K161325 · Stethocloud Pty.Ltd (Clinicloud) · FLL · Jun 22, 2016 · General Hospital
Device Facts
Record ID
K161325
Device Name
CliniCloud Non-contact Thermometer, Model SPL1024
Applicant
Stethocloud Pty.Ltd (Clinicloud)
Product Code
FLL · General Hospital
Decision Date
Jun 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The CliniCloud Non-Contact Thermometer is an infrared thermometer intended for the periodic measurement of human body temperature for persons of all ages, to be used in conjunction with the CliniCloud app.
Device Story
Battery-powered, Bluetooth Low Energy (BLE) enabled infrared thermometer; lacks physical buttons or display. Operates by measuring infrared radiation emitted by patient forehead via Melexis MLX90614ESF-DCC thermopile sensor; ambient temperature measured via internal thermistor. Device pairs with smartphone running CliniCloud app; user initiates measurement via app. Data transmitted via BLE to smartphone; app performs heat loss calculation to convert forehead surface temperature to oral equivalent. Used in home or clinical settings by patients or clinicians. Output displayed on smartphone screen; assists in monitoring body temperature. Benefits include wireless data logging and smartphone-based visualization.
Clinical Evidence
Clinical validation study per ASTM E1965-98 conducted with 60 subjects (85% male, 15% female) across three environments. Compared subject device against predicate and four calibrated reference devices (gold standard). Results demonstrated equivalent clinical accuracy between subject and predicate devices.
Technological Characteristics
Infrared thermometer; Melexis MLX90614ESF-DCC thermopile sensor; Nordic Semiconductor NRF51822 BLE SOC. Powered by 2x AAA batteries. Dimensions: 115 x 36 x 25 mm. Connectivity: BLE 2.4GHz. Standards: ASTM E1965-98, IEC 60601-1, IEC 60601-1-2, FCC Part 15. Biocompatibility: ISO 10993-1, -5, -10. Software level of concern: moderate.
Indications for Use
Indicated for periodic measurement of human body temperature for persons of all ages.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of three faces in profile, stacked on top of each other. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2016
StethoCloud Pty. Ltd. c/o Mr. Dave Yungvirt Third Party Review Group, LLC. The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041
Re: K161325
Trade/Device Name: CliniCloud Non-Contact Thermometer, Model SPL1024 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: May 30, 2016 Received: June 1, 2016
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows a signature of a person named Tina Kiang. The signature is written in black ink and is on a white background. The signature is illegible, but the name Tina Kiang is clearly visible above the signature. The signature appears to be authentic.
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K161325
Device Name
CliniCloud Non-Contact Thermometer, Model SPL1024
Indications for Use (Describe)
The CliniCloud Non-Contact Thermometer is an infrared thermometer intended for the periodic measurement of human body temperature for persons of all ages, to be used in conjunction with the CliniCloud app.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# CliniCloud Non-Contact Thermometer
## K161325
### 1. Submission Sponsor
StethoCloud Pty. Ltd (CliniCloud) Level 5, 141 Flinders Lane Melbourne VIC, 3000 Australia Contact: Andrew Lin Title: CEO
### 2. Submission Correspondent
Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (561)305-5075 Contact: Diane Sudduth, Senior Consultant, RA Email: project.management@emergogroup.com
#### 3. Date Prepared
June 22, 2016
#### 4. Device Identification
| Trade/Proprietary Name: | CliniCloud Non-Contact Thermometer, Model SPL1024 |
|-------------------------|---------------------------------------------------|
| Common/Usual Name: | Forehead Thermometer |
| Classification Name: | Clinical Electronic Thermometer |
| Regulation Number: | 880.2910 |
| Product Code: | FLL – clinical electronic thermometer |
| Device Class: | Class II |
| Review Panel: | General Hospital |
#### Legally Marketed Predicate Device(s)
K131243, Non-contact Infrared Thermometer JPD-FR100, Shenzhen Jumper Medical Equipment Co., Ltd.
The Wireless Thermometer WT701 by Raiing Medical Company (K132761) is used as the Reference Device. The Raiing Medical Thermometer provides an example of a cleared device that communicates wirelessly to a smart phone application as the receiver.
{4}------------------------------------------------
### 5. Device Description
The CliniCloud Non-Contact Thermometer Model SPL1024 (abbreviated 'CliniCloud Non-Contact Thermometer') is a battery powered Bluetooth Low Energy (BLE) enabled portable infrared thermometer intended for the measurement of human body temperature for persons of all ages. It features no physical buttons or a display and must be used in conjunction with the CliniCloud App on a compatible BLE enabled smartphone. The CliniCloud App is free to download.
When power is applied to the device via 2 AAA batteries, the CliniCloud Non-Contact Thermometer starts to advertise via BLE, and can establish a BLE connection with BLE enabled smart phones with the CliniCloud app. The user could then request the CliniCloud non-contact thermometer to take temperature readings from within the CliniCloud app. The measured temperatures are sent back to the smart phone via the established Bluetooth link and displayed on the smartphone's display in the CliniCloud app. The CliniCloud Non-Contact Thermometer uses a medical grade MLX90614ESF-DCC thermopile sensor supplied by Melexis to passively measure object surface temperature using infrared radiation emitted by patients. The sensor also measures the ambient temperature with a built-in thermistor. The sensor is factory calibrated to meet consensus standard ASTM E1965-98. Upon receiving a request from the CliniCloud APP over the established BLE connection the onboard Nordic Semiconductor NRF51822 BLE SOC microcontroller communicates with the thermopile (MLX90614ESF-DCC) over the 12C bus and sends the received temperature readings including the object surface temperature and the ambient temperature to the CliniCloud app via the established BLE connection. On the smartphone app, the well-known heat loss equation is used to perform conversion of forehead surface temperature to oral equivalent temperature.
#### 6. Indication for Use Statement
The CliniCloud Non-Contact Thermometer is an infrared thermometer intended for the periodic measurement of human body temperature for persons of all ages, to be used in conjunction with the CliniCloud app.
## 7. Substantial Equivalence Discussion
The following table compares the CliniCloud Non-Contact Thermometer to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. A reference device was used to support substantial equivalence since the subject device does not have the same technological characteristics (i.e. Bluetooth connectivity and display) as the predicate device. Aside from minor differences in the indications for use statements, i.e. the use environment is not included in the indications for use statement; the CliniCloud Non-Contact Thermometer has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device and conclude that these differences do not alter the intended use of the device.
{5}------------------------------------------------
| Manufacturer | StethoCloud Pty. Ltd (CliniCloud) | Shenzhen Jumper Medical<br>Equipment Co., Ltd. |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CliniCloud Non-Contact<br>Thermometer, Model SPL1024 | Non-contact Infrared Thermometer<br>JPD-FR100 |
| 510(k) Number | K161325 | K131243 |
| Product Code | FLL - clinical electronic<br>thermometer | FLL - clinical electronic thermometer |
| Regulation Number | 880.2910 | 880.2910 |
| Regulation Name | Clinical Electronic Thermometer | Clinical Electronic Thermometer |
| Indications for Use | The CliniCloud Non-Contact<br>Thermometer is an infrared<br>thermometer intended for the<br>periodic measurement of human<br>body temperature for persons of all<br>ages, to be used in conjunction with<br>the CliniCloud app. | The non-contact infrared<br>thermometer, model JPD-FR100, can<br>measure body temperature for<br>infants and adults without contact to<br>human body. It can be used by<br>consumers in household<br>environment and doctor in clinic as<br>reference. |
| Measurement method | Infrared radiation detection | Infrared radiation detection |
| Measurement mode | Forehead measure mode | Forehead measure mode |
| Measuring Range | 32.2°C - 43.3°C (90.0°F - 109.9°F) | 32.2°C - 43.3°C (90.0°F - 109.9°F) |
| Display resolution | 0.1ºC (0.1ºF) | 0.1°C (0.1°F) |
| Measuring accuracy | ± 0.2ºC (0.4ºF) | ± 0.2ºC (0.4ºF) |
| Standards met | ASTM E1965-98 | ASTM E1965-98 |
| C/F switchable | Yes | Yes |
| Recommended<br>measurement distance | 0.1-5cm | 1-6cm |
| Display | Smartphone device Display | LCD display |
| Working voltage | 3V | 3V |
| Memory | Readings are stored on smartphone<br>via app and limited by smartphone<br>hard-drive capacity | 20 sets |
| Battery | 2 x 1.5V AAA | 2 x 1.5V AAA |
| Low battery indication | Yes | Yes |
| Manufacturer | StethoCloud Pty. Ltd (CliniCloud) | Shenzhen Jumper Medical<br>Equipment Co., Ltd. |
| Trade Name | CliniCloud Non-Contact<br>Thermometer, Model SPL1024 | Non-contact Infrared Thermometer<br>JPD-FR100 |
| Waterproof | No | No |
| Dimension | 115 x 36 x 25 mm | 145 x 60 x 50 mm |
| Weight | 80g | 180g |
| Recommended operating<br>condition | 15°C to 40.0 (59°F to 104.0°F)<br><85%, non-condensing | 10°C - 40.0°C (50°F - 104.0°F)<br><95% humidity, non-condensing |
| Signal transmission | Wireless 2.4G Bluetooth LE | None |
{6}------------------------------------------------
As seen in the comparison table, the subject and predicate devices have similar design features and performance specifications. One of the main technological differences between the subject and predicate devices are related to the wireless transmission characteristics in which a reference device was used to support the Bluetooth connectivity. A reference device, K132761, Raiing Wireless Thermometer WT701 was used to demonstrate equivalence for the Bluetooth connectivity. Another technological difference is the display. The subject device uses a smartphone display via the CliniCloud App whereas the predicate device uses a built-in LCD display.
These differences, however, do not raise different questions of safety or effectiveness. As demonstrated in the non-clinical testing, the different technological characteristics do not affect the safety and effectiveness of the CliniCloud Non-Contact Thermometer.
## 8. Non-Clinical Performance Data
Non-clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC60601-1:2005 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2007 Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
- . FCC Part 15 Subpart C test FCC Part 15.247
- ASTM E1965-98: (R 2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- . Biocompatibility testing standards for components that have limited skin contact for the user only, including, "ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation
{7}------------------------------------------------
and testing within a risk management process", "ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity" and "ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization"
- ISO 14971:2007 Medical devices - Application of Risk Management to Medical Devices
The proposed device also passed performance testing, which included wireless coexistence testing, transmission distance testing, measuring distance testing, device accuracy testing and comparison testing to the predicate device under use case environment. The software was assessed according to "Guidance for the Content of Premarket Submissions for Software in Medical Devices, issued on May 11, 2005". The overall level of concern was found to be moderate, with verification and validation testing confirming that the software performs as intended and is compliant to requirement and specification documents.
### 9. Clinical Performance Data
Validation study was performed to clinically validate the CliniCloud thermometer, Model SPL1024, per the specification of ASTM E1965-98, under three (3) different environment settings. This validation study includes data from 60 subjects, 1 subject device, 1 predicate device and 4 calibrated reference devices (same model) used as gold standard. Study included 85% males and 15% females. The validation study demonstrated that the clinical accuracy of the CliniCloud thermometer and the predicate devices yielded equivalent accuracy.
Both the clinical and non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 10. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics, but can be demonstrated that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device(s).
The CliniCloud Non-Contact Thermometer, as designed and manufactured, is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.