← Product Code [KCT](/productcode/KCT) · K161347

# Signia sterilization tray (K161347)

_Covidien · KCT · Sep 7, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161347

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Sep 7, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Signia™ sterilization tray is intended to provide storage for the Signia™ reusable insertion guide and Signia™ manual retraction tool during sterilization, storage and transportation within the hospital environment. The Signia™ sterilization tray is only intended to maintain sterility of the enclosed if it is used in conjunction with an FDA cleared sterilization wrap and has only been evaluated for a non-stacked configuration. The tray can contain at a maximum: one (1) Signia™ adapter, one (1) Signia™ reusable insertion guide and one (1) Signia™ manual retraction tool. The tray is intended to allow steam sterilization of the enclosed medical devices. The validated sterilization cycle parameters are as follows: STEAM STERILIZATION Pre-vacuum (HiVac) Steam Cycles Item: Tray with components wrapped or in sterilization container system 132 °C Pre-vacuum (HiVac) Steam Cycles 134 °C Pre-vacuum (HiVac) Steam Cycles Exposure temperature 270 °F (132 °C) 273 °F (134 °C) Exposure time 4 minutes 3 minutes Vacuum dry time 20 - 40 minutes 20 - 40 minutes

## Device Story

Stainless steel sterilization tray; holds Signia™ adapter, reusable insertion guide, and manual retraction tool. Perforated lid/base design allows steam penetration; includes internal inserts/strips for accessory organization. Used in hospital Central Supply Room (CSR) by qualified medical personnel. Facilitates steam sterilization of enclosed accessories; requires FDA-cleared sterilization wrap. Protects accessories during storage and transport; ensures sterility maintenance post-autoclave. Does not contact patients.

## Clinical Evidence

No clinical data. Bench testing only. Cytotoxicity testing confirmed non-cytotoxic. Cleaning validation (AAMI TIR30:2011, FDA 2015 guidance) assessed carbohydrate, protein, and hemoglobin residuals; all met acceptance criteria. Sterilization validation (AAMI/ANSI ST77:2013) achieved SAL of 10^-6. Reliability evaluation confirmed material durability under reprocessing stresses.

## Technological Characteristics

Stainless steel construction; perforated base and lid; includes latches and handles. Dimensions: 10.0 x 21.4 x 3.0 inches. Sterilization: Pre-vacuum steam (132°C/134°C). Non-powered, mechanical device. No software or connectivity.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- iDrive™ Ultra sterilization tray ([K130532](/device/K130532.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem that resembles an abstract caduceus, with three stylized human profiles forming the staff and a serpent-like shape winding around them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 7, 2016

Covidien Katherine Choi, RAC Principal Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473

Re: K161347

Trade/Device Name: Signia™ Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: August 5, 2016 Received: August 8, 2016

Dear Katherine Choi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael D. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K161347

Device Name Signia™ Sterilization Tray

#### Indications for Use (Describe)

The Signia™ sterilization tray is intended to provide storage for the Signia™ reusable insertion guide and Signia™ manual retraction tool during sterilization, storage and transportation within the hospital environment. The Signia™ sterilization tray is only intended to maintain sterility of the enclosed if it is used in coniunction with an FDA cleared sterilization wrap and has only been evaluated for a non-stacked configuration. The tray can contain at a maximum: one (1) Signia™ adapter, one (1) Signia™ reusable insertion guide and one (1) Signia™ manual retraction tool. The tray is intended to allow steam sterilization of the enclosed medical devices. The validated sterilization cycle parameters are as follows:

#### STEAM STERILIZATION

#### Pre-vacuum (HiVac) Steam Cycles

| Item: Tray with components wrapped<br>or in sterilization container system | 132 °C Pre-vacuum (HiVac) Steam<br>Cycles | 134 °C Pre-vacuum (HiVac) Steam<br>Cycles |
|----------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------|
| Exposure temperature                                                       | 270 °F (132 °C)                           | 273 °F (134 °C)                           |
| Exposure time                                                              | 4 minutes                                 | 3 minutes                                 |
| Vacuum dry time                                                            | 20 - 40 minutes                           | 20 - 40 minutes                           |

Type of Use (Select one or both, as applicable)

|                                       | Depurative Use (Part 111, 21 CFR 331, Subpart D)       |
|---------------------------------------|--------------------------------------------------------|
| <div> <input type="checkbox"/> </div> | Over-The-Counter Use (Part 111, 21 CFR 331, Subpart C) |

| | Prescription Use {Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K161347 510(k) Summary

# DATE PREPARED:

September 2, 2016

#### SUBMITTER:

Covidien 60 Middletown Avenue North Haven, CT 06473 USA

# CONTACT PERSON:

Katherine Y. Choi, RAC Principal Requlatory Affairs Specialist Telephone: (203) 492-8412 Fax: (203) 492-5029

# IDENTIFICATION OF DEVICE:

| Proprietary/Trade Name: | Signia™ sterilization tray |
|-------------------------|----------------------------|
| Classification Name:    | Sterilization wrap         |
| Regulation Number:      | 21 CFR 880.6850            |
| Product Code:           | KCT                        |
| Device Class:           | Class II                   |
| Review Panel:           | General Hospital           |
| Common Name:            | Sterilization tray         |

# PREDICATE DEVICE:

| Proprietary/Trade Name: | iDrive™ Ultra sterilization tray       |
|-------------------------|----------------------------------------|
| 510(k) Number:          | K130532 (Clearance date: May 20, 2013) |
| Classification Name:    | Sterilization wrap                     |
| Regulation Number:      | 21 CFR 880.6850                        |
| Product Code:           | KCT                                    |
| Device Class:           | Class II                               |
| Review Panel:           | General Hospital                       |
| Common Name:            | Sterilization tray                     |

# DEVICE DESCRIPTION:

The Signia™ sterilization tray is an optional reusable accessory to the Signia™ stapling system, and is intended to provide storage of the Signia™ accessories during sterilization, as well as, storage and transportation within the hospital environment. The Signia™ sterilization tray can simultaneously accommodate at a maximum the following Signia™ accessories:

One (1) Signia™ adapter One (1) Signia™ reusable insertion guide One (1) Signia™ manual retraction tool

The Signia™ sterilization tray is composed of a lid and base made of stainless steel. Both the lid and base are perforated to allow penetration of steam during sterilization. There are two latches on the side to secure the lid to the base, and handles for holding or transporting the tray. Inside of the base, inserts

{4}------------------------------------------------

and strips are placed to hold the respective Signia™ accessories along with the pre-printed layout of the accessory as a visual aid. Each corner has a protective cover for protection.

The Signia™ sterilization tray is designed to facilitate the sterilization process when used in conjunction within the Central Supply Room (CSR) wrapping material (an FDA cleared sterilization wrap). The tray is intended to be reprocessed, which is first to be cleaned either manually or by an automatio washerdisinfector, and then sterilization. The tray is to be handled by medical personnel qualified in the transportation, cleaning, sterilization and use of surgical devices.

# INTENDED USE/INDICATIONS FOR USE:

The Signia™ sterilization tray is intended to provide storage for the Signia™ reusable insertion guide and Signia™ manual retraction tool during sterilization, storage and transportation within the hospital environment. The Signia™ sterilization tray is only intended to maintain sterility of the enclosed devices if it is used in conjunction with an FDA cleared sterilization wrap and has only been evaluated for a non-stacked configuration. The tray can contain at a maximum: one (1) Signia™ adapter, one (1) Signia™ reusable insertion guide and one (1) Signia™ manual retraction tool. The tray is intended to allow steam sterilization of the enclosed medical devices. The validated sterilization cycle parameters are as follows:

# STEAM STERILIZATION

Pre-vacuum (HiVac) Steam Cycles

| Item: Tray with components<br>wrapped or in sterilization<br>container system | 132 °C Pre-vacuum (HiVac)<br>Steam Cycles | 134 °C Pre-vacuum (HiVac)<br>Steam Cycles |
|-------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------|
| Exposure temperature                                                          | 270 °F (132 °C)                           | 273 °F (134 °C)                           |
| Exposure time                                                                 | 4 minutes                                 | 3 minutes                                 |
| Vacuum dry time                                                               | 20 - 40 minutes                           | 20 - 40 minutes                           |

# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:

The Signia™ sterilization tray is designed to withstand the indicated cleaning and sterilization conditions to aid in storage and sterilization of the Signia™ accessories. The tray itself is largely made of stainless steel, and does not require any special storage conditions. Repeated processing at the listed temperatures has minimal effect on the sterilization tray.

# SUBSTANTIAL EQUIVALENCE:

The Signia™ sterilization tray has the same intended use as the legally-marketed iDrive™ Ultra sterilization tray cleared via K130532 (May 20, 2013) since both are metal sterilization trays. Also the predicate device and new Signia™ sterilization tray can undergo autoclave sterilization. However, the enclosed devices are different.

{5}------------------------------------------------

The comparison between the Signia™ sterilization tray and the predicate tray K130532 is highlighted in the table below.

|                                          | Subject Device<br>Signia™ sterilization tray                                                                                                                                                                                                                                                       | Predicate Device K130532<br>iDrive™ Ultra sterilization tray                                                                                                                                             |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Enclosed<br>devices for<br>sterilization | The tray can contain at a maximum:<br>one (1) Signia™ adapter,<br>one (1) Signia™ reusable insertion guide<br>and<br>one (1) Signia™ manual retraction tool.                                                                                                                                       | The tray can contain at a maximum:<br>one (1) iDrive™ Ultra powered handle,<br>two (2) Endo GIA™ adapters,<br>two (2) iDrive™ battery insertion guides and<br>one (1) iDrive™ Ultra manual adapter tool. |
| Dimensions                               | Approx. 10.0 x 21.4 x 3.0 (H) inches                                                                                                                                                                                                                                                               | Approx. 10 x 21 x 3 (H) inches                                                                                                                                                                           |
| Sterilization<br>methods                 | Prevacuum steam                                                                                                                                                                                                                                                                                    | Prevacuum steam                                                                                                                                                                                          |
| Steam<br>Sterilization<br>parameters     | 132 °C Pre-vacuum (Hi Vac) Steam Cycle<br>Exposure temperature: 270 °F (132 °C)<br>Exposure time: 4 minutes<br>Vacuum dry time: 20 - 40 minutes<br>134 °C Pre-vacuum (Hi Vac) Steam Cycle<br>Exposure temperature: 273 °F (134 °C)<br>Exposure time: 3 minutes<br>Vacuum dry time: 20 - 40 minutes | Prevacuum Steam Cycles<br>132°C for 4 minutes<br>134°C for 3 minutes<br>Vacuum Dry: 20 minutes                                                                                                           |
| Base & Lid<br>Materials                  | Stainless steel                                                                                                                                                                                                                                                                                    | Aluminum                                                                                                                                                                                                 |
| Stacking                                 | Do not stack cases and trays in the<br>sterilization chamber.                                                                                                                                                                                                                                      | Do not stack cases and trays in the<br>autoclave chamber.                                                                                                                                                |
| Max. Load<br>Capacity                    | 10 pounds                                                                                                                                                                                                                                                                                          | 15 pounds                                                                                                                                                                                                |

# SUMMARY OF STUDIES:

Non-clinical performance data - Although the Signia™ sterilization tray is not intended for patient contact, all the materials that contact the enclosed Signia™ accessories during sterilization have been evaluated for cytotoxicity as an additional validation of safety to determine if there was any release of the material constituents or leachable chemicals produced from the cleaning and sterilization process that might be potentially toxic, and the testing results confirmed 'non-cytotoxic'.

The cleaning validation was designed to simulate the worst case scenario conforming to several industry standards including AAMI TIR30:2011 and the FDA's 2015 reprocessing quidance document. The cleanliness of the tray was assessed by carbohydrate, protein and hemoglobin residuals, and all residuals met predefined acceptance criteria demonstrating the adequate cleanliness of the tray using the cleaning instructions.

The sterilization validation, designed based on the FDA-recognized standard AAMI/ANSI ST77:2013, demonstrated the minimum sterility assurance level (SAL) of 10° can be achieved if the sterilization instructions in the instructions for use (IFU) were followed.

The reliability evaluation concluded the materials used in the Signia™ sterilization tray have been shown, through use in other devices, to withstand the cleaning and sterilization process without any adverse reactions and the tray will not undergo degradation due to the environmental stresses it is exposed to during reprocessing.

{6}------------------------------------------------

Clinical performance data - No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.

#### CONCLUSION:

Based upon the supporting data summarized above, we concluded that Signia™ sterlization tray is as safe and as effective as the legally marketed device K130532, and does not raise different questions of safety and effectiveness than the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K161347](https://fda-staging.innolitics.com/device/K161347)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com