The EP-TRACER System is an electrophysiology measurement system used to acquire, filter, digitize, amplify, display, and record signals obtained during electrophysiological studies and related procedures. The system allows the user to monitor, display and record the signals. The system incorporates a stimulator intended to be used for diagnostic cardiac stimulation during electrophysiological testing of the heart
Device Story
EP-TRACER System is a computerized electrophysiology measurement system; acquires, filters, digitizes, amplifies, displays, and records cardiac signals; incorporates integrated stimulator for diagnostic cardiac stimulation. Used in clinical electrophysiology labs; operated by trained medical professionals. System hardware includes amplifier/stimulator; software provides user interface for signal monitoring and protocol management. Healthcare providers use real-time signal displays and recorded data to guide diagnostic cardiac testing and electrophysiological procedures. Benefits include precise signal acquisition and controlled cardiac stimulation to facilitate diagnosis of cardiac arrhythmias.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through software validation protocols and electrical safety/EMC testing (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366, IEC 60601-2-34). All tests passed.
Technological Characteristics
Class II programmable diagnostic computer. Hardware: Amplifier/stimulator with 38, 70, or 102 channels. Sensing: 1 kHz sampling rate, >100 dB CMRR, 20 MΩ input impedance. Stimulator: 2 isolated channels, 0-25.5 mA output, 0.1-9.9 ms pulse duration. Connectivity: Standalone system. Software: Multi-threaded, Visual C#/.NET environment. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366, IEC 60601-2-34.
Indications for Use
Indicated for patients undergoing electrophysiological studies and diagnostic cardiac stimulation procedures.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, with flowing lines above them, possibly representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2016
CardioTek, B.V. % Melissa Walker President & CTO Graematter, Inc. 1324 Clarkson Clayton Center #332 St Louis. Missouri 63011
Re: K161245
Trade/Device Name: EP-Tracer System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: June 13, 2016 Received: June 22, 2016
Dear Melissa Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Muda Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161245
Device Name EP TRACER System
#### Indications for Use (Describe)
The EP-TRACER System is an electrophysiology measurement system used to acquire, filter, digitize, amplify, display, and record signals obtained during electrophysiological studies and related procedures.
The system allows the user to monitor, display and record the signals. The system incorporates a stimulator intended to be used for diagnostic cardiac stimulation during electrophysiological testing of the heart
| Type of Use (Select one or both, as applicable) | | | |
|--------------------------------------------------|---------------------------------------------|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5: 510(k) Summary per 21CFR §807.92
# EP TRACER System Summary
| Submitter's<br>information | Contact: Oliver Roth<br>Schwarzer Cardiotek GmbH<br>Im Zukunftspark 3<br>74076 Heilbronn / Germany | Contact: Melissa Walker<br>Graematter, Inc.<br>1324 Clarkson Clayton Ctr #332<br>Ballwin, MO 63011<br>Phone: 636-405-7498<br>Date: 02/14/2016 |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Device/<br>classification<br>name | Device Name:<br>The EP-TRACER System is a Class 2 device (product code DQK).<br>Classification/Common name:<br>• Programmable diagnostic computer, 21 CFR §870.1425,<br>The marketed device(s) to which substantial equivalence is claimed:<br>• EP-TRACER System, K134044 | |
| Device<br>description | The EP-TRACER System is a computerized electrophysiology measurement<br>system designed for both regular and experimental EP studies.<br><br>The EP-TRACER System is comprised of these major components,<br><br>1. EP-TRACER hardware – Amplifier/stimulator<br>2. EP-TRACER Software – Software pre-installed | |
| Indications for<br>use | "The EP-TRACER System is an electrophysiology measurement system used<br>to acquire, filter, digitize, amplify, display, and record signals obtained<br>during electrophysiological studies and related procedures.<br><br>"The system allows the user to monitor, display and record the signals. The<br>system incorporates a stimulator intended to be used for diagnostic cardiac<br>stimulation during electrophysiological testing of the heart".<br><br>Continued on next page | |
{4}------------------------------------------------
## EP TRACER System Summary, Continued
Technological The table below lists the technological characteristics for both the new and characteristics predicate devices.
| Device Characteristic | Predicate Device EP-TRACER V1.05<br>K134044 | New Device EP-TRACER V2<br>K161245 |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Amplifier Dimensions (with<br>integrated stimulator) WxDxH | 38 channels: 28x27x7 in cm<br>70/102 channels: 28x27x12 in cm | Same |
| Environmental Specifications | | |
| Temperature Operating | +10°C to +30°C | Same |
| Temperature Transport/Storage | -29°C to +66°C | Same |
| Humidity Operating | 20 - 80 % rH (non-condensing) | Same |
| Humidity Transport/Storage | < 95 % rH (non-condensing) | Same |
| Power Specifications | | |
| Power Requirements | 100 - 240 V AC,<br>50 - 60 Hz | Same |
| Power Input | 38 channels:<br>+5 V, 0.3 A & +12 V, 0.9 A<br>70 channels:<br>+5 V, 0.3 A & +12 V, 1.5 A<br>102 channels:<br>+5 V, 0.3 A & +12 V, 2 A | Same |
| Design | | |
| Sampling and Hold | Each channel sampled prior the<br>acquisition | Same |
| Sampling Rate | 1 kHz | Same |
| CMRR | > 100 dB | Same |
| Input Impedance | Typical 20 MΩ | Same |
| Leakage Current | | |
| Patient Source | < 10 μA | Same |
| Patient Sink | < 10 μA | Same |
| Patient Sink (measured at patient<br>leads under single fault conditions) | < 50 μA | Same |
| Chassis Leakage | < 100 μA | Same |
| ECG Input | | |
| Outputs | 12 lead ECG produced | Same |
| High Pass Filter | 0.05 Hz, 0.2 Hz | Same |
| Low Pass Filter | 150 Hz | Same |
| RF Filtering | All inputs | Same |
| Gain | Between 0 and 255 mm/mV -<br>continuous | Same |
| Saturation Recovery | < 1 sec (manual reset) | Same |
| Notch Filter | Power line (50/60 Hz) | Same |
| Dynamic Range | ±5 mV | Same |
| Baseline Correction | +300 mV | Same |
| Device Characteristic | Predicate Device EP-TRACER V1.05<br>K134044 | New Device EP-TRACER V2<br>K161245 |
| Input/ Output | | |
| Inputs 32/38 channels | 20 intracardiac channels, 6 auxiliary<br>channels, 12 ECG channels | Same |
| Inputs 64/70 channels | 52 intracardiac channels, 6 auxiliary<br>channels, 12 ECG channels | Same |
| Inputs 102 channels | 84 intracardiac channels, 6 auxiliary<br>(pressure) channels, 12 ECG channels | Same |
| Outputs | No outputs | Same |
| Switching | Each channel can be either bipolar or<br>unipolar with manual switching | Same |
| High Pass Filter | 0.05 Hz, 0.2 Hz, 40 Hz, 80 Hz | Same |
| Low Pass Filter | 350 Hz | Same |
| RF Filtering | All inputs | Same |
| Gain | Between 0 and 255 mm/mV –<br>continuous | Same |
| Saturation Recovery | < 1 s (manual reset) | Same |
| Notch Filter | Power line (50/60 Hz) | Same |
| Dynamic Range | ±5 mV | Same |
| Baseline Correction | ±300 mV | Same |
| Stimulator (integrated into device) | | |
| Isolated Stimulus Channels | 2 | Same |
| Pulse Amplitude | | |
| Range | 0 - 25.5 mA into 1000 Ω load | Same |
| Increment | 0.1 ms | Same |
| Accuracy | ±0.15 ms | Same |
| Pulse Duration | | |
| Range | 0.1 - 9.9 ms | Same |
| Increment | 0.1 ms | Same |
| Accuracy | ±0.15 ms | Same |
| Inter-Stimulus Interval (ISI) | | |
| Range | 10 - 9999 ms | Same |
| Range (Burst) | 10 - 9999 ms | Same |
| Increment | 10 ms | Same |
| Sequential Delay (AV) | | |
| Range | 11 - 250 ms | Same |
| Increment | 1 ms | Same |
| Programmed Protocols | Preprogrammed protocols:<br>BASIC 1 (induction)<br>BASIC 2 (termination)<br>BASIC 3 (backup)<br>ACUTE<br>Multi-Sx | Same |
| Programmed Protocols | Pace<br>Automatic mode<br>Wenkebach mode | Same |
| Device Characteristic | Predicate Device EP-TRACER V1.05<br>K134044 | New Device EP-TRACER V2<br>K161245 |
| Programmable Protocol Key | 10 | Same |
| Number of Extra-Stimuli | 5 (S2 - S6) | Same |
| Sensing (ECG Synchronization) | | |
| Automatic or Manual Trigger<br>Setting - Sensitivity | Internal from any surface or intra-cardiac channel | Same |
| Automatic or Manual Trigger<br>Setting - Trigger lockup<br>(refractory time) | 5 - 5000 ms | Same |
| Automatic or Manual Trigger<br>Setting - ECG Delay | 5 - 5000 ms | Same |
| Additional Outputs | No | Same |
| Power Source | Integrated with amplifier<br>38 channels:<br>+5 V, 0.3 A & +12 V, 0.9 A<br>70 channels:<br>+5 V, 0.3 A & +12 V, 1.5 A<br>102 channels:<br>+5 V, 0.3 A & +12 V, 2 A | Same |
| Pacing Channels | | |
| Isolated Channels | (i) atrial and<br>(ii) ventricular and<br>(iii) emergency fixed pace output to atrium and ventricle | Same |
| Circuit Isolation | Compliant with IEC 60601-1, Class CF,<br>5 kV, common & differential mode | Same |
| Computer Controlled Stimulus Pulses | | |
| Current | 0 - 25.5 mA into 1000 Ω load | Same |
| Current Steps | 0.1 mA | Same |
| Accuracy | ±0.1 mA | Same |
| Pulse duration | Pulse width 0.1 - 9.9 ms, steps of 0.1 ms | Same |
| Accuracy | ±0.1 mA | Same |
| Load Impedance | 1000 Ω | Same |
| Max. Output Voltage | 25 V | Same |
| Inter-Stimulus Intervals | | |
| S1 Range | 10 - 9999 ms | Same |
| Stability | Quartz computer clock, ±30 parts per<br>million at +25°C | Same |
| Extra-Stimuli | 5 (S2 - S6) | Same |
| Coupling Interval | 30 - 9990 ms | Same |
| Accuracy | ±10 ms | Same |
| Protocol Automation | | |
| Auto decrement/<br>increment | Yes | Same |
| Backup Manually Controlled<br>Stimulation | Use external backup stimulator | Same |
| Device Characteristic | Predicate Device EP-TRACER V1.05<br>K134044 | New Device EP-TRACER V2<br>K161245 |
| Compliance with Standards | | |
| Standards | EN 60601-1<br>EN 60601-1-2 | Same |
| Device Directive | European Union Medical Device<br>Directive (CE Marked) | Same |
| Environmental/ Electrical Specifications | | |
| Operating Temperature | +10°C to +30°C | Same |
| Storage Temperature | -29°C to +66°C | Same |
| Operating Humidity | 20 - 80 % rH (non-condensing) | Same |
| Storage Humidity | < 95 % rH (non-condensing) | Same |
| Max. Current Draw | 15 A/115 V, 7 A/230 V | Same |
| Chassis Leakage Current | < 100 μΑ | Same |
| Advanced Features | | |
| Display Ablation Parameters | Connection to RF ablation generator(s) | Same |
| Certification | | |
| MDD Device Class | Class IIb | Same |
| IEC 60601-1 | Certified | Same |
| EMC Compliance | Certified | Same |
| CE Marking | Certified CE 0197 | Same |
| US Regulations | K134044 | Current submission |
| System Dimensions | | |
| W x D x H | Customer Option<br>Mobile Cart: 100x80x170 cm<br>Mobile Desk: 125x80x150 cm<br>Control Desk: customer defined<br>Laptop: customer defined | Same |
| Software | | |
| Programming language | Borland Delphi 5 | Visual C# |
| Component library | VCL Framework (Visual Component<br>Library) | .NET |
| Software architecture | Single Threaded | Multi Threaded |
| Drawing method | 2D | 3D |
| License concept | Text file | Dongle |
| Context sensitive help texts | No | Mouse over context help |
| Operating concept | Menus and short-cuts | Menus, short-cuts and display<br>elements (i.e. buttons) |
| Language support | English (full), partial support for other<br>languages | Full multi language support |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Continued on next page
{8}------------------------------------------------
### EP TRACER System Summary, Continued
Performance The updated version of the EP-TRACER system and software (V2.0) testing maintains the capabilities in the previous version, adding user interface enhancements.
> Testing has been conducted through software and/or user verification/validation protocols to ensure the EP TRACER System meets its intended use and user needs.
> Electrical safety and electromagnetic compatibility testing shown in the table below was performed. All tests passed.
| Test | Results |
|-----------------------------------------------------------------------------------------------------------------------|---------|
| Radiated emission: Conducted emission: Radiated immunity:<br>Conducted immunity | Pass |
| Conducted emission (harmonics); Radiated immunity (PMF);<br>Conducted immunity | Pass |
| IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 +<br>AM1:2012 | Pass |
| Amendment - IEC 60601-1:2005 (Third Edition) + CORR.<br>1:2006 + CORR. 2:2007 + A1:2012 | Pass |
| IEC 60601-1-6:2010 (Third Edition) + A1:2013 | Pass |
| IEC 62366:2007 (First Edition) + A1: 2014 for use in conjunction<br>with IEC 60601-1-6: 2010 | Pass |
| IEC 60601-1-27:2011 (Third Edition) for use in conjunction with<br>IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) | Pass |
| IEC 60601-2-34 (Third Edition): 2011 | Pass |
#### Based on the technical characteristics and the results of the performance Conclusion testing, the EP TRACER V2.0 is substantially equivalent to V1.05.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.