The HyperSound Tinnitus Module is indicated for patient home use in the temporary relief of tinnitus symptoms. The device is available in HyperSound Clear 500P audio systems to generate and deliver personalized sound output to tinnitus patients with or without hearing loss and with or without hearing aids. Any fitting of the device must be done by a hearing care professional. The target population is primarily adults (18 years or older).
Device Story
The HyperSound Tinnitus Module is an embedded firmware option for the HyperSound Clear 500P directed audio system. It functions as a tinnitus masker, generating personalized acoustic stimuli to provide temporary relief. A hearing care professional (HCP) uses the HyperFit software tool to program the device, creating a Tinnitus EQ setting tailored to the patient's audiogram and therapy preferences. The patient uses the 500P system at home, where they can further adjust volume, balance, and EQ settings across nine bands. The system uses directed audio speakers to deliver sound, avoiding direct ear contact. By providing customized sound masking, the device helps patients manage tinnitus perception, potentially reducing loudness and annoyance. The device is intended for use in any environment.
Clinical Evidence
Prospective pre/post-intervention clinical study (n=11 adults with chronic, non-pulsatile tinnitus). Subjects used the device for 60 minutes with customized acoustic stimuli. Primary endpoints were 100-point Visual Analog Scale (VAS) for loudness (VAS-L) and annoyance (VAS-A). Results showed significant reductions: mean VAS-L decreased 61% (59.6 to 23.5, p=0.000053) and mean VAS-A decreased 54% (64.3 to 29.5, p=0.00068).
Technological Characteristics
Firmware module embedded in HyperSound Clear 500P directed audio system. Uses directed audio speakers (ultrasound technology) for sound delivery. Frequency response: 300 Hz to 22 kHz. Maximum output hardware-limited to 85dB SPL at 1.5 kHz/1 meter. Power: External supply (100-250VAC to 48V DC). No patient-contact materials.
Indications for Use
Indicated for adults (18+) with chronic, non-pulsatile tinnitus for temporary symptom relief. Suitable for patients with or without hearing loss or hearing aids. Requires fitting by a hearing care professional.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 23, 2016
Turtle Beach Corporation Mr. James A. Barnes VP Administration 12220 Scripps Summit Drive. Suite 100 San Diego, CA 92131
Re: K161331
Trade/Device Name: Hypersound Tinnitus Module Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: July 19, 2016 Received: July 21, 2016
Dear Mr. Barnes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161331
Device Name HyperSound Tinnitus Module
#### Indications for Use (Describe)
The HyperSound Timitus Module is indicated for patient home use in the temporary relief of tinnitus symptoms. The device is available in HyperSound Clear 500P audio systems to generate and deliver personalized sound output to timitus patients with or without hearing loss and with or without hearing aids. Any fitting of the device must be done by a hearing care professional. The target population is primarily adults (18 years or older).
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for the HyperSound Tinnitus Module 510(k) Number: K161331
| Submitter: | Turtle Beach Corporation |
|----------------------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | James A. Barnes, VP Administration |
| Date Prepared: | May 11, 2016 |
| Device Name: | HyperSound Tinnitus Module |
| Device Class: | Class II |
| Classification Name: | Tinnitus Masker |
| Classification Regulation: | 21 C.F.R. 874.3400 |
| Product Code: | KLW |
| Predicate Devices: | K110932 GN Resound Tinnitus Sound Generator<br>K111293 SoundCureTM SerenadeTM Tinnitus<br>Treatment System |
## INDICATION FOR USE
The HyperSound Tinnitus Module is indicated for patient home use in the temporary relief of tinnitus symptoms. The device is available in HyperSound Clear 500P audio systems to generate and deliver personalized sound output to tinnitus patients with or without hearing loss and with or without hearing aids. Any fitting of the device must be done by a hearing care professional. The target population is primarily adults (18 years or older).
### DEVICE DESCRIPTION
The HyperSound Tinnitus Module is a tool to a cleared device that provides the means to create personalized sound output to provide temporary relief to patients suffering from tinnitus. It is available as a firmware option embedded in the commercially available over the counter HyperSound Clear 500P system ("500P"), a Class II, 510(k)- cleared group hearing aid (K133352). The 500P consists of an amplifier and one or two emitters (directed audio speakers) and accompanying mains power supply.
Like the 500P, the HyperSound Tinnitus Module may be used by individuals with or without hearing loss and with or without hearing aids. As part of the prescription activation of the HyperSound Tinnitus Module option, a hearing care professional ("HCP") programs a Tinnitus EQ setting for the patient which may be a standard flat EQ or a custom EQ to fit to a patient's hearing loss and/or to shape frequency output to the user's tinnitus therapy preferences.
The activation and programming by the HCP is accomplished using the HyperFit fitting software tool only available to HCPs for use with the 500P. The user may select amongst relief sounds, fine-tune up to nine EQ bans to their liking and adjust common audio settings on the 500P device such as volume and balance (left and right speaker).
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE
## DEVICES
The HyperSound Tinnitus Module is a firmware functionality option to a cleared device. It has similar features as compared to the predicate devices, as indicated in the table below.
{4}------------------------------------------------
| | New | Predicate | Predicate |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>Device Name | Turtle Beach<br>HyperSound Tinnitus Module | GN ReSound A/S<br>Tinnitus Sound Generator Module | SoundCure, Inc.<br>SoundCure Serenade Tinnitus Treatment System |
| 510(k) | K161331 | K110932 | K111293 |
| Indications for Use | The HyperSound Tinnitus Module is indicated for patient home use in the temporary relief of tinnitus symptoms. The device is available in HyperSound Clear 500P audio systems to generate and deliver personalized sound output to tinnitus patients with or without hearing loss and with or without hearing aids. Any fitting of the device must be done by a hearing care professional.<br><br>The target population is primarily adults (18 years or older). | The Tinnitus Sound Generator Module is a tool to generate sounds to be used in a Tinnitus Management Program to relieve patients suffering from tinnitus. The target population is primarily the adult population over 18 years of age. This product may also be used with children 5 years of age or older. The Tinnitus Sound Generator Module is targeted for healthcare professionals, which are treating patients suffering from tinnitus, as well as conventional hearing disorders. The fitting of the Tinnitus Sound Generator Module must be done by a hearing professional participating in a Tinnitus Management Program. | The SoundCure Serenade Tinnitus System is indicated for use in the temporary relief of tinnitus symptoms. The device is a tool to generate customized sounds to relieve patients suffering from tinnitus and can be used in a tinnitus management program.<br><br>The target population is adults (18 years or older).<br><br>This is a medical device and should only be used with the advice of a physician, audiologist or other hearing healthcare professional. |
| User Population | Adult population (18 years or older) | Adult population (>18 yrs)<br>May also be used with children >5 yrs of age | Adult population (>18 yrs) |
| Where Used | Anywhere | Anywhere | Anywhere |
| Schedule of Use | All day | All day in all environments | All day |
| Mechanism of Action | Uses sound stimulus that can be frequency shaped and customized to the patient. Level of sound can be adjusted by a user volume control. Independent volume parameters per ear (inherent in the device through speaker balance control) Stimulus designed to be placed in the background and ignored | Uses noise that can be broad band to narrow band customized to the patient. Level of sound can be adjusted by a user volume control. Independent volume parameters per ear (inherent to ear specific device) Stimulus designed to be placed in the background and ignored | Uses noise that can be configured from broad band to narrow band, and pure tones customized to the patient. Level of sound can be adjusted by a user volume control. Independent volume parameters per ear Stimulus designed to be placed in the background and ignored |
| Technological Characteristics | Firmware functionality option embedded into a group hearing device. | Software module embedded into a digital hearing instrument. | Sounds encoded into a handheld device with earphones. |
| Maximum Output Characteristics | Maximum output hardware limited at 85dB SPL @ 1.5 kHz/1 meter<br>Response: 300 Hz to 22 kHz | Maximum output fixed at 93dB SPL<br>Output Frequency Response: Cutoffs at 500-6000 Hz | Maximum output fixed at 92dB SPL<br>Output Frequency Response: 1 kHz to 14 kHz |
| Design Features | Tinnitus Sound generator of tones and nature sounds with frequency shaping<br>Sounds customized to the patient<br>Option to a directed audio group hearing aid | Tinnitus Sound Generator of tones and nature sounds with frequency shaping<br>Sounds customized to the patient<br>Option to a behind the Ear hearing aid<br>4 sound programs / tracks | Patient Device Sound Generator with frequency shaped sounds<br>Sounds customized to the patient<br>Handheld device / earphones<br>4 sound programs / tracks |
| Target Anatomy | Ear | Ear | Ear |
| Patient Contact<br>Materials | None | Silicone domes in open<br>configuration | Silicone Earphones |
| Transducer | Directed audio speaker | Hearing aid transducer | Earphones |
| Power | External power supply (100-<br>250VAC to 48V DC) | Battery, Hearing Aid Battery | Rechargeable Lithium-ion<br>Battery<br>Also includes external power<br>supply (100-250VAC to 5V DC)<br>with power cord for recharging |
| Components | Firmware option embedded in<br>500P directed audio system<br>(amplifier, speakers and<br>cabling) with external mains<br>power supply.<br>500P User's Guide<br>HyperSound 500P Tinnitus<br>Module User's Guide | Hearing Aid<br>User Manual | SoundCure Serenade Patient<br>Device (audio player)<br>Earphones<br>External power supply for<br>recharging<br>User's Manual |
| Meets Applicable<br>Standards testing | Yes | Yes | Yes |
# Comnarison Tahle
{5}------------------------------------------------
The subject and predicate devices produce individually customizable sounds designed to provide masking relief to patients suffering from tinnitus. The subject is embedded firmware similar to GN ReSound A/S and like the Serenade device is not limited to users with hearing loss or listeners wearing aid. The addition of the tinnitus firmware module to the cleared 500P device does not affect the safety or effectiveness of the 500P device.
The subject device uses one or more directed audio speakers rather than a hearing aid or earbuds to deliver audio. There is no contact with the body or the ear and the device does not touch the patient during use. This technological difference does not present any new issues of safety or effectiveness.
## PERFORMANCE DATA
### Software Verification and Validation
The HyperSound Tinnitus Module is embedded firmware that has been verified and validated according to the relevant standards for medical device software and risk management procedures conducted and documented as recommended by FDA's Guidance for Industry and Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." All testing has demonstrated that the design of the HyperSound Tinnitus Module meets the established specifications necessary for consistent performance during its intended use.
The HCP HyperFit software tool is used to customize the Tinnitus Module and its application was verified and validated with commercialization of the 500P. The software for the HyperSound Tinnitus Module was considered as a "minor" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or any operator or lead to delay in delivery of appropriate medical care that would likely lead to a minor injury. The output is limited by the 500P system and is less than the output of the predicate devices, which produce sound directly at the ear.
### Clinical Study - HyperSound Tinnitus Support System
A clinical study was conducted by the California Hearing and Balance Center in La Jolla, California under a formal, written, Institutional Review Board (IRB) approved protocol. The clinical study was performed in a controlled audiology clinical office as a pre and post
{6}------------------------------------------------
intervention clinical outcomes study using adult subjects with chronic, non- pulsatile tinnitus. This study is summarized as follows:
Objective: To assess whether high-frequency directed audio, using ultrasound technology, reduces the perception of tinnitus in subjects with chronic, non- pulsatile tinnitus. Study Design: Pre and post-intervention clinical outcomes study. Setting: Tertiary referral center.
Patients: Eleven adult patients with chronic, non-pulsatile tinnitus. Four subjects had unilateral chronic tinnitus and seven patients had bilateral chronic tinnitus. Intervention(s): After being exposed to a quiet environment for 5 minutes to establish baseline tinnitus, subjects completed a validated 100 point Visual Analog Scale (VAS) for tinnitus loudness (VAS-L) and annoyance (VAS-A). Subjects, in consultation with the Investigator (a HCP), selected from a choice of customized acoustic stimuli available to play on the HyperSound 500P. The HyperSound 500P device was tuned or shaped by the Investigator to the patient's audiogram using a standard fitting algorithm. Subjects, with consultation from the Investigator, then fine-tuned (EQ) settings to customize the frequency response of the HyperSound directed audio system output and adjusted volume to their personal preference. Subjects were then exposed (listened) to the selected and customized acoustic stimulus for duration of 60 minutes. After exposure, subjects again completed the Visual Analog Scale (VAS) for tinnitus loudness (VAS-L) and annoyance (VAS-A) to indicate the decree of annoyance and loudness while in the customized acoustic stimulus. Primary Outcome Measure: 100 point validated Visual Analog Scale (VAS) for tinnitus loudness (VAS-L) and annoyance (VAS-A).
Results: Highly significant reduction in tinnitus loudness and annoyance was observed in all subjects during exposure to a customized acoustic stimulus delivered with high frequency directed audio. The mean VAS-L scores decreased by 61% from a pre-exposure mean score of 59.6 to a post-exposure mean score of 23.5 (p=0.000053). The mean VAS-A scores decreased by 54% from a pre-exposure mean score of 64.3 to a post-exposure mean score of 29.5 (p=0.00068).
Conclusion: Delivery of a customized acoustic stimulus via high frequency directed audio using ultrasound technology demonstrated reduction in tinnitus loudness and annovance during exposure to the stimulus.
# SUMMARY OF DATA
The HyperSound Tinnitus Module has been carefully compared to legally marketed devices with respect to intended use and technological characteristics. In addition, nonclinical testing was conducted to measure output and to validate the performance of the software and ensure the module performs as intended and meets the design specifications. The clinical performance information provides data on the masking relief of the HyperSound Tinnitus Module.
The performance testing and comparison to the predicate devices demonstrate that the HyperSound Tinnitus Module is substantially equivalent to the predicate devices and the delivery of masking sound through directed audio does not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.