K161259 · KLS Martin L.P. · HWC · Dec 19, 2016 · Orthopedic
Device Facts
Record ID
K161259
Device Name
KLS Martin Cannulated Headless Screws
Applicant
KLS Martin L.P.
Product Code
HWC · Orthopedic
Decision Date
Dec 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
KLS Martin Cannulated Headless Screws are intended for the treatment of fractures, osteotomies, and nonunions of small bones in the hand, wrist, foot, and ankle.
Device Story
System of headless cannulated bone screws; used for fixation in fractures, non-unions, osteotomies, or small joint fusions. Screws feature proximal and distal threads with different pitches to provide compression across bone fragments; cannulated design allows insertion over guide wire for precise placement. Screws are self-drilling and self-tapping. Used by qualified physicians in clinical settings (OR/clinic). Output is mechanical fixation of bone segments to facilitate reduction, stability, and healing. Benefits include improved bone alignment and stabilization for small bone procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical properties (torsional, driving torque, axial pullout) per ASTM F543-13 and pyrogenicity testing per AAMI ANSI ST72.
Technological Characteristics
Material: Ti-6Al-4V (ASTM F136:2013). Design: Cannulated, headless, self-drilling, self-tapping bone screws. Diameters: 2.5 mm and 3.0 mm. Lengths: 10-40 mm. Principle: Compression via differential thread pitch. Sterile packaging.
Indications for Use
Indicated for patients requiring treatment of fractures, osteotomies, arthrodeses, and non-unions of small bones in the hand, wrist, foot, and ankle.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Synthes 3.0 mm Headless Compression Screws (K050636)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a family of three figures standing together.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 19, 2016
KLS Martin LP Gary Moore Quality Management and Regulatory Affairs Manager 11201 Saint Johns Industrial Parkway S Jacksonville, Florida 32246
Re: K161259
Trade/Device Name: KLS Martin Cannulated Headless Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: November 18, 2016 Received: November 21, 2016
Dear Gary Moore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| K161259 Page 1 of 1 | |
|-------------------------------------------------------------------------|--|
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | |
| Indications for Use | |
| Form Approved: OMB No. 0910-0120 | |
| Expiration Date: January 31, 2017 | |
| See PRA Statement below. | |
| 510(k) Number (if known) | |
| Device Name | |
KLS Martin Cannulated Headless Screws
Indications for Use (Describe)
KLS Martin Cannulated Headless Screws are intended for the treatment of fractures, osteotomies, and nonunions of small bones in the hand, wrist, foot, and ankle.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K161259 Page 1 of 3 P.O. Box 16369 · Jacksonville, FL 32245-6369 904-641-7746 or 800-625-1557 · Fax 904-641-7378 www.klsmartinnorthamerica.com
## Section 5 510(k) Summary 21 CFR 807.92(c)
| Submitter: | KLS Martin LP<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 |
|-----------------|--------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gary Moore<br>Quality Management & Regulatory Affairs Manager<br>Phone: 800-625-1557<br>Fax: 904-641-7378 |
| Date Prepared: | December 12, 2016 |
| Trade Name: | KLS Martin Cannulated Headless Screws |
| Common Name: | Screw, Fixation, Bone |
| Classification: | Smooth or threaded metallic bone fixation fastener<br>Class II, 21 CFR 888.3040, Product Code HWC |
| Predicates: | Tiger Headless Cannulated Screws (K112737) (Primary)<br>Synthes 3.0 mm Headless Compression Screws (K050636) |
#### Device Description:
The KLS Martin Cannulated Headless Screws (CHS) system is comprised of headless cannulated screws intended for bone fixation in the treatment of fractures, non-unions, osteotomies, or to aid in small joint fusions of the hand, wrist, foot, and ankle. Bone fixation is achieved by proximal and distal threads designed with different pitches that, when inserted into the bone, cause compression of the bone fragments for bone reduction, stability, and healing. Cannulation of the screws is designed to allow the user to insert the screw over the guide wire for proper placement prior to compression. The CHS system offers screws in various diameters, overall lengths, and thread lengths to accommodate different sizes and types of bone reduction, such as scaphoid fractures and non-unions. The screws are self-drilling and self-tapping to eliminate the need for drilling a pilot hole prior to implantation. All screws are manufactured from Ti-6Al-4V (ASTM F136:2013). This system also includes the necessary and appropriate instrumentation to facilitate implantation of the screws by qualified and trained physicians.
{4}------------------------------------------------
#### Indications for Use:
KLS Martin Cannulated Headless Screws are intended for the treatment of fractures, osteotomies, arthrodeses, and non-unions of small bones in the hand, wrist, foot, and ankle.
#### Technological Characteristics/Substantial Equivalence Discussion:
#### Similarities to Predicate Devices
The subject and predicate devices include cannulated headless screws with the same intended use. All screw systems are available in titanium alloy, are implanted using the same methods of fixation, and are offered in sterile packaging. The subject and predicate devices provide headless, cannulated, self-tapping screws that have threaded heads which are inserted over the guide wire and countersunk into the bone for compression of bone segments and are available in short and long thread lengths. The subject device is identical in screw length to the predicate devices, with the exception of the Tiger Headless Cannulated Screws (K112737).
#### Differences to Predicate Devices
The subject device differs from the Synthes 3.0 Headless Compression predicate device in that it is only provided in titanium alloy. The Synthes 3.0 Headless Compression screws are provided in both titanium alloy and stainless steel. The fact that the subject device will not be provided in stainless steel does not have any impact on the safety and effectiveness of the device. Both devices are manufactured from titanium alloy, which is a well-known material used in bone fixation devices.
In addition, the subject device manufactures screws with thread diameters of 2.5 mm and 3.0 mm, while the Synthes predicate device manufactures screws with a thread diameter of only 3.0 mm, and the Tiger predicate device manufactures screws with thread diameters of 2.0 mm and 2.4 mm, in addition to the 3.0 mm. The difference in availability of one thread diameter versus another in comparison to the predicate devices does not impact the safety and effectiveness of the subject device. The subject device provides thread diameters that are identical to the Synthes screw and falls within the primary predicate device (Tiger Headless Cannulated Screws) thread diameter range, which is between 2.0 and 3.0 mm.
Supplementary, another difference between the subject device and the predicate devices is the subject device has a screw length ranging from 10-40 mm, whereas the primary predicate has a screw length ranging only from 10-34 mm. This difference in screw length range has no impact on the safety and effectiveness of the subject device. Having a larger screw length range allows for more availability in different types of bone fixation where a longer screw might be needed for sufficient reduction.
{5}------------------------------------------------
### Non-clinical Performance Data:
### Mechanical Properties Testing
The mechanical properties of the KLS Martin Cannulated Headless Screws were tested in accordance with ASTM F543-13:2013 Standard Specification and Test Methods for Metallic Medical Bone Screws. The torsional properties, driving torque, and axial pullout strength were tested on the various screw lengths and thread diameters, in which tests were successfully completed without any signs of device failure. Comparative testing of the KLS Martin 3.0 mm cannulated screw to the Synthes 3.0 mm cannulated screw demonstrates substantial equivalence in performance testing.
### Pyrogenicity Testing
LAL endotoxin testing was conducted according to AAMI ANSI ST72 on the subject device screws to address the presence of bacterial endotoxins and ensure they meet pyrogen limit specifications. The results of the testing demonstrate that the KLS Martin Cannulated Headless Screws contain endotoxin levels below the USP allowed limit for medical devices and meet pyrogen limit specifications.
#### Clinical Performance Data:
Clinical testing was not necessary for the determination of substantial equivalence.
#### Conclusion:
The KLS Martin Cannulated Headless Screws have the same intended use, same principles of operation, similar technological features, and similar performance testing compared to the predicate devices. The information presented supports substantial equivalence of the KLS Martin Cannulated Headless Screws to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.