K161452 · MicroVention, Inc. · HCG · Aug 31, 2016 · Neurology
Device Facts
Record ID
K161452
Device Name
MicroPlex Coil System (MCS) - HyperSoft 3D
Applicant
MicroVention, Inc.
Product Code
HCG · Neurology
Decision Date
Aug 31, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MicroPlex Coil System (MCS) are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Device Story
MicroPlex Coil System (MCS) consists of platinum alloy implant coils in 3D spherical structures. Coils attach to a V-Trak or V-Trak Advanced Delivery Pusher via a polyolefin elastomer filament. The delivery pusher is a variable stiffness stainless steel hypotube. During procedure, physician inserts the pusher into a handheld, battery-powered V-Grip Detachment Controller. Activation of the controller delivers electrical current to heat the filament, causing detachment of the coil into the target vessel or aneurysm. Used in clinical settings for permanent obstruction of blood flow to neurovascular or peripheral vascular abnormalities. Benefits include minimally invasive endovascular treatment of aneurysms and malformations.
Clinical Evidence
Bench testing only. Verification and validation included visual inspection, dimensional measurement, advancement/retraction force testing, simulated use in a cerebrovascular benchtop model, and spring constant testing. Biocompatibility testing performed per ISO 10993 standards (cytotoxicity, sensitization, irritation, hemocompatibility, systemic toxicity, genetic toxicology, and intramuscular implantation) confirmed the device is non-toxic, non-irritant, and non-hemolytic.
Technological Characteristics
Implant: Platinum/Tungsten alloy coil (92/8). Coupler: Platinum/Iridium (90/10). Adhesive: Ultraviolet curing. Stretch-resistant member: Polyethylene Terephthalate or Polyolefin Elastomer (Engage™). Delivery: Stainless steel hypotube pusher. Detachment: Electrical current via battery-powered controller. Sterilization: E-beam. MRI compatible. Dimensions: 1-6 mm diameter, 6-20 cm length. Connectivity: None.
Indications for Use
Indicated for patients requiring endovascular embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae, and vascular occlusion of blood vessels within the neurovascular system or peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
Predicate Devices
MicroPlex Coil System (MCS) - HyperSoft® 3D (K131948, K153594)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 31, 2016
MicroVention, Inc. Ms. Sapna Singh, MS. RAC Regulatory Affairs Project Manager 1311 Valencia Avenue Tustin, California 92780
Re: K161452
Trade/Device Name: MicroPlex Coil System (MCS) Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: August 1, 2016 Received: August 2, 2016
Dear Ms. Singh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J.Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161452
Device Name MicroPlex Coil System (MCS)
#### Indications for Use (Describe)
The MicroPlex Coil System (MCS) are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 10pt;"> <b> </b> </span> <span style="font-size: 10pt;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
| | <span> <span style="font-size: 10pt;"> <b> </b> </span> <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
This 510(k) summary for the MicroPlex Coil System (MCS) – HyperSoft® 3D is submitted in accordance with the requirements of 21 CFR 807.87(h) and 807.92 and following the recommendations outlined in FDA Guidance, The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], dated 28 July, 2014.
#### SUBMITTER [807.92(a)(1)]
MicroVention, Inc. 1311 Valencia Avenue Tustin, California U.S.A.
| Telephone: | (714) 247-8162 |
|------------|----------------|
| Fax: | (714) 247-8014 |
| Contact Person: | Sapna Singh |
|-----------------|------------------------------|
| Email: | sapna.singh@microvention.com |
| Date Prepared: | May 25, 2016 |
#### DEVICE [807.92(a)(2)]
| Name of Device: | MicroPlex Coil System (MCS) |
|---------------------------|-----------------------------------------------------------------------------------------------|
| Common or Usual Name: | MicroPlex Platinum Coils |
| Classification Name: | Neurovascular Embolization Device |
| Product Code: | HCG, KRD |
| Regulatory Class: | Class II |
| Submission Type: | Special 510(K) |
| Regulation Number: | 21 CFR 882.5950 |
| Reviewing Product Branch: | Division of Neurological and Physical Medicine Devices<br>(Office of Device Evaluation, CDRH) |
### PREDICATE DEVICE [807.92(a)(3)]
MicroPlex Coil System (MCS) - HyperSoft® 3D (K131948, K153594)
### DEVICE DESCRIPTION [807.92(a)(4)]
HyperSoft® 3D in the MicroPlex Coil System (MCS) consist of an implant coil made of platinum alloy.
{4}------------------------------------------------
The coils are designed in 3D spherical structure in various loop sizes and lengths. The coil is attached to V-Trak™ or V-Trak™ Advanced Delivery Pusher via polyolefin elastomer filament. The Delivery Pusher is a variable stiffness stainless steel hypotube with platinum and stainless steel coils at the distal end. The proximal end of the Delivery Pusher is inserted into a hand held battery powered V-Grip™ Detachment Controller. When the Detachment Controller is activated, the flow of electrical current heats the polyolefin elastomer filament. resulting in detachment of the implant segment.
# INDICATIONS FOR USE [807.92(a)(5)]
The MicroPlex Coil System (MCS) are intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The MCS are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS [807.92(a)(6)]
The Table I compares the technological characteristics of the existing HyperSoft® 3D coils (K131948, K153594) with the additional models presented in this 510(k) submission. The devices.
- Have the same intended use ●
- Use the same operating principle ●
- Incorporate the same basic coil design ●
- Use same construction and material
- Are packaged and sterilized using same material and processes .
The line extension of the HyperSoft® 3D coils (includes addition of sizes from 1mm to 6 mm secondary wind diameter with lengths from 6 cm to 20 cm) and change in the Stretch Resistance Member Material from Polyethylene Terephthalate (PET) to PET or Polyolefin Elastomer (Engage™) does not change the indications for use of the coils and is not a change to the fundamental scientific technology.
The performance data below shows the device will perform as well as the previously marketed device.
{5}------------------------------------------------
| | Existing HyperSoft® 3D (Predicate Device,<br>(K131948, K153594) | HyperSoft® 3D Line<br>Extension (Subject<br>Device |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Intended Use | | |
| Intended Use<br>Statement | The MicroPlex Coil System (MCS) are<br>intended for the endovascular embolization of<br>intracranial aneurysms and other<br>neurovascular abnormalities such as<br>arteriovenous malformations and<br>arteriovenous fistulae. The MCS are also<br>intended for vascular occlusion of blood<br>vessels within the neurovascular system to<br>permanently obstruct blood flow to an<br>aneurysm or other vascular malformation and<br>for arterial and venous embolizations in the<br>peripheral vasculature. | Same |
| Performance | | |
| Function | The coils are used for the endovascular<br>embolization of aneurysms, other<br>neurovascular abnormalities such as<br>arteriovenous malformations, arteriovenous<br>fistulae and arterial and venous embolizations<br>in the peripheral neurovasculature. | Same |
| Anatomical Location | General intravascular use, including the neuro<br>and peripheral vasculature. | Same |
| Implantable Embolization Coil | | |
| Coil Shape | 3D - Spherical | Same |
| Primary Coil Wire OD | 0.00125 inch | 0.00125 inch and 0.0015<br>inch |
| Coil Implant Diameter | 1 - 5 mm | 1 - 6 mm |
| Coil Restrained Length | 2 - 15 cm | 6 - 20 cm |
| Coil Gap | Closed | Closed |
| Delivery pusher length | 195 cm | Same |
| Material | | |
| Main Coil Wire | Platinum/Tungsten Alloy (Pt/W: 92/8) | Same |
| Coil-to-Pusher Coupler | Platinum/Iridium (Pt/Ir: 90/10) | Same |
| Adhesive | Ultraviolet Curing Adhesive | Same |
| Implant to Pusher<br>Filament | Polyolefin Elastomer | Same |
| Stretch resistant (SR) | Polyethylene Terephthalate (K131948) | Polyethylene<br>Terephthalate or<br>Polyolefin Elastomer<br>(Engage™) |
| Other Attributes | | |
| Detachment System | Detachment Controller; stand alone, hand<br>held battery operated unit detaches the coil<br>implant | Same |
| Catheter compatibility | Compatible with 10-system microcatheters<br>(minimum ID of 0.0165") | Same |
| MRI compatibility | Yes | Same |
| Method of Supply | Sterile and single use (e-beam) | Same |
| Package Configuration | Placed in Introducer Sheath, Dispenser Coil,<br>Pouch, and Shipping Carton | Same |
#### Table I: Predicate Device vs Subject Device Comparison Table
{6}------------------------------------------------
## PERFORMANCE DATA [807.92(b)]
Results of the verification and validation testing (Table II) indicate that the product meets established performance requirements, and is substantially equivalent for its intended use.
| Table II: Design Verification and Validation Test Summary | | |
|-----------------------------------------------------------|--|--|
|-----------------------------------------------------------|--|--|
| Bench Testing | Result |
|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Visual Inspection: Using a microscope, inspect HyperSoft® 3D per<br>device drawing, PDM-ATP. | All test samples passed testing. |
| Dimensional Measurement: Using a microscope, inspect<br>HyperSoft® 3D Coil's secondary wire diameter. | All test samples passed testing. |
| Advancement/Retraction Force: The test represents the maximum<br>force required to advance and retract the coil through the<br>microcatheter. | All test samples passed testing. |
| Simulated Use: The test simulate the use of HyperSoft® 3D coils in-<br>vitro using a cerebrovascular benchtop model. | All test samples passed testing. |
| Spring Constant: The spring constant force (determination of<br>maximum force to break monofilament) of the coil is measured. | All test samples passed testing. |
#### Biocompatibility Summary – HyperSoft® 3D Implant
| Biocompatibility | Test Standard | Results |
|--------------------------------------------|---------------|------------------------|
| Cytotoxity | | |
| MEM Elution Test | ISO 10993-5 | Non-toxic |
| ISO Cell Culture Agar Overlay | ISO 10993-5 | Non-toxic |
| Sensitization | | |
| Sensitization-Guinea Pig Maximization Test | ISO 10993-10 | No sensitizer response |
| Irritation | | |
{7}------------------------------------------------
| ISO Intracutaneous Reactivity Evaluation<br>Test | ISO 10993-10 | Non-irritant | | |
|--------------------------------------------------|--------------|---------------------------------------|--|--|
| Hemocompatibility | | | | |
| Hemolysis | ISO 10993-4 | Non-hemolytic | | |
| Prothrombin Time Assay - ISO | ISO 10993-4 | No adverse effect on coagulation time | | |
| Systemic Toxicity | | | | |
| Systemic toxicity<br>(IV injection) | ISO 10993-11 | Non-toxic | | |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 | Non-pyrogenic | | |
| Genetic Toxicology | | | | |
| Bacteria Reverse Mutation Assay (Ames<br>Test) | ISO 10993-3 | Negative response for mutagenicity | | |
| Intramuscular Implantation | | | | |
| 7-day Muscle Implantation | ISO 10993-6 | Non-irritant | | |
| 13-week Intramuscular Implantation Test | ISO 10993-6 | Non-irritant | | |
| 26-week Intramuscular Implantation Test | ISO 10993-6 | Non-irritant | | |
## Biocompatibility Summary – V-Trak™ or V-Trak™ Advanced Delivery Pusher
| Biocompatibility | Test Standard | Results |
|--------------------------------------------------|---------------|---------------------------------------|
| Cytotoxity | | |
| MEM Elution Test | ISO 10993-5 | Non-toxic |
| ISO Cell Culture Agar Overlay | ISO 10993-5 | Non-toxic |
| Sensitization | | |
| Sensitization-Guinea Pig Maximization Test | ISO 10993-10 | No sensitizer response |
| Irritation | | |
| ISO Intracutaneous Reactivity Evaluation<br>Test | ISO 10993-10 | Non-irritant |
| Hemocompatibility | | |
| Hemolysis | ISO 10993-4 | Non-hemolytic |
| Prothrombin Time Assay - ISO | ISO 10993-4 | No adverse effect on coagulation time |
| Systemic Toxicity | | |
| Systemic toxicity<br>(IV injection) | ISO 10993-11 | Non-toxic |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 | Non-pyrogenic |
{8}------------------------------------------------
#### CONCLUSIONS
Based on the 510(k) summary and information provided herein, we conclude the subject device, HyperSoft® 3D in the MCS, is substantially equivalent in its intended use, design, material, performance, and the underlying fundamental scientific technology used, to the predicate HyperSoft® 3D, K131948, K153594.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.