K161241 · Alber GmbH · ITI · Oct 5, 2016 · Physical Medicine
Device Facts
Record ID
K161241
Device Name
e-fix E35/E36
Applicant
Alber GmbH
Product Code
ITI · Physical Medicine
Decision Date
Oct 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The e-fix is an add-on drive for wheelchairs. The e-fix E35/E36 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.
Device Story
Add-on drive system converting manual wheelchairs to powered wheelchairs; consists of two power wheels, battery pack, control unit, and mounting brackets. User operates via joystick on control unit; system provides motorized propulsion, steering, and braking. Microprocessor-controlled speed, acceleration, and deceleration; includes regenerative braking and down-slope speed regulation. Battery pack is removable and rechargeable. Intended for use by disabled individuals to enhance mobility. Healthcare providers use device to facilitate patient independence; output is physical movement of wheelchair. Benefits include increased patient mobility and reduced physical exertion for propulsion.
Clinical Evidence
Bench testing only. Performance verified against ISO 7176 series standards for stability, braking, energy consumption, speed/acceleration, climatic testing, obstacle climbing, and electromagnetic compatibility. Biocompatibility testing performed per ISO 10993-5.
Technological Characteristics
Add-on drive system; 36V Li-ion battery; 2x 110W (E35) or 2x 150W (E36) motors. Microprocessor-controlled power and control system. Connectivity: joystick interface. Sterilization: N/A. Standards: ISO 7176-1, -2, -3, -4, -6, -8, -9, -10, -14, -15, -21; ISO 10993-5.
Indications for Use
Indicated for disabled persons requiring powered propulsion for manually propelled wheelchairs to increase mobility and flexibility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 5, 2016
Alber Gmbh % Michael Vent Official Correspondent for Alber Gmbh Beo Medconsulting Berlin Gmbh Helmholtzstr. 2-9 Berlin, 10587 GERMANY
Re: K161241
Trade/Device Name: e-fix E35/E36 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: August 31, 2016 Received: September 7, 2016
Dear Mr. Vent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161241
Device Name e-fix E35/E36
Indications for Use (Describe)
The e-fix is an add-on drive for wheelchairs. The e-fix E35/E36 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
Device 510k number: K161241
| Applicant: | Alber GmbH<br>Vor dem Weißen Stein 21<br>72461 Albstadt, Germany<br>Phone: +49 7432 2006-0<br>Fax: +49 7432 2006-299<br>Email: info@alber.de |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Michael Vent<br>BEO MedConsulting Berlin GmbH<br>Helmholtzstr. 2-9<br>10587 Berlin, Germany<br>Phone: +49 30 318 045 3<br>Email: m.vent@beoberlin.de |
| Device: | Proprietary: e-fix E35/E36<br>Common Name: add-on drive for wheelchairs<br>Classification Name: Powered wheelchair<br>Device Class: II, 21 CFR 890.3860<br>Classification Panel: Physical Medicine<br>Product Code: ITI |
#### Predicate Device Information:
K142770
SOLO/SOLO+
(AAT Alber Antriebstechnik GmbH)
#### Indications for Use:
The e-fix is an add-on drive for wheelchairs. The e-fix E35/E36 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.
#### Device Description:
The subject device e-fix E35/E36 is an add-on drive, which converts a manual wheelchair into an electrical wheelchair.
The drive system e-fix E35/E36 includes two power wheels, a battery mount, a battery pack, a control unit, a power supply and optional devices. The wheelchair is not part of the drive system and is provided by the customer. For initial use mounting brackets are required. If the mounting brackets are fitted to the wheelchair, the power wheels, the control unit and the battery pack can be adapted. The main principle of the power wheel is that the motor unit is fixed against turning and the wheel hub is rotating. The fixing of the motor unit to the mounting brackets is realized with the wheel receiver. The wheel receiver supports the torque which the motor applies and is the interface to the wheel chair.
There are different variants of the e-fix E35 is the conventional version of the e-fix. The e-fix E36 provides a more robust drive with a higher torque. The battery pack has more capacity than the normal one. The e-fix E36 is mainly intended for heavy-weighted persons. The e-fix E35 and E36 are available with 22" or 24" wheels. The control unit is the steering device of the system.
{4}------------------------------------------------
lt allows a safe and comfortable ride with the wheelchair.The battery pack is easily removable and rechargeable.
| | K161241 (Subject device) | | K142770 (Predicate device) | | Rationale |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technical Data | | | | | |
| | E35 | E36 | SOLO | SOLO+ | |
| Indications for Use | The e-fix is an add-on drive for wheelchairs. The e-fix E35/E36 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility. | The e-fix is an add-on drive for wheelchairs. The e-fix E35/E36 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility. | SOLO+, SOLO and SERVO turn a manually propelled wheelchair into a powered wheelchair. They are intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility | SOLO+, SOLO and SERVO turn a manually propelled wheelchair into a powered wheelchair. They are intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility | N/A |
| Wheel size (in.) | 22" or 24" | 22" or 24" | 24" | 24" | Some users may choose smaller wheels to reach the floor with their feet for foot propulsion. The wheel size has no influence on the safety of the powered propulsion |
| Weight (Set) | 18.9 kg | 18.9 kg | 22.9 kg | 22.9 kg | The subject device is lighter due to the battery weight. However, it is has no negative influence on the safety of the device |
| Max. user weight | 120 kg | 160 kg | 160 kg | 200 kg | The subject device has a lower user weight capacity and has no negative influence on the safety of the device |
| Total weight approved | 170 kg | 210 kg | 210 kg | 250 kg | The subject device has a lower weight because of lower product and user weight. It has no negative influence on the safety of the device |
| Braking technologies | Manual parking brake<br>Electromotive brake<br>Electromechanical brake | | Manual parking brake<br>Electromotive brake<br>Electromechanical brake | | Same |
| Speed, acceleration, deceleration | Preselected max-speed, microprocessor-controlled speed, acceleration and deceleration. The user controls these parameter with the joystick. | | Preselected max-speed, microprocessor-controlled speed, acceleration and deceleration. The user controls these parameter with the joystick. | | Same |
| | | | | | |
| Forward/backwar<br>d | Joystick & push buttons | | Joystick & push buttons | | Same |
| Locks to prevent<br>unintended/<br>unanticipated<br>movement | The control unit has an<br>immobilizer-key<br>If the electrical current is<br>turned off, the brake is<br>engaged | | The control unit has an<br>immobilizer-key<br>If the electrical current is<br>turned off, the brake is<br>engaged | | Same |
| Foldable<br>wheelchair still<br>foldable? | Foldable, after detaching<br>the battery | | Foldable, after detaching<br>the battery | | Same |
| Quick Release<br>Axle | Provided | | Provided | | Same |
| Wheel<br>synchronization | Provided | | Provided | | Same |
| Down slope<br>speed regulation | Provided | | Provided | | Same |
| Regenerative<br>brake | Provided | | Provided | | Same |
| Performance | | | | | |
| Range<br>(ISO 7176-14) | Up to 10 miles<br>(up to 13miles with the<br>optionally available high<br>capacity battery pack) | | Up to 22<br>miles | 9.3 miles | Subject device has less<br>range than SOLO but it<br>does not adversely affect<br>safety or effectiveness of<br>the subject device |
| Speed:<br>Forward/backward | 3.7 mph/2.6 mph | | 3.8 mph/1.9 mph | | Similar |
| Electrical data | | | | | |
| Battery | 36 V (Li-lon 10S3P) | | 28.8V:<br>16Ah(Pb) | 24V:<br>8.5Ah(Li-Ion) | Li ion battery recharges<br>faster, is less heavy and<br>suitable for more<br>recharges. The<br>differences have no<br>influence on the safety of<br>the devices |
| Motor | 36 V; 2x<br>110W<br>(microprocesso<br>r controlled<br>adaption to the<br>power source<br>(Li-lon)) | 36V; 2x 150W<br>(microprocesso<br>r controlled<br>adaption to the<br>power source<br>(Li-lon)) | 24-28.8V DC,<br>150W<br>(microprocesso<br>r controlled<br>adaption to the<br>power source<br>(PB or Li-Ion)) | 24-28.8V DC.<br>70W<br>(microprocesso<br>r controlled<br>adaption to the<br>power source<br>(PB or Li-Ion)) | The differences in motor<br>performance do not affect<br>safety of the subject<br>device |
## Comparison to predicate device:
{5}------------------------------------------------
### The device is in conformity with Standards and Guidance Documents:
Guidance Document for the Preparation of PreMarket Notification 510k Applications for Mechanical and Powered Wheelchairs and Motorized Three-wheeled vehicles
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
ISO 7176-1:2014, wheelchairs - part 1: determination of static stability. (Physical Medicine)
{6}------------------------------------------------
ISO 7176-2:2001, wheelchairs - part 2: determination of dynamic stability of electric wheelchairs. (Physical Medicine)
ISO 7176-3:2012, wheelchairs - part 3: determination of effectiveness of brakes. (Physical Medicine)
ISO 7176-4:2008, wheelchairs - part 4: energy consumption of electric wheelchairs and scooters for determination of theoretical distance range. (Physical Medicine)
ISO 7176-6:2001, wheelchairs - part 6: determination of maximum speed, acceleration and deceleration of electric wheelchairs. (Physical Medicine)
ISO 7176-8:2014, Wheelchairs Part 8: Requirements and test methods for static, impact and fatique strengths
ISO 7176-9:2009, wheelchairs - part 9: climatic tests for electric wheelchairs. (Physical Medicine)
ISO 7176-10:2008, wheelchairs - part 10: determination of obstacle-climbing ability of electrically powered wheelchairs. (Physical Medicine)
ISO 7176-14:2014, wheelchairs - part 14: power and control systems for electrically powered wheelchairs and scooters - requirements and test methods. (Physical Medicine)
ISO 7176-15:1996, wheelchairs - part 15: requirements for information disclosure, documentation and labeling. (Physical Medicine)
ISO 7176-21:2009, wheelchairs - part 21: requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers. (Physical Medicine)
ISO 10993-5:2009, biological evaluation of medical devices -- part 5: tests for in vitro cytotoxicity. (Biocompatibility)
#### Quality Assurance and Manufacturing Controls:
Alber GmbH operates to certified quality managementsystem according to EN ISO 13485.
#### Conclusion:
Conclusions drawn from the performance, bench and non-clinical tests demonstrate that the subject device substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.