Vein Clamps, Artery Clamps

K161315 · Aros Surgical Instruments Corporation · DXC · Aug 18, 2016 · Cardiovascular

Device Facts

Record IDK161315
Device NameVein Clamps, Artery Clamps
ApplicantAros Surgical Instruments Corporation
Product CodeDXC · Cardiovascular
Decision DateAug 18, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4450
Device ClassClass 2

Indications for Use

A surgical instrument used to occlude a blood vessel temporarily.

Device Story

Vein and Artery Clamps are small surgical instruments used to temporarily occlude blood vessels during surgical procedures. Operated by surgeons in clinical settings, these manual devices apply mechanical pressure to vessels to control blood flow. The device modification includes expanded closing pressure ranges (15 to 120 gms), the addition of a curved design option, and a transition to radiation sterilization. The device functions as a mechanical clamp; it does not involve electronic inputs, software, or automated processing. By providing controlled occlusion, the clamps assist surgeons in maintaining a clear surgical field and managing vascular access during procedures.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical performance testing, including biocompatibility testing per ISO 10993, sterilization validation per ISO 11137, and mechanical compression studies to verify closing pressures.

Technological Characteristics

Manual vascular clamps; expanded closing pressure range of 15 to 120 gms; straight and curved design options; sterilization via radiation (ISO 11137); biocompatibility per ISO 10993.

Indications for Use

Indicated for use as a surgical instrument to temporarily occlude blood vessels in patients undergoing surgery.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 18, 2016 AROS Surgical Instruments Corporation % Mr. Greg Holland Regulatory Specialist Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, California 92606 Re: K161315 Trade/Device Name: Vein Clamp and Artery Clamp Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II Product Code: DXC Dated: July 18, 2016 Received: July 19, 2016 Dear Mr. Greg Holland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Brian D. Pullin -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K161315 Device Name Vein Clamp and Artery Clamp | Indications for Use ( <i>Describe</i> ) | A surgical instrument used to occlude a blood vessel temporarily. | |-----------------------------------------|-------------------------------------------------------------------| |-----------------------------------------|-------------------------------------------------------------------| | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| | | | X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K161315 ### 5. 510(k) Summary Image /page/3/Picture/2 description: The image shows the logo for AROSurgical Instruments Corporation. The word "AROSurgical" is in a stylized font with horizontal lines running through the letters. Below the word is the phrase "INSTRUMENTS CORPORATION" in a smaller, sans-serif font. | Submission Date | May 9, 2016 | |----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | AROS Surgical Instruments Corporation<br>2111 Descanso<br>Newport Beach, CA 92660<br>TEL: 949-640-9932<br>FAX: 949-640-9934 | | Contact Person | Greg Holland<br>Regulatory Consultant to AROS Surgical<br>Instruments Corporation<br>Regulatory Specialists, Inc.<br>3722 Ave. Sausalito<br>Irvine, CA 92606<br>TEL: 949.422.3853<br>FAX: 949.552.2821<br>EMAIL: greg@regulatoryspecialists.com | | Establishment Registration | 2085887 | | Common Name | Vascular Clamp | | Trade Name | Vein Clamp and Artery Clamp | | Classification Name | Clamp, Vascular | | Regulation | 870.4450 | | Class | II | | Panel | Cardiovascular | | Product Code | DXC | | Associated FDA Numbers | K961100 | | Reason for Submission | Device Modification | | Predicate | K961100, Micro Anastomosis Clamps | | Manufacturing Site | Bear Medic Corporation<br>1361 Chika-Machi, Daigomachi<br>Kuju Gun Ibaraki | {4}------------------------------------------------ 319-35 Japan TEL: 81-295 72 1811 FAX: 81-295 72 4157 Special Controls No applicable mandatory performance standards or special controls exist for this device Description - The Vein Clamps and Artery Clamps are small clamps used in surgery on small blood vessels. This Special 510(k) is being done to expand on the pressures, add a curved design, and change the sterilization process. There are no other changes and the original products cleared under K961100 remain unchanged and are still being sold. Intended Use - A surgical instrument used to occlude a blood vessel temporarily. Indications for Use - A surgical instrument used to occlude a blood vessel temporarily Technological Characteristics - This Special 510(k) is being done to expand on the pressures, add a curved design, and change the sterilization process. Changes from the original 510(k), K961100 | | K961100 | Changes | |-------------------|-----------------------------|--------------------| | Closing Pressures | 19.8 to 60.1 gms | 15 to 120 gms | | Design | All clamps were<br>straight | Adds curved design | | Sterilization | Ethylene Oxide | Radiation | - Testing Biocompatibility using ISO 10993 and FDA guidance document were successfully completed. Closing pressures were measured. Sterilization validation was done to ISO 11137. Conclusions from non-clinical performance data After performing non-clinical performance studies, Biocompatibility to ISO 10993, Sterilization Validations and Compression Studies, the data shows that the Vein Clamps and Artery Clamps are substantially equivalent to the predicate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...