 hours | 4.25 / 24 hours | 5.51 / 24 hours |
| Virus | Phi-X174 | 4.37 / 24 hours | 4.37 / 24 hours | 4.37 / 24 hours |
| Mold endospore | Aspergillus Niger | 3.91 / 72 hours | 3.99 / 72 hours | 4.22 / 72 hours |
| Bacterial endospore | Bacillus globigii | 4.11 / 72 hours | 4.41 / 72 hours | 4.41 / 72 hours |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "transformair" in a stylized font, with a series of blue dots and a yellow arc above it. Below the word, the address "3802 Spectrum Blvd. Suite 143, Tampa, FL 33612" is printed in a smaller, simpler font. The address provides a specific location in Tampa, Florida.
# 510(k) SUMMARY
# Transformair Air Purifier
# Submitter:
| Company Name: | Transformair, Inc. |
|------------------|-----------------------------------------------------|
| Company Address: | 3802 Spectrum Blvd.<br>Suite 143<br>Tampa, FL 33612 |
Telephone Number: 352-871-3803 Contact Person: Jaya Rao Date Prepared: 1/19/2017
- Device:
| Proprietary Name: | Transformair Indoor Air Purifier |
|-----------------------|------------------------------------------------------|
| Common or Usual Name: | Air purifier ultraviolet or ultraviolet air purifier |
| Classification Name: | Medical Ultraviolet Air Purifier |
| Regulation Number: | 21 CFR 880.6500 |
| Product Code: | FRA |
| Device Class: | Class 2 |
| Category: | General Hospital |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "transformair" in a muted green color. Above the logo is a curved line of blue dots and a yellow line. Below the logo is the address "3802 Spectrum Blvd. Suite 143, Tampa, FL 33612".
#### Predicate Device:
The predicate device is shown in the following table:
#### Table 1: Predicate Device
| Predicate Device | ABRACAIR Air Cleaners QTZ300-60 and TI 100-30P |
|---------------------|------------------------------------------------|
| Manufacturer | Abracair, Inc. |
| 510(k) Number | K052732 |
| Regulation Number: | 21 CFR 880.6500 |
| Regulation Name: | Medical Ultraviolet Air Purifier |
| Regulatory Class | II |
| Product Code : | FRA |
| Review Panel: | General Hospital |
| Combination Product | No |
#### Intended Use / Indications for Use:
The Transformair Indoor Air Purifier, In Duct Model 16108 is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
Transformair Indoor Air Purifier, In Duct Model 16108 has been demonstrated to destroy Staphylococcus epidermidis, Escherichia coli, MS2, Phi-X174, Aspergillus Niger and Bacillus globigii entrained on the filter of the subject device under the following exposure conditions:
Table 2. Transformair Lethality Testing Results
| Organism | Name | Average Maximum log reduction /exposure time (hours) | | |
|---------------------|----------------------------|------------------------------------------------------|------------------|------------------|
| | | Test temperature | | |
| | | 45°F | 72°F | 110°F |
| Bacteria | Staphylococcus epidermidis | 4.88/ 24 hours | 4.02 /0.33 hours | 4.20 /0.33 hours |
| Bacteria | Escherichia coli | 4.31/ 24 hours | 4.79 /24 hours | 4.40 /0.33 hours |
| Virus | MS2 bacteriophage | 4.13/ 24 hours | 4.25/ 24 hours | 5.51/ 24 hours |
| Virus | Phi-X174 bacteriophage | 4.37/ 24 hours | 4.37/ 24 hours | 4.37/ 24 hours |
| Mold endospore | Aspergillus Niger | 3.91/ 72 hours | 3.99/ 72 hours | 4.22/ 72 hours |
| Bacterial endospore | Bacillus globigii | 4.11/ 72 hours | 4.41/ 72 hours | 4.41/ 72 hours |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "transformair" in a muted gray color. Above the logo is a curved line of blue dots and a yellow line. Below the logo is the address "3802 Spectrum Blvd. Suite 143, Tampa, FL 33612".
The Transformair Indoor Air Purifier is for Over-The-Counter use.
# Device Description:
Transformair® is a patented photo-electrochemical or photo-electrocatalyst (PEC) ultraviolet air purification technology that destroys bacteria in air in medical facilities. It consists of a pre-filter, black light lamps, and a catalytic filter coated with a photocatalyst. The Transformair device has a metal housing with electronic controls and outer dimensions of 28.5 inches by 21 inches by 12.25 inches that is installed in the duct of a buildings HVAC system. The Transformair device has a pre-filter and a chamber equipped with 6 low energy ultraviolet lights (UV-A), which emit light with wavelengths between 320-400 nm, and a catalytic filter.
# Technological Characteristics and Principles of Operation:
The Transformair Air Purifier is a medical ultraviolet air purifier unit that traps microorganisms on a filter where they are destroyed by UV photo-catalyzed generation of free radicals in the validated time periods. The air cleaning chamber consists of a metal housing that contains a non-woven polyester prefilter, glass UV-A lamps (~320-400 nm), a catalytic filter. The metal housing has electronics that plug into a standard outlet (120/220 Volt) to power the UV-A lamps.
The Transformair Air Purifier destroys microbiological contaminants in the air through a photochemical reaction that produces hydroxyl free radicals when UV-A light rays are shone on the catalytic surface.
# Substantial Equivalence:
The Transformair Air Purifier is substantially equivalent to the ABRACAIR Air Cleaner (K052732) since the devices have the same general intended use and similar indications, technological characteristics, and principles of operation. A reference table is provided below comparing the Transformair Air Purifier to the predicate.
| Element of<br>Comparison | Transformair | ABRACAIR (K052732) |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Transformair, Inc. | Abracair, Inc. |
| Device type | Medical Ultraviolet Air Purifier | Medical Ultraviolet Air Purifier |
| Element of<br>Comparison | Transformair | ABRACAIR (K052732) |
| Regulation Number | 21 CFR 880.6500 | 21 CFR 880.6500 |
| Intended Use /<br>Indications for Use | The Transformair Indoor Air Purifier, In<br>Duct Model 16108 is a device intended<br>for medical purposes that is used to<br>destroy bacteria in the air by exposure to<br>ultraviolet radiation.<br><br>Transformair Indoor Air Purifier, In Duct<br>Model 16108 has been demonstrated to<br>destroy Staphylococcus epidermidis,<br>Escherichia coli, MS2, Phi-X174,<br>Aspergillus Niger and Bacillus globigii<br>entrained on the filter of the subject<br>device under the exposure conditions<br>described in Table 2.<br><br>The Transformair Indoor Air Purifier is<br>for Over-The-Counter use. | The Abracair Air Cleaner is intended for<br>the reduction of aerosolized mold and<br>bacteria within hospitals, nursing homes,<br>medical facilities. The device may be<br>used in occupied spaces within hospital<br>environments such as baby and/or<br>neonatal nurseries, hospital rooms,<br>operating rooms, mortuaries, embalming<br>rooms. |
| Mechanism of<br>action | UV light of sufficient energy (UV-A)<br>activates photocatalyst that destroys<br>microorganisms entrained on the filter<br>through a photochemical reaction that<br>is enhanced by a metal wire mesh. | UV light of sufficient energy (UV-C)<br>activates photocatalyst that destroys<br>microorganisms through a<br>photochemical reaction; UV-C light<br>directly destroys microorganism. |
| Model(s) | Model 16108 | QTZ300-60 and QTZ100-24 |
| Installation | In-Duct | Free standing |
| Filter(s) | Pre-filter made with polyester and<br>catalytic filter made with polyester and<br>coated with titanium dioxide (TiO2) | Filter made with quartz fibers coated with<br>titanium dioxide (TiO2) |
| Photocatalyst | Catalytic filter coated with TiO2 and<br>metal wire mesh | Filter coated with TiO2 |
| Light source | UV-A lamps (~320-400 nm) | UV-C Xenon Flash lamps (~100-285nm) |
| Element of<br>Comparison | Transformair | ABRACAIR (K052732) |
| Voltage | Up to 120 Volt / 220 Volt (plugs into<br>standard outlet) (See Footnote1) | 240 Volt AC / 480 Volt AC |
| Current | Up to 1.25 amps (See Footnote1) | 30 Amps at 240 Volts |
| Power Consumption | Up to 240 Watts (See Footnote1) and<br>Footnote2) | Up to 7020 Watts |
| Air cleaning<br>chamber | Particulate pre-filter, UV-A lamps and a<br>catalytic filter coated with TiO2 and<br>metal wire mesh | UV-C lamps and a catalytic filter coated<br>with TiO2 |
| Electronics | UL compliant | UL compliant |
| Dimensions | 28.5 inches by 21 inches by 12.25 inches | Not available |
#### TABLE 3: SUBSTANTIAL EQUIVALENCE COMPARISON CHART
{6}------------------------------------------------
{7}------------------------------------------------
3802 Spectrum Blvd. Suite 143, Tampa, FL 33612
ً This does not raise any new issues of safety and efficacy since the Transformair Air Purifier complies with the applicable UL standards for an air duct mounted device.
ً Transformair Air Purifier is different from the primary predicate device since it uses upto 240 W of UV-A light (320-400 nm) with a TiO-coated catalytic filter and a metal wire mesh to enhance activation of the TiO2 photocatalyst instead of upto 7020 W of UV-C light (~100-285nm) used by the primary predicate. This does not raise any new issues of efficacy since the ability of the Transformair device to reduce the concentration of airborne microorganisms is substantiated by the performance testing.
The Transformair Air Purifier and the predicate ABRACAIR Air Cleaners use the action of UV light on a catalytic filter coated with a Titanium Oxide (TiO2) photocatalyst to destroy microbiological contaminants in the air. Both devices use conventional UV-light lamps and filter materials but, unlike the predicate, the Transformair Air Purifier uses UV-A lamps instead of UV-C lamps and has a metal wire mesh on the catalytic filter. Both devices are designed to operate in a medical facility.
The Transformair air purifier has the same intended use of reducing airborne bacteria and mold spores as the predicate device ABRACAIR Air Cleaner (K052732). In addition, the Transformair air purifier is intended to reduce viruses (MS2 and Phi-X174).
The differences in the indications for use and technological characteristics are supported by the performance testing and UL electrical safety testing and do not raise new questions of safety or efficacy.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Transformair, a company located at 3802 Spectrum Blvd. Suite 143, Tampa, FL 33612. The logo features the company name in a stylized font, with a series of blue dots and a yellow wave above the name. The address is printed in a smaller font below the logo.
# Non-Clinical Performance Testing Summary:
No performance standards have been established for such devices under Sections 514 of the Federal Food, Drug and Cosmetic Act. However, the Transformair Air Purifier has been tested to demonstrate that the device does not generate Ozone. It has also been certified compliant with the following UL standards:
- । ANSI/UL 1995-2011 & CAN-CSA C22.2 No. 236-11, Heating and Cooling Equipment and
- -UL 2043, Heat and Visible Smoke Release For Discrete Products and Their Accessories Installed in Air-Handling Spaces.
Reference the Table below for specific UL tests conducted.
| Test Conducted | UL Reference Standard | Test Results |
|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|--------------|
| Input Test (Circuits – Other than Electric<br>Heaters) (+) | UL 1995, 4th Ed., Section 40 | PASS |
| Starting Test | UL 1995, 4th ed. Section 69 | PASS |
| Leakage Current Test | UL 1995, 4th ed. Section 68 | PASS |
| Temperature Operation Test (Without Any<br>Supplementary Heating Means)(+) | UL 1995, 4th Ed., Section 41 | PASS |
| Dielectric Voltage-Withstand Test | UL 1995, 4th Ed., Section 54 | PASS |
| Fire Test for Heat and Visible Smoke Release<br>For Discrete Products and Their Accessories<br>Installed in Air-Handling Spaces | UL 1995, 4th Ed., Section 18 and UL<br>2043, 3rd Ed. | PASS |
#### Table 4: Underwriters Laboratories Testing Summary
Third party testing was also performed to evaluate filter performance using aerosolized suspensions of the claimed microorganisms. This testing demonstrated retention of the test organisms on the filter. This testing was done in compliance with FDA Good Laboratory Practices (GLP) as defined in 40 CFR, Part 160.
Lastly, third party testing was performed to establish the kill kinetics for bacteria (Staphylococcus epidermidis and Escherichia coli), viruses (MS2 and Phi-X174), and mold spores (Aspergillus Niger) and bacterial spores (Bacillus globigii) over the range of temperatures expected to occur in an HVAC duct. This testing was done in compliance with FDA Good Laboratory Practices (GLP) as defined in 40 CFR, Part 160. Reference Table below for the lethality testing results.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Transformair, a company located at 3802 Spectrum Blvd. Suite 143, Tampa, FL 33612. The logo features the company name in a stylized font, with a series of blue dots and a yellow line above the name. The address is printed in a smaller font below the logo.
# Table 5: Transformair Device Lethality Testing Results
| Category | Aerosolized<br>Biological | Test Temperature<br>(°F) | Exposure time<br>(hours) | Average Maximum log<br>reduction |
|--------------------|-------------------------------|--------------------------|--------------------------|----------------------------------|
| Bacteria | Staphylococcus<br>epidermidis | 45°F | 24 hours | 4.88 |
| | | 72°F | 0.33 hour | 4.02 |
| | | 110°F | 0.33 hour | 4.20 |
| Bacteria | Escherichia<br>coli | 45°F | 24 hours | 4.31 |
| | | 72°F | 24 hours | 4.79 |
| | | 110°F | 0.33 hour | 4.40 |
| Virus | MS2<br>bacteriophage | 45°F | 24 hours | 4.13 |
| | | 72°F | 24 hours | 4.25 |
| | | 110°F | 24 hours | 5.51 |
| Virus | Phi-X174<br>bacteriophage | 45°F | 24 hours | 4.37 |
| | | 72°F | 24 hours | 4.37 |
| | | 110°F | 24 hours | 4.37 |
| Mold<br>spore | Aspergillus<br>Niger spores | 45°F | 72 hours | 3.91 |
| | | 72°F | 72 hours | 3.99 |
| | | 110°F | 72 hours | 4.22 |
| Bacterial<br>spore | Bacillus<br>globigii spores | 45°F | 72 hours | 4.11 |
| | | 72°F | 72 hours | 4.41 |
| | 110°F | 72 hours | 4.41 | |
| | | | | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for "transformair" in a stylized font. Below the logo is the address "3802 Spectrum Blvd, Suite 143, Tampa, FL 33612". The logo has a series of blue dots and a yellow swoosh above the word "transformair".
# Clinical Testing Summary:
Clinical testing has not been performed for the Transformair Air Purifier. Non-clinical bench testing is sufficient to demonstrate substantial equivalence to the predicate ABRACAIR Air Cleaner.
#### Conclusion:
The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.