← Product Code [GEH](/productcode/GEH) · K161337

# CryoDose H or Similar (K161337)

_Nuance Medical, LLC · GEH · Jul 12, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161337

## Device Facts

- **Applicant:** Nuance Medical, LLC
- **Product Code:** [GEH](/productcode/GEH.md)
- **Decision Date:** Jul 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

CryoDose H is indicated for use in the treatment of actinic keratosis, genital warts, lentigo, mulluscum contagiosum, sebortheic keratosis, skin tags, verruca vulgaris, and verruca plana.

## Device Story

CryoDose H is a cryosurgical device used by dermatologists and clinicians to treat benign skin lesions. The device consists of an aerosol-filled canister containing a cryogen mixture (dimethyl ether, propane, isobutane) and foam-tipped applicators. The clinician sprays the cryogen onto the foam tip until saturated, then applies the cold tip directly to the lesion for 15-40 seconds. This process removes thermal energy from the skin, freezing the lesion to facilitate removal. The device is a manual, non-powered tool; it does not involve electronic processing or software. It provides a standardized method for cryotherapy consistent with established dermatological practices.

## Clinical Evidence

Bench testing only. No clinical data presented. Evidence includes biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993) for foam-tipped applicators, chemical composition verification, and comparative bench testing against the predicate. Bench tests measured output temperature at the nozzle and saturated applicator tips, and confirmed ice-ball formation on test pads. Stability and shelf-life testing were also conducted.

## Technological Characteristics

Cryosurgical spray system; aerosol canister with valve, actuator, and foam-tipped applicators. Cryogen: 95% dimethyl ether, 2% propane, 3% isobutane. Manual operation; no energy source required. Biocompatibility per ISO 10993. Non-electronic; no software or connectivity.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- Histofreezer ([K990877](/device/K990877.md))

## Reference Devices

- Histofreezer ([K911420](/device/K911420.md))
- Histofreezer ([K924114](/device/K924114.md))
- Histofreezer ([K971392](/device/K971392.md))
- Histofreezer ([K023487](/device/K023487.md)-OTC)
- CryoNize ([K103310](/device/K103310.md))
- Dr. Scholl's Freeze Away ([K031697](/device/K031697.md) - OTC)
- Wartner Wart Removal ([K011708](/device/K011708.md) - OTC)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 12, 2016

Nuance Medical, LLC Mr. Marc Lieberman CEO 5931 Sea Lion Place, Suite 113 Carlsbad, California 92010

Re: K161337

Trade/Device Name: CryoDose H Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: May 7, 2016 Received: May 13, 2016

Dear Mr. Lieberman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K161337

Device Name CryoDose H

Indications for Use (Describe)

The CryoDose H indications for use as follows: CryoDose H is indicated for use in the treatment of actinic keratosis, genital warts, lentigo, mulluscum contagiosum, sebortheic keratosis, skin tags, verruca vulgaris, and verruca plana.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510(k) Summary

# Nuance Medical, LLC - CryoDose H

This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c).

#### I. SUBMITTER

| Owner:          | Nuance Medical, LLC.<br>5931 Sea Lion Place, Suite 113<br>Carlsbad, CA 92010<br>(760) 585-4849                             |
|-----------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Marc S. Lieberman<br>Nuance Medical, LLC.<br>5931 Sea Lion Place, Suite 113<br>Carlsbad, CA 92010<br>760-585-9548 (Direct) |
| Date Prepared:  | May 7, 2016                                                                                                                |

#### II. DEVICE

| Name of Device:      | CryoDose H                                           |
|----------------------|------------------------------------------------------|
| Common Name:         | Cryogen Spray:<br>Dimethyl Ether, Propane, Isobutane |
| Classification Name: | Unit, Cryosurgical Accessories (21CFR 878.4350)      |

Product Code: GEH

#### III. PREDICATE DEVICE

Primary:

Histofreezer Orasure Technologies, Inc. K990877 Legally marketed medical device This predicate has not been subject to a design-related or other recall.

{4}------------------------------------------------

References: As a reference for the reviewer, other predicate devices utilizing the identical ingredients (95% dimethyl ether, 2% propane, 3% isobutane) or similar ingredients and cleared by the FDA include: Histofreezer (K911420, K924114, and K971392), Histofreezer (K023487-OTC), CryoNize (K103310), Dr. Scholl's Freeze Away (K031697 - OTC), Wartner Wart Removal (K011708 - OTC).

#### IV. DEVICE DESCRIPTION

The predicate is a legally marketed device.

The proposed device, CryoDose H, is used in the practice of dermatology in the treatment of benign skin lesions using a cryogen spray system. The device consists of an aerosol-filled canister and parts within the canister. The canister is sold within a kit with accessories including: Directions for Use, swabs, and packaging materials.

The treatment methodology is simple and has remained virtually unchanged for decades. The aerosol is sprayed onto a foam covered cotton swab that is then placed upon the benign lesion (e.g. wart) for 15-40 seconds. This procedure is used and accepted by physicians using commonly accepted procedures and techniques. It utilizes a decades-old, standard cryogen composition profile to freeze common skin lesions.

#### V. INDICATIONS FOR USE

The Indications for Use is identical to the predicate device.

The CryoDose H indications for use as follows: CryoDose H is indicated for use in the treatment of the following:

- actinic keratosis -
- genital warts -
- lentigo -
- mulluscum contagiosum
- seborrheic keratosis -
- skin tags
- verruca plantaris -
- verruca vulgaris -
- verruca plana

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The proposed device has the same technical characteristics as the predicate device including materials, design, and energy source. There are no technological differences between the predicate and the submitted device.

{5}------------------------------------------------

The cryogen used in the proposed device. CryoDose H, is dimethyl ether (95%), propane (2%), isobutane (3%) is identical in chemical composition as the predicate device, Histofreezer.

All other elements of CryoDose H have the same technical characteristics as the predicate device including:

- । Canister
- -Valve
- Actuator -
- Cover Cap -
- Foam Tipped Swabs -

The method for delivering the cryogen is identical to the predicate.

## Substantial Equivalence:

The predicate device, Histofreezer is a legally marketed device. The proposed device (CryoDose H) has the same intended use (Indications for Use) as the predicate device (Histofreezer). CFR 800.100(b)(1)

The proposed device (CryoDose H) has the same technological characteristics as the predicate device (Histofreezer). There are no technological differences, including no changes in the materials, design, energy source, or other features of the device from those of the predicate device. CFR 800.100(b)(2)(i)

The proposed device and the predicate device use an identical (substantially equivalent) chemical composition by type and percent of components: dimethyl ether (95%), propane (2%), and isobutane (3%).

The proposed device and the predicate device use the same design, energy source, materials, and other features. When the liquid is dispensed, it saturates the foam tip. The foam tip is then applied directly to the desired area of treatment. Subsequently, the thermal energy is removed from the freezing of the skin lesion. Kits include a canister filled with the cryogen and foam tip applicators.

Differences, if any, would be limited to discussion and promotion of product, marketing materials, cosmetic labeling, etc. only.

The proposed device does not raise questions of safety and effectiveness efficacy different from the predicate device.

## Labeling:

The labeling of CryoDose H has been prepared to ensure the medical professional has adequate and clear instructions for safety and usage. It includes canister labeling, instructions for use, safety and warning statements and are substantially equivalent to the predicate.

{6}------------------------------------------------

Thus, CryoDose H is substantially equivalent to the predicate device.

## VII. PERFORMANCE DATA

Tests were performed to ensure the proposed device's output performance was substantially similar to the predicate.

### Biocompatibility Testing

Biocompatibility testing was performed on all patient contact device accessories: the foam tipped swab applicators to ensure that the finished device is biocompatible. Studies were conducted based on the requirements of ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. The battery of test included cytotoxicity, sensitization and irritation. The foam tip swab applicators are categorized and were tested as an external, short term contact with the skin (< 24 hours).

### Chemical Composition Confirmation

Both systems use identical aerosols profiles: dimethyl ether (95%), propane (2%), and isobutane (3%). The proposed device's aerosol is checked and verified upon receipt from the aerosol supplier to ensure the same chemical profile as the predicate.

### Structural and Parts Composition

Engineering verification measurements were taken and visual inspections were made to determine the canisters, valves, and caps were identical or substantially equivalent between the predicate and the proposed device.

## Instructions for Use (Clinical Use) Application and Methodology:

Side-by-side Instructions for Use from the predicate device and the proposed device clinical use and product claims made to determine substantial equivalence. No substantial differences exist between the predicate and proposed device's Instructions for Use.

### Side-by-Side Temperature Bench Testing

Comparative Cryogen Performance testing was conducted to determine the output temperature of the predicate and proposed devices. The output temperature was measured in three tests: (1) the output nozzle exit of the cryogen from the canister, (2) the cryogen-saturated applicator foam tips, and (3) visual confirmation for "ice ball" on neoprene-backed practice pads. Practical tests assessments to determine substantial equivalence were conducted.

{7}------------------------------------------------

### Stability Protocol and Shelf Life Testing

A stability protocol was developed to ensure that the identity, strength, quality, and purity of the product is maintained throughout its labeled dating period. Testing assessments conducted under controlled conditions at room temperature and under accelerated conditions.

### Summary

Based on the testing performance conducted the proposed device, CryoDose H, was found to have a safety and effectiveness profile that is similar, if not identical, to the predicate device, Histofreezer.

#### CONCLUSION VIII.

Based on the information presented above and within this submission, it is concluded that the CryoDose H is safe and effective for its intended use and is substantially equivalent to the predicated device. Also, the labeling is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K161337](https://fda-staging.innolitics.com/device/K161337)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
