← Product Code [JAK](/productcode/JAK) · K161285

# Ablation Confirmation (K161285)

_Neuwave Medical, Inc. · JAK · Jun 29, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161285

## Device Facts

- **Applicant:** Neuwave Medical, Inc.
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Jun 29, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Ablation Confirmation™ (AC), is a Computed Tomography (CT) image processing software package available as an optional feature for use with the Certus® 140 2.45 GHz Ablation System. AC is controlled by the user via an independent user interface on a second monitor separate from the Certus 140 user interface. AC imports images from CT scanners and facility PACS systems for display and processing during ablation procedures. AC assists physicians in identifying ablation targets, assessing proper ablation probe placement and confirming ablation zones. The software is not intended for diagnosis.

## Device Story

AC is CT image processing software integrated with Certus® 140 2.45 GHz Ablation System; operated by physicians via dedicated second monitor/mouse in procedure room. Inputs: CT/MR images via DICOM from PACS/scanners. Processing: semi-automatic segmentation of target lesions/ablation zones; 3D rendering; registration of pre- and post-procedure scans; tissue contraction modeling. Outputs: 2D/3D visualizations of targets, probes, and ablation zones. Clinical utility: enables assessment of probe placement and ablation margins relative to target lesions; supports remote consultation. Benefits: improved procedural accuracy and technical success confirmation.

## Clinical Evidence

Bench testing only. Verification and validation testing performed per 21 CFR 820.30 and FDA software guidance. System passed all pre-determined acceptance criteria.

## Technological Characteristics

Software-based image processing; DICOM connectivity; USB mouse interface; runs on Certus® 140 system hardware; utilizes semi-automatic segmentation and registration algorithms.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Ablation Confirmation Software ([K150313](/device/K150313.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2016

NeuWave Medical, Inc. % Mr. Dan Kosednar Director of Regulatory Affairs and Ouality Assurance 3529 Anderson Street MADISON WI 53704

Re: K161285

Trade/Device Name: Ablation Confirmation™ Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 6, 2016 Received: May 9, 2016

Dear Mr. Kosednar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K161285

Device Name Ablation Confirmation™

#### Indications for Use (Describe)

Ablation Confirmation™ (AC), is a Computed Tomography (CT) image processing software package available as an optional feature for use with the Certus® 140 2.45 GHz Ablation System. AC is controlled by the user via an independent user interface on a second monitor separate from the Certus 140 user interface. AC imports images from CT scanners and facility PACS systems for display and processing during ablation procedures. AC assists physicians in identifying ablation targets, assessing proper ablation probe placement and confirming ablation zones. The software is not intended for diagnosis.

| Type of Use (Select one or both, as applicable)                                                                                           |                                                                                                                                          |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Date:              | 06/09/2016                                                                                                                       |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Subject:           | 510(k) Summary of Safety and Effectiveness Information for NeuWave<br>Medical's Ablation Confirmation™ image processing software |
| 510(k) Number:     | K161285                                                                                                                          |
| Company:           | NeuWave Medical, Inc.<br>3529 Anderson Street<br>Madison, WI 53704                                                               |
| FDA Establishment# | 3008769756                                                                                                                       |
| Contact:           | Dan Kosednar, Director of Regulatory Affairs and Quality Assurance<br>P – 608-512-1592<br>F – 608-512-1509                       |
| Proprietary:       | Ablation Confirmation™ Common: Computed Tomography X-ray System                                                                  |
| Classification:    | Radiology, JAK, 21 CFR 892.1750                                                                                                  |

The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 1992.

## Predicate Devices

Ablation Confirmation™ is substantially equivalent to the following currently marketed device:

- Ablation Confirmation Software – Class II – 21CFR892.1750 which has been the subject of a cleared 510(k) with the FDA log number K150313.
### Indications For Use

Ablation Confirmation™ (AC), is a Computed Tomography (CT) image processing software package available as an optional feature for use with the Certus® 140 2.45 GHz Ablation System. AC is controlled by the user via an independent user interface on a second monitor separate from the Certus 140 user interface. AC imports images from CT scanners and facility PACS systems for display and processing during ablation procedures. AC assists physicians in identifying ablation targets, assessing proper ablation probe placement and confirming ablation zones. The software is not intended for diagnosis.

### Device Description

AC is resident on the Certus® 140 system and is accessible to the physicians via a second, dedicated monitor with its own user interface separate from the ablation user interface. AC functions are controlled via a USB connected mouse. AC connects to a facility PACS system and CT scanner and receives and sends CT and MR images via the DICOM protocol.

AC contains a wide range of image processing tools, including:

- 2D image manipulation
- 3D image generation (from 2D images) ●
- 3D image manipulation
- Region of interest (ROI) identification, segmentation and measurement ●
- Automatic identification of ablation probes
- . Registration of multiple images into a single view

{4}------------------------------------------------

Prior to an ablation procedure, physicians can use AC to semi-automatically segment and visualize ablation target lesions in soft tissue including liver, lung and kidney. The physician initiates the segmentation with tools provided on the screen. AC then uses segmentation algorithms to construct a 2-D visualization of the target lesion selected. The physician can accept the initial segmentation results or use AC tools to manually adjust the defined target lesion. Once accepted, the identified target is rendered into a 3D image.

Upon the placement of ablation probes, taking and importing the CT scan, AC can process the image and identify up to three ablation probes. AC can then perform a registration of the initial CT scan, containing the identified target with the second scan containing the ablation probe(s) in place. The resulting image allows the physician to visualize the ablation probe(s) in relation to the identified target. This enables physicians to ensure proper probe(s) placement prior to starting the ablation. Following the ablation procedure and a post-procedure CT scan, AC allows the physician to semi-automatically segment and visualize the ablation zone using the same process as in the initial target segmentation. AC then performs a registration of the initial CT scan, containing the identified target, with the final CECT scan containing the segmented ablation zone. The physician also has the option to evaluate the effect of potential tissue contraction to help determine the technical success (ablation zone covers target lesion with desired amount of margin) of the ablation procedure.

All AC processing and viewing is accomplished at the Certus® 140 Ablation System without the physician having to leave the procedure area to utilize separate image processing tools.

Additionally, AC allows for the images to be viewed by a remote physician for time-saving clinical consultation on the current procedure.

### Modifications

This 510(k) was submitted to update the Ablation Confirmation software with the following modifications:

- . Add the option for the physician to select a tissue contraction effect to the identified target on the "Evaluate Ablation" screen
### Performance Data

Ablation Confirmation™ was tested in accordance with a test plan that fully evaluated all functions performed by the software. The system passed all pre-determined acceptance criteria identified in the test plan.

Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30, which included testing that fulfills the requirements of FDA "Guidance on Software Contained in Medical Devices". Potential risks arising from the addition of the Tissue Contraction feature were analyzed and satisfactorily mitigated in the device design and labeling.

### Substantial Equivalence Discussion

Ablation Confirmation™ is substantially equivalent in design concepts, technologies and materials to the identified predicate. AC does not present any new questions of safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K161285](https://fda-staging.innolitics.com/device/K161285)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).