K161403 · A.R.C Laser GmbH · GEX · Aug 16, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161403
Device Name
A.R.C. Laser Surgical Fibers and Probes
Applicant
A.R.C Laser GmbH
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The A.R.C. laser surgery fibers & probes are intended for delivery of laser energy to soft tissue in the contact and non-contact mode during surgical procedures, including via endoscopes. The fibers are indicated for use in general surgical applications for incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue. The A.R.C. Laser Surgery Fibers & Probes are indicated for use in General Surgery, Urology, Gynecology and ENT with a cleared compatible laser marketed for the mentioned intended uses and using an SMA 905 connector.
Device Story
A.R.C. Laser Surgical Fibers and Probes are single-use, sterile fiber optic delivery systems; designed to transmit laser energy from a compatible external laser source to target soft tissue. The device consists of a quartz core fiber (2-3 meters long) with an SMA 905 connector at the proximal end for attachment to a cleared laser system. The distal end features a bare tip (flat or spherical) for contact or non-contact delivery of laser energy in pulsed or continuous wave modes. Used by surgeons in clinical settings (General Surgery, Urology, Gynecology, ENT), including via endoscopes. The device facilitates precise tissue interaction—incision, excision, vaporization, ablation, hemostasis, or coagulation—benefiting patients by enabling minimally invasive surgical interventions. Healthcare providers operate the device by connecting it to a compatible laser and manipulating the distal tip to apply energy to the surgical site.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via functional aging tests, energy transmission validation, sterile packaging integrity (ASTM F1886, F1980, F1608), and sterilization validation (DIN EN ISO 11135-1).
Technological Characteristics
Quartz core fiber optic delivery system; 2-3 meter length; 450nm-2100nm wavelength range; SMA 905 connector. Single-use; EO sterilized (DIN EN ISO 11135-1). Packaging standards: ASTM F1886/F1886M-09, ASTM F1980-07, ASTM F1608-00, DIN EN ISO 11737-2, DIN EN 868, ISO 10993-1, EN ISO 11607-1.
Indications for Use
Indicated for patients undergoing soft tissue surgical procedures in General Surgery, Urology, Gynecology, and ENT. Used for incision, excision, vaporization, ablation, hemostasis, or coagulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right, with a stylized eagle above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 16. 2016
A.R.C. Laser GmbH Ms. Angela Thyzel CEO Bessemer St. 14 Nurnberg, 90411 Germany
Re: K161403 Trade/Device Name: A.R.C. Laser Surgical Fibers And Probes Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: Mav 20, 2016 Received: May 20, 2016
Dear Ms. Thyzel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Post-market Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161403
### Device Name
A.R.C Laser Surgery Fibers and Probes
### Indications for Use (Describe)
The A.R.C laser surgery fibers and probes are intended for delivery of laser energy to soft tissue in the contact and non-contact mode during surgical procedures, including via endoscopes. The fibers are indicated for use in general surgical applications for incision, excision, vaporization, hemostasis or coagulation of soft tissue.
The A.R.C laser surgery fibers and probes are indicated for use in General Surgery, Urology, Gynecology and ENT with a cleared compatible laser marketed for the mentioned intended uses and using an SMA 905 connector.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image features the logo for A.R.C. Laser, set against a solid blue background. The acronym "A.R.C." is prominently displayed in large, bold white letters, stacked above the word "LASER," which is enclosed in a white box. Below this, the tagline "enlighten your surgery." is written in a smaller, italicized font, completing the logo's design.
# Section 5 - 510(k) SUMMARY
| Title: | A.R.C. Laser Surgical Laser Fibers & Probes |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | A.R.C. Laser GmbH<br>Bessemer St. 14<br>90411, Nurnberg<br>Germany |
| Contact : | Angela Thyzel |
| Tittle : | General Manager |
| Phone : | 0911-21779-0 |
| Fax : | 0911-21779-99 |
| E-Mail: | a.thyzel@arclaser.de |
| Address: | Bessemer St. 14<br>90411, Nurnberg<br>Germany |
| Date Prepared: | May 15, 2016 |
| Device Trade Name: | A.R.C. Laser Surgical Laser Fibers & Probes |
| Common Name: | Fiber Optic laser delivery system |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology |
| Device product code: | GEX |
| Device Classification | 21 CFR 878.4810 |
| Predicate Devices: | Adler MicroMed Surgical Laser Fibers (K152417) |
| Device Description: | The A.R.C. surgical laser fibers & probes are single use laser delivery<br>devices provided EO sterile and intended for medical applications<br>in various fields of laser surgery.<br>The devices are based on a quartz core, have a length of 6.5 feet (2<br>meters) to 9.9 feet (3 meters) and a wavelength range between 450nm to<br>2100nm. |
Page 13
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for A.R.C. Laser. The logo is set against a blue background. The text "A.R.C." is in large, bold, white letters, with "LASER" in a smaller font size directly below it. Underneath the word "LASER" is the tagline "enlighten your surgery" in a smaller, italicized font.
The proximal end of the fibers is connected to a cleared laser system via an SMA 905 connector, while the distal end delivers the laser energy to the target tissue in pulsed and continuous wave mode.
The distal end have a bare tip which may be flat or spherical, and may be used with a variety of sterile hand pieces for the various intended uses and effects.
## Intended
Use:
The A.R.C. laser surgery fibers & probes are intended for delivery of laser energy to soft tissue in the contact and non-contact mode during surgical procedures, including via endoscopes. The fibers are indicated for use in general surgical applications for incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue.
The A.R.C. Laser Surgery Fibers & Probes are indicated for use in General Surgery, Urology, Gynecology and ENT with a cleared compatible laser marketed for the mentioned intended uses and using an SMA 905 connector.
# Substantial
Equivalence:
The A.R.C. laser surgery fibers & probes have the same mechanism of action, same intended use and the same technological characteristics as the predicate devices, the Adler MicroMed Laser Surgery Fibers, K152417.
Thus, we conclude that the A.R.C. Laser Surgery Fibers & Probes are as safe and effective as the predicate devices.
## Performance
Bench Testing : The A.R.C. laser surgery fibers & probes are single use and sterile medical devices, thus the following tests have been performed
| Test | Standard/ Guidelines | Section in submission |
|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Bench, functional tests | Internal methods | Section 11<br>Fibers functional aging test<br>Fibers Energy transmission validation |
| Sterile packaging test:<br>Visual packaging integrity test<br>Sealing strength validation<br>Microbial barrier<br>Peel test | ASTM F1886/F1886M-09: 2013<br>ASTM F1980-07:2001<br>ASTM F1608-00:2009<br>DIN EN ISO 11737-2:2014<br>DIN EN 868:2009<br>ISO 10993-1:2009(R) 2013<br>EN ISO 11607-1 | Section 13<br>Packaging<br>Shelf Life |
| Sterilization validation | DIN EN ISO 11135-1:2014<br>ISO 10993-7: 2009 | Section 14<br>Appendix 8-Sterilization<br>Validation -EO Residuals |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for A.R.C. Laser. The logo is set against a blue background. The text "A.R.C." is in large, bold, white letters, with the word "LASER" directly underneath in a smaller font. Below that, the tagline "enlighten your surgery" is written in a cursive font.
The performance testing demonstrated that the A.R.C. laser surgery fibers & probes are as safe and as effective as the cleared predicate devices.
Clinical Performance None data:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.