← Product Code [EMA](/productcode/EMA) · K161256

# TheraCem (K161256)

_Bisco, Inc. · EMA · Jul 29, 2016 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K161256

## Device Facts

- **Applicant:** Bisco, Inc.
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Jul 29, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Use TheraCem to cement the following: - 1. Metal crowns, bridges, inlays and onlays (includes porcelain-fused-to-metal and composite-to-metal) - 2. Porcelain, Ceramic Crowns, inlays and onlays (includes alumina and zirconia) - 3. Resin crowns, bridges, inlays and onlays (resin-based composite-ceramic hybrid) - 4. Metal (prefabricated or cast) and non-metal/fiber endodontic posts - 5. Implant supported restorations - 6. Orthodontic Appliances (brackets, bands)

## Device Story

TheraCem is a self-etching, self-adhesive, dual-cured resin luting cement. It is a paste/paste system delivered via dual-syringe. It eliminates the need for separate etching, priming, or bonding steps during the cementation of dental restorations (crowns, bridges, inlays, onlays, posts, implant-supported restorations, and orthodontic appliances). The material is fluoride- and calcium-releasing and radiopaque. Used by dental professionals in a clinical setting, the cement provides a bond to dental materials, reducing chair time. The radiopacity allows for identification on radiographs. The device benefits patients by providing a secure, efficient restoration cementation process with potential therapeutic ion release.

## Clinical Evidence

Bench testing only. Physical/mechanical properties tested include bond strength (Modified ISO 29022/Gel-Cap), diametral tensile strength, film thickness (ISO 4049:2009), flexural strength (ISO 4049:2009), radiopacity (ISO 4049:2009), working/setting time, compressive strength, and ion release (calcium/fluoride). All results were equivalent to predicates or met ISO requirements. Biocompatibility evaluated via ISO 7405:2008 and ISO 10993-1; oral toxicity study (10 mice, 14 days) showed no toxicity.

## Technological Characteristics

Self-etching, self-adhesive, dual-cured resin luting cement. Composition: Methacrylate-based resin, amorphous silica, and Portland cement. Delivery: Dual-syringe. Properties: Radiopaque, fluoride and calcium releasing. Complies with ISO 4049:2009 for film thickness, flexural strength, and radiopacity.

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- BisCem by Bisco, Inc. ([K082449](/device/K082449.md))

## Reference Devices

- NuSmile Biocem by NuSmile, LTD / Pulpdent ([K123265](/device/K123265.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 29, 2016

Bisco, Inc. Ryan Hobson RA Registration Coordinator 1100 West Irving Park Road Schaumburg, Illinois 60193

Re: K161256

Trade/Device Name: Theracem Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA, Dated: May 2, 2016 Received: May 4, 2016

Dear Ryan Hobson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/7 description: The image shows the name "Susan Kunna DDS, MA" in a handwritten-like font. The text is in black ink and appears to be a signature or a title. The letters are connected, giving it a cursive appearance.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K161256

Device Name TheraCem

Indications for Use (Describe)

Use TheraCem to cement the following:

- 1. Metal crowns, bridges, inlays and onlays (includes porcelain-fused-to-metal and composite-to-metal)
- 2. Porcelain, Ceramic Crowns, inlays and onlays (includes alumina and zirconia)
- 3. Resin crowns, bridges, inlays and onlays (resin-based composite-ceramic hybrid)
- 4. Metal (prefabricated or cast) and non-metal/fiber endodontic posts
- 5. Implant supported restorations
- 6. Orthodontic Appliances (brackets, bands)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BISCO. The logo consists of five connected hexagons, with the first and last hexagons being red, the second and fourth being a gradient of red and blue, and the middle hexagon being blue. Below the hexagons is the word "BISCO" in a bold, sans-serif font, with the top half of the letters in gray and the bottom half in black.

## 510 (k) SUMMARY

Applicant:

Bisco, Inc. 1100 W. Irving Park Road Schaumburg IL, 60193

Contact Person:

Ryan Hobson Tel: 847-534-6143 Fax: 847-534-6143

02 May 2016

Date Prepared:

Trade Name: Common Name: Product Code: Classification/Name:

TheraCem Self-Adhesive Resin Cement EMA Dental Cement Class II per 21 CFR 872.3275

## Predicate Devices:

#### TheraCem is substantially equivalent to:

| Primary Predicate:   | BisCem by Bisco, Inc. K082449                     |
|----------------------|---------------------------------------------------|
| Reference Predicate: | NuSmile Biocem by NuSmile, LTD / Pulpdent K123265 |

#### Indications for Use:

Use TheraCem to cement the following:

- 1. Metal crowns, bridges, inlays and onlays (includes porcelain-fused-to-metal and composite-to-metal)
- Porcelain, Ceramic Crowns, inlays and onlays (includes alumina and zirconia) 2.
- 3. Resin crowns, bridges, inlays and onlays (resin-based composite-ceramic hybrid)
- 4. Metal (prefabricated or cast) and non-metal/fiber endodontic posts
- 5. Implant supported restorations
- 6. Orthodontic Appliances (brackets, bands)

BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the BISCO company logo. The logo consists of a series of connected, three-dimensional, hexagonal shapes in red and blue. Below the shapes, the word "BISCO" is written in a bold, sans-serif font, with the letters in gray and black.

## 510 (k) SUMMARY (continued)

The indications for use of TheraCem are the same as those for BisCem and NuSmile BioCem and are summarized in the table below:

|  | BisCem<br>(K082449)                                                                                                                                                                                                                                                                                                                       | TheraCem                                                                                                                                                                                                                                                                                                                                                                                                                                                       | NuSmile BioCem<br>(K123265)                                                                                                                                        |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|  | Luting metal crowns, bridges, inlay, and onlays including porcelain-fused-to-metal and composite-to-metal varieties Luting resin crowns, bridges, inlays, onlays, and veneers; Luting metal or non-metal/fiber posts; Luting orthodontic appliances; Luting porcelain inlays, onlays, crowns, and veneers (includes alumina and zirconia) | Use TheraCem to cement the following:<br>Metal crowns, bridges, inlays and onlays (includes porcelain-fused-to-metal and composite-to-metal) Porcelain, Ceramic Crowns, inlays and onlays (includes alumina and zirconia) Resin crowns, bridges, inlays and onlays (resin-based composite/composite-ceramic hybrid) Metal (prefabricated or cast) and non-metal/fiber endodontic posts Implant supported restorations Orthodontic Appliances (brackets, bands) | Pulpdent RMGI Low Viscosity is a resin-modified glass ionomer preparation used by dental professionals as a liner, base or luting material in dental restorations. |

### Description of Applicant Device:

TheraCem is a self-etching, self-adhesive, dual-cured resin luting cement that is exclusively formulated for luting crowns, bridges, inlays, onlays and posts (prefabricated metal and nonmetal/fiber posts, as well as cast posts). TheraCem is a paste/paste, fluoride- and calciumreleasing, luting cement which requires no etching, no priming or bonding of the prepared surfaces. It is easy-to use, requires only a short chair time, and produces a good bond to most dental materials. The cement is available in a Natural shade. It is radiopaque, allowing for easy identification on radiographs.

> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for BISCO. The logo consists of a series of connected, three-dimensional hexagonal shapes in red, light blue, and dark blue. Below the shapes is the word "BISCO" in a bold, sans-serif font, with the top half of the letters in gray and the bottom half in black.

# 510 (k) SUMMARY (continued)

## Technological Characteristics:

All components of TheraCem are based upon industry standard chemistry. The chemical composition of TheraCem is similar to BisCem and BioCem and is summarized in the table below:

| Chemical Composition  | BisCem<br>(K082449)           | NuSmile BioCem<br>(K123265)         | TheraCem                              |
|-----------------------|-------------------------------|-------------------------------------|---------------------------------------|
| Filler                | Amorphous Silica              | Amorphous Silica                    | Amorphous Silica &<br>Portland Cement |
| Resin composition     | Methacrylate based            | Methacrylate based                  | Methacrylate based                    |
| Polymerization Method | Dual cured                    | Dual cured                          | Dual cured                            |
| Method of Application | Bonding agent not<br>required | Bonding agent not<br>required       | Bonding agent not<br>required         |
| Ions Released         | Fluoride                      | Calcium, phosphate, and<br>fluoride | Calcium and fluoride                  |

| Physical Mechanical<br>Property | BisCem<br>(K082449) | NuSmile BioCem<br>(K123265)       | TheraCem                          |
|---------------------------------|---------------------|-----------------------------------|-----------------------------------|
| Radiographic Appearance         | Radiopaque          | Radiopaque                        | Radiopaque                        |
| Ions Released                   | Fluoride releasing  | Fluoride and calcium<br>releasing | Fluoride and calcium<br>releasing |
| Delivery system                 | Dual-syringe        | Dual-syringe                      | Dual-syringe                      |

The difference in filler is TheraCem's use of Portland cement, an industry standard chemical, to facilitate calcium release and is substantially equivalent in performance to amorphous silica. A reference predicate, BioCem (K123265), has been included to demonstrate substantial equivalence for calcium releasing.

> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for BISCO. The logo consists of five connected hexagons, alternating between red and blue. Below the hexagons is the word "BISCO" in a bold, sans-serif font, with the letters in shades of gray and black.

## 510 (k) SUMMARY (continued)

#### Performance Data:

The following physical/mechanical properties of TheraCem were tested:

| Physical / Mechanical Property                              | TheraCem                                                                   |
|-------------------------------------------------------------|----------------------------------------------------------------------------|
| Bond Strength<br>(Modified ISO 29022 and<br>Gel-Cap method) | TheraCem is equivalent to the predicates.                                  |
| Diametral Tensile Strength                                  | TheraCem is equivalent to the predicates.                                  |
| Film Thickness<br>(ISO 4049:2009)                           | TheraCem meets the requirements of ISO 4049:2009 for Film<br>Thickness.    |
| Flexural Strength<br>(ISO 4049:2009)                        | TheraCem meets the requirements of ISO 4049:2009 for<br>Flexural Strength. |
| Radiopacity<br>(ISO 4049:2009)                              | TheraCem meets the requirements of ISO 4049:2009 for<br>Radiopacity.       |
| Working Time / Setting Time                                 | TheraCem is equivalent to the predicates.                                  |
| Compressive Strength                                        | TheraCem is equivalent to the predicates.                                  |
| Calcium Release                                             | TheraCem is equivalent to the predicates.                                  |
| Fluoride Release                                            | TheraCem is equivalent to the predicates.                                  |

#### Biocompatibility:

An evaluation of biocompatibility was conducted using ISO 7405:2008 and ISO 10993-1 to determine the safety of TheraCem. It is concluded from the safety evaluation and the results of the Oral Toxicity Study (10 mice, 14 days) that TheraCem was not toxic in this test.

#### Conclusion:

It is concluded from review of the predicate device indications, chemical composition, biocompatibility, and physical properties that TheraCem is substantially equivalent to the predicate devices.

> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.

---

**Source:** [https://fda-staging.innolitics.com/device/K161256](https://fda-staging.innolitics.com/device/K161256)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).