← Product Code [OMP](/productcode/OMP) · K161410

# VT·One, VT·One 60days (K161410)

_Carilex Medical, Inc. · OMP · Oct 13, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161410

## Device Facts

- **Applicant:** Carilex Medical, Inc.
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Oct 13, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Carilex VT·One/VT·One 60days is indicated for patients who benefit from removal of fluids and excess exudates, infectious material, and tissue debris, which may promote wound healing. The VT-One/VT-One 60days suction therapy unit is indicated on use with patients with the following wounds: Traumatic, Dehisced wounds, Partial thickness burns, Chronic wounds including pressure ulcers, diabetic foot ulcers, and venous leg ulcers, Acute wounds, Flaps and grafts.

## Device Story

VT.One and VT.One 60days are portable, battery-operated or mains-powered suction pumps for negative pressure wound therapy (NPWT). Device connects to wound dressing via tube and disposable canister; applies negative pressure to remove exudates. Used in clinical or home settings to promote wound healing. VT.One is a durable device (3-year life); VT.One 60days is a single-patient-use device programmed to cease operation after 60 days. Healthcare providers monitor wound progress and device function. Benefits include fluid/debris removal to facilitate healing.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety and electromagnetic compatibility, and performance comparison testing against the predicate.

## Technological Characteristics

Powered suction pump; negative pressure range up to -150 mmHg. Power: AC 100-240V or internal Lithium-ion battery. Dimensions: 150x85x50 mm; Weight: 350g. Operating modes: continuous and intermittent. Canister capacity: 150 ml. Software-controlled operation; VT.One 60days includes a 60-day usage lockout. Standards: IEC 60601-1, IEC 60601-1-2.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- VOLTERA Powered Suction Pump, S1001-3 Series ([K112853](/device/K112853.md))

## Reference Devices

- DeRoyal Foam Kits ([K112458](/device/K112458.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 13, 2016

Carilex Medical, Inc. % Maria Griffin Senior Consultant Mdi Consultants, Inc. 55 Northern Blvd. Great Neck, New York 11021

Re: K161410

Trade/Device Name: VT.One, VT.One 60days Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: September 8, 2016 Received: September 9, 2016

Dear Maria Griffin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K161410

Device Name Carilex VT.One/VT.One 60days

Carilex VT One/VT One 60days is indicated for patients who benefit from removal of fluids and excess exudates. infectious material, and tissue debris, which may promote wound healing. The VT One 60days suction therapy unit is indicated on use with patients with the following wounds:

- Traumatic
- · Dehisced wounds
- · Partial-thickness burns
- · Chronic wounds including pressure ulcers, diabetic foot ulcers and venous leg ulcers
- Acute wounds
- Flaps and grafts

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "carilex" in a light blue color. The word is written in a sans-serif font. There is a registered trademark symbol to the right of the word.

## 510(K) SUMMARY

The assigned 510(k) number is: K161410

### 1. Submitter's Identification:

Carilex Medical, Inc. No. 77 Keji 18t Rd.,Guishan Dist. Taoyuan City Taiwan 333 Registration Number: 9710603 Contact: Henry Kao Tel: +886-3-3287882 Fax: +886-3-3288622 Email: henry.kao@carilexmedical.com

Date Summary Prepared: May 20, 2016

- 2. Trade Name of the Devices: Carilex VT.One/ VT.One 60days
- Common or Usual Name: Powered Suction Pump 3.
- Classification Name: 4. Regulation Number: Product code: Panel:

Negative Pressure Wound Therapy Powered Suction Pump 878.4780 OMP General & Plastic Surgery

### ട്. Predicate Device Information:

VOLTERA Powered Suction Pump, S1001-3 Series, K112853

#### 6. Device Description:

Carilex VT One and VT One 60days are suction pumps with collection canisters for negative pressure wound therapy. Carilex VT.One 60days are modified from its precursor VOLTERA Powered Suction Pump, S1001-3 Series (K112853), and they are smaller, portable and lighter in weight. The pump is connected to the wound dressing via a tube connected to a disposable canister. The device provides negative pressure wound therapy to the wound at a range of pressure settings and removes exudates from the wound site to the disposable canister. The device can operate either by a mains power supply or internal battery.

The intended use, performance, and technology for Carilex VT·One and VT·One 60days are mostly identical. However, the VT·One 60days is programmed to stop working after 60 days of use and will not re-start after this time. This feature makes VT One 60days a single patient use NPWT pump that has a lifetime of 60 days; while VT-One is a durable medical device with a minimum lifetime of 3 years.

{4}------------------------------------------------

# carilex®

### Intended Use: 7.

Carilex VT·One/VT·One 60days is indicated for patients who benefit from removal of fluids and excess exudates, infectious material, and tissue debris, which may promote wound healing. The VT-One/VT-One 60days suction therapy unit is indicated on use with patients with the following wounds: Traumatic, Dehisced wounds, Partial thickness burns, Chronic wounds including pressure ulcers, diabetic foot ulcers, and venous leg ulcers, Acute wounds, Flaps and grafts.

| Item                            | Proposed Devices                                                                                                                                                                                                                                                                                 |                 | Predicate Device                                                                                                                                                                                                                                                                                 |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                     | VT-One                                                                                                                                                                                                                                                                                           | VT-One 60days   | VOLTERA Powered Suction Pump, S1001-3 Series                                                                                                                                                                                                                                                     |
| Classification                  | Class II                                                                                                                                                                                                                                                                                         |                 | Class II                                                                                                                                                                                                                                                                                         |
| Code for Federal<br>Regulations | 878.4780                                                                                                                                                                                                                                                                                         |                 | 878.4780                                                                                                                                                                                                                                                                                         |
| Prescription<br>Medical Device  | YES                                                                                                                                                                                                                                                                                              |                 | YES                                                                                                                                                                                                                                                                                              |
| Model                           | S1004-0012                                                                                                                                                                                                                                                                                       | S1004-0022      | S1001-3 Series                                                                                                                                                                                                                                                                                   |
| Compatible<br>NPWT Dressing     | DeRoyal Foam Kits (K112458)                                                                                                                                                                                                                                                                      |                 | DeRoyal Foam Kits (K112458)                                                                                                                                                                                                                                                                      |
| Intended Use                    | Carilex VT-One/VT-One 60days is<br>indicated for patients who benefit from<br>removal of fluids and excess exudates,<br>infectious material, and tissue debris,<br>which may promote wound healing.                                                                                              |                 | The VOLTERA Powered Suction Pump, S1001-<br>3 Series is indicated for patients who would<br>benefit from wound management via the<br>application of negative pressure for removal of<br>fluids and excess exudate, infectious material,<br>and tissue debris which may promote wound<br>healing. |
| Indications for<br>Use          | The VT-One/ VT-One 60days suction<br>therapy unit is indicated on use with<br>patients with the following wounds:<br>Traumatic Dehisced wounds Partial thickness burns Chronic wounds including pressure<br>ulcers, diabetic foot ulcers, and<br>venous leg ulcers Acute wounds Flaps and grafts |                 | The VOLTERA suction pump is indicated on<br>use with patients with the following wounds:<br>Traumatic Dehisced wounds Partial thickness burns Chronic wounds including pressure ulcers,<br>diabetic foot ulcers, and venous leg ulcers Acute wounds Flaps and grafts                             |
| Contraindications               | Presence of necrotic tissue Malignancy Untreated Osteomyelitis Untreated malnutrition Exposed arteries, veins, nerves, or<br>organs. Use over anastomotic sites                                                                                                                                  |                 | Presence of necrotic tissue Malignancy Untreated Osteomyelitis Untreated malnutrition Exposed arteries, veins, nerves, or organs. Use over anastomotic sites                                                                                                                                     |
| Suction Capacity                | 0.8 Liter / min                                                                                                                                                                                                                                                                                  | 0.6 Liter / min | 2.5 Liter / min                                                                                                                                                                                                                                                                                  |
| Max. Vacuum                     | - 150 mmHg                                                                                                                                                                                                                                                                                       |                 | - 200 mmHg                                                                                                                                                                                                                                                                                       |
| Power Input                     | AC 100-240V / 50-60 Hz                                                                                                                                                                                                                                                                           |                 | AC 110-240V / 47-63 Hz                                                                                                                                                                                                                                                                           |

### 8. Technological Comparison to Predicate Devices:

{5}------------------------------------------------

## carilex®

| Power Output                | DC 12V / 1.25A                                                  | DC 9V / 3A                                            |
|-----------------------------|-----------------------------------------------------------------|-------------------------------------------------------|
| Battery Type                | Lithium ion                                                     | Lithium ion                                           |
| Operating Time<br>(Battery) | 16 hours, depend on use                                         | 24~48 hours, depend on use                            |
| Dimensions                  | 150 mm x 85 mm x 50 mm                                          | 170 mm x 160 mm x 90 mm                               |
| Weight                      | 350g                                                            | 1.35 kg                                               |
| Operating Mode              | Continuous & Intermittent                                       | Continuous & Intermittent                             |
| Canister                    | 150 ml                                                          | 300 ml                                                |
| Life-time                   | 3 years<br>                                                     | 3 years                                               |
| Software<br>Setting         | Programmed to<br>be repetitively<br>used during<br>lifetime<br> | Programmed to be repetitively used during<br>lifetime |

### Discussion of Non-Clinical Tests Performed for Determination of Substantial 9. Equivalence are as follows:

The following tests were performed to determine substantial equivalence:

IEC60601-1:2005+COR1(2006)+COR2(2007)+AMD1(2012) Medical Electrical Equipment – Part 1:General Requirements for basic safety and essential performance

IEC60601-1-2:2007 Medical Electrical Equipment – Part 1-2: General Requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility-Requirements and Tests

Performance/Comparison Testing of VT.One 60days with the predicate. The pumps met all pre-determined acceptance criteria and passed the tests.

### Discussion of Clinical Tests Performed: 10.

Clinical tests were not performed.

### 11. Conclusions:

After analyzing intended use, indications for use, technology, bench test reports, shelflife reports, and EMC and electrical safety tests, it can be concluded that Carilex VT·One/ VT·One 60days is substantially equivalent to the predicate device VOLTERA Powered Suction Pump, S1001-3 Series (K112853).

---

**Source:** [https://fda-staging.innolitics.com/device/K161410](https://fda-staging.innolitics.com/device/K161410)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).