Sunlamp products (booths and beds) designed for skin tanning. Devices utilize high-pressure lamps (with blue cobalt and clear glass filters) or low-pressure lamps (separated from user by acrylic panels). User lies in or stands within the device during a tanning session. Safety features include control switches that interrupt power if doors open or filters break. Operated by the user in a clinic or salon setting. Output is ultraviolet radiation for cosmetic tanning. Benefits include aesthetic skin darkening. Devices are identical to previously cleared predicates (K152233).
Clinical Evidence
No clinical data. Bench testing only. Devices were tested for conformity with 21 CFR 1040.20 performance standards, electrical safety, and electromagnetic compatibility (EMC). Biocompatibility of PMMA acrylic panels was established via ISO 10993 (European standard 30993). Usability data provided to support OTC labeling.
Technological Characteristics
Sunlamp booths/beds using high-pressure (1000W/600W/1500W) or low-pressure lamps. Materials: metal frame, plastic thermoformed panels, PMMA acrylic panels (3mm-8mm thickness). High-pressure lamps use double glass filters (blue cobalt inner, clear outer). Connectivity: standalone. Sterilization: N/A. Software: moderate level of concern. Compliance: 21 CFR 1040.20.
Indications for Use
Indicated for individuals seeking to tan the skin using a sunlamp product.
Regulatory Classification
Identification
A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.
Special Controls
*Classification.* Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
*Labeling* —(i)*Sunlamp products.* (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
*Sunlamp products and UV lamps intended for use in sunlamp products.* Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
*Performance standard.* Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 6, 2017
Sportarredo Group S.C. % Rhonda Alexander Senior Regulatory Specialist Registrar Corp. 144 Research Dr. Hampton, Virginia 23666
Re: K161394
Trade/Device Name: HPO Exotic, Lp1, Lp2, Lp3, MasterSun 360, MasterSun 360 New Reflector, Vega Lux Regulation Number: 21 CFR 878.4635 Regulation Name: Ultraviolet Lamp For Tanning Regulatory Class: Class II Product Code: LEJ Dated: May 13, 2016 Received: May 19, 2016
Dear Rhonda Alexander:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161394
Device Name
HPO Exotic, LP1, LP2, LP3, MasterSun 360, MasterSun 360 New Reflector, Vega Lux
Indications for Use (Describe) Sunlamp product intended for use to tan the skin.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (21 CFR 807.92)
#### l. SUBMITTER
Sportarredo Group S.C. Via dell' Industria 7 Gruaro, Venice 30020 Italy
| Contact Person: | Stefano Iridio, Area Sales Manager (s.iridio@sportarredousa.com) |
|-----------------|-------------------------------------------------------------------|
| Phone Number: | +39 0421 767676 |
| Date Prepared: | April 15, 2016 |
# II. DEVICE
| Names of Devices: | HPO Exotic<br>LP1<br>LP2<br>LP3<br>Mastersun 360<br>Mastersun 360 New Reflector<br>Vega Lux |
|-----------------------|---------------------------------------------------------------------------------------------|
| Common or Usual Name: | Booth, Sun Tan |
| Regulatory Class: | Class II |
| Product Code: | LEJ |
#### III. PREDICATE DEVICE
Per the special controls for these devices, the predicate devices which were legally offered for sale on or before September 2, 2014, but which were 510(k) exempt. Thus, the submitter has compared the subject devices to similar models of suntan beds, named as follows (along with the corresponding radiation safety report accession number):
| K152233 | |
|--------------------------------------------------------|--|
| HPO (08-9921019) | |
| LP1 (9922711-002) | |
| LP2 (9721344-006) | |
| LP3 110-140 (9721344-005) | |
| Mastersun 360/Mastersun 360 new Reflector (15-9721345) | |
| Vega Lux Open (09-9921019) | |
No reference devices were used in this submission.
| Prior Submissions: |
|--------------------|
These devices were previously cleared under k-number K152233.
{4}------------------------------------------------
# DEVICE DESCRIPTION HPO Exotic
This model is a high pressure stand up appliance with tanning lamps located in four vertical columns. The high pressure lamps have a double filter: one blue cobalt filter in the inner part and one acrylic filter in the outer. The 1000 W lamps have an additional clear glass filter between the blue filters and the acrylic. The body frame is all in metal and it closes where it is possible all the electrical components. The body frame is covered by plastic thermoformed panels.
# LP1
These suntan beds are three models with the same frame, aesthetic and with many common electrical parts. All the beds use the same low pressure lamps, the difference is the presence or not of the facial lamps.
# There are three versions:
LP1-32/32EL: suntan bed with only low pressure lamps, 16 in the canopy and 16 in the base LP1-35/35EL: suntan bed with 16 low pressure lamps in the base and 16 low pressure lamps in the canopy. In the canopy is mounted one high pressure facial lamp.
LP1-45/45EL: suntan bed with 16 low pressure lamps in the base and 16 low pressure lamps in the canopy. In the canopy are mounted two high pressure facial lamps.
In all models the low pressure lamps are separated from the person by acrylic panels (3mm thickness for the canopy and 5mm thickness for the base). The high pressure lamps have a double glass filter, one blue cobalt filter in the inner part and one clear filter in the outer. Every blue filter has a control switch. If the door is opened during the session or the filters are removed, the control switch interrupts the power supply on the control board and all lamps switch off.
The body frame is all in metal and it closes where it is possible all the electrical components. The body frame is covered by plastic thermoformed panels.
# LP2
This model is a bed with low pressure lamps in the base. The facial has 3 high pressure lamps. The low pressure lamps are separated from the person by acrylic panels (3mm thickness for the canopy and 5 mm thickness for the base). The high pressure lamps have a double glass filter: one blue cobalt filter in the inner part and one clear filter in the outer. Each blue filter has a control switch. If the door is opened during the session or the filter breaks the control switch interrupts the power supply and the lamp switches off.
# LP3
These suntan beds are two models with the same frame, aesthetic and with many common electrical parts. The new models are similar to the current LP3, only the number of lamps was changed.
{5}------------------------------------------------
There are two models:
LP3-110: suntan bed with 42 low pressure lamps, 18 in the base and 24 in the canopy. The facial has four 600W (Kalfasun 630F) high pressure lamps.
LP3-140: suntan bed with 48 low pressure lamps, 20 in the base and 28 in the canopy. The facial has two 600W (Kalfasun 630F) and two 1500W (Kalfasun 1530F) high pressure lamps. Both models are beds with low pressure tanning lamps in the canopy and in the base and high pressure lamps in the facial area. The low pressure lamps are separated from the person by acrylic panels (3mm thickness for the canopy and 8mm thickness for the base). The high pressure lamps have a double glass filter: one blue cobalt filter in the inner part and one clear filter in the outer.
# Mastersun
This model is a bed with a high pressure tanning lamps located in five horizontal columns placed around the person body. Three columns are in the canopy, and two are in the bed, under the acrylic panel (8mm thickness). The high pressure lamps have a double glass filter one blue cobalt filter in the inner part and one clear filter in the outer.
#### Vega
This model is a booth with low pressure lamps: the machine is mainly composed of three curved panels, two fixed and one is the door. Each panel has a set low pressure lamps separated from the person by acrylic panels.
#### The associated sunlamps include:
| HPO Exotic: | BLV - Germany MHLF 630 (Kalfasun 630 F) 23470401<br>BLV - Germany MHLF 1530 (Kalfasun 1530 F)23480601 |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| LP1: | BLV - Germany MHLF 630 (Kalfasun 630 F) 23470401<br>Lighttech - Hungary B23-S White 100W<br>Lighttech - Hungary B14-S White 100W<br>Cosmedico - Germany Cosmostar Longlife 100W |
| LP2: | BLV - Germany MHLF 630 (Kalfasun 630 F) 23470401<br>Lighttech - Hungary B23-S White 100W<br>Lighttech - Hungary B25-S White 160W |
| LP3: | BLV - Germany MHLF 630 (Kalfasun 630 F) 23470401<br>BLV - Germany HLF 1530 (Kalfasun 1530 F) 23480601<br>Lighttech - Hungary B23-S White 100W<br>Lighttech - Hungary B25-S White 160W |
| Mastersun: | BLV - Germany MHLF 630 (Kalfasun 630 F) 23470401<br>BLV - Germany MHLF 640 (Kalfasun 640 F) 234720 01<br>BLV - Germany MHLF 1530 (Kalfasun 1530 F) 23480601<br>BLV - Germany MHLF 1540 (Kalfasun 1540 F) 234820 |
| 01 Vega: | Lighttech - Hungary B23-S White 160W |
{6}------------------------------------------------
The associated protective eyewear:
Ultra Sunglobes – World SunCare Products, Ltd – England
# IV. INDICATIONS FOR USE
Sunlamp product intended for use to tan the skin.
#### V. SUMMARY OF COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATEDEWICE
The subject and predicate devices are identical.
#### VI. PERFORMANCE
The subject devices were tested to be in conformity with Performance Standard 21 CFR 1040.20. They are identical to the predicate.
Usability data is included with the submission to support the conversion of indications from Rx to OTC.
#### Biocompatibility testing
The subject devices are categorized as surface devices which only come into contact with intact skin for a duration of less than 24 hours. The only material that comes into contact with the patient's skin is the polymethyl methacrylate (PMMA), which has been shown to be biocompatible via European standard 30993 (which corresponds to ISO 10993).
No additional biocompatibility testing was done, as the devices are identical to the predicate in every way.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on all devices, which were shown to pass the tests conducted.
No additional electrical or electromagnetic testing was performed, as the devices are identical to the predicate in every way.
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
{7}------------------------------------------------
#### VII. CONCLUSIONS
Since the subject and predicate devices are identical technologically, with few minor aesthetic differences, we conclude that the change in intended use from prescription to over-the-counter does not negatively affect a finding of substantial equivalence between the devices and predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.