K161333 · Spectranetics, Inc. · DRE · Jul 8, 2016 · Cardiovascular
Device Facts
Record ID
K161333
Device Name
TightRail Sub-C Rotating Dilator Sheath
Applicant
Spectranetics, Inc.
Product Code
DRE · Cardiovascular
Decision Date
Jul 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TightRail Sub-C Rotating Dilator Sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads, indwelling catheters, and foreign objects.
Device Story
Mechanical, intra-operative device for percutaneous tissue dilation. Consists of proximal handle drive mechanism and distal dilation catheter; packaged with optional outer support sheath. Operator (physician) advances, withdraws, and rotates device over target lead, catheter, or foreign object. Trigger actuation on handle translates rotation to axial movement of distal dilation mechanism via internal cam path. Distal mechanism dilates fibrous tissue attachments surrounding target object to facilitate removal. Used in clinical settings for lead/catheter extraction. Benefits patient by enabling removal of indwelling objects through minimally invasive percutaneous approach.
Clinical Evidence
No clinical data required. Substantial equivalence demonstrated via bench-top design verification and validation testing, including dimensional verification, tensile, torque, axial load, weld integrity, radio-detectability, corrosion resistance, and simulated use testing. Biocompatibility testing performed per standard requirements.
Technological Characteristics
Mechanical rotating dilator sheath; 9F to 13F diameters; 15.5 cm effective length. Features proximal handle trigger, distal dilation mechanism, and laser-cut hypotube shaft. Sterilization via AAMI TIR:28-2009 adoption. Biocompatibility testing includes cytotoxicity, sensitization, reactivity, systemic toxicity, complement activation, hemolysis, thrombosis, and pyrogenicity.
Indications for Use
Indicated for patients requiring percutaneous tissue dilation to facilitate removal of cardiac leads, indwelling catheters, and foreign objects.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three heads facing right, representing the department's mission to protect the health of all Americans. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 08, 2016
Spectranetics, Inc. Stephanie Byrum Regulatory Affairs Associate 9965 Federal Drive Colorado Springs, Colorado 80921
Re: K161333
Trade/Device Name: TightRail Sub-C Rotating Dilator Sheath Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel Dilator for Percutaneous Catheterization Regulatory Class: Class II Product Code: DRE Dated: May 11, 2016 Received: May 12, 2016
Dear Stephanie Byrum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. G. Hilleman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161333
Device Name
TightRail Sub-C Rotating Dilator Sheath
Indications for Use (Describe)
The TightRail Sub-C Rotating Dilator Sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads, indwelling catheters, and foreign objects.
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the Spectranetics logo. On the left is a blue diamond shape with a white circle in the center. To the right of the diamond is the word "Spectranetics" in a bold, blue font. Below the company name is the tagline "Always Reaching Farther" in a smaller, lighter font.
## 510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92(c) Prepared on May 9, 2016.
| 510(k) Submitter / Holder: | Spectranetics<br>9965 Federal Drive<br>Colorado Springs, CO 80921.3617 |
|----------------------------|------------------------------------------------------------------------|
| | Establishment Registration No: 3007284006 |
| Contact: | Stephanie Byrum |
| | Regulatory Affairs Associate |
| | Office: 719.447.2671 |
| | Mobile: 719.482-4765 |
| | Fax: 719.447.2070 |
| | Email: Stephanie.byrum@spnc.com |
#### Subject Device
| Device Trade Name: | TightRail Sub-C Rotating Dilator Sheaths |
|----------------------------|-------------------------------------------------|
| Device Common Name: | Sheath |
| Device Class: | II |
| Classification Regulation: | 21 CFR 870.1310 |
| Regulation Description: | Vessel dilator for percutaneous catheterization |
| Product Code: | DRE |
| 510(k) Type: | Traditional |
| Model Numbers: | 560-009, 560-011, 560-013 |
## Predicate Device
| Predicate Device | |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| The TightRail Sub-C Rotating Dilator Sheaths were compared to the following legally marketed predicate devices: | |
| 510(k) Number: | K142546 (cleared 25 September 2014) and K150360<br>(cleared 4 March 2015) |
| Manufacturer: | Spectranetics |
| Trade Name: | TightRail and TightRail Mini Rotating Dilator Sheaths |
| Device Common Name: | Sheath |
### Device Description
The TightRail Sub-C Rotating Dilator Sheaths are mechanical, intra-operative devices. The devices consist of a proximal handle drive mechanism with a distal dilation catheter. The sheaths are packaged with an optional outer support sheath. The dilator sheath is advanced, withdrawn, and rotated about the lead, catheter or foreign object to be removed. Actuating the trigger on the proximal handle activates a rotary dilation mechanism sheathed at the distal terminus of the catheter. Rotation of the inner shaft is translated to axial actuation of the dilation mechanism via a cam path contained within the distal components. Actuation of the distal dilation mechanism causes dilation of tissue and fibrous attachments surrounding the object targeted for removal, thereby facilitating removal of said object. The diameter sizes range from 9 French (F) to 13 F. The nominal effective length of the TightRail Sub-C is 15.5 cm.
### Intended and Indications for Use
The TightRail Sub-C Rotating Dilator Sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads, indwelling catheters, and foreign objects.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Spectranetics logo. The logo consists of a blue diamond shape with a white circle inside, followed by the word "Spectranetics" in blue, and the tagline "Always Reaching Farther" in smaller, black font below. The logo is clean and modern, and the colors are bright and eye-catching.
# Technological Characteristics
The TightRail Sub-C Rotating Dilator Sheath features the same performance characteristics as the predicate devices (K142546 and K150360). There are no significant changes to the function of the device. Changes have been made to the inner and outer shaft. The predicate devices featured a tri-coil design while the subject device features a laser-cut hypotube.
### Performance Data
The following testing was conducted to validate and verify that the subject device met all specifications and was substantially equivalent to the predicate device:
# Design Verification and Validation Testing
- . Dimensional Verification
- Tensile Test .
- Torque to Deform Test
- Axial Load Test ●
- Weld Integrity Test ●
- Outer Sheath Axial Load Test
- . Radio-Detectability Test
- . Corrosion Resistance Characterization
- Simulated Use Testing
- Dimensional Verification at 2 years
- Outer Sheath Axial Load Test at 2 years ●
- Simulated Use Test at 2 years ●
## Sterilization
- Product adoption equivalency per AAMI TIR:28-2009 .
- Biocompatibility:
- Cytotoxicity ●
- Sensitization
- Intracutaneous Reactivity ●
- Acute Systemic Toxicity ●
- C3a Complement Activation ●
- SC5b-9 Complement Activation
- . Direct Hemolysis
- Indirect Hemolvsis ●
- Thrombosis Partial Thromboplastin Time (PTT) ●
- Material Mediated Pvrogenicity ●
# Preclinical and Clinical Data:
Preclinical and clinical data was not required to demonstrate substantial equivalence. The design characteristics of the subject device are similar to the predicate. The design verification and validation test results demonstrated that the subject device is as safe and clinically effective as the predicate device.
### Substantial Equivalence
Based on the similarities in design between the subject and predicate devices, and the performance data, the TightRail Sub-C are substantially equivalent to the previously cleared versions of the TightRail and TightRail Mini (K142546 and K150360, respectively).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.