Just Like Me Personal Lubricant

K161385 · Wtfn-Dba Classic Erotica/Holiday Products · NUC · Aug 17, 2016 · Obstetrics/Gynecology

Device Facts

Record IDK161385
Device NameJust Like Me Personal Lubricant
ApplicantWtfn-Dba Classic Erotica/Holiday Products
Product CodeNUC · Obstetrics/Gynecology
Decision DateAug 17, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2

Indications for Use

Just like Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural latex, polyisoprene, and polyurethane condoms.

Device Story

Just Like Me Personal Lubricant is a non-sterile, water-based, clear, non-greasy, odorless liquid. It is applied topically to the penis or vagina to supplement natural lubrication and enhance comfort during intimate sexual activity. The device is intended for over-the-counter use by consumers. It is compatible with natural rubber latex, polyisoprene, and polyurethane condoms. The product is supplied in 4 oz. packaging with a 2-year shelf life.

Clinical Evidence

No clinical data. Evidence consists of bench testing: biocompatibility (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity), condom compatibility with natural rubber latex, polyisoprene, and polyurethane condoms per ASTM D7661-10, and accelerated aging shelf-life studies confirming physical/microbiological specifications.

Technological Characteristics

Water-based personal lubricant. Non-sterile. Formulated as a clear, non-irritating, non-greasy, odorless liquid. Biocompatibility per ISO 10993. Condom compatibility per ASTM D7661-10. Shelf-life testing per USP <51>, <61>, and <62>.

Indications for Use

Indicated for penile and/or vaginal application to moisturize and lubricate, enhancing comfort during intimate sexual activity and supplementing natural lubrication. Compatible with natural latex, polyisoprene, and polyurethane condoms. For over-the-counter use.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 17, 2016 WTFN - dba Classic Erotica/Holiday Products % Louie Goryoka Regulatory Consultant Med-Device Consulting, Inc. 5804 Rainbow Hill Road Agoura Hills, CA 91301 Re: K161385 > Trade/Device Name: Just Like Me Personal Lubricant Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: May 12, 2016 Received: May 19, 2016 Dear Louie Goryoka, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161385 Device Name Just Like Me Personal Lubricant #### Indications for Use (Describe) Just like Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural latex, polyisoprene, and polyurethane condoms. Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY Just Like Me Personal Lubricant | Company Name:<br>Company Address: | WTFN - dba Classic Erotica<br>20950 Lassen Street<br>Chatsworth, California 91311 | |-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Med-Device Consulting Inc.<br>Louie Goryoka - Regulatory Consultant<br>Phone 1 (818) 735-0488<br>mdci@m-dci.us | | Summary Preparation Date: | August 16, 2016 | | Device Trade Name:<br>Common Name:<br>Classification Name(s):<br>Classification Regulation:<br>Device Code:<br>Advisory Panel:<br>Predicate Devices | Just Like Me Personal Lubricant<br>Personal Lubricant<br>Condom<br>21 CFR §884.5300, Class II<br>NUC<br>Obstetrics and Gynecology<br>Love Liquid® Personal Lubricant<br>510(k) number K152168 | ### Device Description The Just Like Me Personal Lubricant is a non-sterile, water based, over-the- counter personal lubricant formulated to be a clear, non-irritating, non-greasy, odorless liquid. The proposed device contains a blend of ingredients similar to ingredients found in the predicate device. The device is designed to supplement the body's own natural lubrication fluids and is compatible for use with or without natural rubber latex, polyisoprene, and polyurethane condoms during intimate sexual activity as evidenced by condom compatibility testing. #### Indication for Use Statement The Just Like Me Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural latex, polyisoprene, and polyurethane condoms. #### Technological Characteristics The Just Like Me Personal Lubricant has the same intended use as the predicate device. The Just Like Me Personal Lubricant has similar technological characteristics to the predicate device. It contains ingredients that are similar to those found in the predicate device. {4}------------------------------------------------ # Performance Data # Biocompatibility The device contact duration with the user is A-limited <24 hours, in accordance with ISO 10993. Biocompatibility testing was performed in accordance with ISO 10993. The results demonstrate that the device is safe for its intended use. | Test Performed | Results | |-----------------------------------------------------------------------------------------|-----------------------------------------------------------| | Cytotoxicity (Direct Contact) | Product is Non-toxic | | ISO GUINEA PIG MAXIMIZATION<br>SENSITIZATION TEST- (Method for Liquid<br>Test Articles) | Product does not elicit a sensitization<br>response | | Vaginal Mucosa Irritation with<br>Histopathology | Product is considered a non-irritant to<br>vaginal tissue | | Acute Systemic Toxicity | There is no evidence of system toxicity | # Condom Compatibility Compatibility Testing was performed in accordance with ASTM D7661-10; "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms" on three marketed brands of Natural Rubber Latex condoms, one brand of Polyisoprene condoms, and one brand of Polyurethane condoms. This product is compatible with natural rubber latex, polyisoprene, and polyurethane condoms as demonstrated by the condom compatibility testing. # Shelf Life Testing The Just Like Me Personal Lubricant has a 2-years shelf life based on the results of an accelerated aging study. The accelerated aging study evaluated the single version of packaging (3.8 oz.) 4 oz. packaging. Following aging, the product met its specifications for viscosity, specific gravity, appearance, color, odor passed the testing, Preservative Effectiveness and Antimicrobial Effectiveness Testing based on USP <51>, Total Microbial Aerobic Count (TAMC) and Total Yeast and Mold Count (TYMC) based on USP <61>, and Absence of Pathogenic organism based on USP <62>. # Conclusion The results of performance testing demonstrate that the Me Personal Lubricant is substantially equivalent to predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...