CardioConfirm is intended to be used by qualified clinicians to finalize cardiology reports for adult and pediatric patients. To aid the clinician's interpretation of resting ECGs, CardioConfirm supports the display and onscreen measurement of ECG waveforms. CardioConfirm also supports the algorithmic comparison of serial adult ECGs and allows the user to edit the automatic comparison statements. CardioConfirm is intended to be controlled by a host application that manages user authentication, user permissions, and secure persistent storage of the tests.
Device Story
CardioConfirm is a software component (DLL or executable) embedded in host applications (e.g., EMR) to manage diagnostic cardiology reports. It receives DICOM 12-lead ECG waveforms and encapsulated PDF objects from Mortara or third-party equipment. Clinicians use the device to view, measure, and perform serial comparisons of resting ECGs; edit patient demographics, interpretations, and conclusions; and electronically sign reports. The VERITAS SERIAL COMPARISON algorithm automatically generates comparative statements for adult ECGs based on prior interpretations and waveforms. The device provides tools like an onscreen magnifying loupe and conclusion templates. It facilitates clinical decision-making by allowing physicians to review, amend, and finalize diagnostic tests, which are then stored in the host application. It is intended for use by qualified clinicians in clinical settings.
Clinical Evidence
No clinical data was required to support substantial equivalence. Evidence is based on non-clinical software verification and validation testing conducted in accordance with FDA guidance for software in medical devices.
Technological Characteristics
Windows-based software (DLL or executable) designed per 21 CFR Part 820. Functions as a data management and analysis tool for ECG, stress, and Holter reports. Supports DICOM 12-lead ECG waveform and encapsulated PDF objects. Features include onscreen measurement, magnifying loupe, and serial comparison algorithms. Connectivity is managed via host application integration. No specific hardware materials or energy sources; software-only device.
Indications for Use
Indicated for adult and pediatric patients requiring cardiology report finalization, including resting ECG, stress, and Holter reports.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Public Health Service
September 3, 2016
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Mortara Instrument, Inc. Margaret Mucha Director of Global Regulatory Affairs 7865 North 86th Street Milwaukee. Wisconsin 53224
Re: K161465
Trade/Device Name: CardioConfirm Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: August 1, 2016 Received: August 2, 2016
Dear Margaret Mucha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K161465
Device Name CardioConfirm
Indications for Use (Describe)
CardioConfirm is intended to be used by qualified clinicians to finalize cardiology reports for adult and pediatric patients. To aid the clinician's interpretation of resting ECGs, CardioConfirm supports the display and onscreen measurement of ECG waveforms. CardioConfirm also supports the algorithmic comparison of serial adult ECGs and allows the user to edit the automatic comparison statements.
Cardio Confirm is intended to be controlled by a host application that manages user authentication, user permissions, and secure persistent storage of the tests.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo above the text "Traditional 510(k) Notification K161465". The logo is a white heart shape with a blue background. The text is centered below the logo.
# Section 5
#### 510(k) Summary Statement
#### 1. Submitter
Primary Contact Mortara Instrument, Inc. 7865 North 86th Street Margaret Mucha Milwaukee, WI 53224 Director of Global Requlatory Affairs Meg.mucha@mortara.com Telephone 414-354-1600 Secondary Contact Fax 414-354-4760 Sarah Weber Senior Regulatory Affairs Manager sarah.weber@mortara.com
## 2. Product Names
CardioConfirm Device Trade Name
Common/ Usual Name
Classification
Electrocardiograph
870.2340 DPS Electrocardiograph 870.1425 DQK Computer, diagnostic, programmable Cardiovascular
Note: There are no previous submissions for this device
## 3. Predicate Devices to which this is Substantially Equivalent
| E-Scribe (primary) | K930558 |
|-------------------------------|---------|
| Pyramis ECG Management System | K032038 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with a heart shape at the top. Below the logo, the text "Traditional 510(k) Notification" is displayed. The number "K161465" is printed below the notification text. The logo and text are centered on a white background.
These predicate devices have not been subject to a design-related recall
#### 4. Device Description
CardioConfirm is a software component that is embedded into a host application (e.g. EMR) as a Dynamic Link Library (DLL) or launched as an executable application. CardioConfirm allows the user to display, edit and finalize diagnostic cardiology ECG, Stress and Holter reports. A Clinician can open a diaqnostic cardiology test from a workstation application and pass it to CardioConfirm as a DICOM object for viewing or editing. CardioConfirm is able to open and edit diagnostic cardiology ECG, Stress and Holter reports from Mortara Instrument, Inc. devices and offers an optional version which can edit diagnostic cardiology ECG, Stress and Holter reports from multiple diagnostic cardiology equipment vendors.
CardioConfirm allows the clinician the ability to view, measure, and compare previously acquired resting ECGs as well as perform visual and textual serial comparison of acquired ECGs. With visual serial comparison, the clinician can view current and previous ECG's simultaneously superimposed on the screen. Textual serial comparison allows the clinician to enter or edit text that describes the significant differences between the ECG's. For adult ECG's the initial text can be generated by the VERITAS SERIAL COMPARISON algorithm. The algorithm generates comparative statements automatically on the basis of the VERITAS INTERPRETATION statements originally made by the Mortara acquiring devices, measurements and the ECG waveforms. The clinician can edit the patient demographics, previous measurements and interpretations and the serial comparison statements, and then sign final ECG reports.
The clinician can also view and edit stress and Holter reports that includes editing patient demographics, summary statistics and physician conclusions. The clinician can electronically sign both stress and Holter final reports in CardioConfirm and print final reports or generate PDF copies to store in the host application. If a physician wants to amend a previously finalized report with additional information, CardioConfirm has an option to perform amendments to final reports and store the updated test in the host application.
CardioConfirm is available in two versions. The standard version is intended to be used with tests generated by Mortara equipment. The professional version is intended to be used with tests generated by any
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo above the text "Traditional 510(k) Notification K161465". The logo is a blue shape that resembles a heart. The text is in a simple, sans-serif font and is centered below the logo. The text "K161465" is on the line below the notification.
equipment that produces DICOM 12-lead ECG Waveform and Encapsulated PDF objects.
#### 5. Intended Use
CardioConfirm is intended to be used by qualified clinicians to finalize cardiology reports for adult and pediatric patients. To aid the clinician's interpretation of resting ECGs, CardioConfirm supports the display and onscreen measurement of ECG waveforms. CardioConfirm also supports the algorithmic comparison of serial adult ECGs and allows the user to edit the automatic comparison statements.
CardioConfirm is intended to be controlled by a host application that manages user authentication, user permissions, and secure persistent storage of the tests.
#### 6. Technological characteristics
CardioConfirm employs the same functional scientific technology as its predicate device E-Scribe (K930558) and Pyramis ECG Management System (K032038). At a high level, all three devices employ a Windows based tool to transfer electrocardiographic data from resting ECGs, stress, and Holter exams to further process and analyze the data, prepare final reports, and archive the data and reports.
CardioConfirm was designed and manufactured by Mortara Instrument according to 21 CFR Part 820. CardioConfirm is substantially equivalent to E-Scribe (Predicate K930558) and Pyramis ECG Management System (Predicate K032038) with the following technological differences:
- CardioConfirm is imbedded as a DLL in a host application such as an ● EMR or EHR
- CardioConfirm can be made available as a Windows executable file . that can be controlled by a host application
- CardioConfirm can print reports and waveforms ●
- CardioConfirm can accept demographic updates outside of ● CardioConfirm
- For Resting ECG Reports: .
- o CardioConfirm reads DICOM 12-Lead ECG waveform objects
- o CardioConfirm allows the user to edit previously-signed interpretations
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a white heart shape inside a blue square. Below the logo, the text "Traditional 510(k) Notification" is displayed, followed by the code "K161465" on the next line. The text and code appear to be related to a regulatory notification process, possibly in the medical field.
- CardioConfirm provides an onscreen magnifying loupe for Resting o ECG waveforms
- o CardioConfirm allows the user to edit the statement prediction in reports
- For Stress and Holter Reports: .
- o CardioConfirm will read DICOM Encapsulated PDF objects
- o CardioConfirm will write DICOM Encapsulated PDF objects
- o CardioConfirm will allow the user to use conclusion templates
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with a white heart shape inside a blue square. Below the logo, the text "Traditional 510(k) Notification" is displayed, followed by the code "K161465" on the next line. The text is in a simple, sans-serif font and is centered below the logo.
| Data<br>Management<br>System | CardioConfirm | E-scribe (Primary) | Pyramis ECG<br>Management System | Change<br>explanation | |
|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|----------------------------------|
| BRAND | Mortara | Mortara | Quinton | | |
| COMPANY | Mortara Instrument, Inc. | Mortara Instrument, Inc. | Quinton Inc. | | |
| 510 (K) Number | Present Application | K930558 | K032038 | | |
| Intended Use | CardioConfirm is intended to<br>be used by qualified<br>clinicians to finalize<br>cardiology reports for adult<br>and pediatric patients. To<br>aid the clinician's<br>interpretation of resting<br>ECGs, CardioConfirm<br>supports the display and<br>onscreen measurement of<br>ECG waveforms.<br>CardioConfirm also supports<br>the algorithmic comparison<br>of serial adult ECGs and<br>allows the user to edit the<br>automatic comparison<br>statements.<br>CardioConfirm is intended to<br>be controlled by a host<br>application that manages<br>user authentication, user<br>permissions, and secure | This system is designed to<br>be used as a storage<br>system for ECG data in<br>hospitals and large<br>medical clinics. The E-<br>Scribe allows the user to<br>store, edit and retrieve<br>patient ECG information | The Pyramis ECG<br>Management System is an<br>ECG data management<br>and information system.<br>Its primary function is to<br>store records of biologic<br>origin such as ECG,<br>Stress and Holter records<br>received from those<br>respective recording<br>devices in a database and<br>subsequently allow the<br>user to select, view, edit,<br>print/fax/email and export<br>(distribute) those records | Equivalent | |
| | persistent storage of the | | (distribute) those records | | |
| Data<br>Management<br>System | CardioConfirm | E-scribe (Primary) | Pyramis ECG<br>Management System | Change<br>explanation | |
| BRAND | Mortara | Mortara | Quinton | | |
| COMPANY | Mortara Instrument, Inc. | Mortara Instrument, Inc. | Quinton Inc. | | |
| 510 (K) Number | Present Application<br>tests. | K930558 | K032038 | | |
| Operating<br>System | Windows 7, Windows 8.1, etc. | Windows 2008 R2, Windows 7 | Windows 2008, Windows 7 | Equivalent | |
| Form: DLL for<br>embedding in a<br>host application | Yes | Not applicable | Not applicable | New feature | |
| Form: executable<br>controlled by a<br>host application | Yes | Not applicable | Not applicable | New feature | |
| IO: option to limit<br>to Mortara device<br>data only | Yes | Yes | Yes | Equivalent | |
| IO: offline PDF<br>generation of<br>reports | Yes | Yes | Yes | Equivalent | |
| IO: report printing | Yes | Yes | Yes | Equivalent | |
| Data<br>Management<br>System | CardioConfirm | E-scribe (Primary) | Pyramis ECG<br>Management System | Change<br>explanation | |
| BRAND | Mortara | Mortara | Quinton | | |
| COMPANY | Mortara Instrument, Inc. | Mortara Instrument, Inc. | Quinton Inc. | | |
| 510 (K) Number | Present Application | K930558 | K032038 | | |
| User<br>Permissions:<br>view only | Yes | Yes | Yes | Equivalent | |
| | User<br>Permissions:<br>view and save | Yes | Yes | Yes | Equivalent |
| | | User<br>Permissions:<br>view and<br>electronically<br>sign | Yes | Yes | Yes |
| User<br>Permissions: edit<br>demographics | | | Yes | Yes | Yes |
| | User<br>Permissions: edit<br>measurements | | Yes | Yes | Yes |
| | | Data<br>Management<br>System | CardioConfirm | E-scribe (Primary) | Pyramis ECG<br>Management System |
| BRAND | | Mortara | Mortara | Quinton | |
| COMPANY | Mortara Instrument, Inc. | Mortara Instrument, Inc. | Quinton Inc. | | |
| 510 (K) Number | Present Application | K930558 | K032038 | | |
| View: patient<br>identity always<br>visible | Yes | Yes | Yes | Equivalent | |
| View: pan and<br>zoom | Yes | Yes | Yes | Equivalent | |
| Edit: patient<br>demographics | Yes | Yes | Yes | Equivalent | |
| Edit: accepts<br>demographic<br>updates outside<br>of CardioConfirm | Yes | Not applicable | Not applicable | New feature | |
| Sign:<br>electronically<br>sign | Yes | Yes | Yes | Equivalent | |
| Resting ECG | | | | | |
| IO: option to<br>support non- | Yes for professional | No | Yes | Equivalent to | |
| Data<br>Management<br>System | CardioConfirm | E-scribe (Primary) | Pyramis ECG<br>Management System | Change<br>explanation | |
| BRAND | Mortara | Mortara | Quinton | | |
| COMPANY | Mortara Instrument, Inc. | Mortara Instrument, Inc. | Quinton Inc. | | |
| 510 (K) Number | Present Application | K930558 | K032038 | | |
| Mortara device<br>data | version | | | Pyramis | |
| IO: reads DICOM<br>12-Lead ECG<br>Waveform<br>objects | Yes | Not applicable | Not applicable | Equivalent to<br>Pyramis | |
| IO: writes DICOM<br>12-lead ECG<br>Waveform<br>objects | Yes | Yes | No | Equivalent to E-<br>Scribe | |
| User<br>Permissions: edit<br>interpretation<br>including serial<br>comparison and<br>conclusion | Yes | Yes | Yes | Equivalent | |
| View: waveform<br>zoom | Optimizes for display of<br>full 10s waveforms over a | Fit page, Fit width, 75%,<br>100%, 150%, 200%, | Fit width, Fit height, 25%,<br>50%, 75%, 100%, 125%,…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.