K161430 · Natus Manufacturing Limited · IKT · Mar 2, 2017 · Physical Medicine
Device Facts
Record ID
K161430
Device Name
Myoject Luer Lock Needle Electrode
Applicant
Natus Manufacturing Limited
Product Code
IKT · Physical Medicine
Decision Date
Mar 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.1385
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The disposable hypodermic monopolar needles are FOR SINGLE USE ONLY in electromyography (EMG) in situations wherein it is desired to insert an electrode, into a patient to locate a particular muscle and then inject a medication into that muscle. The hypodermic needle with an open lumen is designed for muscle stimulation, motor unit action potential recording and drug delivery. Once the physician is satisfied with the location, he/she injects a drug therein via the lumen of the needle.
Device Story
Disposable hypodermic monopolar needle electrode; used in electromyography (EMG). Input: electrical signals from muscle tissue (motor unit action potentials); output: electrical stimulation to muscle and delivery of medication via open lumen. Operated by physicians in clinical settings. Device features Luer Lock/Slip hub for syringe attachment; stainless steel 304 needle with PTFE coating; single contact touch-proof DIN 42802 connector. Physician inserts needle to locate muscle, records potentials, stimulates muscle, and injects medication through lumen. Benefits include reduced patient discomfort via improved tip sharpness and secure drug delivery via Luer Lock mechanism.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included tensile testing, coating thickness, leak testing, tip geometry measurements, penetration testing, and Luer Lock design verification. Biocompatibility testing conducted per ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 10993-11 (systemic toxicity) confirmed compliance.
Indicated for patients requiring electromyography (EMG) procedures where muscle localization, motor unit action potential recording, muscle stimulation, and subsequent drug delivery into the muscle are necessary.
Regulatory Classification
Identification
A diagnostic electromyograph needle electrode is a monopolar or bipolar needle intended to be inserted into muscle or nerve tissue to sense bioelectrical signals. The device is intended for medical purposes for use in connection with electromyography (recording the intrinsic electrical properties of skeletal muscle).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a symbol that resembles three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2017
Natus Manufacturing Limited Michael Galvin Manager. Ouality and Regulatory Affairs IDA Business Park Gort, Co. Galway Ireland
Re: K161430
Trade/Device Name: Myoject™ Luer Lock Needle Electrode Regulation Number: 21 CFR 890.1385 Regulation Name: Diagnostic Electromyograph Needle Electrode Regulatory Class: Class II Product Code: IKT Dated: February 9, 2017 Received: February 13, 2017
Dear Mr. Galvin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
- Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161430
#### Device Name
Myoject(TM) Luer Lock Needle Electrode
#### Indications for Use (Describe)
The disposable hypodermic monopolar needles are FOR SINGLE USE ONLY in electromyography (EMG) in situations wherein it is desired to insert an electrode, into a patient to locate a particular muscle and then inject a medication into that muscle. The hypodermic needle with an open lumen is designed for muscle stimulation, motor unit action potential recording and drug delivery. Once the physician is satisfied with the location, he'she injects a drug therein via the lumen of the needle.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
28 February 2017
Natus Manufacturing Limited Traditional 510(k) Myoject™ Luer Lock Needle Electrode
Image /page/3/Picture/3 description: The image shows the word "natus" in a teal color. The font is sans-serif and the letters are bold. There is a registered trademark symbol to the right of the "s".
Natus Manufacturing Limited, IDA Business Park Gort, Co. Galway 353 091 630043 tel. 353 091 630050 fax.
# Section 5: 510(k) Summary
| Manufacturer's Name: | Natus Manufacturing Limited<br>IDA Business Park<br>Gort, County Galway<br>Ireland |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Michael Galvin<br>Manager, Quality and Regulatory Affairs<br>Natus Manufacturing Limited<br>IDA Business Park<br>Gort, County Galway<br>Ireland |
| Telephone Number: | +353-(0)91-647451 |
| Fax Number: | +353-(0)91-630050 |
| Summary Date: | 17 May 2016 |
| Trade Names: | Myoject™ Luer Lock Needle Electrode |
| Common or Usual Name: | Electrode, Needle, Diagnostic Electromyograph |
| Classification Name: | Diagnostic Electromyograph Needle Electrode |
| Device Class: | Class II |
| Product Code: | IKT |
| Classification Regulation: | 21.CFR.890.1385 |
{4}------------------------------------------------
28 February 2017
Natus Manufacturing Limited Traditional 510(k) Myoject™ Luer Lock Needle Electrode
| Classification Panel: | Neurology |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device: | K973444 Teca Myoject Disposable Needle Electrode |
| Device Description: | The Myoject™ Luer Lock Needle Electrode is a Disposable<br>Hypodermic Needle Electrode |
| Intended Use: | The disposable hypodermic monopolar needles are FOR SINGLE<br>USE ONLY in electromyography (EMG) in situations wherein it is<br>desired to insert an electrode, in the form of a probe, into a patient<br>to locate a particular muscle and then inject a medication into that<br>muscle. The hypodermic needle with an open lumen is designed<br>for muscle stimulation, motor unit action potential recording and<br>drug delivery. Once the physician is satisfied with the location,<br>he/she injects a drug therein via the lumen of the needle. |
| Comparison: | The Myoject™ Luer Lock Needle Electrode employ the same<br>technological characteristics as the predicate device. |
| Nonclinical Tests: | Tensile testing Coating thickness Leak Testing Tip geometry measurements Penetration testing Luer Lock Design |
| Substantial<br>Equivalence: | The Myoject™ Luer Lock Needle Electrode is equivalent to the device cleared under K973444 |
It has been shown in this 510(k) submission that the differences between the Myoject™ Luer Lock and the predicate device do not raise any questions regarding its safety and effectiveness. The Myoject™ Luer Lock Needle electrode device is substantially equivalent to the predicate device as it has the same intended use and similar technological characteristics as the previously cleared predicate devices. The Myoject™ Luer Lock Needle Electrode, as designed and manufactured is determined to be substantially equivalent to the referenced predicate devices.
Substantial Equivalence Comparison Table.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "natus" in a teal sans-serif font. The letters are bold and slightly rounded. A small registration mark is located to the right of the "s".
| Table 5A: Substantial Equivalence Comparison Table | | |
|----------------------------------------------------|--|--|
| | | |
| Year | 2016 | 20 Nov 1997 (original submission<br>date) | Discussion of Differences |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Natus Manufacturing Limited | TECA Corporation<br>3 Campus Drive<br>Pleasantville<br>New York 10570 | Minor changes to the labelling include change of legal manufacturer name, address and branding. In addition the new device labelling incorporates warnings and harmonized symbols. However, these changes do not raise questions of safety and effectiveness as they provide additional information to the user for the safe and effective use of the device. |
| Trade Name | Myoject™ Luer Lock Needle<br>Electrodes | Teca Myoject Disposable Needle<br>Electrode | |
| 510(k) number | K161430 | K973444 | |
| Product Code | IKT | IKT | |
| Labelling | Packaging Labels | Packaging Labels | |
| Intended Use | The disposable hypodermic monopolar needles are FOR SINGLE USE ONLY in electromyography (EMG) in situations wherein it is desired to insert an electrode, in the form of a | The disposable hypodermic monopolar needles are FOR SINGLE USE ONLY in electromyography (EMG) in situations wherein it is desired to insert an electrode, in the form of a probe, into a patient to | |
| Year | 2016 | 20 Nov 1997 (original submission date) | Discussion of Differences |
| | | | |
| | Manufacturer | Natus Manufacturing Limited | TECA Corporation |
| | 3 Campus Drive | | |
| | Pleasantville | | |
| | New York 10570 | | |
| Trade Name | | | |
| | Myoject™ Luer Lock Needle<br>Electrodes | Teca Myoject Disposable Needle<br>Electrode | |
| 510(k) number | | | |
| | K161430 | K973444 | |
| Product Code | IKT | IKT | |
| | probe, into a patient to locate a<br>particular muscle and then inject a<br>medication into that muscle. The<br>hypodermic needle with an open<br>lumen is designed for muscle<br>stimulation, motor unit action<br>potential recording and drug<br>delivery. Once the physician is<br>satisfied with the location, he/she<br>injects a drug therein via the lumen<br>of the needle. | locate a particular muscle and then<br>inject a medication into that muscle.<br>The hypodermic needle with an open<br>lumen is designed for muscle<br>stimulation, motor unit action<br>potential recording and drug delivery.<br>Once the physician is satisfied with<br>the location, he/she injects a drug<br>therein via the lumen of the needle.<br><br>Note: TECA Corp. does not supply<br>any drugs with needle electrodes nor<br>does TECA Corp. offer for sale any<br>form of drugs. | can be included in labeling.<br>The needles are sold on their<br>own as a medical device and<br>are not specifically designated<br>for use with a specific drug.<br>Therefore, this statement is<br>not appropriate for the IFU. |
| Physical<br>Characteristics | Monopolar Needle | Monopolar Needle | Same |
| Connector | Single Contact Touch Proof DIN42<br>802 | Single Contact Touch Proof DIN42<br>802 | Same |
| Year | 2016 | 20 Nov 1997 (original submission date) | Discussion of Differences |
| Manufacturer | Natus Manufacturing Limited | TECA Corporation<br>3 Campus Drive<br>Pleasantville<br>New York 10570 | |
| Trade Name | Myoject™ Luer Lock Needle<br>Electrodes | Teca Myoject Disposable Needle<br>Electrode | |
| 510(k) number | K161430 | K973444 | |
| Product Code | IKT | IKT | |
| Material of Needle | Stainless 304 | Stainless 304 | Same |
| Needle Size (mm) | Needle Diameter:<br>0.30<br>0.41<br>0.46<br>0.51<br>0.71 | Needle Diameter:<br>0.30<br>0.41<br>0.46<br>0.51<br>0.71 | Same<br>The dimensions relate to needle diameter |
| Needle Coating material | PTFE (Xylan 8820 G4075) | PTFE (Ultralon R-605/T-5) | This change resulted from the<br>Accessories development<br>Department's request to<br>reduce the initial penetration<br>force and mitigate against the<br>obsolescence of the Ultralon<br>coating. This change does not<br>raise questions of safety and<br>effectiveness as the needle<br>coating material remains as<br>PTFE however just the |
| Year | 2016 | 20 Nov 1997 (original submission date) | Discussion of Differences |
| Manufacturer | Natus Manufacturing Limited | TECA Corporation<br>3 Campus Drive<br>Pleasantville<br>New York 10570 | |
| Trade Name | Myoject™ Luer Lock Needle<br>Electrodes | Teca Myoject Disposable Needle<br>Electrode | |
| 510(k) number | K161430 | K973444 | |
| Product Code | IKT | IKT | |
| | | | formulation has changed to<br>reduce the initial penetration<br>force to aid easier insertion of<br>the needle. As a result, this<br>reduces patient discomfort<br>when performing EMG<br>examinations. |
| Hub | Luer Lock/Slip | Luer Slip | This change originated from<br>market place demand for the<br>needles to fit a Luer Lock<br>Syringe. The Luer Lock<br>needle locks into place and as<br>a result is more likely to be<br>seepage free. Therefore, this<br>reduces the risk of medication<br>accidentally leaking from the<br>syringe and possibly causing<br>harm to the patient and user.<br>This does not affect the safety |
| Year | 2016 | 20 Nov 1997 (original submission date) | Discussion of Differences |
| Manufacturer | Natus Manufacturing Limited | TECA Corporation<br>3 Campus Drive<br>Pleasantville<br>New York 10570 | and effectiveness of the<br>device as the overall function<br>of the device remains<br>unchanged, the Luer lock<br>simply adds an additional<br>safety mechanism to the<br>device. |
| Trade Name | MyojectTM Luer Lock Needle<br>Electrodes | Teca Myoject Disposable Needle<br>Electrode | |
| 510(k) number | K161430 | K973444 | |
| Product Code | IKT | IKT | |
| Hub Material | Polypropylene (Exxonmobil)<br>Escorene PP9074 MED | Polycarbonate | The material for this device<br>has changed compared to the<br>predicate device, however<br>comparison with the predicate<br>device demonstrates the<br>material performance is<br>equivalent and they are both<br>variants of polymer materials.<br>There is an insignificant |
| Year | 2016 | 20 Nov 1997 (original submission date) | Discussion of Differences |
| Manufacturer | Natus Manufacturing Limited | TECA Corporation<br>3 Campus Drive<br>Pleasantville<br>New York 10570 | |
| Trade Name | Myoject™ Luer Lock Needle<br>Electrodes | Teca Myoject Disposable Needle<br>Electrode | |
| 510(k) number | K161430 | K973444 | |
| Product Code | IKT | IKT | materials; as a result product<br>performance, safety and<br>effectiveness are not affected.<br>In addition, Biocompatibility<br>testing demonstrates there are<br>no new concerns of safety or<br>effectiveness. |
| Needle grind angle | 10-16 degrees conical | 10-20 degrees conical | The grind angles have similar<br>specifications; the slight<br>change in the upper range of<br>the grind angle is not a<br>significant change and<br>therefore does not raise new<br>concerns of safety and<br>effectiveness. The change<br>does not alter the diagnostic<br>effects of the device. |
| Recording area<br>(sqmm) | .19(mm≤) fine and then increases to<br>.89(mm≤) depending on needle size | 0.25 | There are some differences in<br>recording areas which vary<br>depending on needle size |
| Year | 2016 | 20 Nov 1997 (original submission date) | Discussion of Differences |
| Manufacturer | Natus Manufacturing Limited | TECA Corporation<br>3 Campus Drive<br>Pleasantville<br>New York 10570…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.