← Product Code [KPI](/productcode/KPI) · K161349

# Everyway Incontinence Stimulation System (K161349)

_Everyway Medical Instrument Co., Ltd. · KPI · Jul 7, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161349

## Device Facts

- **Applicant:** Everyway Medical Instrument Co., Ltd.
- **Product Code:** [KPI](/productcode/KPI.md)
- **Decision Date:** Jul 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5320
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Everyway Incontinence Stimulation System is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.

## Device Story

Dual-channel powered muscle stimulator; delivers electrical stimulation via vaginal probes (PR-02A, PR-03A, PR-04A, PR-14A) to pelvic floor muscles. User-operated via intensity control knob; LCD icons indicate operational status. Six pre-programmed stimulation modes (varying phase duration, rate, on/off times, timer) provide neuromuscular re-education. Intended for over-the-counter (OTC) use. Output parameters (current/power density) designed to match predicate device performance. Benefits include rehabilitation of weak pelvic floor muscles and treatment of urinary incontinence.

## Clinical Evidence

Usability study conducted to support device design and labeling for over-the-counter use. No clinical efficacy trials for incontinence treatment were required; substantial equivalence supported by bench testing and comparative analysis of output parameters.

## Technological Characteristics

Dual-channel powered muscle stimulator; electrical stimulation via vaginal probes. Six pre-programmed modes; max output 75mA. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10. Software verification performed per FDA guidance.

## Regulatory Identification

A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).

## Predicate Devices

- Yarlap Nonimplanted electrical continence device ([K141643](/device/K141643.md))

## Reference Devices

- Life-Care Vaginal Probe models PR-02A, PR-03A, PR-04A, PR-14A ([K122194](/device/K122194.md))
- Everyway Comfy EMS/Model EV-805 ([K071951](/device/K071951.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 7, 2017

Everyway Medical Instrument Co., Ltd. Robert Tu President and Operator Owner 3FL., No. 5, Lane 155, Section 3, Beishen Rd. Shenkeng Dist, New Taipei City, Taiwan 22203 R.O.C.

Re: K161349 Trade/Device Name: Everyway Incontinence Stimulation System Regulation Number: 21 CFR& 876.5320 Regulation Name: Nonimplanted electrical continence device Regulatory Class: II Product Code: KPI Dated: June 6, 2017 Received: June 7, 2017

Dear Robert Tu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for EVERYWAY MEDICAL INSTRUMENTS CO., LTD. The logo consists of the word "EVERYWAY" in a stylized font, with horizontal lines above and below the word. The text "EVERYWAY MEDICAL INSTRUMENTS CO.,LTD." is written in a simple, sans-serif font to the right of the stylized logo.

3F1., No. 5, Lane 155, Sec. 3,Beishen Rd, Shenkeng Dist, New Taipei City, Taiwan,

# Indications For Use

510(k) Number (if known): K161349

Device Name: Everyway Incontinence Stimulation System

Indications For Use:

Everyway Incontinence Stimulation System is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of _ 1

{3}------------------------------------------------

3Fl., No. 5, Lane 155, Sec. 3,Beishen Rd, Shenkeng Dist, New Taipei City, Taiwan,

### 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 at July 06, 2017.

The assigned 510(k) number is: K161349 .

- 1. Submitter's Identifications: Establishment: EVERYWAY MEDICAL INSTRUMENTS CO., LTD. Address: 3Fl., No. 5, Lane 155, Sec. 3, Beishen Rd., Shenkeng Dist, New Taipei City 222, Taiwan Registration Number: 9616877 Operations: Manufacturer
Owner/Operator: EVERYWAY MEDICAL INSTRUMENTS CO., LTD. Address : 3FI., No. 5, Lane 155, Sec. 3, Beishen Rd., Shenkeng Dist, New Taipei City 222, Taiwan Contact Person: Robert Tu Phone : 886-2-2662-0038 Fax No: 886-2-2664-5566 e-mail : robert@everyway-medical.com

- 2. Name of the Device: Everyway Incontinence Stimulation System
- 3. Information of the 510(k) Cleared Device (Predicate Device): K141643: Yarlap Nonimplanted electrical continence device. K122194: Life-Care Vaginal Probe models PR-02A, PR-03A, PR-04A, PR-14A (for reference) K071951:Everyway Comfy EMS/Model EV-805(for reference)

#### 4. Classification Information:

Trade/Device Name: Everyway Incontinence Stimulation System Regulation Number: 21 CFR Part 876.5320 Regulation Name: Stimulator, Electrical, Non-implantable, For Incontinence. Regulatory Class: II Product Code: KPI

#### 5. Device Description:

The Everyway Incontinence Stimulation System is a dual channel powered muscle stimulator intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.

The EV-805I stimulator is basically an Everyway's existing K071951-510(K) cleared model for power muscle stimulator. To make it delivery the same stimulation output as that of chosen predicate device, we made some change in software so as to generate muscle stimulation current that is discharged through the PR-02A, PR-04A, PR-14A Life-Care Vaginal Probe on the vaginal skin for the treatment of urinary incontinence.

{4}------------------------------------------------

3F1., No. 5, Lane 155, Sec. 3,Beishen Rd, Shenkeng Dist, New Taipei City, Taiwan,

The device is operated in such a way that turn on the intensity control knob and adjust the intended intensity or turn off the device during operation through the same control knob. While device is turned on for operation, the related LCD icons then illuminate for indicating the device is in operation condition. User then follow the instruction for use to get appropriate stimulation treatment for urinary incontinence.

With the combination of the main device parts as above mentioned , the device can be used as recommended in manual for the treatment of urinary incontinence for over-the-counter (OTC) use as that of the chosen 510(K) predicate device.

#### 6. Intended Use:

The Everyway Incontinence Stimulation System is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.

#### 7. Comparison to the 510(k) Cleared Device (Predicate Device):

The main reason for choosing these two predicate 510(K) clear model are because that all of them bear the same intended use as well as almost the same accessories. For the comparison concerning the device features, please see the following comparison table:

| Features               | 510(K) Proposed Model                                                                                                                                                                                                                                                                                 | New Model                                                                                                                                                                                                                                                                                                                                  |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model Name             | Yarlap<br>Nonimplanted electrical continence<br>device                                                                                                                                                                                                                                                | Everyway Incontinence Stimulation<br>System                                                                                                                                                                                                                                                                                                |
| 510(K) No.             | K141643                                                                                                                                                                                                                                                                                               | K161349                                                                                                                                                                                                                                                                                                                                    |
| Prescription or<br>OTC | OTC                                                                                                                                                                                                                                                                                                   | OTC                                                                                                                                                                                                                                                                                                                                        |
| FDA product<br>code    | KPI                                                                                                                                                                                                                                                                                                   | KPI                                                                                                                                                                                                                                                                                                                                        |
| Indication for<br>Use  | The device is intended to provide<br>electrical stimulation and<br>neuromuscular re-education for the<br>purpose of rehabilitation of weak<br>pelvic floor muscles for the treatment<br>of stress, urge and mixed urinary<br>incontinence in women and to<br>maintain urinary continence in<br>women. | The Everyway Incontinence<br>Stimulation System is intended to<br>provide electrical stimulation and<br>neuromuscular re-education for the<br>purpose of rehabilitation of weak<br>pelvic floor muscles for the treatment<br>of stress, urge and mixed urinary<br>incontinence in women and to<br>maintain urinary continence in<br>women. |
| Electrode Used         | Incontinence probe<br>PC Electrode area: $6.4cm²$ X2                                                                                                                                                                                                                                                  | Incontinence probe<br>Electrode area:<br>PR-02A: $7.65cm²$ X2<br>PR-03A: $7.87cm²$ X2<br>PR-04A: $6.25cm²$ X2<br>PR-14A: $9.05cm²$ X2                                                                                                                                                                                                      |

{5}------------------------------------------------

3F1., No. 5, Lane 155, Sec. 3,Beishen Rd, Shenkeng Dist, New Taipei City, Taiwan,

In addition to the significant comparison feature, we also provided the comparison for the output characteristics as the following table :

For the operation mode, both predicate and new device provided the following six program modes:

| New Device (EV-805I) |      |                |      |         |          |         |
|----------------------|------|----------------|------|---------|----------|---------|
| Program              | Max. | Phase Duration | Rate | ON Time | OFF Time | Timer   |
| P1                   | 75mA | 200us          | 12Hz | 5 sec   | 5 sec.   | 15 min  |
| P2                   | 75mA | 250us          | 20Hz | 8 sec.  | 8 sec.   | 20 min  |
| P3                   | 75mA | 200us          | 12Hz | 5 sec.  | 10 sec.  | 15 min  |
| P4                   | 75mA | 200us          | 10Hz | 6 sec.  | 12 sec.  | 20 min  |
| P5                   | 75mA | 250us          | 12Hz | 5 sec.  | 15 sec.  | 15 min  |
| P6                   | 75mA | 200us          | 35Hz | 6 sec.  | 18 sec.  | 20 min  |
| K141643 YARLAP       |      |                |      |         |          |         |
| Program              | Max. | Phase Duration | Rate | ON Time | OFF Time | Timer   |
| P1                   | 80mA | 200us          | 12Hz | 5 sec   | 5 sec.   | 15 min. |
| P2                   | 80mA | 250us          | 20Hz | 8 sec.  | 8 sec.   | 20 min. |
| P3                   | 80mA | 200us          | 12Hz | 5 sec.  | 10 sec.  | 15 min. |
| P4                   | 80mA | 200us          | 10Hz | 6 sec.  | 12 sec.  | 20 min. |
| P5                   | 80mA | 250us          | 12Hz | 5 sec.  | 15 sec.  | 15 min. |

#### 8. Discussion of Non-Clinical Tests Verification Activities Performed to Determine the Safety and Performance of the devices are as the followings:

35Hz

6 sec.

18 sec.

20 min.

P6

80mA

200us

Compliance to applicable voluntary standards includes IEC 60601-2-10, as well as IEC 60601-1. IEC60601-1-11 and IEC 60601-1-2 requirement. In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.

Discussion: The Compliance to applicable voluntary standards as above mentioned indicates that the new device in this submission used the same standards as that of predicate device. Therefore; we consider that the compliance of standards included in our submission is adequate for the determination of substantial equivalence.

#### 9. Discussion of Clinical Test Validation Activities Performed to Determine the Effectiveness of Device are as the followings:

The Usability Study was conducted for the Everyway Incontinence Stimulation System to support the device design and labeling for over-the-counter use of the device.

{6}------------------------------------------------

3Fl., No. 5, Lane 155, Sec. 3,Beishen Rd, Shenkeng Dist, New Taipei City, Taiwan,

#### 10. Summary for the technology comparison.

Basically the Everyway Incontinence Stimulation System has the similar technological characteristics with the predicate device in the product design, material, energy source type, main program mode and the main output waveform...etc. There exists some difference in the detailed output parameters (mainly in the output intensity and electrode sizes). Through the detailed calculation comparison of stimulation output energy for each operation mode(in particular the output current density and power density), we found the output level in each operation mode for our Everyway Incontinence Stimulation System and predicate device are very close and within the acceptable range. So we believe the difference in detailed output parameters does not affect the determination of substantial equivalence.

#### 11. Conclusions

The Everyway Incontinence Stimulation System has the same intended use and the similar technological characteristics as the cleared devices. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.

---

**Source:** [https://fda-staging.innolitics.com/device/K161349](https://fda-staging.innolitics.com/device/K161349)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
