HyperView Hyperspectral Tissue Oxygenation Measurement System
K161237 · Hypermed Imaging, Inc. · MUD · Dec 16, 2016 · Cardiovascular
Device Facts
Record ID
K161237
Device Name
HyperView Hyperspectral Tissue Oxygenation Measurement System
Applicant
Hypermed Imaging, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Dec 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The HyperView™ Hyperspectral Tissue Oxygenation Measurement System is intended for use by physicians/healthcare professionals as a noninvasive tissue oxygenation measurement system that reports an approximate value of: - oxygen saturation (O2Sat), - oxyhemoglobin level (Oxy), and - deoxyhemoglobin (Deoxy) level in superficial tissue. The HyperView™ system displays two-dimensional, color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions. The HyperView™ system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.
Device Story
Handheld hyperspectral imaging system; captures images of superficial tissue using visible light (400-700nm) via flash illumination. Device analyzes light absorption signatures at 8 specific wavelengths to calculate oxyhemoglobin, deoxyhemoglobin, and oxygen saturation levels. Operates via resistive touchscreen and image capture button; requires no patient contact or contrast agents. Produces 2D color-coded maps and quantitative measurements. Used by clinicians to assess vascular viability and wound healing potential. Data stored in DICOM format. Benefits include noninvasive, rapid assessment of tissue perfusion in clinical settings.
Clinical Evidence
Prospective validation study (n=20) comparing HyperView to OxyVu™-1. Primary endpoint: mean O2Sat difference between devices. Results showed mean O2Sat difference of 3.22% ± 3.38%, within the pre-specified ±7% equivalence margin. Secondary endpoints (Oxy, Deoxy) showed mean differences of 5.5% and 0.8% respectively. Stratification by skin tone, ulcer presence, and peripheral artery disease (PAD) confirmed consistent performance across subgroups. No statistically significant differences in O2Sat measurements were noted between PAD and non-PAD limbs.
Technological Characteristics
Handheld device; visible light (400-700nm) illumination using filtered CREE LEDs. Imaging sensor, thermal sensor, and flash illumination. Connectivity via USB and docking station. Software-based spectral analysis of 8 wavelengths. Powered by internal battery. Electrical safety per ANSI/AAMI ES60601-1; EMC per IEC 60601-1-2; Laser safety per IEC 60825-1; Photobiological safety per IEC 62471.
Indications for Use
Indicated for physicians/healthcare professionals to determine superficial tissue oxygenation levels in patients with potential circulatory compromise.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles forming a wing-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2016
Hypermed Imaging, Inc. % Julie Powell Vice President, Quality Assurance Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, Texas 78701
Re: K161237
Trade/Device Name: Hyperspectral Tissue Oxygenation Measurement System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: November 11, 2016 Received: November 14, 2016
Dear Julie Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark Jellison
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K161237
#### Device Name
HyperView™ Hyperspectral Tissue Oxygenation Measurement System
#### Indications for Use (Describe)
The HyperView™ Hyperspectral Tissue Oxygenation Measurement System is intended for use by physicians/healthcare professionals as a noninvasive tissue oxygenation measurement system that reports an approximate value of.
- · oxygen saturation (O2Sat),
- · oxyhemoglobin level (Oxy), and
- deoxyhemoglobin (Deoxy) level
in superficial tissue. The HyperView™M system displays two-dimensional, color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions.
The HyperView™ system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
# Hyperview Hyperspectral Tissue Oxygenation Measurement System
# K161237
#### 1. Submission Sponsor
HyperMed Imaging, Inc.
8700 Trail Lake Drive West, Suite 200
Memphis, TN 38125
USA
Office number: (901) 453-3154
Contact: Carlyn Reynolds
Title: Vice President Regulatory Compliance & Quality Assurance
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
816 Congress Avenue, Suite 1400
Austin, TX 78701
Cell Phone: (269) 720-1329
Office Phone: (512) 327-9997
Contact: Julie Powell, Vice President, QA/RA
Email: project.management@emergogroup.com
#### 3. Date Prepared
April 29, 2016
{4}------------------------------------------------
#### 4. Device Identification
| Trade/Proprietary Name: | HyperView™ Hyperspectral Tissue Oxygenation Measurement System |
|-------------------------|----------------------------------------------------------------|
| Common/Usual Name: | Tissue Saturation Oximeter |
| Classification Name: | Oximeter |
| Regulation Number: | 21 CFR 870.2700 |
| Product Code: | MUD, Oximeter |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
#### 5. Legally Marketed Predicate Device(s)
OxyVu™-1, 510(k) K073656, HyperMed, Inc.
#### 6. Device Description
HyperMed Imaging Inc.'s HyperView handheld hyperspectral imaging system consists of the HyperView Camera and the HyperView Accessories Kit that includes a Docking Station, Calibration Verification System, Power Cord and USB Cable Set.
The HyperView system provides noninvasive measurement of approximate tissue oxygenation, sensitive to local, regional, and systemic low/no-flow ischemia. HyperView uses multi-wavelength imaging technology to analyze the signature of light as it is absorbed by the patient's tissue. The HyperView system analyzes these light absorption signatures to measure oxyhemoglobin and deoxyhemoglobin levels in the tissue, and creates a color-coded image that illustrates relative concentrations of oxygenated hemoglobin and deoxygenated hemoglobin.
Quantitative data is provided for oxyhemoglobin and deoxyhemoglobin levels as well as oxygen saturation. The technology is suitable for surface assessment, as the visible spectrum light used penetrates no more than a few millimeters deep.
The HyperView is a handheld product consisting of several subassemblies including precision optical components, electro-optics, imaging sensor, a thermal sensor, flash illumination, flex circuits, printed circuit boards, a battery, plastics as well as image processing and graphics user interface software. Users will interact with the device using a built-in, resistive, touchscreen display and an image capture button similar to a digital camera. The product does not come in contact with the patient and no injectable contrast is required. The camera uses visible light to momentarily illuminate the patient's skin for photographic imaging. Such information is useful in determining viability of vascular delivery in the superficial skin and correspondingly, such information can assist the physician in determining the ability of the patient's vascular system to support normal tissue health and wound healing.
{5}------------------------------------------------
#### 7. Indication for Use Statement
The HyperView™ Hyperspectral Tissue Oxygenation Measurement System is intended for use by physicians/healthcare professionals as a noninvasive tissue oxygenation measurement system that reports an approximate value of:
- oxygen saturation (O2Sat),
- oxyhemoglobin level (Oxy), and
- deoxyhemoglobin (Deoxy) level
in superficial tissue. The HyperView™ system displays two-dimensional, color-coded images of tissue oxygenation of the scanned surface and reports hyperspectral tissue oxygenation measurements for selected tissue regions.
The HyperView™ system is indicated for use to determine oxygenation levels in superficial tissues for patients with potential circulatory compromise.
#### 8. Substantial Equivalence Discussion
The following table compares the HyperView system to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| | SUBJECT DEVICE | PREDICATE | Significant<br>Differences |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Manufacturer | HyperMed Imaging, Inc. | HyperMed, Inc. | |
| Trade Name | HyperView™ System | OxyVu™-1 | |
| Product Code | MUD | MUD | Same |
| Regulation Number | 21 CFR 870.2700 | 21 CFR 870.2700 | Same |
| Regulation Name | Tissue Saturation Oximeter | Tissue Saturation Oximeter | Same |
| Indications for Use | The HyperView™<br>Hyperspectral Tissue<br>Oxygenation<br>Measurement System is<br>intended for use by<br>physicians/healthcare<br>professionals as a<br>noninvasive tissue<br>oxygenation measurement<br>system that reports an<br>approximate value of:<br>• oxygen saturation<br>(O2Sat), • oxyhemoglobin<br>level (Oxy), and | The OxyVu™-1<br>Hyperspectral Tissue<br>Oxygenation (HTO)<br>Measurement System<br>is intended for use by<br>healthcare professionals as<br>a noninvasive tissue<br>oxygenation measurement<br>system that reports an<br>approximate value of:<br>• oxygen saturation (HT-<br>Sat),<br>• oxyhemoglobin level (HT -<br>Oxy), and | Same |
{6}------------------------------------------------
K161237
| | SUBJECT DEVICE | PREDICATE | Significant |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HyperMed Imaging, Inc. | HyperMed, Inc. | Differences |
| Trade Name | HyperViewTM System | OxyVuTM-1 | |
| | • deoxyhemoglobin<br>(Deoxy) level<br>in superficial tissue. The<br>HyperViewTM system<br>displays two-dimensional,<br>color-coded images of<br>tissue oxygenation of the<br>scanned surface and<br>reports hyperspectral<br>tissue oxygenation<br>measurements for<br>selected tissue regions.<br>The HyperViewTM system is<br>indicated for use to<br>determine oxygenation<br>levels in superficial tissues<br>for patients with potential<br>circulatory compromise. | • deoxyhemoglobin (HT-<br>Deoxy) level<br>in superficial tissue. The<br>OxyVuTM-1 system displays<br>two-dimensional color<br>coded<br>images of tissue<br>oxygenation of the scanned<br>surface and reports<br>hyperspectral tissue<br>oxygenation measurements<br>for selected tissue regions.<br>The OxyVuTM-1 system is<br>indicated for use to<br>determine oxygenation<br>levels in superficial tissues<br>for patients with potential<br>circulatory compromise. | |
| System Components | • Hand-held, battery<br>operated device<br>including built in<br>resistive touchscreen<br>control and user<br>interface software<br>• PC software application<br>for reports<br>• Desktop docking station<br>to provide battery<br>charging and USB to<br>computer data interface<br>• USB computer interface<br>• Optionally tripod<br>mountable<br>• Calibration verification<br>cards are only required<br>to be used by user for<br>periodic verification of<br>calibration<br>• Laser verification cards<br>are used for periodic<br>verification of laser<br>alignment | • System Console: Cart,<br>System Electronics, CPU,<br>Monitor, Keyboard,<br>Pointing Device and<br>Printer (AC powered)<br>• Hyperspectral imaging<br>instrument head with<br>support arm: broadband<br>illuminator, camera and<br>spectral filter for<br>collecting hyperspectral<br>imaging cube<br>• Single use OxyVuTM Check<br>Pads for calibration | Different; the update of<br>the system from the<br>cart-based unit to a<br>hand-held device<br>involved changes to the<br>product dimensions,<br>user interface, power<br>source, and packaging.<br>The change from cart-<br>based unit to hand-held<br>unit had no effect on<br>performance. There<br>were no additional<br>safety or effectiveness<br>concerns as the<br>hyperspectral<br>technology used is the<br>same. |
| Calibration | Unit is factory calibrated | Calibration is required | Similar; both require |
{7}------------------------------------------------
| | SUBJECT DEVICE | PREDICATE | Significant |
|---------------------------------------------------------------|---------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HyperMed Imaging, Inc. | HyperMed, Inc. | Differences |
| Trade Name | HyperView™ System | OxyVu™ -1 | |
| | and does not require<br>calibration before each<br>use. | before each use | calibration.<br>The change in<br>calibration frequency<br>had no effect on<br>performance. The<br>change in calibration<br>frequency does not<br>present any additional<br>safety or effectiveness<br>concerns. |
| Time per patient image | < 1 second capture time<br>5 — 10 minutes use | 3 minute capture time<br>10 to 15 minutes use | Similar method to<br>capture images.<br>Multiple images are<br>captured simultaneously<br>to reduce imaging time.<br>The change in image<br>time had no effect on<br>performance. The<br>patient image time<br>change does not present<br>any additional safety or<br>effectiveness concerns. |
| Spectrum of Light Used | Visible Spectrum Light<br>(400nm to 700nm) | Visible Spectrum Light<br>(400nm to 700nm) | Same |
| Wavelengths of Light<br>Used for Data Analysis<br>and Imaging | 8 Wavelengths | 15 Wavelengths | Similar; subject device<br>uses a subset of 8 of the<br>15 wavelengths used by<br>the predicate.<br>The change in number of<br>wavelengths had no<br>effect on performance.<br>The change in number of<br>wavelengths does not<br>present any additional<br>safety or effectiveness<br>concerns. |
| LEDs | CREE LEDs (filtered) | CREE LEDs (3.0 mW/cm²) | Similar; both use LEDs. |
{8}------------------------------------------------
| | SUBJECT DEVICE | PREDICATE | Significant<br>Differences |
|----------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HyperMed Imaging, Inc. | HyperMed, Inc. | |
| Trade Name | HyperViewTM System | OxyVuTM-1 | |
| | Illuminator 1: 11.0<br>mW/cm2<br>Illuminator 2: 6.1 mW/cm2 | (unfiltered) | The change in LEDs and<br>the filtering of their<br>output had no effect on<br>performance. The LED<br>change does not present<br>any additional safety or<br>effectiveness concerns. |
| LED/Laser<br>Classification | Class II Laser | Class I LED | Different; LED/Laser<br>classification changed<br>from Class I to Class II.<br>The classification<br>difference had no effect<br>on performance and<br>does not present any<br>additional safety or<br>effectiveness concerns. |
| Software | C, C++, Java and .net (for PC<br>application) | Matlab and Java | Similar as both use<br>software.<br>The change in<br>programming languages<br>had no effect on<br>performance. This<br>change does not present<br>any additional safety or<br>effectiveness concerns. |
| Region of Interest<br>(ROI) Area | Support for freehand<br>drawing ROI selection<br>(mm2) | Support for freehand<br>drawing ROI selection<br>(mm2) | Same |
| Image Type | Diagnostic | Diagnostic | Same |
| Measures | Oxygen saturation<br>Oxyhemoglobin level<br>Deoxyhemoglobin level | Oxygen saturation<br>Oxyhemoglobin level<br>Deoxyhemoglobin level | Same |
| Method of<br>Measurement | Spectral analysis at specific<br>wavelengths of light<br>returned from target (8<br>wavelengths, which are a | Spectral analysis at specific<br>wavelengths of light<br>returned from target (15<br>wavelengths) | Similar; as both use<br>wavelengths of light for<br>measurement.<br>The wavelengths for the |
{9}------------------------------------------------
| | SUBJECT DEVICE | PREDICATE | Significant<br>Differences |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HyperMed Imaging, Inc. | HyperMed, Inc. | |
| Trade Name | HyperViewTM System<br>wavelengths) | OxyVuTM-1 | |
| | | | hand-held subject device<br>are a subset of the<br>wavelengths used in the<br>predicate device and the<br>change from 15 to 8<br>wavelengths had no<br>effect on performance.<br>This change does not<br>present any additional<br>safety or effectiveness<br>concerns. |
| Output Display | Numeric<br>Two-dimensional color<br>map of approximate tissue<br>oxygenation | Numeric<br>Two-dimensional color map<br>of approximate tissue<br>oxygenation | Same |
| Graphical Images | • VIS - Visible<br>• Sat – Saturation Map<br>(new map based upon<br>the oxygen saturation<br>level)<br>• Oxy - Tissue<br>oxygenation map<br>• Deoxy - Tissue<br>deoxyhemoglobin map<br>• Note: The HSI map<br>replaced with Saturation<br>Map. | • VIS - Visible<br>• HSI - Tissue oxygenation<br>map of oxyhemoglobin<br>and deoxyhemoglobin<br>• Oxy - Tissue oxygenation<br>map<br>• Deoxy - Tissue<br>deoxyhemoglobin map<br>Note: The Oxygen<br>Saturation value displayed<br>without a map. | Similar; both use same<br>graphical images for VIS,<br>Oxy, Deoxy. The HSI<br>map was replaced with<br>the Saturation Map.<br>The change from an HSI<br>to an oxygen saturation<br>map had no effect on<br>performance. This<br>change does not present<br>any additional safety or<br>effectiveness concerns. |
| Selection Tools | Circle<br>Region<br>Perimeter with optional<br>second region | Circle<br>Region<br>Perimeter | Similar; both devices use<br>the same Selection<br>Tools, with added<br>feature for the<br>Perimeter Tool.<br>The additional feature in<br>the Perimeter Selection<br>Tool had no effect on<br>how the device<br>performed. This change<br>does not present any<br>additional safety or |
{10}------------------------------------------------
K161237 p.8/12
| | SUBJECT DEVICE | PREDICATE | Significant<br>Differences |
|-------------------------------------|-------------------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HyperMed Imaging, Inc. | HyperMed, Inc. | |
| Trade Name | HyperViewTM System | OxyVuTM-1 | effectiveness concerns. |
| Data Storage | DICOM images stored on device | Images stored in Matlab files | Similar; as both have data storage capabilities.<br>The change in data storage type (to an industry standard format) had no effect on performance of the device. This change does not present any additional safety or effectiveness concerns. |
| Operating Distance<br>from Patient | Approximately 15" | Approximately 17" | Similar; both products have approximately the same working distance from the patient.<br>The change in operating distance from the patient had no effect on device performance. This change does not present any additional safety or effectiveness concerns. |
| Imaged Area | 100 mm x 100 mm | 80 mm x 80 mm | Similar; imaged area is increased with hand-held subject device.<br>The increase in image area had no effect on performance. This change does not present any additional safety or effectiveness concerns. |
| Electrical Safety<br>Testing Passed | Yes<br>ANSI/AAMI ES60601-<br>1:2005/(R)2012<br>IEC 60601-1-2:2007 | Yes<br>IEC 60601-1<br>IEC 60601-1-2 | Same |
{11}------------------------------------------------
K161237 p.9/12
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the HyperView system, and in showing substantial equivalence to the predicate device of this 510(k) submission, HyperMed a number of tests. The HyperView system meets all requirements for design characteristics, non-clinical performance testing, EMC/EMI testing, electrical safety, laser safety and battery safety testing to confirm that the output meets the design inputs and specifications for the device.
The HyperView system passed all testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
- . Software verification and validation testing per FDA Guidance and IEC 62304: conformance of software development life cycle for the HyperView System and compliance to the requirements of the FDA guidance document for software contained in a medical device.
- Electrical safety testing per ANSI/AAMI ES60601-1: PASS
- . Electromagnetic compatibility testing per IEC 60601-1-2: PASS
- Laser safety testing per IEC 60825-1:2007: PASS
- . Photobiological safety of lamps and lamp systems per IEC 62471:2006: PASS
- . Secondary cells and batteries safety testing per IEC 62133:2012 : PASS
- . Transportation safety testing of Lithium Batteries per UN 38.3:2009: PASS
- Transit testing per ASTM D4169:2014: PASS
The HyperView passed all the testing in accordance with internal requirements, national standards, and international standards shown above to support substantial equivalence of the subject device.
#### 10. Clinical Performance Data
A validation study was performed with the HyperView system for assessing the hyperspectral imaging in persons with vascular disease and / or diabetes mellitus and potentially compromised tissue oxygenation. The study was conducted to evaluate the HyperView handheld system with the predicate OxyVu™-1 console system device. The study was performed in compliance with Good Clinical Practices (GCP) with subjects enrolled in an IRB approved study that were consented for participation according to the intended use of the device, defined inclusion criteria, and defined exclusion criteria; with the purpose of meeting the study objectives.
The study objectives were defined as the following:
{12}------------------------------------------------
#### p.10/12
- Comparison of hyperspectral imaging tissue oxygenation as the difference in the mean tissue . oxygen saturation (O2Sat) measurements of the OxyVu™-1 compared to the HyperView handheld system.
- . Secondary endpoints for this study include testing of mean oxyhemoglobin (Oxy) and deoxyhemoglobin (Deoxy) levels with stratification by the following:
- Skin pigmentation and melanin concentration levels O
- Non-ulcerated and ulcerated areas O
- Degree of peripheral artery disease in the study limb o
The clinical study was planned for a minimum of 20 (twenty) subjects to be enrolled, which was met. The subjects had images taken of target areas with both the HyperView handheld system and the predicate device, during a single session. No follow up visits were required.
The outcomes measurements for the clinical study are summarized as follows:
#### Primary Efficacy Endpoint Results
The pre-specified primary efficacy endpoint for the validation study was to compare the O2Sat measurement difference between the investigational HyperView and predicate OxyVu™-1 device ((HyperView O2Sat minus OxyVuTM-1 O2Sat).
A Student's t-tests was performed using SAS version 9.4 (SAS Institute Inc., Cary, NC) to compare the mean differences with non-adjusted results. A difference that is not greater than 7% nor less than -7% between the mean O2Sat values for each device was deemed substantially equivalent.
A mean O2Sat difference of 3.22 +/- 3.38% was demonstrated between the HyperView and the OxyVu™-1 devices in this study. Lower confidence intervals and upper confidence intervals were 2.58% and 3.87% respectively.
A second analysis was performed comparing the study limbs by image sequence. The conclusions remained the same. The mean differences in O2Sat between the devices were 3.17%, 3.0% and 3.5% respectively.
#### Secondary Efficacy Endpoint Results
Secondary objectives entailed assessing O2Sat, Oxy, and Deoxy differences stratified by skin tone, presence of ulcers, and an ABI indicative of peripheral artery disease.
#### Stratification by Skin Tone:
The mean difference was 2.42% for the light skin tone and 4.64% for medium skin tone. While the means of these two groups were statistically different (P>.0002), the mean values and confidence intervals were all within a 5% difference between the two cameras.
An analysis on the index forearm region was also performed, as all three skin tone groups were represented for this location (light, medium, and dark). In this location, there were no
{13}------------------------------------------------
statistically significant differences noted between the skin tone groups (P>0.1968). The mean differences were 4% or less for all three groups.
#### Stratification by Non-ulcerated and Ulcerate Study Limbs
Eleven study limbs (31%) had a foot ulcer present. Each study limb had 3 sets of paired images for a total of 33 paired data sets. The mean difference in O2Sat between the devices was 1.91% with lower and upper confidence intervals of 0.55% and 3.26% in study limbs with a foot ulcer and 3.8% when ulcers were not present. The mean O2Sat difference was significantly smaller between the devices when ulcers were present on the foot (P>0.0068).
#### Stratification by Peripheral Artery Disease (PAD) of the Study Limb
There were 15 study limbs that met this definition. The mean difference in O2Sat for this group was 2.6% with lower and upper confidence intervals of 1.14% and 4.06% respectively.
There were no statistically significant differences noted between the device measurements of O2Sat between study limbs with PAD compared to study limbs without PAD (P>0.4449).
### Oxy Differences
An assessment was made of the difference in Oxy for paired images. Oxy values are on a numeric scale of 0 to 255. A mean Oxy difference of 14.07 (5.5%) was observed between the two devices.
Oxy differences by skin tone on the plantar surface of the foot demonstrated mean difference of 11.94 (4.7%) in light skin toned feet and 17.85 (7.0%) in medium skin toned feet.
Oxy differences by skin tone on the forearm demonstrated a mean difference of 12.62 (4.9%) in light toned skin, 13.64 (5.3%) in medium toned skin, and 17.5 (6.8%) in dark toned skin.
Differences in Oxy between the two devices with an ulcer present was 11.3 (4.4%).
Difference in Oxy between for subjects with peripheral artery disease was 12.27 (4.8%).
## Deoxy Differences
An assessment was made of the difference in Deoxy for paired images. Deoxy values are on a numeric scale of 0 to 255. A mean difference of 1.95 (0.8%) was observed.
Deoxy differences by skin tone on the plantar surface of the foot demonstrated mean difference of 2.9 (1.1%) in light skin toned feet and 0.28 (0.1%) in medium skin toned feet.
Deoxy differences by skin tone of the forearm demonstrated a mean difference of 2.47 (1.0%) in light toned skin, 1.1 (0.4%) in medium toned skin, and 2.38 (0.9%) in dark toned skin.
Deoxy values between the two devices for study limbs with ulcerated tissue (challenged tissue) demonstrated a mean difference of 3.09 (1.2%).
The difference in Deoxy between the two devices in study limbs with PAD was 3.27 (1.3%).
The validation study concluded that the HyperView system is safe and effective for its intended use and the outcomes of the study met the stated objectives. The validation study supports the indication for use as a noninvasive tissue oxygenation measurement system that reports approximate values of
{14}------------------------------------------------
#### K161237
## p.12/12
oxygen saturation (O2Sat), oxyhemoglobin level (Oxy), and deoxyhemoglobin (Deoxy) level in superficial tissue.
## 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device or the device has the same intended use and different technological characteristics provided it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.
The HyperView™ Hyperspectral Tissue Oxygenation Measurement System, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.